The Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI) are intended to treat infants to the elderly that suffer from a traumatic brain injury such as trauma, stroke, disease, tumor, infection, etc. whereby the patient's original bone will not be used.
Device Story
Patient-specific implantable prosthetic plates; fabricated from PEEK thermoplastic polymer with biphasic calcium phosphate (BCP PEEK). Input: patient CT scan imaging data. Transformation: custom sizing/shaping based on CT data to match patient anatomy; non-load bearing design. Output: patient-specific cranial/craniofacial implant. Used in surgical settings; operated by surgeons. Attached to native bone via standard cranioplasty hardware. Benefits: fills bony voids/defects; BCP PEEK formulation facilitates enhanced bone ingrowth. Provided non-sterile; requires steam sterilization by end-user.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993 series: genotoxicity, cytotoxicity, sensitization, irritation, systemic toxicity), bacterial endotoxin testing, steam sterilization validation (ANSI/AAMI ST79), and mechanical testing validation using a worst-case scenario.
Technological Characteristics
Implant grade PEEK formulated with biphasic calcium phosphate (BCP PEEK). Dimensions: 2-10mm thick, 25-250mm wide/long. Patient-specific geometry derived from CT scan data. Non-load bearing. Provided non-sterile; steam sterilizable. No integrated electronics or software.
Indications for Use
Indicated for patients from infants to elderly suffering from traumatic brain injury (trauma, stroke, disease, tumor, infection) requiring filling of bony voids or defects in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone) where original bone is unavailable.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 25, 2025
Kelyniam Global Inc.
Elise Bozzuto
Quality Director
97 River Road, Suite A
Canton, Connecticut 06019
Re: K250334
Trade/Device Name: Fusion Craniofacial Implant; Fusion Skull Implant
Regulation Number: 21 CFR 882.5320
Regulation Name: Preformed Alterable Cranioplasty Plate
Regulatory Class: Class II
Product Code: GWO
Dated: February 5, 2025
Received: February 5, 2025
Dear Elise Bozzuto:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250334 - Elise Bozzuto
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250334 - Elise Bozzuto
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by
Adam D. Pierce -S
Date: 2025.07.25
15:32:38 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250334 | |
| Device Name Fusion Craniofacial Implant (FCI); Fusion Skull Implant (FSI) | |
| Indications for Use (Describe) Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI) are intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250334, Page 1 of 4
510(k) Summary
Date Prepared: 06/23/2025
Submitter Information:
Submitter:
Kelyniam Global Inc. (KGI)
97 River Road, Suite A
Canton, Connecticut 06019
Tel: (800) 280-8192
Fax: (501) 641-2000
Official Correspondent:
Elise Bozzuto
Quality Director
Tel: (800) 280-8192
Email: ebozzuto@kelyniam.com
Device Information:
Trade Name: Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI)
Common Name: Fusion Cranial/Skull Implant or Fusion Craniofacial Implant
Classification Information: Panel: Neurology / Product Code: GWO / Classification: Class II in accordance with 21 CFR 882.5320, Preformed Alterable Cranioplasty Plate.
Predicate Devices:
- Primary Predicates - K103582 Kelyniam Global Inc. (KGI) Customized Skull Implant (CSI); K121755 Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI); K182711 Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI).
Device Description: Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI) are individually sized and shaped implantable prosthetic plates intended to fill a bony void or defect area in a specific patient's cranial and craniofacial skeleton, ranging from 2 mm to 10 mm thick (typically 4mm thick based on the patient CT Scan imaging data) x 25mm to 250 mm wide x 25 mm to 250 mm long.
A Fusion Skull Implant (FSI) is intended to fill a bony void or defect area in a specific patient's skull, whereas a Fusion Craniofacial Implant (FCI) is intended to fill a bony void or defect area in a specific patient's facial region of the skull, excluding the Maxilla (upper jaw area surrounding the teeth only) and Mandible, both considered load bearing areas of the facial region of the skull.
The size, asymmetrical shape, thickness, contour, and edge profile are design elements of the
{5}
K250334, Page 2 of 4
non-load bearing patient-specific base implant. These design elements are used to support the base implant in the bony void or defect area to provide a "Precise Fit."
The single-use alterable base implant (1) is fabricated from a billet block of implant grade Polyether ether ketone (PEEK) Thermoplastic Polymer formulated with biphasic calcium phosphate (BCP PEEK), using a patient's CT scan imaging data, (2) is provided clean (non-sterile) for steam sterilization prior to implantation at a hospital or surgical site, and (3) are attached to the native bone using commercially available cranioplasty hardware and fasteners.
**Indications for Use:** The Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI) are intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone).
**Intended Use:** The Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI) are intended to treat infants to the elderly that suffer from a traumatic brain injury such as trauma, stroke, disease, tumor, infection, etc. whereby the patient's original bone will not be used.
**Statement of Substantial Equivalence:** The Kelyniam subject devices are intended to fill a bony void or defect area in a patient's specific Cranial and Craniofacial skeleton and are substantially equivalent in indications for use, technology, and material to:
K103582, K121755, and K182711 Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI) (Primary).
**Comparison of Technological Characteristics with the Predicate Device(s):** The Kelyniam subject devices are manufactured from implant grade Polyether ether ketone Thermoplastic Polymer material formulated with biphasic calcium phosphate, known hereafter as BCP PEEK, sold non-sterile, and intended to fill a bony void or defect area in a patient's specific Cranial and Craniofacial skeleton. This material is subject to the same design phase and manufacturing process as the Kelyniam cleared devices. This material is intended to facilitate enhanced bone ingrowth following implantation, while not introducing any new technological issues. The Kelyniam subject device is identical to the cleared primary predicate devices in every way except for the proposed base material change and proposed labeling changes.
See Device Comparison Chart below for details.
| | Subject Device | Primary Predicates | | |
| --- | --- | --- | --- | --- |
| Manufacturer | Kelyniam Global Inc. (KGI) | Kelyniam Global Inc. (KGI) | | |
| 510(k) # | | K103582 | K121755 | K182711 |
| Description | Fusion Craniofacial Implant (FCI), Fusion Skull Implant (FSI) | Customized Skull Implant (CSI) | Customized Craniofacial Implant (CCI) | Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI) |
| Intended Use | Is intended to fill a bony void or defect area in a patient's specific cranial/craniofacial skeleton | Is intended to fill a bony void or defect area in a patient's specific cranial/craniofacial skeleton | | |
| Material | Evonik Vestakeep iC4800R® | Evonik Vestakeep i4® or Invibio Inc PEEK-Optima LT-1® | | |
{6}
K250334, Page 3 of 4
| Technical Specification | Plate: Custom sized to each patient using CT scan data | Plate: Custom sized to each patient using CT scan data |
| --- | --- | --- |
| Sterilization | Provided Non-sterile | Provided Non-sterile |
| Product Code | GWO (882.5320, Preformed Alterable Cranioplasty Plate) | GWO (882.5320, Preformed Alterable Cranioplasty Plate) |
| Classification | Class II | Class II |
# Performance Data:
Biocompatibility Testing: Evonik Industries has conducted the following biocompatibility testing on Vestakeep iC4800R® Polyether ether ketone (PEEK) material formulated with biphasic calcium phosphate, BCP PEEK. Results are included with this submission.
ISO 10993-3 Genotoxicity: Ames Test, ISO 10993-3 Genotoxicity: Mouse Lymphoma test, ISO 10993-5 Cytotoxicity, ISO 10993-6 Test for local effects after Implantation in bone (4 weeks), ISO 10993-6 Test for local effects after Implantation in bone (12 weeks), ISO 10993-6 Test for local effects after Implantation in bone (26 weeks), ISO 10993-10 Sensitization: Maximization test according to Magnusson and Kligman, ISO 10993-10 Irritation: Intracutaneous Reactivity, ISO 10993-11 Acute Systemic Toxicity, ISO 10993-11 Subchronic Systemic Toxicity, ISO 10993-12 GC/MS Fingerprint, Material-Mediated Pyrogens, USP Class VI Acute Systemic Toxicity, Intracutaneous Reactivity, Muscle Implantation.
Additional cytotoxicity testing and hemocompatibility testing were conducted separately on the subject devices.
Performance Testing: Evonik Vestakeep iC4800R® PEEK (BCP PEEK) Certification of Analysis/Certification of Compliance are reviewed and accepted when received against approved specifications, as part of receiving inspection activities.
Cleaning Validation: Testing was conducted for the submission of the Kelyniam cleared devices. The cleaning process of the subject devices are identical to that of the Kelyniam cleared devices. During the cleaning process, isopropyl alcohol is used as an antiseptic wipe. An accelerated ageing study performed by Evonik (raw material manufacturer) shows that no substances were reported from the gas-chromatography prepared from the extract using isopropanol, indicating that there are no deleterious effects from the use of isopropyl alcohol. The cytotoxicity tests performed on the materials confirm that no changes in the chemical composition of the materials could be detected. Neither the unaged material nor the sterilized and/or aged material showed signs of cytotoxicity.
Bacterial endotoxin testing was performed for the subject device. The testing determined endotoxins detected on finished devices were well below the maximum acceptable limits.
Steam Sterilization Validation: Testing was conducted in accordance with the following: ANSI/AAMI ST79. The finished product is shipped non-sterile, to be sterilized by the end-user. Sterilization recommendations are provided within the Instructions For Use (IFU).
{7}
K250334, Page 4 of 4
Mechanical Testing Validation: There is no industry accepted standard governing mechanical testing for non-load bearing implantable prosthetic plates. A Mechanical Testing Validation Protocol was developed by Kelyniam to validate the subject devices.
Ship Testing Validation: Testing was conducted for the submission of the Kelyniam cleared devices. The packaging and shipping process of the subject devices is identical to that of the Kelyniam predicate devices.
**Summary on Non-Clinical Testing:** The testing performed on the subject devices demonstrates substantial equivalence to the predicate devices. The subject devices performed equivalent to the predicate devices during mechanical testing using a worst-case scenario.
**Summary on Clinical Testing:** Clinical Testing was determined not applicable for Kelyniam Global Inc. Fusion Craniofacial Implant (FCI) and Fusion Skull Implant (FSI).
**Conclusion:** The biocompatibility testing, steam sterilization testing, and mechanical testing of the subject devices were performed. The results of these tests, as well as additional product related validations (all samples passed the acceptance criteria) support the safety and effectiveness of the device and are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.