The Innovita Flu A/B Antigen Rapid Test is a rapid immunochromatographic assay for the qualitative detection of influenza A and B viral nucleoprotein antigens directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. The test is intended for use as an aid in the differential diagnosis of acute influenza A and B viral infections. The test is not intended for the detection of influenza C viruses. Negative test results are presumptive and should be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions.
Device Story
Rapid chromatographic immunoassay for qualitative detection of influenza A and B nucleoprotein antigens. Input: nasopharyngeal swab specimen. Principle: double antibody sandwich immunoassay; specimen applied to well; antigen forms immune complex with latex-labeled monoclonal antibodies; complex migrates to test zone; captured by specific antibodies forming red-purple line if concentration exceeds LoD. Internal control line confirms test validity. Used in clinical settings; performed by healthcare professionals. Output: visual presence/absence of colored lines on test cassette. Results aid differential diagnosis of acute influenza; negative results are presumptive and require molecular confirmation. Benefits: rapid point-of-care identification of influenza A/B to guide patient management.
Clinical Evidence
Prospective clinical study (n=1101) compared device against FDA-cleared molecular assay during 2023-2024 season. Influenza A: PPA 85.7% (95% CI: 80.6%-89.5%), NPA 99.5% (95% CI: 98.8%-99.8%). Influenza B: PPA 85.7% (95% CI: 72.2%-93.3%), NPA 100% (95% CI: 99.6%-100%). Analytical performance included LoD, inclusivity, cross-reactivity, and microbial interference testing.
Technological Characteristics
Qualitative lateral flow immunoassay. Components: test cassette (nitrocellulose membrane), extraction diluent, flocked swabs. Principle: double antibody sandwich using latex microsphere-labeled monoclonal antibodies. Visual readout. No instrumentation. Sterilization via ethylene oxide (ISO 11135). Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for symptomatic patients of all ages with signs/symptoms of respiratory infection for qualitative detection of influenza A and B viral nucleoprotein antigens from nasopharyngeal swabs. Not for influenza C detection. Negative results are presumptive and require confirmation by molecular assay or culture.
Regulatory Classification
Identification
An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY
I Background Information:
A 510(k) Number
K250398
B Applicant
Innovita (Tangshan) Biological Technology CO., LTD
C Proprietary and Established Names
Innovita Flu A/B Antigen Rapid Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PSZ | Class II | 21 CFR 866.3328 - Influenza Virus Antigen Detection Test System | MI - Microbiology |
II Submission/Device Overview:
A Purpose for Submission:
The purpose of this submission is to request premarket notification 510(k) clearance for the Innovita Flu A/B Antigen Rapid Test for prescription use.
B Measurand:
Influenza type A and type B nucleoprotein antigens
C Type of Test:
Qualitative lateral flow immunoassay
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
{1}
The Innovita Flu A/B Antigen Rapid Test is a rapid immunochromatographic assay for the qualitative detection of influenza A and B viral nucleoprotein antigens directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection.
The test is intended for use as an aid in the differential diagnosis of acute influenza A and B viral infections. The test is not intended for the detection of influenza C viruses. Negative test results are presumptive and should be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for influenza A were established during the December 2023 and July 2024 influenza season when influenza A/H1N1pdm09, influenza A/H3N2 and influenza B/Victoria viruses were the predominant influenza viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If an infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
Not applicable. There is no associated instrumentation as the test is visually read.
## IV Device/System Characteristics:
### A Device Description:
The Innovita Flu A/B Antigen Rapid Test is intended for the qualitative detection of Influenza Virus type A and Influenza Virus type B antigen directly from nasopharyngeal swab (NPS) specimen of symptomatic patients. The Innovita Flu A/B Antigen Rapid Test is validated for testing nasopharyngeal swab without transport media. The nasopharyngeal swab does not use biotin-streptavidin/avidin chemistry.
The test cassette in the test kit is assembled with a test strip in a plastic housing that contains a nitrocellulose membrane with three lines: two test lines (Flu A line and Flu B line) and a control line (C line).
Briefly, the NPS specimen is collected from symptomatic individuals using the supplied flocked swabs. Then, the lid of the tube containing the extraction diluent is unsealed and the swab is inserted into the tube. The swab tip should be completely immersed in the solution, and then stirred 10 - 15 times. The tube is squeezed when removing the swab to extract the contents of the swab. The tube containing the diluted specimen with the extraction diluent is then closed with a lid. Three (3) drops of the extracted specimen is added into the specimen well on the test cassette. After 10 - 30 minutes, the results are read visually.
{2}

Figure 1: Pictorial description of the steps for Innovita Flu A/B Antigen Rapid Test
# Kit Components
The Innovita Flu A/B Antigen Rapid Test consists of the following components:
Test cassette
- Extraction diluent
Swab
Nasopharyngeal swab
Package Insert
- Quick Reference Instructions (QRI)
External Controls
$\mathrm{O}$ Influenza A Positive Control
$\mathrm{O}$ Influenza B Positive Control
Negative Control
# B Principle of Operation:
The Innovita Flu A/B Antigen Rapid Test is a double antibody sandwich immunoassay-based test intended to detect influenza A and B viral nucleoprotein antigens in nasopharyngeal samples. The test device consists of the specimen zone and the test zone.
- Specimen zone: The specimen zone contains monoclonal antibody against the Flu A/Flu B antigen labeled with latex microspheres and chicken IgY antibody conjugated with latex microspheres.
- Test zone: The test line contains the other monoclonal antibody against Flu A/Flu B antigen. The control line contains rabbit anti-chicken IgY antibody.
# A Interpretation of Results
The qualitative results of the Innovita Flu A/B Antigen Rapid Test are visually interpreted by the user.
{3}
# 1. Positive results:
The presence of two red-purple lines at the positions of Flu A and Control (C) in the result window indicates positive for Flu A antigen. The presence of two red-purple lines at the positions of Flu B and Control (C) indicates positive for Flu B antigen. The presence of three red-purple lines appear at the positions of Flu A, Flu B and Control (C) indicates positive for both Flu A antigen and Flu B antigen positive.
# 2. Negative results:
Only one red-purple line appearing at the control line (C) indicates negative result.
# 3. Invalid
If the control line (C) fails to appear, no matter whether the Flu A/Flu B line is visible or not, the test is invalid. Review the procedure and repeat the test with a new test device.
Examples of the positive, negative, and invalid results interpretations are provided below.

Figure 3: Pictorial representation of the Interpretation of Results for Innovita Flu A/B Antigen Rapid Test (In this image, $C =$ Control Line, Flu A = Influenza A Test Line, and Flu B = Influenza B Test Line)
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
CareStart Flu A&B Plus
# B Predicate 510(k) Number(s):
K191514
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K191514 - CareStart Flu A&B Plus | K250398 - Innovita Flu A/B Antigen Rapid Test |
| --- | --- | --- |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The CareStart Flu A&B Plus is an in vitro rapid immunochromatographic assay for the qualitative detection of influenza virus type A and B nucleoprotein | The Innovita Flu A/B Antigen Rapid Test is a rapid immunochromatographic assay for the qualitative detection of influenza A and B viral nucleoprotein antigens |
{4}
| | antigens directly nasopharyngeal swab specimens of symptomatic patients. | directly from nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. |
| --- | --- | --- |
| | The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and B viral infections. This test is intended to distinguish between influenza type A and/or B virus in a single test. This test is not intended to detect influenza type C viral antigens. Negative test results are presumptive and should be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the basis for treatment or other patient management decisions. | The test is intended for use as an aid in the differential diagnosis of acute influenza A and B viral infections. The test is not intended for the detection of influenza C viruses. Negative test results are presumptive and should be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza viral infection and should not be used as the sole basis for treatment or other patient management decisions. |
| | Performance characteristics for influenza A and B were established during the 2018-2019 influenza season when influenza A/H3N2 and A/H1N1pdm09, and B/Victoria were the predominant influenza viruses in circulation. When other influenza viruses are emerging, performance characteristics may vary. | Performance characteristics for influenza A were established during the December 2023 and July 2024 influenza season when influenza A/H1N1pdm09, influenza A/H3N2 and influenza B/Victoria viruses were the predominant influenza viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. |
| | If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be | If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel |
{5}
VI Standards/Guidance Documents Referenced:
| Document Title | Issued by | Applicable study |
| --- | --- | --- |
| Guidance: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Assays - Class II Special Controls Guidance for Industry and FDA Staff | FDA/CDRH | All Studies |
| Guidance: Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses | FDA/CDRH | All Studies |
| Guidance: Testing for Biotin Interference in In Vitro Diagnostic Devices | FDA/CDRH | Interference |
| Guideline: CLSI EP12 Evaluation of Qualitative, Binary Output Examination Performance | CLSI | All Studies |
| Guideline: CLSI EP07 Interference Testing in Clinical Chemistry | CLSI | Interference |
| Guideline: CLSI EP37-Supplemental Tables for Interference Testing in Clinical Chemistry | CLSI | Interference |
| Standard: ISO 14971-2019 Medical devices - Application of risk management to medical devices | ISO | All Studies Risk Management |
| Standard: ISO 11135:2014, Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices | ISO | Sterility |
{6}
| Standard: ISO 11737-2:2020 Sterilization of health care products — Microbiological methods | ISO | Sterility |
| --- | --- | --- |
| Standard: ISO 11135:2014, Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices | | |
| Standard: ISO 10993-7, Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals | | |
| Standard: ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process | ISO | Biocompatibility |
| Standard: ISO 10993-5, Third edition, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity | | |
| Standard: ISO 10993-10, Third Edition, Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization | | |
| Standard: ISO 10993-11:2017, Biological evaluation of medical devices — Part 11: Tests for systemic toxicity | | |
| Standard: ISO 10993-23:2021, Biological evaluation of medical devices — Part 23: Tests for irritation USP-NF (151) Pyrogen test | | |
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
A combined precision and reproducibility study was conducted for the Innovita Flu A/B Antigen Rapid Test. Three concentrations each of Flu A: H1N1 A/Dominican Republic/7293/2013 pdm09 and Flu B: Victoria/ B/Michigan/01/2021 were spiked into pooled nasal fluid of which 50 microliters (μL) of each sample were applied to dry nasal swabs to prepare the following test sample panel members:
1. Negative Sample
2. High Negative Sample at 0.1xLoD for each strain
3. Low Positive Sample at 1xLoD for each strain
4. Positive Sample at 3xLoD for each strain
The study was conducted at three (3) different sites to assess variability between days, runs, operators, reagent lots, and site, with the following study design: 3 sites x 2 operators per site x 1 lot per site x 5 days x 1 run x 5 replicates yielding 150 results/sample panel. Samples were blinded and randomized before allotting them to the operators and processed per the IFU of the candidate device.
The results were in 100% agreement with the expected results for negative, 1xLoD, and 3xLoD samples by lot/site, by operator, by day, and overall for both analytes. All results were negative when tested with high negative samples. The results are summarized below by lot/site and are presented as percent positive agreement.
{7}
Table 1.1. Precision - Reproducibility Study results
| Sample | Positive Percent Agreement | | | | | | Total Positive Percent Agreement (95% Confidence Intervals) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1/Lot 1 | | Site 2/Lot 2 | | Site 3/Lot 3 | | | |
| | Flu A | Flu B | Flu A | Flu B | Flu A | Flu B | Flu A | Flu B |
| Negative sample | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/150 (0%) (0%-2.5%) | 0/150 (0%) (0%-2.5%) |
| Flu A high negative (0.1×LoD) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/150 (0%) (0%-2.5%) | 0/150 (0%) (0%-2.5%) |
| Flu A low positive (1×LoD) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 150/150 (100%) (97.5%-100%) | 0/150 (0%) (0%-2.5%) |
| Flu A moderate positive (3×LoD) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 150/150 (100%) (97.5%-100%) | 0/150 (0%) (0%-2.5%) |
| Flu B high negative (0.1×LoD) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/50 (0%) | 0/150 (0%) (0%-2.5%) | 0/150 (0%) (0%-2.5%) |
| Flu B low positive (1×LoD) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/150 (0%) (0%-2.5%) | 150/150 (100%) (97.5%-100%) |
| Flu B moderate positive (3×LoD) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/50 (0%) | 50/50 (100%) | 0/150 (0%) (0%-2.5%) | 150/150 (100%) (97.5%-100%) |
A separate precision study was conducted to assess the precision of the external controls that will be packaged with the test kit. Three lots of the external controls were tested with the following study design: 3 operators (1 lot per operator) x 1 day x 20 replicates.
Results were in $100\%$ agreement with the expected results and are summarized below as positive percent agreement by lot/operator.
Table 1.2. Precision Study results for external controls
| Sample | Positive Percent Agreement | | | | | | Total Positive Percent Agreement (95% Confidence Intervals) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Lot 1 | | Lot 2 | | Lot 3 | | | |
| | Flu A | Flu B | Flu A | Flu B | Flu A | Flu B | Flu A | Flu B |
| Negative control swab | 0/20 (0%) | 0/20 (0%) | 0/20 (0%) | 0/20 (0%) | 0/20 (0%) | 0/20 (0%) | 0/60 (0%) (0%, 6.02%) | 0/60 (0%) (0%, 6.02%) |
| Flu A positive control swab | 20/20 (100%) | 0/20 (0%) | 20/20 (100%) | 0/20 (0%) | 20/20 (100%) | 0/20 (0%) | 60/60 (100%) (93.98%,100%) | 0/60 (0%) (0%, 6.02%) |
| Flu B positive control swab | 0/20 (0%) | 20/20 (100%) | 0/20 (0%) | 20/20 (100%) | 0/20 (0%) | 20/20 (100%) | 0/60 (0%) (0%, 6.02%) | 60/60 (100%) (93.98%,100%) |
{8}
2. Linearity:
Not applicable; the device is a qualitative assay with binary visually read results.
3. Analytical Specificity/Interference:
a) Cross Reactivity and Microbial Interference:
Cross reactivity and microbial interference studies were conducted to assess potential false positives or interference from other respiratory pathogens in nasal samples. A total of 49 potential cross reactive or interfering microorganisms were individually spiked into pooled nasal fluid with or without live influenza A or B at 2.5x LoD. Fifty (50) microliters of each sample was directly applied to the test swab and processed per the IFU. Each organism was tested in replicates of 3 using 3 different reagent lots.
Results are summarized in the table below. Neither cross-reactivity nor microbial interference was observed for any of the tested microorganisms at the concentration used in the study.
Table 2. Cross reactivity and microbial interference results
| Microorganism | Concentration tested | Flu A positive swabs | | Flu B positive swabs | | Negative swabs | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Positive rate | | Positive rate | | Positive rate | |
| | | Flu A | Flu B | Flu A | FluB | Flu A | Flu B |
| Human rhinovirus 1A | 6.6×10^6 PFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Human parainfluenza virus 1 (HPIV-1) | 3.2×10^6 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Acinetobacter calcoaceticus | 6.2×10^7 CFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Human adenovirus 1 | 3.2×10^6 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Human mastadenovirus B7 | 1.78×10^6 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Bordetella pertussis | 3.08×10^6 CFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Candida albicans | 7.3×10^7 CFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Chlamydophila pneumoniae | 4.6×10^7 CFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Human coronavirus 229E | 3.2×10^5 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Beta coronavirus 1 (Human coronavirus OC43) | 1.0×10^5 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Corynebacterium diphtheriae | 1.28×10^8 CFU/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| Human Coxsackievirus B4 | 3.2×10^6 TCID50/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
{9}
| Cytomegalovirus | 1.4×105TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Enterococcus faecalis | 1.78×109CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Escherichia coli | 2.88×107CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Gardnerella vaginalis | 9.92×106CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Haemophilus influenzae | 3.36×107CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Klebsiella pneumoniae subsp. pneumoniae | 1.04×109CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Lacticaseibacillus casei | 4.32×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Legionella pneumophila subsp. pneumophila | 1.12×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Listeria monocytogenes | 2.28×109CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Measles virus | 1.78×105TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Human Metapneumovirus 27 (hMPV-27) Type A2 | 7.6×105TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Moraxella (Branhamella) catarrhalis | 8.8×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Mumps virus | 9.84×106TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Mycobacterium tuberculosis | 3.32×107CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Mycoplasma pneumoniae | 6.6×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Neisseria gonorrhoeae | 2.8×107CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Neisseria meningitidis | 1.44×106CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Neisseria sicca | 8.4×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Human parainfluenza virus 2 | 1.78×106TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Human parainfluenza virus 3 | 3.2×107TCID50/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Proteus vulgaris | 1.04×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
| Pseudomonas paraeruginosa | 6.4×108CFU/mL | 9/9(100%) | 0/9(0%) | 0/9(0%) | 9/9(100%) | 0/9(0%) | 0/9(0%) |
{10}
# b) Endogenous/Exogenous Interfering Substances Study
The performance of the Innovita Flu A/B Antigen Rapid Test was evaluated in the presence of potentially interfering substances that might be present in respiratory specimens. Each potential interferent was prepared in pooled nasal fluid in the presence or absence of $2.5\mathrm{xLoD}$ concentration of influenza virus and was tested by 3 operators using 3 different reagent lots. No cross-reactivity or interference was observed amongst the substances tested. The results are summarized below.
Table 3. Endogenous/exogenous interfering substances study results
| Interfering Substance | Final Concentration | Without Analytes | | With Analytes | |
| --- | --- | --- | --- | --- | --- |
| | | Positive rate | | Positive rate | |
| | | Flu A | Flu B | Flu A | Flu B |
| Acetaminophen | 10 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Acetyl salicylic acid | 15 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Beclomethasone | 15% | 0/3 | 0/3 | 3/3 | 3/3 |
| Citalopram | 100 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Citalopram + 2,5-Dimethyl-2,5-dimethyl-1,1-dioxane | 100 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Citalopram + 2,5-Dimethyl-1,1-dioxane + 2,5-dimethyl-1,1-dioxane + 2,5-dimethyl-2,5,6-trimethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6 | 100 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Citalopram + 2,5-Dimethyl-1,1-dioxane + 2,5-dimethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6 | 100 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
| Citalopram + 2,5-Dimethyl-1,1,2,3,4,5,6-tetramethyl-1,1,2,3,4,5,6 | 100 mg/mL | 0/3 | 0/3 | 3/3 | 3/3 |
{11}
# c) Biotin Interference
To detect if biotin interferes with the Innovita Flu A/B Antigen Rapid Test, biotin concentrations up to $4000\mathrm{ng / mL}$ were tested. Each dilution of biotin prepared in nasal fluid was tested by 3 operators using 3 different reagent lots with contrived positive samples containing $2.5\mathrm{x}$ LoD analytes for influenza A, and influenza B in pooled nasal fluid. No interference was observed. Results are summarized below.
Table 4. Biotin interference results summary
| Concentration of biotin interference | Flu A positive swabs | | Flu B positive swabs | | Negative swabs | |
| --- | --- | --- | --- | --- | --- | --- |
| | Positive rate | | Positive rate | | Positive rate | |
| | Flu A | Flu B | Flu A | Flu B | Flu A | Flu B |
| 0 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (00%) | 0/9 (00%) |
| 250 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (100%) | 0/9 (00%) |
| 500 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| 1000 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| 2000 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
{12}
| 4000 ng/mL | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) | 9/9 (100%) | 0/9 (0%) | 0/9 (0%) |
| --- | --- | --- | --- | --- | --- | --- |
## d) Competitive Interference
A competitive interference study was conducted to evaluate potential for a high concentration of one target analyte to interfere with the detection of another target analyte at low concentration. Testing was conducted in triplicates with different combinations of concentrations at 6x LoD and high concentrations (2000x LoD) of influenza A, and influenza B prepared in pooled nasal fluid.
The table below summarizes the results of the competitive interference study. There was no competitive interference observed.
| Operator | Lot No. | Low concentration sample | High concentration sample | # of positive results/total reps (concordance rate) | |
| --- | --- | --- | --- | --- | --- |
| | | | | Flu A | Flu B |
| 1 | Lot 1 | Flu A low | Flu B high | 3/3 (100%) | 3/3 (100%) |
| | | Flu B low | Flu A high | 3/3 (100%) | 3/3 (100%) |
| 2 | Lot 2 | Flu A low | Flu B high | 3/3 (100%) | 3/3 (100%) |
| | | Flu B low | Flu A high | 3/3 (100%) | 3/3 (100%) |
| 3 | Lot 3 | Flu A low | Flu B high | 3/3 (100%) | 3/3 (100%) |
| | | Flu B low | Flu A high | 3/3 (100%) | 3/3 (100%) |
## 4. Assay Reportable Range:
Not applicable; the device is a binary qualitative assay that is visually read.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### a) Controls
**Internal procedural control:**
A built-in internal procedural control is needed to indicate the test device is functioning properly and to ensure correct use of the device. The internal control is part of the test strip membrane and is therefore automatically run within the development time of each test. The control line yields a red-purple control line (C).
**External Controls:**
External Quality Control materials are supplied with the kit and consist of two positive swabs (influenza A and B) and 1 negative swab (negative for both influenza A and B).
### b) Stability Studies
#### i. Real Time Stability
A real-time stability study was conducted to evaluate stability and determine the shelf-life of the unopened test kit. Three lots of unopened Innovita Flu A/B Antigen Rapid Test were stored under two conditions: $30 \pm 2^{\circ}\mathrm{C}$ (which is the upper end of the $15 - 30^{\circ}\mathrm{C}$ room temperature range and representative of the worst-case scenario) and $4 \pm 2^{\circ}\mathrm{C}$. At defined intervals, each lot was assessed with following three-membered sample panel: negative clinical matrix, and co-spiked positive samples
{13}
with inactivated influenza A (H1N1), or live influenza B (Victoria) viruses, each at 3x LoD.
Fifty (50) μL of each sample were applied to the swab and tested according to the instructions for use. Three replicates of each sample were tested for each time point.
Baseline testing was performed within one month of each manufactured lot. Subsequent time testing time points were every three (3) months. At the time of clearance, all study data have met the protocol defined acceptance criteria, and support storage of the test kits at 4-30 °C for 12 months.
## ii. Transport Stability
Transport stability under simulated summer and winter shipping conditions was tested to evaluate worst-case shipping and handling conditions over an extended period of time. Performance of unopened test kits was assessed by comparing pre- and post-transport results, using the same sample panel described for real-time stability study (above). Samples were tested in triplicates each of three (3) device lots. Each kit was stored at the following temperature profiles for a five-cycle period to simulate transport conditions:
- Summer profile (Ranging from 22 - 40°C)
- Winter profile (Ranging from -10 – 18°C)
All results were as expected and demonstrated reagent stability in all transport conditions tested.
## 6. Detection Limit:
### a) Limit of Detection
A limit of detection (LoD) study was conducted to determine the lowest detectable concentration of four (4) influenza A virus strains (including H3N2, and H1N1 subtypes) and four (4) influenza B virus strains (representing the Victoria and Yamagata lineage) at which at least 95% of all true positive replicates tested positive. Fifty (50) μL sample of each virus subtype was diluted in nasal fluid and pipetted onto the dry swab and thereafter processed per the instructions for use. Testing was conducted on three (3) lots of test devices.
A preliminary LoD was first determined by testing serial 10-fold dilutions of viral stocks diluted in nasal fluid. The last concentrations that produced 3 positive test results from a 1:10 dilution was further evaluated using a 2-fold dilution series, in triplicate for each level, to refine the tentative LoD. The lowest concentration at which all 3 replicates were positive was treated as the tentative LoD. This LoD was then confirmed by testing 20 replicates/lot using 3 different lots with concentrations at the tentative limit of detection for each virus strain.
As per the acceptance criteria for confirmation of the LoD, at least 95% of the replicates (≥ 19/20) should be positive to be considered as the confirmed LoD. The confirmed LoDs observed for all strains were identical for the 3 lots tested. The results of the two-fold dilutions and confirmatory LoD for each strain are included below. The final LoD for each strain is presented in the bold in the table.
Table 6. Single analyte LoD results
{14}
| Virus strain | Virus concentration | | Positive replicates | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | TCID50/mL | TCID50/Swab | Preliminary LoD Results (Representative Lot) | Confirmatory LoD Results (3 lots) | | |
| Influenza A Flu A virus (H3N2) A/Virginia/ATCC6/2012 | 8.90E+03* | 4.45E+02 | 3/3 | Not tested | | |
| | 4.45E+03 | 2.23E+02 | 3/3 | | | |
| | 2.23E+03 | 1.11E+02 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 1.11E+03 | 5.56E+01 | 0/3 | Not tested | | |
| Influenza A Flu A virus (H3N2) (A/Baltimore/JH-0586/2022) | 5.00E+03* | 2.50E+02 | 3/3 | Not tested | | |
| | 2.50E+03 | 1.25E+02 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 1.25E+03 | 6.25E+01 | 2/3 | Not tested | | |
| Flu A virus (H1N1) (A/Dominican Republic/7293/2013 pdm09) | 5.00E+02* | 2.50E+01 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 2.50E+02 | 5.00E+03 | 2/3 | Not tested | | |
| Flu A virus (H1N1) A/Virginia/ATCC1/2009 TC isolate | 1.60E+03* | 8.00E+01 | 3/3 | Not tested | | |
| | 8.00E+02 | 4.00E+01 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 4.00E+02 | 2.00E+01 | 2/3 | Not tested | | |
| Flu B virus (Victoria lineage) (B/New Jersey/1/2012) | 8.80E+03* | 4.40E+02 | 3/3 | Not tested | | |
| | 4.40E+03 | 2.20E+02 | 3/3 | | | |
| | 2.20E+03 | 1.10E+02 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 1.10E+03 | 5.50E+01 | 1/3 | Not tested | | |
| Flu B virus (Victoria lineage) (B/Michigan/01/2021) | 5.70E+03* | 2.85E+02 | 3/3 | Not tested | | |
| | 2.85E+03 | 1.43E+02 | 3/3 | | | |
| | 1.43E+03 | 7.13E+01 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 7.13E+02 | 3.56E+01 | 1/3 | Not tested | | |
| Flu B virus (Yamagata lineage) (B/Brisbane/9/14) | 1.26E+02* | 6.30E+00 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 6.30E+01 | 3.15E+00 | 1/3 | Not tested | | |
| Flu B virus (Yamagata lineage) (Panama/45/90) | 3.80E+03* | 1.90E+02 | 3/3 | Not tested | | |
| | 1.90E+03 | 9.50E+01 | 3/3 | | | |
| | 9.50E+02 | 4.75E+01 | 3/3 | 20/20 | 20/20 | 20/20 |
| | 4.75E+02 | 2.38E+01 | 2/3 | Not tested | | |
*Last concentrations that produced three positive test results from 1:10 dilution
# b) Inclusivity
Analytical reactivity testing for the Innovita Flu A/B Antigen Rapid Test was performed to determine that the device can adequately detect the target analytes across a variety of strains. A selection of temporally, geographically, and genetically diverse influenza strains were tested for inclusivity, including 29 influenza A strains and 12 influenza B strains. In this study, each influenza virus strain was serially diluted to a concentration near the LoD and tested in replicates of 10 with the Innovita Flu A/B Antigen Rapid Test. The lowest concentrations at which 10/10 replicates were detected per lot with three different lots at the concentrations shown below.
Table 7. Analytical reactivity with variants of Influenza A and B
| Virus | Viral strain | Inclusivity |
| --- | --- | --- |
| Flu A virus (H1N1) | A/Baltimore/JH-22377/2022 | 1.6×105TCID50/mL |
| | A/Bangladesh/3002/2015 (H1N1) pdm09 | 1.3×104CEID50/mL |
{15}
| | A/Bretagne/06091/2023 | 5.5×103PFU/mL |
| --- | --- | --- |
| | A/California/04/2009 (H1N1) pdm09 | 1.1×102TCID50/mL |
| | A/California/07/2009 pdm09 | 2.3×104CEID50/mL |
| | A/California/08/2009 pdm09 | 1.6×104CEID50/mL |
| | A/Iowa/53/2015 (H1N1) pdm09 | 2.9×104CEID50/mL |
| | A/Massachusetts/15/2013 (H1N1) pdm09 | 1.6×104CEID50/mL |
| | A/Michigan/272/2017 | 9.6×103TCID50/mL |
| | A/Michigan/45/2015(H1N1) pdm09 | 1×104CEID50/mL |
| | A/New Hampshire/02/2010 (H1N1) pdm09 | 1.8×104CEID50/mL |
| | A/South Carolina/2/2010 (H1N1) pdm09 | 2.5×105CEID50/mL |
| | A/St. Petersburg/61/2015 | 9.3×105CEID50/mL |
| Flu A virus (H3N2) | A/Aichi/2/68 | 2.3×103CEID50/mL |
| | A/Alaska/232/2015 | 2.6×104CEID50/mL |
| | A/Baltimore/JH-0440/2022 | 2.8×104TCID50/mL |
| | A/Baltimore/JH-286/2021 | 2.8×105TCID50/mL |
| | A/Baltimore/JH-335/2022 | 8.9×104TCID50/mL |
| | A/California/2/2014 | 5.8×102TCID50/mL |
| | A/Hong Kong/4801/2014 | 9.6×105CEID50/mL |
| | A/Hong Kong/8/68 | 1.58×103CEID50/mL |
| | A/Indiana/08/2011(H3N2) | 8.1×102TCID50/mL |
| | A/Michigan/15/2014 | 9.3×103FFU/mL |
| | A/Minnesota/11/2010(H3N2)v | 2.2×104CEID50/mL |
| | A/Perth/16/2009 | 1.1×105TCID50/mL |
| | A/Singapore/INFIMH-16-0019/2016 | 2.2×104CEID50/mL |
| | A/Switzerland/9715293/2013 | 1.6×105CEID50/mL |
| | A/Texas/71/2017 | 9.3×103FFU/mL |
| | A/Wisconsin/67/2005 (H3N2) | 1.5×105CEID50/mL |
| Flu B virus (Victoria lineage) | B/Baltimore/JH002/2021 | 1.6×104TCID50/mL |
| | B/Brisbane/60/2008 | 1×104CEID50/mL |
| | B/Colorado/6/2017 | 1.6×104CEID50/mL |
| | B/Florida/78/2015 | 2.8×104TCID50/mL |
| | B/Hong Kong/286/2017 | 2.7×103TCID50/mL |
| | B/Maryland/15/2016 | 1.3×103TCID50/mL |
| | B/New Hampshire/01/2021 | 1.3×102TCID50/mL |
| Flu B virus (Yamagata lineage) | B/Florida/4/2006 | 5×104CEID50/mL |
| | B/Guangdong/120/00 | 1.68×102TCID50/mL |
| | B/Massachusetts/2/2012 | 1.0×104TCID50/mL |
| | B/Texas/6/11 | 3.8×102TCID50/mL |
| | B/Wisconsin/1/2010 BX-41A | 1.8×105CEID50/mL |
{16}
7. High Dose Hook Effect:
The device was tested to determine if the device was affected by a high dose hook effect at high concentrations of the analytes. A high dose of inactive influenza A and B were tested in this study. Fifty (50) microliters of each sample was added directly to the head of the swabs. Swabs were processed per the test's IFU/QRI with one device lot. No high dose hook effect was observed for the high concentration tested.
Table 8. High-dose hook effect study results
| Sample | Concentration | Positive Results / Total Replicates | |
| --- | --- | --- | --- |
| | | Influenza A | Influenza B |
| Flu A virus (H1N1) A/Baltimore/JH-22377/2022 | 1.6×10^{9} TCID_{50}/mL | 9/9 (100%) | 0/9 (0%) |
| Flu A virus (H3N2) A/Switzerland/9715293/2013 | 1.6×10^{9} TCID_{50}/mL | 9/9 (100%) | 0/9 (0%) |
| Flu B virus (Victoria Lineage) B/Colorado/6/2017 | 1.6×10^{8} CEID_{50}/mL | 0/9 (0%) | 9/9 (100%) |
| Flu B virus (Yamagata Lineage) B/Wisconsin/1/2010 BX-41A | 1.8×10^{9} CEID_{50}/mL | 0/9 (0%) | 9/9 (100%) |
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable. See "C. Clinical Studies." for performance comparison with a clinical comparator.
2. Matrix Comparison:
This device is only intended for use with nasopharyngeal swab specimens. As no other specimen or sample type are claimed for us with this device, a matrix comparison study is not applicable.
C Clinical Studies:
1. Clinical Sensitivity and Specificity
A prospective clinical study was conducted to assess the performance of the candidate test when compared to a 510(k)-cleared influenza RT-PCR assay with an extraction step. The study prospectively enrolled symptomatic subjects at three (3) clinical study sites between December 2023, and July 2024.
A nasopharyngeal sample for the RT-PCR comparator was collected first from one nostril and a nasopharyngeal sample for the candidate test was then collected from the other nostril. This order of sample collection was consistent across the entire study. Comparator test samples were collected at the clinical study site and inserted into viral transport media per the IFU of the comparator test and transported to a central laboratory for testing. Samples for the candidate device were collected and were immediately tested with the Innovita Flu A/B Antigen Rapid Test according to the instructions for use.
{17}
A total of 1109 subjects were enrolled, of which 8 samples were excluded due to being lost during transit to the laboratory for comparator testing. Hence, 1101 subjects were evaluable.
The number of positives and positivity rate of influenza A and influenza B specimens per age group are presented in Table below.
Table 9. Prospective clinical study during the 2023/2024 influenza season
| Age Group | Positivity Rates for influenza A based on an FDA-cleared molecular test during the Clinical Study | | |
| --- | --- | --- | --- |
| | Number of Subjects | Number of Influenza A Positives | Influenza A Positivity Rate |
| ≤5 Years of Age | 463 | 121 | 26.1% |
| 6-21 Years of Age | 399 | 95 | 23.8% |
| ≥22 Years of Age | 239 | 21 | 8.8% |
| Total | 1101 | 237 | 21.5% |
| Age group | Positivity Rates for influenza B based on an FDA-cleared molecular test during the Clinical Study | | |
| | Number of Subjects | Number of Influenza B Positives | Influenza B Positivity Rate |
| ≤5 Years of Age | 463 | 14 | 3.0% |
| 6-21 Years of Age | 399 | 19 | 4.8% |
| ≥22 Years of Age | 239 | 9 | 3.8% |
| Total | 1101 | 42 | 3.8% |
Results from the Innovita Flu A/B Antigen Rapid Test were compared with those from the result obtained by highly sensitive influenza RT-PCR comparator assay, leading to the following performance estimates:
Table 10. Clinical performance estimates - influenza A
| | Comparator Positives | Comparator Negatives | Total |
| --- | --- | --- | --- |
| Candidate Positives | 203 | 4 | 207 |
| Candidate Negatives | 34 | 860 | 894 |
| Total | 237 | 864 | 1101 |
| Positive Percent Agreement (PPA) = 85.7% (95% CI: 80.6%-89.5%) | | | |
| Negative Percent Agreement (NPA) = 99.5% (95% CI: 98.8%-99.8%) | | | |
Table 11. Clinical performance estimates - influenza B
| | Comparator Positives | Comparator Negatives | Total |
| --- | --- | --- | --- |
| Candidate Positives | 36 | 0 | 36 |
| Candidate Negatives | 6 | 1059 | 1065 |
| Total | 42 | 1059 | 1101 |
| Positive Percent Agreement (PPA) = 85.7% (95% CI: 72.2%-93.3%) | | | |
| Negative Percent Agreement (NPA) = 100% (95% CI: 99.6%-100%) | | | |
{18}
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
When the test is valid, it produces binary values, positive or negative for influenza A or B antigens.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.