ProVu Single Use Video Stylet with Reinforced ET Tube (Size 6.5) (038-995-065U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.0) (038-995-070U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.5) (038-995-075U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 8.0) (038-995-080U)
K250243 · Flexicare Medical Limited. · BTR · May 20, 2025 · Anesthesiology
Device Facts
Record ID
K250243
Device Name
ProVu Single Use Video Stylet with Reinforced ET Tube (Size 6.5) (038-995-065U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.0) (038-995-070U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.5) (038-995-075U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 8.0) (038-995-080U)
Applicant
Flexicare Medical Limited.
Product Code
BTR · Anesthesiology
Decision Date
May 20, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProVu Endotracheal Tube with Video Stylet is intended for intubation procedures. The ProVu Endotracheal Tube with Video Stylet is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The ProVu Endotracheal Tube with Video Stylet is indicated for viewing during non-difficult and difficult intubation procedures and for verifying endotracheal tube placement.
Device Story
System consists of sterile, single-use reinforced endotracheal tube (ETT) and video stylet with removable malleable rod. Used in hospital environments by clinicians for oral/nasal intubations. Video stylet captures real-time color images from tracheal lumen aperture; transmits video to external 8-inch display. Clinician views display to guide tube placement, verify positioning, and navigate airway during difficult or non-difficult intubations. Provides visual confirmation of tube placement; assists in securing temporary artificial airway for mechanical ventilation. Benefits include improved visualization during intubation procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including biocompatibility (cytotoxicity, sensitization, reactivity, pyrogenicity, implantation, chemical characterization), sterilization validation (ISO 11135), electrical safety/EMC (IEC 60601-1, 60601-2-18, 60601-1-2), and mechanical performance testing per ISO 5361 and ISO 5356-1. Usability validation conducted per FDA human factors guidance.
Technological Characteristics
System includes PVC/PP reinforced ETT and video stylet. Features high-volume, low-pressure PVC cuff and two Murphy eyes. Imaging via video camera with real-time color output. Connectivity via cable to 8-inch display. Sterilized via EtO (ISO 11135). Complies with ISO 5356-1, ISO 5361, ISO 10993-1, ISO 18562, and IEC 60601 standards.
Indications for Use
Indicated for adults requiring mechanical ventilation as a temporary artificial airway. Used for oral and nasal intubations, viewing during non-difficult and difficult intubation procedures, and verifying endotracheal tube placement.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 20, 2025
Flexicare Medical Limited
Rebecca Funston
Director of Regulatory Affairs and Quality Assurance
Cynon Valley Business Park Mountain Ash.
Rhondda Cynon Taf, Wales CF45 4ER
United Kingdom
Re: K250243
Trade/Device Name: ProVu Single Use Video Stylet with Reinforced ET Tube (Size 6.5) (038-995-065U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.0) (038-995-070U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.5) (038-995-075U); ProVu Single Use Video Stylet with Reinforced ET Tube (Size 8.0) (038-995-080U)
Regulation Number: 21 CFR 868.5730
Regulation Name: Tracheal Tube
Regulatory Class: Class II
Product Code: BTR, BSR
Dated: April 23, 2025
Received: April 23, 2025
Dear Rebecca Funston:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
2K250243 - Rebecca Funston
Page
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
3K250243 - Rebecca Funston
Page
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250243
Device Name
ProVu Single Use Video Stylet with Reinforced ET Tube (Size 6.5) (038-995-065U);
ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.0) (038-995-070U);
ProVu Single Use Video Stylet with Reinforced ET Tube (Size 7.5) (038-995-075U);
ProVu Single Use Video Stylet with Reinforced ET Tube (Size 8.0) (038-995-080U)
Indications for Use (Describe)
The ProVu Endotracheal Tube with Video Stylet is intended for intubation procedures. The ProVu Endotracheal Tube with Video Stylet is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations.
The ProVu Endotracheal Tube with Video Stylet is indicated for viewing during non-difficult and difficult intubation procedures and for verifying endotracheal tube placement.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025/03/20
1. Submission sponsor
Name: Flexicare Medical Limited
Address: Cynon Valley Business Park Mountain Ash. CF45 4ER. United Kingdom
Contact person: Rebecca Funston
Title: Director of Regulatory Affairs and Quality Assurance
Tel: +44 (0)1443 474647
2. Subject Device Information
| Trade/Device Name | ProVu Video Stylet and Reinforced Endotracheal Tube |
| --- | --- |
| Common Name | Endotracheal Tube |
| Regulatory Class | Class II |
| Classification | 21CFR 868.5730 / tube, tracheal (w/wo connector)/ BTR |
| Product code | BTR, BSR |
| Submission type | Traditional 510(K) |
3. Predicate Device
VIVASIGHT-SL (TVTTM) device manufactured by “ETView Ltd” at date of clearance (Now owned by Ambu) and cleared under “K152438”.
4. Device Description
The ProVu Video Stylet with Reinforced ET Tube is a sterile, single-use, video-enabled stylet and ETT intubation system with malleable and directional control intended to assist with oral and nasal intubations and designed for use in adults in a hospital environment.
The device comprises of two main components:
- The ProVu Reinforced Endotracheal tube
- The ProVu Video Stylet (with a removable malleable rod)
5. Intended use & Indication for use
The ProVu Endotracheal Tube with Video Stylet is intended for intubation procedures. The ProVu Endotracheal Tube with Video Stylet is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations.
The ProVu Endotracheal Tube with Video Stylet is indicated for viewing during non-difficult and difficult intubation procedures and for verifying endotracheal tube placement.
1 / 5
{5}
6. Comparison to the Predicate Device
| | ProVu Video Style with Reinforced ET Tube | ETView Ltd.'s VIVASIGHT-SL (TVT™) | Comparison |
| --- | --- | --- | --- |
| 510(k) | K250243 | K152438 | / |
| Indications for use | The ProVu Video Style with Reinforced ET Tube is intended for intubation procedures. The ProVu Video Style with Reinforced ET Tube is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The ProVu Video Style with Reinforced ET Tube is indicated for viewing during non-difficult and difficult intubation procedures and for verifying endotracheal tube placement. | The ETView VivaSight-SL (TVTTM) is intended for intubation procedures. The ETView VivaSight-SL (TVTTM) is indicated for use as a temporary artificial airway in adults requiring mechanical ventilation. It is intended for oral and nasal intubations. The VivaSight-SL (TVTTM) System is indicated for viewing during non-difficult and difficult intubation procedures, for verifying endotracheal tube and endobronchial blocker placement and repositioning, for viewing during suctioning and for general inspection of the airway. | Equivalent |
| Product Code | BTR, BSR | BTR | Same |
| Regulation No. | 21 CFR 868.5730 | 21 CFR 868.5730 | Same |
| Classification | Class II | Class II | Same |
| Supplied/Use | Single use | Single use | Same |
| Method of sterilization | EtO | EtO | Same |
| Basic structure | Standard ETT | Standard ETT | Same |
| Endotracheal Tube Material | - Polyvinyl Chloride (PVC)
- Polypropylene (PP) | - Polyvinyl Chloride (PVC)
- Polypropylene (PP) | Same |
2 / 5
{6}
| Imaging Sensor | Video camera | Video camera | Same |
| --- | --- | --- | --- |
| Video transfer | Colored real time image acquisition from the aperture of the tracheal lumen | Colored real time image acquisition from the aperture of the tracheal lumen | Same |
| Video Format | Motion Joint Photographic Experts Group | Composite Video Baseband Signal | Similar |
| Bevel | Opening to the bottom. The camera situates centrally within lumen. | Opening to the bottom with the camera on the top | Similar |
| Display system | Cable connection to ProVu 8” display | Cable connection to VivaSight-Max Monitor-7” monitor | Equivalent |
| Tube Dimensions: Internal Diameter | Four models: • ID 6.5 mm • ID 7.0 mm • ID 7.5 mm • ID 8.0 mm | Three models: • ID 7.0 mm • ID 7.5 mm • ID 8.0 mm | Similar |
| Tube Dimensions: Outer Diameter | Four models: • 9.0mm • 9.4mm • 10.1mm • 10.8mm | Three models: • OD 10.0 mm • OD 10.5 mm • OD 11.0 mm | Similar |
| Cuff | Medical grade PVC, High volume, low pressure | Medical grade PVC, High volume, low pressure | Same |
| Number of Murphy Eyes | 2 | 2 | Same |
| Shelf life | 0.5 year | 3 Years | Similar |
| Packaging | Rigid Blister tray, EO Permeable paper top | EO Permeable peel pouch | Same |
| Target population | Adult | Adult | Same |
| Emergency use | Emergency and elective | Emergency and elective | Same |
| Environment of use | Hospital | Hospital | Same |
| Duration of use | <30d | <30d | Same |
| Standards met | ISO 5356-1 ISO 5361 ISO 14971 ISO 11135 IEC 60601-1 | ISO 5356-1 ISO 14971 ISO 11135 IEC 60601-1 IEC 60601-1-2 | Same |
{7}
| | IEC 60601-1-2
IEC 60601-2-18 | IEC 60601-2-18 | |
| --- | --- | --- | --- |
| Biocompatibility | ISO 10993-1
ISO 18562 | ISO 10993-1 | Same |
7. Non-clinical Data
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
Biocompatibility of the subject device was evaluated in accordance with the FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” and ISO 18562-1”. Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process” as recognized by FDA. The subject device is an externally communicating device with tissue contact for a prolonged duration (up to 30 days) with indirect gas pathway contact. The following tests are provided.
- Cytotoxicity
- Sensitization
- Intracutaneous reactivity
- Material Mediated Pyrogenicity
- Subcutaneous Implantation
- Chemical Characterisation
- Toxicological Risk Assessment
Per ISO 18562-1 “Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process”, Gas Pathway Testing using the methods in ISO 18562-2 “Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter” and ISO 18562-3 “Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic compounds (VOCs)” was completed as part of the biocompatibility testing.
Sterilization and shelf-life testing
The subject device is provided sterile, and its shelf-life is 0.5 year. Sterilization Process has been validated according to ISO 11135:2014. EO/ECH residual test was performed according to ISO 10993-7:2008.
{8}
5 / 5
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.
# Bench performance testing
The following bench tests were performed:
- ISO 5361 Third edition 2016-09-01 Anaesthetic and respiratory equipment - Tracheal tubes and connectors
- ISO 5356-1 Third edition 2004-05-15 Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and sockets
# Usability validation
Usability validation was conducted according to FDA guidance “Applying Human Factors and Usability Engineering to Medical Devices”.
# 8. Clinical study
Not applicable.
# 9. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.