← Product Code [ONE](/productcode/ONE) · K250224

# handLITE (TN19S) (K250224)

_iSMART Developments, Ltd. · ONE · Apr 18, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250224

## Device Facts

- **Applicant:** iSMART Developments, Ltd.
- **Product Code:** [ONE](/productcode/ONE.md)
- **Decision Date:** Apr 18, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The handLITE is intended to emit energy in the red and near infrared region of the light spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat contact dermatitis of the hand and wrist.

## Device Story

The handLITE is a wearable, home-use LED phototherapy device designed to treat contact dermatitis of the hand and wrist. The device consists of a flexible silicone glove with integrated red (630nm) and near-infrared (830nm) LEDs, powered by a rechargeable lithium-ion polymer battery. The device operates via an integrated controller that manages treatment duration; it delivers 30mW/cm² intensity for a 10-minute session, resulting in an 18J/cm² dose per treatment. Patients operate the device at home, applying it to the dorsal and palmar aspects of the hand. The device uses a timer to automatically switch off after the session. Clinical output is measured by the patient's reduction in signs and symptoms of dermatitis (SCORAD index). The device provides a non-pharmacological treatment option, potentially reducing itchiness and scaling associated with contact dermatitis.

## Clinical Evidence

Prospective clinical study (n=25) evaluated safety and effectiveness for mild-to-moderate contact dermatitis. Subjects treated one hand 5x/week for 6 weeks (30 treatments). Primary endpoint: SCORAD index improvement. Results showed mean improvement of 12.2 for treated hands vs. 3.5 for untreated control hands. 78% of subjects reported improvement; 65% reported reduced itch/scaling. No adverse events reported. Usability testing (n=19) confirmed all subjects could correctly follow labeling instructions.

## Technological Characteristics

Flexible silicone glove with integrated red (630nm) and near-infrared (830nm) LEDs. Intensity: 30mW/cm². Energy source: Rechargeable lithium-ion polymer battery. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-2-83, IEC 62133-2, IEC 62471, IEC 62366-1. Software: Embedded controller for treatment timing.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Philips BlueControl ([K171055](/device/K171055.md))
- faceLITE LED Mask ([K191629](/device/K191629.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 18, 2025

iSMART Developments LTD
Susan D'arcy
Managing Director
129 Green Lanes, Wylde Green
Birmingham, B73 5LT
United Kingdom

Re: K250224
Trade/Device Name: handLITE (TN19S)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: ONE
Dated: January 15, 2025
Received: January 27, 2025

Dear Susan D'arcy:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K250224 - Susan D'arcy
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250224 - Susan D'arcy
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

YAN FU -S
Digitally signed by YAN FU -S
Date: 2025.04.18 17:22:49
-04'00'

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250224  |   |
|  Device Name handLITE (TN19S)  |   |
|  Indications for Use (Describe) The handLITE is intended to emit energy in the red and near infrared region of the light spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat contact dermatitis of the hand and wrist.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

handLITE LED Device

510(k) Summary

iSmart Development

iSmart.health

# 510(k) Summary -K250224

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).

|  Submitter's Name: | iSMART Developments LTD  |
| --- | --- |
|  Submitter's Address: | 129 Green Lanes, Wylde Green, Birmingham, B73 5LT United Kingdom  |
|  Contact Person: | Susan D'Arcy  |
|  Telephone: | +44 (0) 7880313315  |
|  Date Prepared: | March 28, 2025  |
|  Device Trade Name: | handLITE Light-Emitting Diode (LED) Device (TN19S)  |
|  Common Name | Powered Light-Based Non-Laser Surgical Instrument  |
|  Classification Name | Laser surgical instrument for use in general and plastic surgery and in dermatology  |
|  Regulation | 21 CFR 878.4810  |
|  Product Code | ONE  |
|  Review Panel | General & Plastic Surgery  |
|  Device Class | Class II  |

## Predicate Devices

K171055
K191629
Philips Blue Control
faceLITE LED Mask

## Intended Use

The handLITE is intended to emit energy in the red and near infrared region of the light spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat contact dermatitis of the hand and wrist.

## Device Description

The handLITE device consists of:

1. Flexible silicone glove with integrated controller (contains rechargeable lithium-ion polymer batter)
2. Power supply and country specific adaptors

handLITE is a home use wearable light emitting diode (LED) phototherapy device.

The device consists of a flexible silicone glove that contains red (630nm) and near infrared (830nm) light emitting diodes (LEDs) and an integrated controller. The LEDs generate the light at an intensity of $30\mathrm{mW/cm^2}$, delivering $18\mathrm{J/cm^2}$ per treatment.

The power supply is used to charge the integrated Lithium battery in the handLITE.

Page 1 of 10

{5}

handLITE LED Device

510(k) Summary

^{}[]

ismart

development

ismart.health

# Substantial Equivalence Discussion

# Substantial Equivalence

The handLITE LED Device is substantially equivalent to the Philips BlueControl (K171055) and faceLITE LED Mask (K191629).

|  Description | Subject device
handLITE (TN19S) | Predicate
Philips BlueControl
K171055 | Predicate
faceLITE LED mask
K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Device manufacturer | ISMART DEVELOPMENTS LTD | Philips Electronics
Nederland B.V. | iSMART Marketing SVCS Ltd | N/A  |
|  Device classification name | Powered Light-Based Non-Laser Surgical Instrument | Powered Light-Based Non-Laser Surgical Instrument | Light Based Over the Counter Wrinkle Reduction | BlueControl: Same
faceLITE: Different due to indication  |
|  Device product code | ONE | ONE | OHS | BlueControl: Same
faceLITE: Different code due to indication  |
|  Regulation | 878.4810
Laser surgical instrument for use in general and plastic surgery and in dermatology. | 878.4810
Laser surgical instrument for use in general and plastic surgery and in dermatology. | 878.4810
Laser surgical instrument for use in general and plastic surgery and in dermatology. | BlueControl: Same
faceLITE: Same  |

Page 2 of 10

{6}

handLITE LED Device

510(k) Summary

i

ismart

development

ismart.health

|  Description | Subject device
handLITE (TN19S) | Predicate
Philips BlueControl
K171055 | Predicate
faceLITE LED mask
K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Device classification | Class II | Class II | Class II | BlueControl: Same
faceLITE: Same  |
|  Prescription/OTC | Prescription use | Prescription use | Over the counter | BlueControl: Same
faceLITE: Different  |
|  Use environment | Home use | Home use | Home use | BlueControl: Same
faceLITE: Same  |
|  Intended user | Lay user | Lay user | Lay user | BlueControl: Same
faceLITE: Same  |
|  Intended patient population | Adults with mild to moderate contact dermatitis on the hands or wrists | Adults over the age of 18 with mild to moderate psoriasis vulgaris | Adults with facial wrinkles | BlueControl: Similar
faceLITE: Similar  |

Page 3 of 10

{7}

handLITE LED Device

510(k) Summary

^{}[]

|  Description | Subject device handLITE (TN19S) | Predicate Philips BlueControl K171055 | Predicate faceLITE LED mask K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Intended Use and Indications for Use | The handLITE is intended to emit energy in the red and near infrared region of the light spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat contact dermatitis of the hand and wrist. | The Philips BlueControl is intended to emit energy in the blue region of the spectrum, is generally indicated to treat dermatological conditions and specifically indicated to treat mild psoriasis vulgaris. | The faceLITE LED mask is an over-the-counter device that is intended for the use in the treatment of full-face wrinkles. | **BlueControl:** Same use and general indication, similar specific indication
**faceLITE:** Same use, different indications  |
|  Intended location of use | Hand and wrist (dorsal and palmar aspects) | Extremities^{1} | Face | **BlueControl:** Similar, subject device treats specific area that is within the intended areas of predicate
**faceLITE:** Different  |
|  Energy source | Light emitting diodes | Light emitting diodes | Light emitting diodes | **BlueControl:** Same
**faceLITE:** Same  |

1 Treatment area was not specified in K171055, treatment area as per Kleinpenning, et al. (2012), Efficacy of blue light vs. red light in the treatment of psoriasis: a double-blind, randomized comparative study. Journal of the European Academy of Dermatology and Venereology, 219-225 and manufacturer instructions for use.

Page 4 of 10

{8}

handLITE LED Device

510(k) Summary

i

smart

development

smart.health

|  Description | Subject device handLITE (TN19S) | Predicate Philips BlueControl K171055 | Predicate faceLITE LED mask K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Peak wavelength (FWHM) | Red: 630nm ± 10nm
NIR: 830nm ± 10nm | Blue: 452nm ± 7nm | Red: 630nm ± 10nm
NIR: 830nm ± 10nm | BlueControl: Different wavelength/color band
faceLITE: Same  |
|  Intensity | 30mW/cm² total | 40mW/cm² ± 5mW/cm² | 30mW/cm² total | BlueControl: Similar intensity
faceLITE: Same  |
|  Treatment time | 600 seconds (10 minutes) | 1800 seconds (30 minutes) | 600 seconds (10 minutes) | BlueControl: Subject device treatment time is shorter
faceLITE: Same  |

Page 5 of 10

{9}

handLITE LED Device

510(k) Summary

i

ismart

development

ismart.health

|  Description | Subject device handLITE (TN19S) | Predicate Philips BlueControl K171055 | Predicate faceLITE LED mask K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Treatment protocol | 5× weekly, 6 weeks | A full course of the treatment consists of 4 weeks of daily treatment 5-7 times a week followed by 8 weeks of treatment 3 times a week. A single treatment takes approximately 30 minutes with or without interruptions. | 5× weekly, 6 weeks | BlueControl: Subject device protocol is fewer treatments and shorter duration than predicate (30 treatments over 6 weeks vs 52 treatments over 12 weeks)
faceLITE: Same  |
|  Treatment dose | 18J/cm² | 72J/cm² ±9J/cm² | 18J/cm² | BlueControl: Subject device dosage is significantly less than predicate
faceLITE: Same  |
|  Cumulative dose | 540J/cm² | 3744J/cm² | 540J/cm² | BlueControl: Subject device dosage is significantly less than predicate
faceLITE: Same  |

Page 6 of 10

{10}

handLITE LED Device

510(k) Summary

^{}[]

|  Description | Subject device handLITE (TN19S) | Predicate Philips BlueControl K171055 | Predicate faceLITE LED mask K191629 | Significant differences  |
| --- | --- | --- | --- | --- |
|  Timers | Devices uses a timer and software to control treatment duration. | Yes - 30 minutes treatment per area after which device automatically switches off. | Devices uses a timer and software to control treatment duration. | BlueControl: Similar
faceLITE: Same  |
|  Software controlled | Yes | Yes | Yes | BlueControl: Similar
faceLITE: Same  |
|  Patient contacting materials | Transparent silicone inner glove surface and pocket. Nylon, spandex, and polyester yarn attachment strap. | Clear polycarbonate LED cover
OEKO-TEX®-certified fabric device holder and straps | Transparent silicone inner mask surface. Nylon, spandex, and polyester yarn attachment strap. | BlueControl: Different LED housing materials (flexible silicone vs rigid polycarbonate). Both devices use materials listed in Attachment G of the FDA biocompatibility guidance issued 9/8/2023.
faceLITE: Same  |

Page 7 of 10

{11}

handLITE LED Device

510(k) Summary

Ismart Development Ismart,health

The subject device has been subjected to clinical performance testing to assess the safety and effectiveness of the device in treating contact dermatitis on the hand and wrist, as described below.

## Electrical Safety Testing

handLITE was designed to meet applicable industry standards for electromagnetic compatibility (EMC) and electrical safety.

The device relies on standards identified for Medical Electrical Equipment which ensure that the product is safe and effective for its intended use regarding EMC and electrical safety.

handLITE has been tested and found to comply with the following applicable standards:

1. IEC 60601-1: 2005 + AMD1:2012 + AMD2:2020: Medical electrical equipment. General requirements for basic safety and essential performance
2. IEC 60601-1-2:2014+AMD1:2020. Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests for: Home Healthcare Environment
3. IEC TS 60601-4-2:2024 Medical electrical equipment – Part 4-2: Guidance and interpretation – Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
4. IEC 60601-1-11:2015 + IEC 60601-1-11:2015/A1:2021 - Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
5. IEC 60601-2-57:2011. Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
6. IEC 60601-2-83:2019. Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
7. IEC 62133-2:2017. Secondary Cells and Batteries Containing Alkaline or Other Non-Acid Electrolytes – Safety Requirements for Portable Sealed Secondary Cells, and for Batteries Made from Them, for Use in Portable Applications- Part 2: Lithium systems.

## Non-Clinical Performance Testing

### Eye Safety

In addition to the specific electrical safety standards cited above, handLITE has been tested to and complies with IEC 62471:2006, Photobiological safety of lamps and lamp systems.

Page 8 of 10

{12}

Thermal safety

During normal operation with a treatment time of 10 minutes the change in skin temperature from baseline to the end of treatment is +3.4°C for handLITE, with a mean skin temperature of 36.4°C at the end of normal treatment. The glove surface, when in contact with the skin, increases from baseline by +1.8°C, with a mean surface temperature of 28.7°C at the end of normal treatment.

Light characteristics.

Results of testing by the sponsor verified the output characteristics of the device.

Usability/Human factors

The handLITE labeling was subject to label comprehension testing (IEC 62366-1:2015+A1:2020, Application of usability engineering to medical devices)

Nineteen subjects took part in the study, 6 male and 13 female. Twelve subjects (63%) were white, five subjects (26%) were Asian, one subject (5%) was Black or African American, and one subject (5%) was Native Hawaiian or Other Pacific Islander. Eight of the subjects (42%) identified as Hispanic or Latino. Nine subjects (47%) identified English as a second language. The mean age of the study group was 45.42 years of age, with a range of 27 to 61 years old. The average REALM (Rapid Estimate of Adult Literacy in Medicine) words incorrect was 8, with an average REALM score of 8, which represented 7^{th}-8^{th} grade. Two subjects presented with a REALM score of 4^{th}-6^{th} grade, ten with a 7^{th}-8^{th} grade score and the remaining seven had a score of high school.

All 19 subjects were able to complete all tasks by following the device's labeling, including instructions for use. Per the definitions of IEC 62366-1:2015+A1:2020, all tasks were performed within correct or normal use by all subjects.

Observations of tasks that were not completed exactly as specified in the labeling were within the bounds of normal use and did not pose additional hazards.

Clinical Performance Testing

A clinical study was undertaken to support the safety and effectiveness of the handLITE in the treatment of mild-to-moderate contact/irritant dermatitis of the hand.

The objective of the study was to assess the safety and effectiveness of the handLITE in reducing the signs and symptoms of contact dermatitis of the hand. The study included two groups with different treatment courses: Subjects in group 1 treated one hand at a time five times per week, for six weeks (12 weeks total, 30 treatments per hand); Subjects in group 2 treated one hand at a time, seven times per week, for eight weeks (16 weeks total, 56 treatments per hand). Each treatment session included a 10-minute treatment of the dorsal aspect (back) of the hand and a 10-minute treatment of the palmar aspect (palm). After the initial visit, patients conducted treatment sessions at home. The primary effectiveness endpoint was an improvement in the signs and symptoms of contact dermatitis, measured using the Scoring Atopic Dermatitis (SCORAD) index, with the untreated hand during the first treatment cycle used as a control.

{13}

handLITE LED Device 510(k) Summary

iSmart Development

Twenty-five subjects were enrolled in the study and twenty-three completed the full treatment course. Of the two subjects who did not complete the study, one was lost to follow-up after week 9 and the other became pregnant during the study and was discontinued at week 10 per the study's inclusion/exclusion criteria. Available data from both discontinued subjects were included in the evaluation of safety and efficacy.

The average subject age was 44 (range 24 to 63) and the group comprised 9 male and 16 female patients. All Fitzpatrick skin types were represented except type V. 22 subjects identified as white/Caucasian, 1 as Asian, 1 as Black, and 1 as American Indian or Alaska Native; 8 subjects identified as Hispanic or Latino.

There were no adverse events reported during the study. At the end of treatment, the mean improvement in grading of the signs and symptoms of dermatitis was 12.2 for the treated hand and 3.5 for the untreated (control) hand.

Subjects also completed a subjective assessment during the study. By the end of the study, 18 of 23 subjects (78%) reported an improvement in their dermatitis since starting the treatment, 15 of 23 subjects (65%) responded that their dermatitis was less itchy and scaly since starting the treatment, and 18 of 23 subjects responded that they were "not at all unhappy" about their contact dermatitis. All subjects also responded that the handLITE LED device was either "easy" or "extremely easy" to use.

The results of this study demonstrate that a 6-week course of 30 treatments using the handLITE LED device is effective in treating the signs and symptoms of contact dermatitis.

## Conclusions

iSMART Developments LTD has demonstrated that the handLITE LED device has an equivalent intended use, the same generic classification, and the same basic principles and technologies as the predicate devices. Clinical performance testing demonstrates an improvement in the signs and symptoms of contact dermatitis and no safety concerns. Therefore, the handLITE LED device is substantially equivalent to the Philips Blue Control (K171055) and faceLITE LED mask (K191629).

Page 10 of 10

---

**Source:** [https://fda-staging.innolitics.com/device/K250224](https://fda-staging.innolitics.com/device/K250224)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
