TENS AND EMS (TENS and Muscle Stimulator) (JT8012C,JT8016,JT9020E,JT9026)

K250360 · Shenzhen Jiantuo Electronics Co., Ltd. · NUH · May 11, 2025 · Neurology

Device Facts

Record IDK250360
Device NameTENS AND EMS (TENS and Muscle Stimulator) (JT8012C,JT8016,JT9020E,JT9026)
ApplicantShenzhen Jiantuo Electronics Co., Ltd.
Product CodeNUH · Neurology
Decision DateMay 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

TENS (Transcutaneous Electric Nerve Stimulation): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, abdomen, bottom, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. EMS(Electrical Muscle Stimulation)/PMS (Powered Muscle Stimulation): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Device Story

TENS and EMS device (models JT8012C, JT8016, JT9020E, JT9026) provides transcutaneous electrical nerve stimulation for pain relief and electrical muscle stimulation for muscle performance. Device delivers electrical pulses to user via electrodes applied to skin. Operated by patient in home setting. Output affects sensory nerves to modulate pain perception and motor nerves to induce muscle contraction. Benefits include temporary pain relief and improved muscle performance.

Technological Characteristics

Transcutaneous electrical nerve stimulator; electrical pulse output; OTC use; multi-model series (JT8012C, JT8016, JT9020E, JT9026).

Indications for Use

Indicated for temporary relief of pain in shoulder, waist, back, neck, abdomen, bottom, arms, and legs due to exercise or household strain; also indicated for stimulation of healthy muscles to improve/facilitate muscle performance. Intended for OTC use.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION Build Correspondence Convert to PDF May 11, 2025 Shenzhen Jiantuo Electronics Co., Ltd. % Doris Dong Senior Regulatory Affairs Project Manager Shanghai CV Technology Co., Ltd. Room 805, No. 19 Dongbao Road, Songjiang Area, Shanghai Shanghai, 201613 China Re: K250360 Trade/Device Name: TENS AND EMS (TENS and Muscle Stimulator) (JT8012C, JT8016, JT9020E, JT9026) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: January 27, 2025 Received: February 10, 2025 Dear Doris Dong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250360 - Doris Dong Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K250360 - Doris Dong Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **PAMELA D. SCOTT -S** Digitally signed by PAMELA D. SCOTT -S Date: 2025.05.11 20:34:56 -04'00" Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250360 Device Name TENS AND EMS (TENS and Muscle Stimulator) (JT8012C,JT8016,JT9020E,JT9026) Indications for Use (Describe) TENS (Transcutaneous Electric Nerve Stimulation): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, abdomen, bottom, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. EMS(Electrical Muscle Stimulation)/PMS (Powered Muscle Stimulation): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...