The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.
Device Story
Depth Electrode system comprises slender, cylindrical polyurethane electrodes with platinum-iridium contacts, titanium alloy guidance screws, and connection cables. Used in stereoelectroencephalography (sEEG) procedures to monitor brain electrical signals or deliver stimulation. Electrodes are inserted into the brain via a 2.1 mm burr hole; guidance screws stabilize the electrode at the skull. Operated by neurosurgeons in a neurosurgical operating room. Output consists of electrical signals transmitted to external EEG monitoring/stimulation equipment. Healthcare providers use these signals to localize brain activity, aiding clinical decision-making for neurological conditions. Benefits include precise, temporary subsurface brain access for diagnostic or therapeutic monitoring.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993-5, -10, -11, -23) confirmed non-cytotoxic, non-irritant, non-sensitizing, and non-pyrogenic status. Additional testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), shelf life (ASTM F1980), transit testing (ASTM D4169), and sterilization validation (ISO 11135-1).
Technological Characteristics
0.8 mm diameter polyurethane electrode; platinum-iridium alloy contacts; 8-16 contacts; 26.5-63.0 mm exploration length. Titanium alloy guidance screws (20-35 mm). EO sterilization. Connectivity via connection cables to external EEG monitoring equipment. No active software/algorithm components.
Indications for Use
Indicated for temporary (less than 30 days) recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
November 5, 2025
Beijing Sinovation Medical Technology Co., Ltd.
Lina Wang
RA Specialists
Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road
Fengtai District
Beijing, 100070
China
Re: K250363
Trade/Device Name: Depth Electrode (RSDE-08); Depth Electrode (RSDE-10); Depth Electrode (RSDE-12); Depth Electrode (RSDE-14); Depth Electrode (RSDE-16); Depth Electrode (RSDE-16F); Depth Electrode (RSDE-08S); Depth Electrode (RSDE-10S); Depth Electrode (RSDE-12S); Depth Electrode (RSDE-14S); Depth Electrode (RSDE-16S); Depth Electrode (RSDE-16G); Depth Electrode (RSDE-18); Depth Electrode (RSDE-18F); Depth Electrode (RSDE-18T); Depth Electrode (RSDE-20); Depth Electrode (RSDE-20F); Depth Electrode (RSDE-20T); Guidance Screw (DX-20); Guidance Screw (DX-25); Guidance Screw (DX-30); Guidance Screw (DX-35); Connection Cable (LX-12); Connection Cable (LX-16); Connection Cable (LX-20); Connection Cable (LX-S); Connection Cable (LX-RF2)
Regulation Number: 21 CFR 882.1330
Regulation Name: Depth Electrode
Regulatory Class: Class II
Product Code: GZL
Dated: December 3, 2024
Received: February 10, 2025
Dear Lina Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database
{1}
K250363 - Lina Wang
Page 2
available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K250363 - Lina Wang
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak -S
for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K250363 | ? |
| Please provide the device trade name(s). | | ? |
| Depth Electrode (RSDE-08);
Depth Electrode (RSDE-10);
Depth Electrode (RSDE-12);
Depth Electrode (RSDE-14);
Depth Electrode (RSDE-16);
Depth Electrode (RSDE-16F);
Depth Electrode (RSDE-08S);
Depth Electrode (RSDE-10S);
Depth Electrode (RSDE-12S);
Depth Electrode (RSDE-14S);
Depth Electrode (RSDE-16S);
Depth Electrode (RSDE-16G);
Guidance Screw (DX-20);
Guidance Screw (DX-25);
Guidance Screw (DX-30);
Guidance Screw (DX-35);
Connection Cable (LX-12);
Connection Cable (LX-16);
Connection Cable (LX-20);
Connection Cable (LX-S);
Connection Cable (LX-RF2) | | |
| Please provide your Indications for Use below. | | ? |
| The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and
stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface
level of the brain. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
# 510(k) Summary
## I. Submitter
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China; 100070
Tel: +86 10 80481789
Contact person: Lina Wang
Date Prepared: Oct 26, 2025
## II. Device
Name of Device: Depth Electrode
Common or User name: Depth Electrode
Classification Name: Depth Electrode(21 CFR 882.1330)
Regulatory Class: II
Product Code: GZL
## III. Predicate device
Predicate Device: Evo@sEEG System
Predicate 510(K) number: K222404
Manufacture: NeuroOne Medical Technologies Corp.
The 510(k) summary of the predicate device is provided as K222404.pdf
This predicate has not been subject to a design-related recall.
Predicate Device: DIXI Medical Microdeep Depth Electrode
Predicate 510(K) number: K170959
Manufacture: DIXI Medical.
The 510(k) summary of the predicate device is provided as K170959.pdf
Predicate Device: Depth Electrodes
Predicate 510(K) number: K163355
Manufacture: Ad-tech Medical Instrument Corporation
The 510(k) summary of the predicate device is provided as K163355.pdf
No reference devices were used in this submission.
## IV. Description of Device
The Depth Electrode consists of the Depth Electrode (electrode) and accessories. The accessories consists of Connection cable and Guidance screw. The Depth Electrodes are provided sterile after EO sterilization and are single use only.
The electrodes are slender, cylindrical tubes inserted into the brain. The electrodes have conductive contacts at the distal end to monitor the brain's electrical signals or deliver electrical stimulation.
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{5}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
The electrode is a 0.8 mm diameter Polyurethane electrode with platinum-iridium alloy contacts and are available in varying numbers of contacts: 8 to 16 contacts, with an exploration length (recording depth) from 26.5 to 63.0 mm. The longitudinal width of each contact is 2.0 mm, and are spaced 1.5 mm apart (distance between contact groups is 10 mm).
The Guidance screws are secured in the skull to provide an access point while stabilizing the electrode. It consists of a tube through which the electrode can pass. The proximal end is threaded, allowing the surgeon to screw it into the patient's skull via an appropriately sized burr hole. The distal end is threaded to accept the Sealing Cap of electrodes.
Guidance screws are available in 20 mm, 25 mm, 30 mm, and 35 mm lengths, and can only be used by inserting through a small 2.1 mm burr hole drilled in the skull. Guidance screws should be used only when depth electrodes are warranted.
The Connection cables are used for the transmission of electrical signals that connects to EEG monitoring equipment. When performing EEG monitoring, the connection cable of each model shall be used in conjunction with the electrode of the corresponding channel.
The Depth Electrode is temporarily placed (less than 30 days) at the subsurface level of the brain. The system is designed to be used in stereoelectroencephalography (sEEG) procedures.
## V. Indication for use
The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.
The indications of Use statement for the Depth Electrode is identical to the predicate device.
## VI. Comparison summary
The following table compares the Subject Device to the predicate device with respect to indications for use, and technological characteristics, forming the basis for the determination of substantial equivalence. The subject device does not raise any new question of safety or effectiveness as compared to the predicate device.
| Device Classification Comparison | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Subject device | Predicate Device K222404 | Predicate Device K170959 | Predicate Device K163355 | Comparison |
| Trade name | Depth Electrode | Evo@sEEG System | DIXI Medical Microdeep Depth Electrode | Depth Electrodes | N/A |
| Manufacturer | BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD. | NeuroOne Medical Technologies Corp. | DIXI Medical | Ad-tech Medical Instrument Corporation | N/A |
| Device | 882.1330 | 882.1330 | 882.1330 | 882.1330 | Identical |
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{6}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
| Regulation Number: | | | | | |
| --- | --- | --- | --- | --- | --- |
| Device/Regulation Name: | Electrode, Depth | Electrode, Depth | Electrode, Depth | Electrode, Depth | Identical |
| Regulation Description/Common name: | Depth electrode | Depth electrode | Electrode, Depth | Electrode, Depth | Identical |
| Product Code: | GZL | GZL | GZL | GZL | Identical |
| Device Class/Regulation Classification: | Class II | Class II | Class II | Class II | Identical |
| Intended Use/Indication for Use Comparison | | | | | |
| Intended Use: | The device is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain. | The device is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain. | A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain. | The device are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain | Identical |
| Indication for Use | The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. | The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. | The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. | The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. | Identical |
| Electrode Technological and Performance Characteristics Comparison | | | | | |
| Electrode Point | 0.8 mm | 0.8 mm | 0.8 mm | 0.86-1.96 mm | Identical |
公司地址:北京市丰台区航丰路1号院1号楼1室2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{7}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
| Diameter | | | | | |
| --- | --- | --- | --- | --- | --- |
| Electrode Material | Polyurethane | Polyamide | / | / | Different |
| Number of Contacts | 8-16 | 5-16 | 5-18 | Up to 16 | Minor different |
| Contact Height/Length | 2 mm | 2 mm | 2 mm | / | Same |
| Contact spacing | 1.5mm | 1.5-3.2 mm | 1.5-11 mm | / | Minor different |
| Contact Material | platinum-iridium | Platinum | platinum-iridium | Platinum | Identical |
| Exploration Length (Recording Depth) | 26.5-63 mm | 16-80 mm | 16-80.5 mm | | Minor different |
| Sterile | EO sterilization | EO sterilization | EO sterilization | EO sterilization | Same |
| Single-Use | yes | yes | yes | Yes | Same |
| Environment of Use | Intraoperative and Neurological monitoring locations at subsurface levels of the brain | Intraoperative and Neurological monitoring locations at subsurface levels of the brain | Intraoperative and Neurological monitoring locations | Intraoperative and Neurological | Identical |
| Duration of Use | Less than 30 days | Less than 30 days | Less than 30 days | Less than 30 days | Same |
| Maximum Stimulation Charge Density | ≤ 30 μC/cm² | / | ≤ 30 μC/cm² | ≤ 30 μC/cm² | Identical |
| Where used | Neurosurgical operating room | Neurosurgical operating room | / | / | Identical |
| Principles of Operation | Depth electrode are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signals | Depth electrode are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signals | Depth electrodes are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signals | / | Identical |
| Guidance Screw Technological and Performance Characteristics Comparison | | | | | |
| Depth Electrode Diameter | 0.8 mm | 0.8 mm | 0.8 mm | / | Same |
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{8}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
| Compatibility | | | | | |
| --- | --- | --- | --- | --- | --- |
| Material(s) | Titanium alloys | Titanium | Titanium alloys | / | Same as K170959 |
| Lengths Offered | 20-35 mm | 20-35 mm | 15-35 mm | / | Same as K222404 |
| Guidance Screw ID | 0.9 mm | 0.9 mm | 0.8 mm | / | Same as K222404 |
| Single-Use | yes | yes | yes | / | same |
| Sterile | EO sterilization | EO sterilization | EO sterilization | / | same |
| Environment of Use | Intraoperative and Neurological monitoring locations | Intraoperative and Neurological monitoring locations | Intraoperative and Neurological monitoring locations | / | same |
| Duration of Use | Less than 30 days | Less than 30 days | Less than 30 days | / | same |
| Principles of Operation | Threaded into a pre-drilled hole in the skull | Threaded into a pre-drilled hole in the skull | Threaded into a pre-drilled hole in the skull | / | Identical |
# VII. Performance Data
The Depth Electrode has the identical intended use and similar technological characteristics as the predicate, the minor different technological characteristics are Electrode Material, Number of Contacts, Contact Material, Recording Depth of Depth Electrode and Material(s) of Guidance Screw.
Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device.
The following performance data were provided in support of substantial equivalence.
| Test/Standard | Test Method Summary | | Result |
| --- | --- | --- | --- |
| Biocompatibility Evaluation
aDepth Electrode (Tissue/bone, brain and cerebrospinal fluid contact, >24h to 30 days)
bSliding Seal Tube and Sealing Cap (Indirect Tissue/bone contact, >24h to 30 days)
cGuidance Screw (Tissue/bone and cerebrospinal fluid contact, >24h to 30 days) | | | |
| a,b,cCytotoxicity ISO 10993-5 | MEM Elution using L-929 Cells (GLP) | Pass - Non-cytotoxic | |
| a,b,cIrritation ISO 10993-23 | Intracutaneous Irritation test (GLP) | Pass - Non-irritant | |
| a,b,cSkin sensitization ISO 10993-10 | Guinea pig maximization sensitization test (GLP) | Pass - Non-sensitizing | |
| a,b,cPyrogenicity USP<151> ISO | Rabbit material-mediated | Pass - Non-Pyrogenic | |
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{9}
Sinovation®
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
| 10993-11 | pyrogenicity test (GLP) | |
| --- | --- | --- |
| | Acute systemic toxicity effects in mice (GLP) | Pass - Non-toxic |
| | Rat IV and IP injection study (GLP)
Chemical characterization study with toxicological risk assessment per ISO 10993-17 and ISO 10993-18 | Pass - Non-toxic |
| | 1 week and 30 day Subcutaneous test in Rabbits (GLP) | Pass - Minimal or no reaction |
| | 1 week and 30 day bone implantation in Rabbits (GLP) | Pass - Minimal or no reaction |
| | Brain Implantation Study - 1 week and 30 day Brain implantation in Rats (GLP)
Paper-based risk assessment of materials and manufacturing
In vitro Cytotoxicity Test using neuronal cell model - The activity of differentiated PC-12 Cells and morphology of undifferentiated PC-12 Cells | Pass
Acceptable
Pass |
| | Hemolytic test using rabbit blood (GLP) | Pass - Non-hemolytic |
| | Bacterial reverse mutation – AMES Assay (GLP) | Pass - Non-mutagenic |
| | In Vitro mouse lymphoma assay (GLP) | Pass - non-mutagenic, non-clastogenic |
| Other types of test /verification are as follows: | | |
| Test | Standards/summary | Result |
| bench test | Verify the mechanical, electrical and functional of the device | Pass |
| Electrical safety | IEC 60601-1 | Pass |
| EMC | IEC 60601-1-2 | Pass |
| shelf life | ASTM F1980 | Pass |
| transit testing | ASTM D4169 | Pass |
| EO Sterilization validation | ISO 11135-1 | Pass |
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
{10}
Sinovation
华科恒生
北京华科恒生医疗科技有限公司
BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
| EO, ECH residue | ISO 10993-7 | Pass |
| --- | --- | --- |
| Bacterial Endotoxin validation | usp<161> | Pass |
| Cleaning validation | ISO 19227 | Pass |
## VIII. Conclusion
The Depth Electrode has the identical intended use and similar technological characteristics as the predicate device, the minor different technological characteristics are Electrode Material, Number of Contacts, Contact Material, Exploration Length of Depth Electrode and the Material(s) of Guidance Screw. Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device. The Depth Electrode has undergone bench test, electrical, biocompatibility, packaging and sterilization testing to demonstrate the differences in the sizes and materials do not raise questions of safety or effectiveness.
公司地址:北京市丰台区航丰路1号院1号楼1至2层101内2层218室
Address: Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China
Tel:+86 10 80481789 Email: sales@hkhs.cn www.sinovationmed.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.