← Product Code [FDS](/productcode/FDS) · K250296

# Disposable Distal Attachment D-201 Series (K250296)

_Olympus Medical Systems Corp. · FDS · May 1, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250296

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** May 1, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Disposable Distal Attachment D-201 Series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope’s view field within the gastrointestinal tract.

## Device Story

Disposable Distal Attachment D-201 Series consists of seven sterile, single-use, transparent plastic tubes; attached to distal end of endoscope via medical tape; maintains suitable depth of field during gastrointestinal endoscopic procedures; subset of models includes side drainage hole for egress of water and blood; used by physicians in clinical settings; provides physical spacing between endoscope lens and tissue; facilitates improved visualization; single-use design eliminates need for reprocessing.

## Clinical Evidence

Bench testing only. Includes biocompatibility (ISO 10993-1:2018), sterilization validation (ISO 11135:2014), EO residuals (ISO 10993-7:2008), packaging/shelf-life (ISO 11607-1:2019, ASTM F1980-21), mechanical performance testing, and human factors testing.

## Technological Characteristics

Transparent plastic tube; straight geometry; outer diameters 11.35–15.7 mm; attached via medical tape; sterile (EtO); single-use; biocompatible per ISO 10993-1:2018; includes side drainage holes (except model D-201-11802).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Olympus Distal Attachment, MODELS MH-462, -463, -464, -465, -466, -483, -587, -588, -589, -590, -591, -592, -593, -594, -595, -596, -597, -598 ([K984358](/device/K984358.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

May 1, 2025

Olympus Medical Systems Corp.
% Roshana Ahmed
Program Manager, Regulatory Affairs
Olympus Corporation of the Americas
800 West Park Drive
Westborough, Massachusetts 01581

Re: K250296
Trade/Device Name: Disposable Distal Attachment D-201 Series
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDS, FDF
Dated: January 31, 2025
Received: January 31, 2025

Dear Roshana Ahmed:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250296 - Roshana Ahmed
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K250296 - Roshana Ahmed
Page 3

Sincerely,

**Shanil P. Haugen -S**

Shanil P. Haugen, PhD
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
K250296
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250296

Device Name
Disposable Distal Attachment D-201 Series

Indications for Use (Describe)
The Disposable Distal Attachment D-201 Series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope’s view field within the gastrointestinal tract.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

&gt; “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

{4}

K250296

OLYMPUS

Traditional 510(k)

Disposable Distal Attachment D-201 Series

# 510(k) Summary

## I. Submitter

Olympus Medical Systems Corporation

2951 Ishikawa-cho

Hachioji-shi

Tokyo 192-8507

Japan

Contact Person: Seiko Yunoki

Phone: +81 42-642-2111

Date Prepared: January 20, 2025

## II. Device

|  Device Proprietary Name | Disposable Distal Attachment D-201 Series  |
| --- | --- |
|  Common or Usual Name | Distal Attachment  |
|  Classification Name | Endoscope and accessories  |
|  Regulation Number | 876.1500  |
|  Product Code | FDS/FDF  |
|  Device Classification | Class II  |

## III. Predicate Device

Substantial equivalence is claimed to the following device:

- Olympus Distal Attachment, MODELS MH-462, -463, -464, -465, -466, -483, -587, -588, -589, -590, -591, -592, -593, -594, -595, -596, -597, -598, K984358, Olympus America, Inc.

## IV. Device Description

The Disposable Distal Attachment D-201 Series is comprised of seven (7), sterile, single-use, distal attachments of varying diameters. Each distal attachment is a short transparent

Page 1 of 4

{5}

OLYMPUS
Traditional 510(k)
Disposable Distal Attachment D-201 Series

plastic tube which is attached to the distal end of an endoscope to facilitate observation of tissues during endoscopic procedures. A subset of the models includes a drainage side hole that allows egress of water and blood during use.

The Disposable Distal Attachment D-201 Series is to be used with compatible endoscopes.

## V. Indications for Use

The Disposable Distal Attachment D-201 Series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope's view field within the gastrointestinal tract.

## VI. Comparison of Technological Characteristics

The subject device has the same intended use as the predicate device. The devices are intended to be attached to compatible endoscopes to keep the suitable depth of endoscope's view field within the gastrointestinal tract.

The subject device and predicate device are straight transparent tubes which are attached to the distal end of endoscopes using medical tape to maintain a suitable depth of the field of view during endoscopic procedures. Except for variant D-201-11802, the subject device includes a side drainage hole to facilitate egress of fluids during endoscopic procedures. Furthermore, the subject device is provided sterile (via EtO) and is intended for single use. The subject and predicate device are offered in similar outer diameters; however, the subject device features a shorter length from the distal end of the endoscope than the predicate device.

Bench testing, biological safety evaluation, sterilization validation, and shelf-life studies were undertaken on the subject device and predicate device to demonstrate substantial equivalence. A comparison of the subject and predicate device technological characteristics is provided below.

Page 2 of 4

{6}

OLYMPUS
Traditional 510(k)
Disposable Distal Attachment D-201 Series

|   | Subject Device
Disposable Distal Attachment D-201 Series | Predicate Device
Distal attachment MH series | Analysis  |
| --- | --- | --- | --- |
|  Manufacturer | Olympus Medical Systems Corp. | Olympus Medical Systems Corp. | -  |
|  Indications for use | The Disposable Distal Attachment D-201 series have been designed to be attached to the distal end of the endoscope to keep the suitable depth of endoscope's view field within the gastrointestinal tract. | Olympus Distal Attachment has been designed to be attached to the distal end of the endoscope to facilitate endoscopic therapy.
Olympus Distal Attachment is used for the followings
1.Gastrointestinal mucosal resection (endoscopic mucosal resection)
2.Keeping the suitable depth of endoscope's view field
3.Helping the endoscope with being inserted into the gastrointestinal tract | Substantially equivalent  |
|  Models | D-201-10704
D-201-11304
D-201-11804
D-201-12704
D-201-14304
D-201-15004
D-201-11802 | MH-463
MH-598 | Substantially equivalent  |
|  Type | Straight | MH-463: Straight
MH-598: Straight with rim | Substantially equivalent  |
|  Outer diameter | Ø 11.35 – 15.7 mm | MH-463: Ø 13.5 mm
MH-598: Ø 19.2 mm | Substantially equivalent  |
|  Drainage Side Holes | Yes
D-201-10704
D-201-11304
D-201-11804
D-201-12704
D-201-14304
D-201-15004 | - | Different  |
|   |  No
D-201-11802 | No | Identical  |
|  Attachment to endoscope | Medical tape | Medical tape | Identical  |

Page 3 of 4

{7}

OLYMPUS
Traditional 510(k)
Disposable Distal Attachment D-201 Series

|   | Subject Device
Disposable Distal Attachment D-201 Series | Predicate Device
Distal attachment MH series | Analysis  |
| --- | --- | --- | --- |
|  Single use/Re-usable | Single use | Reusable | Different  |
|  Sterile/non-sterile | Sterile | Non-sterile | Different  |
|  Sterilization method | ETO | Non sterilized
(Autoclave by the user before use) | Different  |
|  Transparent | Yes | Yes | Identical  |
|  Biocompatible | Yes
ISO 10993-1:2018 | Yes
ISO 10993-1 | Identical  |

## VII. Performance Data

The following performance data were provided to demonstrate substantial equivalence:

- Biocompatibility testing per ISO 10993-1:2018 including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP &lt;51&gt;)
- Sterilization validation per ISO 11135:2014
- Ethylene oxide residuals per ISO 10993-7:2008
- Packaging validation and shelf life testing in accordance with ISO 11607-1:2019 and ASTM F1980-21
- Mechanical testing and comparative testing to verify device performance
- Human Factors Testing

Clinical data is not required to demonstrate substantial equivalence.

## VIII. Conclusion

The non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device. Therefore, it is concluded that the Disposable Distal Attachment D-201 Series is substantially equivalent to the identified predicate device.

Page 4 of 4

---

**Source:** [https://fda-staging.innolitics.com/device/K250296](https://fda-staging.innolitics.com/device/K250296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).