K250217 · Agnes Medical Co., Ltd. · GEI · Sep 3, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250217
Device Name
AGNES Ultra
Applicant
Agnes Medical Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AGNES RF (DL) handpiece, AGNES RF (F) handpiece and AGNES RF (B) handpiece of AGNES Ultra are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
AGNES Ultra is a monopolar radiofrequency (RF) electrosurgical device used in dermatology and general surgery for tissue coagulation and hemostasis. The system comprises a main console with an LCD interface, handpieces (DL, F, and B models), disposable micro-needle electrodes, a neutral electrode pad, and a footswitch. The operator selects parameters (output levels 0-20, 2-32W; duration 50-2000ms) via the console. RF energy is delivered through the micro-needle electrode into the patient's tissue. The device includes a negative pressure function (levels 1-3) to assist in tissue contact. The output is an oscillating rectangular wave at 1MHz. The device is operated by healthcare professionals in clinical settings. By delivering controlled RF energy, the device achieves localized coagulation, aiding in hemostasis during surgical procedures.
Clinical Evidence
No clinical data was performed. Evidence consists of bench testing, electrical safety/EMC testing, biocompatibility testing (ISO 10993), and an ex vivo study on liver, skin, and muscle tissue to verify thermal performance in accordance with FDA guidance for electrosurgical devices.
Technological Characteristics
Monopolar RF electrosurgical device; 1MHz output frequency; oscillating rectangular waveform; 0-32W output power; 50-2000ms duration. Materials: Stainless steel (STS 304) electrodes, some with gold coating. Negative pressure function (0 to -45kPa). Connectivity: Standalone console. Sterilization: EO gas for disposable electrodes. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
Indications for Use
Indicated for use in dermatologic and general surgical procedures requiring electrocoagulation and hemostasis. No specific patient population age or gender restrictions are provided.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 3, 2025
Agnes Medical Co., Ltd
% Sanghwa Myung
Regulatory Affair Specialist
E&m
D1474, PyeongCheon Arco Tower, 230, Simin-Daro, Dongan-gu
Anyangsi, Gyeonggi 14067
Korea, South
Re: K250217
Trade/Device Name: AGNES Ultra
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: March 10, 2025
Received: August 4, 2025
Dear Sanghwa Myung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250217 - Sanghwa Myung
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250217 - Sanghwa Myung
Page 3
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.09.03
10:46:19 -04'00'
Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250217
Device Name
AGNES Ultra
Indications for Use (Describe)
AGNES RF (DL) handpiece, AGNES RF (F) handpiece and AGNES RF (B) handpiece of AGNES Ultra are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250217
# 510(k)Summary
# K250217
This summary of 510(K) – safety and effectiveness information are being submitted in accordance with requirements of 21 CFR Part 807.92.
Submitter: AGNES MEDICAL CO., LTD.
(Seohyeon-Dong, 5f Cocoplaza), 20, Seohyeon-Ro 210beon-Gil, Bundang-Gu, Seongnam-Si, Gyeonggido, Korea (ZIP code: 13591)
Telephone: +82-31-8020-9702
Fax: +82-31-701-5162
E-mail: phee3928@agnesmedical.com
Contact: E&M
Regulatory Affair/Sanghwa (Terri) Myung
Telephone: +82-70-7807-0550
FAX: +82-31-388-9263
Cellphone: +82-10-4952-6638
E-mail: mshenmc@gmail.com
Primary Contact: Sanghwa(Terri) Myung
Date 510(k) summary prepared: August 28th 2025
1. Proposed Device Identification
Product Name: RF Electrosurgical Device
Trade Name (Model): AGNES Ultra
Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories
Regulation Number: 21 CFR 878.4400
Device Class: Class II
Product Code: GEI
Review Panel: General & Plastic Surgery
510(k) Summary.
1 / 8 page
{5}
K250217
## 2. Predicate Device Identification
510(k) Number: K242469
Device Name: RFMagik Lite
Manufacturer: AGNES MEDICAL CO., LTD.
Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories
Regulation Number: 21 CFR 878.4400
Device Class: Class II
Product Code: GEI
Review Panel: General & Plastic Surgery
## 3. Reference Device
510(k) Number: K233120
Manufacturer: Chungwoo Co., Ltd.
Device Name: CWM-930S
Common Name: Electrosurgical cutting and coagulation device and accessories
Classification Name: Electrosurgical cutting and coagulation device and accessories
Classification Product Code and Regulation: GEI, 21CFR 878.4400
Device Class: II
## 4. Device Description
AGNES Ultra is used to coagulate the patient's tissue. AGNES Ultra is a medical device combined with RF current, to function as an electrosurgical device for use in dermatology and general surgical procedures. It is possible to select and change modes, parameters, outputs, etc. using the panel on the main body. It consists of the main device, LCD screen, handpieces, single use electrodes, Neutral electrode pad, NE pad cable, food switch.
RF energy handpiece is AGNES RF (DL) handpieces, AGNES RF (F) handpiece and AGNES RF (B) handpiece.
that delivers RF energy through the disposable electrode in the handpiece.
## 5. Indication for use
AGNES RF (DL) handpiece, AGNES RF (F) handpiece and AGNES RF (B) handpiece of AGNES Ultra are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
## 6. Summary of Technological Characteristics Comparison
The following comparison table is presented to demonstrate substantial equivalence.
1) Predicate Device
510(k) Summary.
{6}
K250217
| Descriptive Information | | Subjective Device | Predicate Device 1 |
| --- | --- | --- | --- |
| Manufacturer | | AGNES MEDICAL CO., LTD | AGNES MEDICAL CO., LTD |
| Device Name | | AGNES Ultra | RFMagik Lite |
| 510(k) number | | K250217 | K242469 |
| Classification Product Code / Regulatory Number | | GEI / 878.4400 | GEI / 878.4400 |
| Regulatory Class | | II | II |
| Indications for Use | | AGNES RF (DL) handpiece, AGNES RF (F) handpiece and AGNES RF (B) handpiece of AGNES Ultra are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. | RO handpiece and AGNES RF handpiece of RFMagik Lite are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. |
| Prescription or OTC | | Prescription | Prescription |
| Mode of Operation | | Monopolar | Monopolar |
| Waveform | | Oscillating rectangular wave | Oscillating rectangular wave |
| Output frequency | | 1MHz | 1MHz |
| Output operating time | | Min 50ms / Max 2,000ms | Min 50ms / Max 2,000ms |
| Output power levels | | 0~20 levels (0, 2 to 32W) | 0~20 levels (0, 2 to 32W) |
| Max. output power | | 0, 2 ~ 32W at 200 ohm | 0, 2 ~ 32W at 200 ohm |
| Negative Pressure | | Off, Lv.1~Lv.3
- Off: 0kPa
- Lv.1: -25kPa
- Lv.2: -35kPa
- Lv.3: -45kPa | N/A |
| Electrical Safety Standards Complies | | with IEC 60601-1, IEC 60601-1-2, IEC60601-2-2 | with IEC 60601-1, IEC 60601-1-2, IEC60601-2-2 |
| Connected Handpiece | | Four Type of Handpiece
1) AGNES RF (DL) Handpiece
2) AGNES RF (F) Handpiece or AGNES RF (B) Handpiece | Four Type of Handpiece
1) RO Handpiece
2) AGNES RF Handpiece |
| Connect Patient contact Electrode Tip | Sterilization | EO Gas | EO Gas |
| | Shelf Life | 3 Years | 3 Years |
| | - | 1) K223805
(GA-B, GA-C, GA-E, GA-I, GA-S, GA- | 1) K171707
(GA-B, GA-C, GA-E, GA-I, GA-S, GA- |
510(k) Summary.
3 / 8 page
{7}
K250217
| | | W3A, GA-F3A, GA-IL, GA-SL, GA-W3B, GA-F1A, AG-CN18G, AG-CN19G, AG-CN20G, AG-CN21G, AG-CN23G) | W3A, GA-F3A, GA-IL, GA-SL, GA-W3B, GA-F1A) |
| --- | --- | --- | --- |
| | | 2) K242469 (RO-25S, RO-25D) | 2) K242469 (RO-25S, RO-25D) |
| | | 3) RO-25S(G), RO-25D(G), ROS-25S(G), ROS-25D(G) | |
| | | 4) AN-B, AN-C, AN-E, AN-I, AN-S, AN-W3A, AN-F3A, AN-IL, AN-SL, AN-W3B, AN-F1A, AN-F3B, AN-B3A, AN-B4A | |
| Electrode type | | Micro needle type | Micro needle type |
| Needle thickness | 1) GA-B, GA-C, GA-E, GA-I, GA-S, GA-W3A, GA-F3A, GA-IL, GA-SL, GA-W3B, GA-F1A : 0.2mm, 0.25mm | | |
| | | 2) AG-CN18G, AG-CN19G, AG-CN20G, AG-CN21G, AG-CN23G : 0.64mm(23G) ~ 1.28mm (18G) | 1) GA-B, GA-C, GA-E, GA-I, GA-S, GA-W3A, GA-F3A, GA-IL, GA-SL, GA-W3B, GA-F1A : 0.2mm, 0.25mm |
| | | 2) RO-25S, RO-25D : 0.2mm | |
| | | 3) RO-25S(G), RO-25D(G), ROS-25S(G), ROS-25D(G) : 0.2mm | 2) RO-25S, RO-25D : 0.2mm |
| | | 4) AN-B, AN-C, AN-E, AN-I, AN-S, AN-W3A, AN-F3A, AN-IL, AN-SL, AN-W3B, AN-F1A, AN-F3B, AN-B3A, AN-B4A : 0.2mm, 0.25mm, 0.3mm | |
| Max insertion depth | | 5mm | 5mm |
| Min insertion depth | | 0.4mm | 0.4mm |
| Treatment Area | | RO-25S: 8.5mm × 9.3mm
RO-25D: 8.5mm × 9.3mm
RO-25S(G) : 8.5mm × 9.3mm
RO-25D(G) : 8.5mm × 9.3mm | RO-25S: 8.5mm × 9.3mm
RO-25D: 8.5mm × 9.3mm |
510(k) Summary.
4 / 8 page
{8}
K250217
2) Reference Device
| | Subject device | Reference Device 1 |
| --- | --- | --- |
| Manufacturer | AGNES Medical Co., Ltd. | Chungwoo Co., Ltd. |
| Device Name | AGNES Ultra | CWM-930S |
| 510(k) number | K250217 | K233120 |
| Classification Name | Electrosurgical cutting and coagulation device and accessories (21 CFR Part 878.4400) | Electrosurgical cutting and coagulation device and accessories (21 CFR Part 878.4400) |
| Product code | GEI | GEI |
| Indication for use | AGNES RF (DL) handpiece, AGNES RF (F) handpiece and AGNES RF (B) handpiece of AGNES Ultra are intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. | This device is intended for use in dermatologic and general surgical procedures for electro coagulation. |
| Mode of operation | Monopolar | Bipolar and Monopolar |
| Output energy | Radiofrequency | Radiofrequency |
| Output Frequency | 1MHz | 1MHz, 2 MHz |
| Rated Input | AC 100-240V, 50/60Hz | AC 100-240V, 50/60Hz |
| Output Power | Max. 32W | 1MHz: 35W± 20% (500ohm)
2MHz: 25W± 20% (500ohm) |
| RF Intensity | 1~20 Level | 1~10 Level |
| RF Duration | 50ms ~ 2000ms | 10ms ~ 900ms |
| Needles | RO-25S(G)
RO-25D(G)
ROS-25S(G)
ROS-25D(G)
RO-25S, RO-25D | 25pin (Insulated, Bipolar)
49pin (Non-insulated, Bipolar)
14pin (Non-insulated, Bipolar)
1pin (Insulated, monopolar) |
510(k) Summary.
{9}
K250217
| Materials | RO-25S/D: Stainless Steel (STS 304)
The others: Stainless Steel (STS 304) + gold coating | Gold coating |
| --- | --- | --- |
| Needle length | - RO-25S & RO-25S(G) & ROS-25S(G): 0.4 mm
- RO-25D(G) & ROS-25D(G): 0.4 mm & 0.8 mm
- RO-25D: 0.4 mm & 1.2 mm | Bipolar: 0.5mm~3.5mm(14pin, 25pin, 49pin) / 0.5mm~7.0mm(36pin) ±15% / Adjustment unit 0.1mm
monopolar: 0.5mm~5.0mm (0.1mm step) |
| Number of pins | 25 pin | Bipolar: RF Microneedle 14pin, 25pin, 36pin, 49pin
Monopolar: 1pin |
| Needle application range | RO-25S: 8.5mm × 9.3mm
RO-25D: 8.5mm × 9.3mm
RO-25S(G) : 8.5mm × 9.3mm
RO-25D(G) : 8.5mm × 9.3mm
ROS-25S(G): 8.5mm × 9.3mm
ROS-25D(G): 8.5mm × 9.3mm | 1pin(1.0 x 1.0mm),
25Pin, 49Pin(7.8 x7.8mm),
14pin(7.8 x 2.0mm),
36pin(11x11mm) |
| Negative Pressure function | Off: 0kpa
Lv.1: -25Kpa (= -0.25bar)
Lv.2: -35kPa (= -0.35bar)
Lv.3: -45kPa (= -0.45bar)
(Only ROS-25S(G), ROS-25D(G)) | Max 0.78 bar |
- Discussion
The subject devices are substantially equivalent to the predicate devices with respect to indications for use, technology and construction. The differences between the predicate devices and the subject devices are minor and any risks have been mitigated through testing. The different characteristics do not raise different questions of safety and effectiveness.
7. Non-Clinical Testing
1) Electrical Testing and EMC Testing.
The electrical and EMC tests were performed with the compatible radio frequency electrosurgical devices in accordance with the FDA recognized standards,
- AAMI ES60601-1:2005/AMD2:2021, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
510(k) Summary.
{10}
K250217
- IEC60601-2-2:2017/AMD1:2023, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of Radio frequency surgical equipment and Radio frequency.
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- TR IEC 60601-4-2: 2024 Medical electrical equipment. Guidance and interpretation. Electromagnetic immunity: performance of medical electrical equipment and medical electrical system
2) Biocompatibility
The biocompatibility tests were performed in accordance with the FDA recognized standards,
ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Test for irritation
3) Sterility
The EO gas sterilization was verified and validated in accordance with the FDA recognized standards
ISO 11135: 2014, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
ISO 11138-1:2017, Sterilization of health care products - Biological Indicators - Part 1: General requirements
ISO 11138-2: 2017, Sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes
ISO 10993-7: 2008, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
4) Shelf life
ISO 11607-1: 2019, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
ISO 11607-2:2019, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1929-15, Standard test method for detecting seal leaks in porous medical packaging by dye Penetration
- ASTM F88M-21, Standard test method for seal strength of flexible barrier materials
5) Performance Testing
AGNES MEDICAL conducted bench testing to assure that the operates safely and within the predefined design specifications.
510(k) Summary.
{11}
K250217
## 6) Ex Vivo Study
Ex Vivo testing conducted on three types of tissue – Liver, skin and Muscle under GLP Thermal testing in accordance with FDA’s "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", August 15, 2014
## 8. Clinical Testing
Clinical testing is not a requirement and has not been performed.
## 9. Conclusion
The AGNES Ultra and the predicate device have same indication for use, Technology, mechanism of actions, operational principles and performance for the proposed indications for use.
The non-clinical data for proposed device, including biocompatibility, bench testing, hardware, and software documentation shows that the device should perform as intended in the specified use. In addition, the Electromagnetic Compatibility and Electrical Safety testing shows that the device is safe. The comparison between the subject devices and the predicate devices shows that the indication for use and technological characteristic is substantially equivalent although there are some differences, the performance test reports are supported to the substantial equivalence of the subject device, the performance test reports are provided to demonstrate substantial equivalence of the subject devices. Therefore, we conclude that the different characteristics do not raise different questions of safety and effectiveness, and thus the subject devices are substantially equivalent to the predicate devices.
510(k) Summary.
8 / 8 page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.