K250306 · Propper Manufacturing Co., Inc. · FRG · Oct 31, 2025 · General Hospital
Device Facts
Record ID
K250306
Device Name
Chex-All Sterilization Pouches and Tubes
Applicant
Propper Manufacturing Co., Inc.
Product Code
FRG · General Hospital
Decision Date
Oct 31, 2025
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using: Hospital size sterilizer cycles - Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time - Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time - Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time Tabletop sterilizer cycles: - Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time Ethylene Oxide sterilization cycles: - Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following: Metal lumen: 7 inch long, 3.2mm internal diameter Plastic lumen: 5 inch long, 2.0 mm internal diameter. Suitable Lumen cycles: Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time). Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time). Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time). The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide. If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL≥10⁻⁶). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization. The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13"x18".
Device Story
Sterilization pouches/tubes composed of porous paper and plastic film; used by healthcare providers to enclose medical devices for steam or EO sterilization. Input: medical devices placed inside pouch; device provides physical barrier to maintain sterility (SAL 10^-6) post-sterilization. Includes integrated chemical process indicators (crème-to-gray/black for steam; beige-pink-to-green for EO) to distinguish processed from unprocessed items. Used in hospital and tabletop sterilizers; supports specific lumen dimensions and double-pouch configurations. Output: sterilized, protected medical device. Benefits: ensures sterility maintenance for 30 months (steam) or 12 months (EO) post-processing, facilitating safe clinical reuse.
Clinical Evidence
Bench testing only. Validated per ISO 11607-1, ISO 11140-1, and ANSI/AAMI/ISO 17665-1/14937. Testing included tensile/burst strength, microbial barrier (DIN 58953-6), seal integrity (ASTM F1929/F1886), peel-open characteristics, and sterilization efficacy (spore growth assays) for hospital/tabletop steam and EO cycles. Shelf-life and sterility maintenance validated via real-time exposure (30 months steam/12 months EO).
Technological Characteristics
Materials: porous paper, PET/CPP clear compound film, high-temp adhesive, medical double-sided tape. Sensing: chemical process indicators (ink-based). Dimensions: 3"-18" width, various lengths/rolls. Connectivity: N/A. Sterilization: steam/EO compatible. Standards: ISO 11607-1, ISO 1924-2, ISO 3781, ISO 2758, ISO 3689, ISO 9197/9198, ASTM D882, ASTM F1929, ASTM F88, ASTM F1886, DIN 58953-6.
Indications for Use
Indicated for healthcare providers to enclose medical devices for sterilization via gravity steam, pre-vacuum steam (hospital/tabletop), or ethylene oxide. Supports single or double pouch configurations and specific metal/plastic lumen devices. Not restricted by age/gender.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Safe Secure Packaging Co. Sterilization Pouch with Steam and EO Process Indicators (K153540)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 31, 2025
Propper Manufacturing Co, Inc.
Andrew Sharavara
Chief Technical Officer
36-04 Skillman Avenue
Long Island City, New York 11101
Re: K250306
Trade/Device Name: Chex-All® Sterilization Pouches and Tubes
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: FRG, JOJ
Dated: January 31, 2025
Received: February 3, 2025
Dear Andrew Sharavara:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250306 - Andrew Sharavara
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250306 - Andrew Sharavara
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S
Date: 2025.10.31
16:42:53 -04'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K250306
Device Name
Chex-All® Sterilization pouches and tubes
Indications for Use (Describe)
Indications for use:
Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
Hospital size sterilizer cycles
- Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
- Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
Tabletop sterilizer cycles:
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time
Ethylene Oxide sterilization cycles:
- Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following:
Metal lumen: 7 inch long, 3.2mm internal diameter
Plastic lumen: 5 inch long, 2.0 mm internal diameter.
Suitable Lumen cycles:
Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time).
Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time).
Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time).
The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide.
If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL≥10⁻⁶). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization.
The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13"x18".
FORM FDA 3881 (6/20)
Page 1 of 3
PSC Publishing Services (301) 443-6740
{4}
The following tables list the available model numbers of the Chex-All sterilization pouches and tubes.
| Cat. Numbers | External Dimensions | | Max load weight |
| --- | --- | --- | --- |
| Chex-All Heat-seal sterilization pouch, flat | Width | Length | lb |
| 02500800 | 3” | 8” | 0.99 |
| 02501000 | 5” | 10” | 0.99 |
| 02501200 | 7” | 12” | 1.10 |
| 02501400 | 8” | 16” | 1.10 |
| 02501800 | 12” | 18” | 2.65 |
| 02502400 | 18” | 24” | 2.65 |
| Chex-All Heat-seal sterilization tube, flat, sold in rolls | | | |
| 02600200 | 2” | 100’ | 0.99 |
| 02600300 | 3” | 100’ | 0.99 |
| 02600400 | 4” | 100’ | 0.99 |
| 02600600 | 6” | 100’ | 1.10 |
| 02600900 | 9” | 100’ | 1.10 |
| 02601200 | 12” | 100’ | 2.65 |
| Chex-All II Self-seal sterilization pouch, flat | | | |
| 02400800 | 3” | 8” | 0.99 |
| 02401000 | 5” | 10” | 0.99 |
| 02401200 | 7” | 12” | 1.10 |
| 02401400 | 8” | 16” | 1.10 |
| 02401500 | 5” | 15” | 0.99 |
| 02401800 | 12” | 18” | 2.65 |
| 02402800 | 3.5” | 24” | 0.99 |
| Chex-All III Self-seal sterilization pouch, flat | | | |
| 02430900 | 3.5” | 9.5” | 0.99 |
| 02431000 | 5.25” | 10” | 0.99 |
| 02431300 | 7.5” | 12” | 1.10 |
| 02431400 | 9” | 16” | 1.10 |
| 02431800 | 13” | 18” | 2.65 |
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (6/20)
Page 2 of 3
{5}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 3 of 3
{6}
510(k) Summary (K250306)
Submitted by: Propper Manufacturing Company, Inc.
Address: 36-04 Skillman Avenue, Long Island City, New York 11101
Contact Name: Andrew Sharavara, Ph.D., Chief Technical Officer
Telephone: (800) 832-4300 x149
Fax: (718) 482-8909
E-mail: as@proppermfg.com
Date Submitted: October 31, 2025
**Device information:**
Device Trade Name: Chex-All® Sterilization Pouches and Tubes
Classification Name: Sterilization wrap
Common Name: Sterilization pouch
Product Code: FRG, JOJ
Classification: Class II (21 C.F.R. 880.6850, 21 C.F.R. 880.2800)
Predicate Device: K153540
**Description of the Device**
The Chex-All® Sterilization pouches and tubes are made from a medical grade porous paper thermally sealed to a plastic film on the left, right, and bottom of the pouch. The top side is open to insert medical devices to be sterilized and to seal the pouch. They are available either pre-manufactured self-seal or heat-seal sizes or in a roll of heat-seal tubing for the customer to cut the pouches to their required size.
The pouches also contain chemical process indicators used to demonstrate that the pouches have been a subject of either a steam or ethylene oxide sterilization process.
All Chex-All sterilization pouches and tubes are flat and include the following products:
| Category name | Type | Steam and EO Indicator | Cat. Numbers | External Dimensions | | Max load, lb |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | Width | Length | |
| Chex-All pouches | Heat-seal sterilization pouch, flat | External, Internal | 02500800 | 3” | 8” | 0.99 |
| | | | 02501000 | 5” | 10” | 0.99 |
| | | | 02501200 | 7” | 12” | 1.10 |
| | | | 02501400 | 8” | 16” | 1.10 |
| | | | 02501800 | 12” | 18” | 2.65 |
| | | | 02502400 | 18” | 24” | 2.65 |
| Chex-All tubes | Heat-seal sterilization | External, Internal | 02600200 | 2” | 100’ | 0.99 |
| | | | 02600300 | 3” | 100’ | 0.99 |
Page 1 (K250306)
{7}
Page 2 (K250306)
| | tube, flat, sold in rolls | | 02600400 | 4” | 100’ | 0.99 |
| --- | --- | --- | --- | --- | --- | --- |
| | | | 02600600 | 6” | 100’ | 1.10 |
| | | | 02600900 | 9” | 100’ | 1.10 |
| | | | 02601200 | 12” | 100’ | 2.65 |
| Chex-All II pouches | Self-seal sterilization pouch, flat | External, Internal | 02400800 | 3” | 8” | 0.99 |
| | | | 02401000 | 5” | 10” | 0.99 |
| | | | 02401200 | 7” | 12” | 1.10 |
| | | | 02401400 | 8” | 16” | 1.10 |
| | | | 02401500 | 5” | 15” | 0.99 |
| | | | 02401800 | 12” | 18” | 2.65 |
| | | | 02402800 | 3.5” | 24” | 0.99 |
| Chex-All III pouches | Self-seal sterilization pouch, flat | External | 02430900 | 3.5” | 9.5” | 0.99 |
| | | | 02431000 | 5.25” | 10” | 0.99 |
| | | | 02431300 | 7.5” | 13” | 1.10 |
| | | | 02431400 | 9” | 16” | 1.10 |
| | | | 02431800 | 13” | 18” | 2.65 |
## Indications for Use
Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
### Hospital size sterilizer cycles
- Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
- Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
### Tabletop sterilizer cycles:
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time
### Ethylene Oxide sterilization cycles:
- Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following:
- Metal lumen: 7 inch long, 3.2mm internal diameter
- Plastic lumen: 5 inch long, 2.0 mm internal diameter.
### Suitable Lumen cycles:
- Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time).
- Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time).
{8}
Page 3 (K250306)
Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time).
The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide.
If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL ≥ 10⁻⁶). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization.
The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13"x18".
The following tables list the available model numbers of the Chex-All sterilization pouches and tubes.
| Category name and type | Cat. Numbers | External Dimensions | | Max load, lb |
| --- | --- | --- | --- | --- |
| | | Width | Length | |
| Chex-All Heat-seal sterilization pouch, flat | 02500800 | 3" | 8" | 0.99 |
| | 02501000 | 5" | 10" | 0.99 |
| | 02501200 | 7" | 12" | 1.10 |
| | 02501400 | 8" | 16" | 1.10 |
| | 02501800 | 12" | 18" | 2.65 |
| | 02502400 | 18" | 24" | 2.65 |
| Chex-All Heat-seal sterilization tube, flat, sold in rolls | 02600200 | 2" | 100' | 0.99 |
| | 02600300 | 3" | 100' | 0.99 |
| | 02600400 | 4" | 100' | 0.99 |
| | 02600600 | 6" | 100' | 1.10 |
| | 02600900 | 9" | 100' | 1.10 |
| | 02601200 | 12" | 100' | 2.65 |
| Chex-All II Self-seal sterilization pouch, flat | 02400800 | 3" | 8" | 0.99 |
| | 02401000 | 5" | 10" | 0.99 |
| | 02401200 | 7" | 12" | 1.10 |
| | 02401400 | 8" | 16" | 1.10 |
| | 02401500 | 5" | 15" | 0.99 |
| | 02401800 | 12" | 18" | 2.65 |
| | 02402800 | 3.5" | 24" | 0.99 |
| Chex-All III Self-seal sterilization pouch, flat | 02430900 | 3.5" | 9.5" | 0.99 |
| | 02431000 | 5.25" | 10" | 0.99 |
| | 02431300 | 7.5" | 13" | 1.10 |
| | 02431400 | 9" | 16" | 1.10 |
| | 02431800 | 13" | 18" | 2.65 |
{9}
Summary of Device Technological Characteristics with Comparison to the Predicate Device
| | Predicate Device, K153540 | Subject Device, K250306 | Comparison |
| --- | --- | --- | --- |
| Manufacturer | Safe Secure Packaging Co. | Propper Manufacturing Co., Inc. | N/A |
| Class of Medical Device | II | II | Identical |
| Code | FRG | FRG, JOJ | Identical |
| Regulation | 21 CFR 880.6850 | 21 CFR 880.6850 | Identical |
| Product name | Sterilization Pouch with Steam and EO Process Indicators | Chex-All Sterilization Pouches and Tubes | Similar |
| Type of packaging device | Flat pouch and roll (tube) | Flat pouch and roll (tube) | Identical |
| Pouch dimensions | Length 8-25.5” Width 3-18” | Length 8-24” Width 3-18” | Similar |
| Tube roll dimensions | Width 2-12”, 100 Ft length | Width 2-12”, 100 Ft length | Identical |
| Single use | Yes | Yes | Identical |
| Indications for use | Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
•Gravity Steam at 121°C (250°F) for 30 minutes, 25 min dry time
•Pre-vacuum Steam at 132°C (270°F) for 4 minutes, 20 min dry time
•Pre-vacuum Steam at 134°C (273°F) for 3 minutes, 20 min dry time
•Pre-vacuum steam at 135°C (275°F) for 3 minutes, 16 min dry time
•Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes.
Aeration time of 8 hours at 60°C (140°F). | Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
Hospital size sterilizer cycles
Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time
Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
Tabletop sterilizer cycles:
Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time
Ethylene Oxide sterilization cycles:
• Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes.
Aeration time of 8 hours at 60°C (140°F). | Intended use: identical
Cycles: similar |
Page 4 (K250306)
{10}
| The device is not intended and has not been validated for sterilization of devices that contain lumens. | Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following: Metal lumen: 7 inch long, 3.2mm internal diameter Plastic lumen: 5 inch long, 2.0 mm internal diameter. Suitable Lumen cycles: Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time). Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time). | Use with lumen devices: different |
| --- | --- | --- |
| The external chemical indicators on the pouches/rolls are intended to demonstrate that the device has been exposed to the Steam or EO sterilization process and to distinguish between processed and unprocessed devices. The chemical indicators change from green to purple after exposure to steam and from yellow to brown after exposure to ethylene oxide. | Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time). | Intended use of process indicators: identical |
| If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 5 years from the date of manufacture. | The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide. | Color change of indicators: different |
| The pouches are intended to allow sterilization of the enclosed medical device(s) and also, to maintain sterility (SAL=10^{-6}). | If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL ≥ 10^{-6}). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization. | Shelf life: similar |
| The subject device is intended and has been validated to maintain sterility of the enclosed device for 6 months after steam sterilization and 24 months after EO sterilization. | | Sterility: identical |
| | | Sterility maintenance Similar |
Page 5 (K250306)
{11}
| | The maximum validated pouch load is 2.64 pounds (1.2kg). | The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13x18”. | Max load weight: identical |
| --- | --- | --- | --- |
| Material composition | Porous paper or Tyvek, PET/CPP high-temperature blue compound film, high temperature adhesive paper, medical double side Tape, EO gas indicator ink, Steam indicator ink | Porous paper, PET/CPP high-temperature clear compound film, high temperature paper, medical double side Tape, EO gas indicator ink, Steam indicator ink | Identical except for film color and Tyvek |
| Air permeance | The maximum equivalent pore size diameter shall not exceed 50 um | The maximum equivalent pore size diameter shall not exceed 50 um | Identical |
| Shelf life before sterilization | 5 years | 2 years | Similar |
| Shelf life after sterilization | 6 months after steam sterilization (gravity or pre-vacuum), or 24 months after EO sterilization. | 30 months after steam sterilization (gravity or pre-vacuum), or 12 months after EO sterilization. | Similar |
| Film Thickness | Film: +0.002 mm | Film: +0.002 mm | Identical |
| Seal strength before sterilization | 2.5-5.5N/15mm | 2.5 N/15mm-5.5 N/15mm | Identical |
| Burst Strength | 2.5-5.5N/15mm | 390-400 kPa. | Different units |
| Peel-open | No splitting of paper more than 10 mm from heat seal lines. | Conform to 5.1.9 d) of AAMI/ISO 11607-1:2019 | Similar |
| Tensile Strength | Paper ≥220N/50mm | Paper ≥4.40 kN/m (≥220N/50mm) | Identical |
| Seal integrity | Visual examination of the seal area through the transparent side of the package shows no channels present | Visual examination of the seal area through the transparent side of the package shows no channels present | Identical |
| Seal width | More than 8 mm | More than 8 mm | Identical |
| EO residues limits | Less than 1 mg/pouch/day | Less than 1 mg/pouch/day | Identical |
| Indicator color change | Steam sterilization: Green to purple EO sterilization: yellow to brown | Steam sterilization: crème to gray/black EO sterilization: beige-pink to green | Different |
Page 6 (K250306)
{12}
Summary of non-clinical testing
Provided below is the summary of non-clinical testing that was performed to demonstrate that Chex-All sterilization pouches and tubes met the acceptance criteria for each standard or test method.
| Parameter | Standard | Acceptance criteria | Results |
| --- | --- | --- | --- |
| Tensile strength MD | ISO 1924-2 | Not less than 4,40 kN/m | Pass |
| Tensile strength CD | ISO 1924-2 | Not less than 2,20 kN/m | Pass |
| Wet Tensile Strength MD | ISO 3781 | Not less than 0,90 kN/m | Pass |
| Wet Tensile Strength CD | ISO 3781 | Not less than 0,45 kN/m | Pass |
| Burst strength | ISO 2758 | Not less than 230 kPa | Pass |
| Paper wet burst strength | ISO 3689 | Not less than 35 kPa | Pass |
| Paper chlorides | ISO 9197 | Less than 0.05% | Pass |
| Paper sulphates | ISO 9198 | Less than 0.25% | Pass |
| Tensile Strength (MD) N/15mm | ASTM D882 | Not less than 20N/15mm | Pass |
| Tensile Strength (CD) N/15mm | ASTM D882 | Not less than 20N/15mm | Pass |
| Testing for microbial barrier | DIN 58953-6 | CFU=0 | Pass |
| Dye penetration test-seal | ASTM F1929 | No leaks in edge seals | |
| Seal strength | ASTM F88 | Not less than 1,5N/15mm | Pass |
| Shelf life | Real time 2 years exposure | Must pass performance tests | Pass |
| Integrity of seals - visual | ASTM F1886 | No visible seal defects | Pass |
| Peel-open characteristics | ISO 11607-1 | As per 5.1.9 d) | Pass |
| Sterility maintenance | 30- and 12- months real time exposure of sterilized pouches Ref. USP | No microbial growth | Pass |
| Sterilant penetration and attainment of sterility in hospital Steam sterilization cycles | ANSI/AAMI/ISO 17665-1:2006/(R)2013, Annex D. Loaded pouches, half cycle | No spore growth | Pass |
| Sterilant penetration and attainment of sterility in tabletop Steam sterilization cycles | ANSI/AAMI/ISO 17665-1:2006/(R)2013, Annex D. Loaded pouches, full cycle | No spore growth | Pass |
| Sterilant penetration and attainment of sterility in EO cycles | ANSI/AAMI/ISO 14937:2009/(R)2013. Loaded pouches, USP protocol | No spore growth | Pass |
| Double-pouch packaging in steam cycles | AAMI/ST-79:2017/(R)2022, 9.5.4. | No spore growth | Pass |
| Drying time | ISO/TS 17665-2:2009, Annex A | Cut-off - no greater than 0.2%. | Pass |
Page 7 (K250306)
{13}
| Steam penetration - lumen devices. Hospital and table-top steam sterilizers, pre-vacuum cycles | Half-cycle test in hospital sterilizer and full cycle test in table-top sterilizer, USP protocol | No spore growth | Pass |
| --- | --- | --- | --- |
| Steam indicator | FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003 | Performance criteria for Process indicators | Pass |
| Steam indicator | FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003, Typical approved cycle | Color change in cycles used in healthcare | Pass |
| Steam indicator | ISO 11140-1:2014., 7.2 | Ink transference | Pass |
| EO indicator | FDA Chemical Indicator Guidance | Performance criteria for Process indicators | Pass |
| EO indicator | FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003, Typical approved cycle | Color change in cycles used in healthcare | Pass |
| EO indicator | ISO 11140-1:2014., 7.2 | Ink transference | Pass |
# Conclusion
The conclusion drawn from the nonclinical test demonstrates that the Chex-All sterilization pouches and tubes are as safe, as effective, and perform as well as or better than the legally marketed predicate device, K153540.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.