Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual.
Device Story
Automatic Upper Arm Blood Pressure Monitor; non-invasive oscillometric measurement; uses electronic capacitive pressure sensor to detect cuff pressure oscillations; converts pressure changes to electrical signals; calculates systolic/diastolic BP and pulse rate. Device includes Bluetooth for data transmission to mobile app; app enables remote measurement control and data display. Used by medical professionals or patients at home; cuff wrapped around left upper arm. Provides visual feedback on device display (BP, pulse, WHO classification bar, battery status). Assists in routine BP monitoring; aids clinical decision-making by providing objective hemodynamic data.
Clinical Evidence
No new clinical data provided. Substantial equivalence is supported by bench testing (electrical safety, EMC, biocompatibility) and reliance on clinical performance data from predicate K193627 and reference K193624, which were tested per ISO 81060-2.
Technological Characteristics
Oscillometric measurement; electronic capacitive pressure sensor. Power: 4x AA batteries or DC 6V adapter. Connectivity: Bluetooth for app integration. Biocompatibility: ISO 10993 series. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-30, ISO 81060-1. Form factor: Arm-type and tunnel-type units with multiple cuff sizes.
Indications for Use
Indicated for adult patients requiring monitoring of systolic/diastolic blood pressure and pulse rate. Intended for use by medical professionals or at home.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Automatic Upper Arm Blood Pressure Monitor (K193627)
Reference Devices
Automatic Upper Arm Blood Pressure Monitor (K193624)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 3, 2025
Dongguan E-test Technology Co., Ltd
Wan Victor
Official Correspondent
Room 201,301, Building 1, Changping Section No.1,
Dongshen Road, Changping Town
Dongguan, Guangdong 523570
China
Re: K250231
Trade/Device Name: Automatic Upper Arm Blood Pressure Monitor (BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X, BA-835, BA-836, BA-839, BA-840, BA-856)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: July 28, 2025
Received: August 1, 2025
Dear Wan Victor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250231 - Wan Victor
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250231 - Wan Victor
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
for LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250231 | |
| Device Name Automatic Upper Arm Blood Pressure Monitor model: BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X, BA-835, BA-836, BA-839, BA-840, BA-856) | |
| Indications for Use (Describe) Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250231
# 510(k) Summary for K250231
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
Company Name: DONGGUAN E-TEST TECHNOLOGY CO., LTD
Establishment Registration Number: 3017896172
Address: Room 201,301, Building 1, Changping Section No.1, Dongshen Road, Changping Town, Dongguan City, Guangdong, China Dongguan Guangdong 523560 China
Contact Person (including title): Wan Victor (General Manager)
Tel: 0769-81158038
Fax: /
Post code: 523570
E-mail: victor@sinoetest.com
## Application Correspondent:
Contact Person: Mr. Wan Victor
DONGGUAN E-TEST TECHNOLOGY CO., LTD.
Address: Room 201,301, Building 1, Changping Section No.1, Dongshen Road, Changping Town, Dongguan City, Guangdong, China Dongguan Guangdong 523560 China
Tel: 0769-81158038
Email: regulatory@glomed-info.com
## 2. Subject Device Information:
Type of 510(k): Traditional
Common Name: Noninvasive blood pressure measurement system
Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
Trade Name: Automatic Upper Arm Blood Pressure Monitor
Model Name: BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X, BA-835, BA-836, BA-839, BA-840, BA-856
Review Panel: Cardiovascular
Product Code: DXN
Regulation Number: 870.1130
Regulatory Class: II
## 3. Predicate Device Information
### Predicate Device
Sponsor: DONGGUAN E-TEST TECHNOLOGY CO., LTD.
Trade Name: Automatic Upper Arm Blood Pressure Monitor
Model: BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X, BA-818, BA-819
Common Name: Noninvasive blood pressure measurement systems
Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
510(K) Number: K193627
Review Panel: Cardiovascular
Product Code: DXN
Regulation Number: 870.1130
Regulation Class: II
### Reference Device
Sponsor: Dongguan E-Test Technology Co., Ltd.
1 / 6
{5}
K250231
Trade Name: Automatic Upper Arm Blood Pressure Monitor
Model: BA-815, BA-816
Common Name: Noninvasive blood pressure measurement systems
Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
510(K) Number: K193624
Review Panel: Cardiovascular
Product Code: DXN
Regulation Number: 870.1130
Regulation Class: II
## 4. Device Description
Automatic Upper Arm Blood Pressure Monitor is designed to measure the systolic, diastolic and pulse rate of an individual by using a non-invasive technique which an inflatable cuff is wrapped around upper arm. Our method to define systolic and diastolic pressures is similar to the auscultatory method but using an electronic capacitive pressure sensor rather than stethoscope and mercury manometer. The sensor converts tiny alteration in cuff pressure to electrical signals; by analyzing those signals to define the systolic, diastolic and calculating pulse rate is a well-known technique in the market so called "oscillometric method". The device also has low voltage indication, which will be triggered when the battery is low.
The product includes Bluetooth transmission functionality which can connect to APP and transfer data for APP, and the measuring data, including systolic diastolic pressures and pulse rate can be displayed on APP. The APP can control the device measuring when connecting device.
There are 18 models and 2 types (Arm type and tunnel type). Arm type has models BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X. They are all the same except for appearance and buttons. Tunnel type has models BA-835, BA-836, BA-839, BA-840, BA-856. They are all the same except for appearance and cuff circumference. Arm type has cuff as accessory and there are 6 sizes, you can select suitable size according to your arm circumference. The product provides 4 AA batteries or DC 6.0V adapter for power supply for power supply.
## 5. Intended Use / Indications for Use
Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual.
## 6. Comparison to predicate devices
| Elements of Comparison | Subject Device | Primary Predicate Device | Reference Device | Verdict |
| --- | --- | --- | --- | --- |
| Company | DONGGUAN E-TEST TECHNOLOGY CO., LTD | DONGGUAN E-TEST TECHNOLOGY CO., LTD | DONGGUAN E-TEST TECHNOLOGY CO., LTD | -- |
| Trade Name | Automatic Upper Arm Blood Pressure Monitor | Automatic Upper Arm Blood Pressure Monitor | Automatic Upper Arm Blood Pressure Monitor | -- |
| Model | BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X, BA-835, BA-836, BA-839, BA-840, BA-856 | BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X, BA-818, BA-819 | BA-815, BA-816 | -- |
{6}
K250231
| Elements of Comparison | Subject Device | Primary Predicate Device | Reference Device | Verdict |
| --- | --- | --- | --- | --- |
| 510(k) Number | K250231 | K193627 | K193624 | -- |
| Medical Specialty | Cardiovascular | Cardiovascular | Cardiovascular | Same |
| Regulation | 21 CFR 870.1130 | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Product Code | DXN | DXN | DXN | Same |
| Regulation Class | II | II | II | Same |
| Intended Use / Indications for Use | Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual. | Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual. | Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual. | Same |
| Type for use | OTC | OTC | OTC | Same |
| Power Source | DC 6V (4 X AA 1.5V alkaline batteries)
For arm type adapter: Output DC 6V/1A
For tunnel type adapter: Output DC 6V/1.5A | DC 6V (4 X AA 1.5V alkaline batteries) | DC 6V (4 X AA 1.5V alkaline batteries) | Similar Note 3 |
| Measurement Site | Upper Arm | Upper Arm | Upper Arm | Same |
| Measuring Range | Pressure: 0~294mmHg
Pulse: 40~199 beats/minute | Pressure: 0~280 mmHg
Pulse: 40-199 beats/minute | Pressure: 0~280 mmHg
Pulse: 40-199 beats/minute | Similar Note 1 |
| Measuring Accuracy | Pressure: ±3 mmHg
Pulse: ±5% | Pressure: ±3 mmHg
Pulse: ±5% | Pressure: ±3 mmHg
Pulse: ±5% | Same |
3 / 6
{7}
K250231
| Elements of Comparison | Subject Device | Primary Predicate Device | Reference Device | Verdict |
| --- | --- | --- | --- | --- |
| Resolution | 1 mmHg | 1 mmHg or 0.1kPa | 1 mmHg or 0.1kPa | Same |
| Cuff Circumference | Arm type:
size A: 17cm--22cm (SMALL ADULT CUFF)
size B: 22cm--30cm (ADULT CUFF-1)
size C: 24cm--34cm (ADULT CUFF-2)
size D: 22cm--42cm (L-LARGE ADULT CUFF)
size E: 30cm--42cm (LARGE ADULT CUFF)
size F: 42cm--50cm (EXTRA LARGE ADULT CUFF)
Tunnel type:
BA-835, BA-839, BA-840: 22-34cm
BA-836, BA-856: 28-42cm | There are 6 size:
size A: 17cm--22cm (SMALL ADULT CUFF)
size B: 22cm--30cm (ADULT CUFF-1)
size C: 24cm--34cm (ADULT CUFF-2)
size D: 22cm--42cm (L-LARGE ADULT CUFF)
size E: 30cm--42cm (LARGE ADULT CUFF)
size F: 42cm--50cm (EXTRA LARGE ADULT CUFF) | For BA-815: 22-34 cm;
For BA-816: 28-42 cm; | Same |
| Inflation and Deflation | Automatic | Automatic | Automatic | Same |
| Measuring Method | Oscillometry | Oscillometry | Oscillometry | Same |
| Patient Population | Adult | Adult | Adult | Same |
| Display | Blood Pressure (Systolic and Diastolic), Pulse, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Blood Pressure (Systolic and Diastolic), Pulse, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Blood Pressure (Systolic and Diastolic), Pulse, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Same |
| Operation condition | Temperature: 5°C ~ 40°C
Humidity: 10~90%RH
Atmospheric Pressure:86 kPa~106 kPa | Temperature: 5°C ~ 40°C
Humidity: 15~90%RH
Atmospheric Pressure:86 kPa~106 kPa | Temperature: 5°C ~ 40°C
Humidity: 15~90%RH
Atmospheric Pressure:86 kPa~106 kPa | Similar Note 2 |
| Transport/storage | Temperature: -20°C ~ +65°C | Temperature: -20°C ~ +65°C | Temperature: -20°C ~ +65°C | Same |
4 / 6
{8}
K250231
| Elements of Comparison | Subject Device | Primary Predicate Device | Reference Device | Verdict |
| --- | --- | --- | --- | --- |
| environment | Humidity: 10~95%RH
Atmospheric Pressure:86
kPa~106 kPa | Humidity: 10~95%RH
Atmospheric Pressure:86
kPa~106 kPa | Humidity: 10~95%RH
Atmospheric Pressure:86
kPa~106 kPa | |
| Safety and EMC | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 80601-2-30 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 80601-2-30 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 80601-2-30 | Same |
| Biocompatibility | ISO 10993 series | ISO 10993 series | ISO 10993 series | Same |
Comparison in Detail(s):
Note 1:
The Pressure Measuring Range is slightly different. The IEC 80601-2-30 can demonstrate that the subject device can maintain the safety and performance within the measurement range. Thus, the difference does not raise different questions of safety and effectiveness
Note 2:
Compared with the predicate device, the subject device requires less strict operation environment. The IEC 60601-1-11 test report can demonstrate that the subject device can maintain the safety and performance within the operation and storage environment. Thus, this difference does not raise different questions of safety and effectiveness.
Note 3:
The Power Source is slightly different. The subject device meets the requirements of IEC60601-1 Thus, the difference does not raise different questions of safety and effectiveness
7. Test Summary
7.1 Non-Clinical Tests Performed
1) Electrical safety, and electromagnetic compatibility Test
Electrical safety and EMC testing were conducted on the proposed device. The device was shown to comply with IEC 60601-1-2:2020 and IEC TR 60601-4-2:2016 for electromagnetic compatibility, IEC 60601-1:2020, IEC 60601-1-11:2020, ISO 80601-2-30:2018 and ISO 81060-1:2007 for electrical safety.
2) Biocompatibility Test
We declare that the cuff in the Automatic Upper Arm Blood Pressure Monitor (Model: BA-831X, BA-832X, BA-833X, BA-837X, BA-838X, BA-842X, BA-843X, BA-845X, BA-847X, BA-849X, BA-851X, BA-852X, BA-855X) is identical to those used in the Automatic Upper Arm Blood Pressure Monitor K193627, in formulation, processing, and cleaning, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
And We declare that the cuff cover in the Automatic Upper Arm Blood Pressure Monitor (model: BA-835, BA-836, BA-839, BA-840, BA-856) is identical to those used in the Automatic Upper Arm Blood Pressure Monitor K193624 in formulation, processing, and cleaning, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
3) Software verification and validation
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket
5 / 6
{9}
K250231
Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
## 4) Cybersecurity
Cybersecurity software documentation was created for subject device according to FDA guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 18, 2018.
## 7.2 Summary of Clinical Performance
The principles, software, algorithms, sensors, etc. of the blood pressure measurement part of the subject device and the predicate device are the same, so the clinical data of the predicate device, K193627, and reference device, K193624, can represent the subject device, and the predicate device, K193627, and reference device, K193624, have been tested according to the ISO 81060-2 standard and used for the subject device submitted this time.
## 8. Date of the summary prepared: September 1, 2025
## 9. Final Conclusion
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K193627, and reference device, K193624.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.