Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g. - Metastatic tumors - Recurrent glioblastomas - Trigeminal neuralgia - Medically refractory essential tremor - Orbital tumors - Ocular tumors - Optic nerve tumors - Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas) - Skull base tumors - Head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland) - Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors) - Refractory mesial temporal lobe epilepsy in adults - Refractory epilepsy associated with structural changes such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations (adult and pediatric)
Device Story
Leksell Gamma Knife (Esprit, Icon, Perfexion) is a stereotactic radiosurgery system delivering ionizing radiation to intracranial targets. Device uses 192 radiation sources to deliver prescribed doses as one or more shots to target sites. Operated by clinicians in clinical settings; system includes patient fixation (frames/masks), imaging (kV X-ray/CBCT in Esprit/Icon), and motion management (HDMM). Output is radiation dose delivery to specific head structures. System updates include new ECU, kV-generator, and HDMM camera to replace obsolete hardware, supported by Control System SW 11.6. Clinicians use system to treat tumors and neurological conditions; output affects clinical decision-making by enabling precise, non-invasive lesion ablation. Benefits include targeted therapy for refractory conditions and tumors.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence established via non-clinical module, integration, and system-level verification and validation testing, including design and usability validation per IEC 62366-1.
Technological Characteristics
Radionuclide radiation therapy system using 192 radiation sources. Models: Esprit, Icon, Perfexion. Features: kV X-ray/CBCT (Esprit/Icon), HDMM motion management, G-frame/Vantage/Mask fixation. Software: Control System SW 11.6 running on Windows 10. Standards: IEC 60601-1 series, IEC 62304, ISO 10993 series, IEC 62366-1. Biocompatible materials used.
Indications for Use
Indicated for stereotactic irradiation of intracranial structures including metastatic tumors, recurrent glioblastomas, trigeminal neuralgia, essential tremor, various benign/malignant head/neck/orbital/ocular/pediatric tumors, and refractory epilepsy (mesial temporal lobe or structural-related) in adult and pediatric populations.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
Predicate Devices
Leksell Gamma Knife® (Elekta Esprit, Icon™, and Perfexion™) (K222047)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2025
Elekta Solutions AB
Simon Sjöhage
Senior Regulatory Affairs Engineer
Hagaplan 4
Stockholm, 113 68
Sweden
Re: K250391
Trade/Device Name: Leksell Gamma Knife® (Elekta Esprit); Leksell Gamma Knife® (Icon™); Leksell Gamma Knife® (Perfexion™)
Regulation Number: 21 CFR 892.5750
Regulation Name: Radionuclide Radiation Therapy System
Regulatory Class: Class II
Product Code: IWB
Dated: February 12, 2025
Received: February 12, 2025
Dear Simon Sjöhage:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250391 - Simon Sjöhage
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250391 - Simon Sjöhage
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250391
Device Name
Leksell Gamma Knife® (Elekta Esprit);
Leksell Gamma Knife® (Icon™);
Leksell Gamma Knife® (Perfexion™)
Indications for Use (Describe)
Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g.
- Metastatic tumors
- Recurrent glioblastomas
- Trigeminal neuralgia
- Medically refractory essential tremor
- Orbital tumors
- Ocular tumors
- Optic nerve tumors
- Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas)
- Skull base tumors
- Head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland)
- Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors)
- Refractory mesial temporal lobe epilepsy in adults
- Refractory epilepsy associated with structural changes such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations (adult and pediatric)
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Elekta
K250391
# Traditional 510(k) Summary (21 CFR § 807.92)
## I. Submitter
Address
Elekta Solutions AB
Hagaplan 4
113 68 Stockholm
Sweden
Contact
Simon Sjöhage
Senior Regulatory Affairs Engineer
+46707994829
Simon.Sjohage@elekta.com
510(k) Number
K250391
Date Prepared
2025-06-05
## II. Device
Trade Name
Leksell Gamma Knife® – Elekta Esprit
Leksell Gamma Knife® Icon™
Leksell Gamma Knife® Perfexion™
Product Classification
Class II
Common Name
Radionuclide radiation therapy system
Regulation Number
892.5750
Product Code
IWB
## III. Predicate Device
Predicate #
K222047
Predicate Trade Name
Leksell Gamma Knife® (Elekta Esprit, Icon™, and Perfexion™)
Predicate Product Code
IWB
## IV. Device Description Summary
Leksell Gamma Knife® (available models Elekta Esprit, Icon™, and Perfexion™) is a radiosurgery system for use in the stereotactic irradiation of head structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
Elekta Solutions AB
LGK Traditional 510(k)
June 2025
Elekta Solutions AB | Box 7593 | SE-103 93 Stockholm | Sweden
Tel: +46 8 587 254 00 | Fax: +46 8 587 255 00 | www.elekta.com | Reg No. 559157-5286
{5}
Elekta
# V. Intended Use/ Indications for Use
Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g.
- Metastatic tumors
- Recurrent glioblastomas
- Trigeminal neuralgia
- Medically refractory essential tremor
- Orbital tumors
- Ocular tumors
- Optic nerve tumors
- Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas)
- Skull base tumors
- Head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland)
- Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors)
- Refractory mesial temporal lobe epilepsy in adults
- Refractory epilepsy associated with structural changes such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations (adult and pediatric)
# VI. Indications for Use Comparison
The differences are not critical to the intended use and the differences do not affect the safety and effectiveness of the device when used as labeled as the device has the identical intended use (teletherapy to brain structures) as the existing device. A comparison is provided in the below table:
| # | Predicate device (K222047) | Subject device | Comparison discussion |
| --- | --- | --- | --- |
| Intended use | Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g. | Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g. | Identical |
| E.g. Indications for use | - Metastatic tumors
- Recurrent glioblastomas
- Trigeminal neuralgia
- Medically refractory essential tremor
- Orbital tumors
- Ocular tumors
- Optic nerve tumors
- Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas)
- Skull base tumors
- Head and neck tumors (such as | - Metastatic tumors
- Recurrent glioblastomas
- Trigeminal neuralgia
- Medically refractory essential tremor
- Orbital tumors
- Ocular tumors
- Optic nerve tumors
- Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas)
- Skull base tumors
- Head and neck tumors (such as | Substantially equivalent.
The general purpose of the device and its functions is not affected.
The new indication (Refractory mesial temporal lobe epilepsy in adults) falls within the intended use of the predicate device (stereotactic irradiation of head structures).
For certain diseases that can result in epilepsy, such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations, radiation treatment is already on-label for both the pediatric and adult population.
Clinical performance of the Leksell Gamma Knife for all indications (including this new and additional indication) demonstrates a favorable benefit-risk profile. |
Elekta Solutions AB
LGK Traditional 510(k)
June 2025
{6}
Elekta
| | unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland)
- Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors) | unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland)
- Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors)
- Refractory mesial temporal lobe epilepsy in adults
- Refractory epilepsy associated with structural changes such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations (adult and pediatric) | |
| --- | --- | --- | --- |
# VII. Technological Comparison
Leksell Gamma Knife® (available models Elekta Esprit, Icon™, and Perfexion™) is a radiosurgery system for use in the stereotactic irradiation of intra-cranial structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
The major changes of technological characteristics of the Leksell Gamma Knife® (subject device), as compared to the predicate device Leksell Gamma Knife® (K222047) are primarily replacement of obsolete hardware components.
| Feature | Predicate Device (K222047) | Subject Device | Comparison Discussion |
| --- | --- | --- | --- |
| # of radiation sources | 192 | 192 | Identical |
| Operating System | Windows 10 | Windows 10 | Identical |
| LGK treatment control system | Control System SW 11.5 | Control System SW 11.6 | Version 11.6 enables the new hardware released with this version. The ECU hosting the control system is replaced with a new ECU with a modified design. The change in design required a new risk control to be added to ensure the same level of safety as predicate |
Elekta Solutions AB
LGK Traditional 510(k)
June 2025
{7}
Elekta
| | | | device and thus ensuring substantial equivalence. |
| --- | --- | --- | --- |
| User Interface | Model Esprit: Functional KeyPad (FKP) & Isolation and Patient Surveillance Unit (IPSU)
Model Icon and Perfexion :
Operator Control (OPC) | Model Esprit: Functional KeyPad (FKP) & Isolation and Patient Surveillance Unit (IPSU)
Model Icon and Perfexion :
Operator Control (OPC) | Identical |
| Patient Surveillance System (PSS) | Model Esprit: Digital PSS Camera
Model Icon and Perfexion:
Analog Video Camera | Model Esprit: Digital PSS Camera
Model Icon and Perfexion:
Analog Video Camera | Identical |
| Fixation | Model Esprit and Icon:
Frame (G-frame, Vantage), & and Mask
Model Perfexion:
Frame (G-frame) | Model Esprit and Icon:
Frame (G-frame, Vantage), & and Mask
Model Perfexion:
Frame (G-frame) | Identical |
| Imagin Modality | Model Esprit and Icon:
kV X-ray/CBCT
Model Perfexion:
na | Model Esprit and Icon:
kV X-ray/CBCT
Model Perfexion:
na | The old kV-generator used in the CBCT system is replaced with a new one. The clinical workflow, safety, functionality, and performance is unchanged thus ensuring substantial equivalence. |
| Patient Movement Detection System | Model Esprit and Icon:
High Definition Motion Management System (HDMM)
Model Perfexion:
na | Model Esprit and Icon:
High Definition Motion Management System (HDMM)
Model Perfexion:
na | The HDMM camera is replaced with a new one. The clinical workflow, safety, functionality, and performance is unchanged thus ensuring substantial equivalence. |
The fundamental technical characteristics of the subject device Leksell Gamma Knife® have not changed and are substantially equivalent to its predicate device Leksell Gamma Knife® (K222047)
# VIII. Summary of Performance Testing (Non-Clinical)
Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new control system and the new hardware against requirement specification.
Regression test of unchanged functionalities in the developed system was done to ensure that new and updated functionalities did not introduce any undesirable effects.
Performance testing, including design and usability validation of the system, has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and
Elekta Solutions AB
LGK Traditional 510(k)
June 2025
{8}
Elekta
user needs. The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) for the affected functionality, mainly the user interface of the Control Panel were effective.
In relation to the new indications for use, updated labeling was in scope of usability evaluation according to IEC 62366-1. The result demonstrate that the labeling support the safe and effective use of the device. Update of the ECU, kV-generator, HDMM-camera, or the new control system version did not require usability validation and previous data from the predicate device remain valid.
Verification testing has been done to the following standards to ensure safety of the device:
- IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-2-11, IEC 60601-2-28, IEC 60601-2-68
- IEC 62304
- IECO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-18, ISO 10993-23
- IEC 62366-1
Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification, user needs have been met and that the device functions as intended.
Biocompatibility testing has been performed to ensure that all materials are biocompatible in relation to the use of the device.
IX. Summary of Performance Testing (Clinical)
No animal or clinical tests were performed to establish substantial equivalence with the predicate device. The performance data demonstrate that the subject device is as safe and effective and performs as well as the predicate devices (K222047).
X. Substantial Equivalence Conclusion
Leksell Gamma Knife® is claimed to be substantially equivalent (SE) to the predicate device Leksell Gamma Knife® (K222047). The intended use is identical and the difference in indications are not critical to the intended use and the principle of operation remain unchanged.
New hardware has been introduced to replace old obsolete components. A new control system software version has been introduced to support and enable the new components. The subject device does not introduce any new clinical requirements or functions.
The technological characteristics are substantially equivalent to the predicate device; replacement of components do not affect the fundamental scientific technology or raise different questions of safety or effectiveness of the device. The device safety and performance have bene addressed by non-clinical testing in conformance with predetermined performance criteria, FDA guidance, and recognized consensus standards.
The result of verification and validation as well as conformance to relevant safety standards demonstrate that the Leksell Gamma Knife® meets the established safety and performance criteria and is substantially equivalent to the predicate device.
Elekta Solutions AB
LGK Traditional 510(k)
June 2025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.