K250238 · Abanza Tecnomed S.L · MBI · Feb 25, 2025 · Orthopedic
Device Facts
Record ID
K250238
Device Name
WasherCap™ Fixation System
Applicant
Abanza Tecnomed S.L
Product Code
MBI · Orthopedic
Decision Date
Feb 25, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
WasherCap™ Fixation System is intended for fixation of soft tissue grafts, including tendons and ligaments, during surgical procedures such as in Anterior Cruciate Ligament (ACL) reconstruction of the knee.
Device Story
WasherCap™ Fixation System is an implantable, non-active orthopedic device used for soft tissue graft fixation during ACL reconstruction. The system comprises a washer and cap made of PEEK (ASTM 2026) and a titanium alloy screw (ISO 5832-3, ASTM F136). It is used by orthopedic surgeons in a surgical setting. The device is implanted using a dedicated, reusable WasherCap™ Instrument Set. The system provides mechanical fixation of tendons or ligaments to bone, facilitating graft healing. The current submission introduces an 11mm implant size to expand the range of compatible tendon grafts (6mm–11mm). The device is non-degradable and provides stable fixation.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by bench testing and comparison to the predicate device (K212197).
Technological Characteristics
Implantable non-active fixation system. Washer/cap: PEEK (ASTM 2026). Screw: Titanium alloy (Ti-6Al-4V ELI, ISO 5832-3, ASTM F136). Mechanical fixation principle. Reusable surgical instrument set required for implantation. No software or electrical components.
Indications for Use
Indicated for fixation of soft tissue grafts (tendons/ligaments) to bone during surgical procedures, specifically ACL reconstruction of the knee.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2025
Abanza Tecnomed S.1 % Jessica Czamanski Regulatory Consultant Precision Life Science Partners 300 Creek View Road Newark, Delaware 19711
Re: K250238
Trade/Device Name: WasherCap™ Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: January 24, 2025 Received: January 27, 2025
Dear Jessica Czamanski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Thomas Mcnamara -S
For: Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K250238
Device Name
WasherCap™ Fixation System
Indications for Use (Describe)
WasherCap™ Fixation System is intended for fixation of soft tissue grafts, including tendons and ligaments, during surqical procedures such as in Anterior Cruciate Ligament (ACL) reconstruction of the knee.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ABANZA. The logo consists of a stylized blue shape on the left, resembling a curved, elongated form with intersecting lines. To the right of the shape is the word "ABANZA" in a blue, sans-serif font. The overall design is clean and modern.
# WahserCap™ Fixation System – 510(k) Summary
#### l. Submitter
| Submitter Name: | Anbanza Tecnomed, S.L. |
|----------------------|-----------------------------------------------|
| Contact Person: | Ander Menaut<br>RD Manager |
| Address: | Calle Nueva 29<br>Multiva, ES, 31192<br>Spain |
| Telephone: | (+34) 948-044-643 |
| Email: | regulatory@abanzamed.com |
| Date of Preparation: | January 24, 2025 |
#### II. Application Correspondent
| Contact's Name: | Precision Life Sciences Partners LLC |
|-----------------|-------------------------------------------------|
| Contact Person: | Jessica Czamanski |
| Address: | 300 Creek View Road<br>Newark, DE, 19711<br>USA |
| Telephone: | (754) 422-9101 |
| Email: | jczamanski@plsp.pro |
#### lll. Device
| Trade Name: | WasherCap™ Fixation System |
|-------------------------|------------------------------------------------|
| Common Name: | Non-Degradable Soft Tissue Fixation Device |
| Classification Name: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
#### IV. Predicate Device
| Manufacturer: | Abanza Tecnomed, S.L. |
|-------------------------|----------------------------|
| Device Name: | WasherCap™ Fixation System |
| 510(k) Number: | K212197 |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Avanza. The logo consists of a stylized blue icon on the left and the word "ABANZA" in blue on the right. The icon appears to be an abstract shape with curved lines, possibly representing movement or growth. The font used for the word "ABANZA" is a clean, sans-serif typeface.
#### V. Device Description
The WasherCap™ Fixation System is an implantable non-active device consisting of three components: the washer, the cap, and the screw. All components building the WasherCap™ Fixation System are made of high-quality biocompatible materials, with a long-existence use in terms of human safety and patient's clinical performance.
The Washer and the Cap are made from non-degradable organic thermoplastic polymer biomaterial, so-called Polietheretherketone (PEEK by Invibio™ Biomaterial Solutions in accordance with ASTM 2026) and the Screw is made from implantable degree Titanium Alloy (Ti-6Al- 4V ELI) in accordance with ISO 5832-3 and ASTM F136 standards.
To achieve correct implantation and use of the device requires specific accessories: WasherCap™ Instrument Set. The required accessories are product- specific reusable surgical instruments that have been designed for the implantation of the WasherCap™ Fixation System. Use of the WasherCap™ Instrument Set will ensure the surgical procedure, and the use of the WasherCap™ Fixation System are carried out in an appropriate manner.
#### VI. Intended Use
The WasherCap™ Fixation System is intended for fixation of soft tissue grafts to bone.
#### VII. Indications for Use
WasherCap™ Fixation System is intended for fixation of soft tissue grafts, including tendons and ligaments, during surgical procedures such as in Anterior Cruciate Ligament (ACL) reconstruction of the knee.
# VIII. Comparison of Technological Characteristics
The proposed WasherCap™ Fixation System is identical to its predicate, except for a 1mm difference which will allow compatibility with the full range of tendon grafts used in ACL reconstruction surgery, which spans 6mm – 11mm. The subject device includes accessories sized appropriately to effectively use the 11mm implant.
All materials and manufacturing processes are identical to those used for the predicate.
#### IX. Sterilization
There are no differences in the sterilization process between the subject and predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Abanza logo, which consists of a blue abstract shape on the left and the word "ABANZA" in blue capital letters on the right. Below the logo, there is the text "X. Performance Data" in black. The text is left-aligned and appears to be a section heading or title.
All performance data for the WasherCap™ Fixation System was obtained using the same validated methodology used in the predicate submission (K212197).
# Performance Testing - Bench
Mechanical testing verifications were performed on the proposed WasherCap™ Fixation System 11mm implant to ensure device performance was identical to that of the predicate (K212197). Testing was performed using the same acceptance criteria and protocols used in the predicate submission. The proposed device was found to be substantially equivalent in performance compared to its predicate.
# Biocompatibility
There is no change in materials or manufacturing of the WasherCap™ Fixation System. The biocompatibility data in the predicate submission (K212197) support the subject device.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and electromagnetic compatibility have not been evaluated.
# Software Verification and Validation
The device does not contain software/firmware.
# Performance Testing - Animal
This submission does not include any animal performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
# Performance Testing - Clinical
This submission does not include any clinical performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
#### XI. Conclusion
The proposed WasherCap™ Fixation System has the same intended use, indications for use, operating principle, fundamental technology, materials, and manufacturing as the predicate device. Additional testing was performed to ensure performance of the proposed 11mm implant and
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ABANZA. The logo consists of a stylized, abstract shape on the left, resembling a network or interconnected structure, rendered in shades of blue. To the right of the shape is the word "ABANZA" in a sans-serif font, also in blue. The overall design is clean and modern.
associated instruments is identical to the predicate, and that the addition of the new size does not introduce new questions for safety and effectiveness. The information provided in this submission demonstrated that the subject device, WasherCap™ Fixation System, is substantially equivalent to its predicate, WasherCap™ Fixation System K212197.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.