The LapBox Power Tissue Containment System is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50.
Device Story
Single-use sterile device; double-wall inflatable polyurethane chamber mounted on insertion shaft. Procedure: shaft inserted into abdominal cavity; chamber deployed; organ placed inside; chamber inflated via external handpump; sleeve exteriorized; port placed over sleeve; power morcellator inserted through port for tissue morcellation; chamber deflated and removed post-procedure. Used in gynecologic laparoscopic surgery; operated by trained physicians. Maintains pneumoperitoneum during morcellation. Provides physical barrier to contain tissue; prevents potential spread of unsuspected malignancy during morcellation. Output: contained tissue specimen for removal. Benefits: enables contained power morcellation; reduces risk of tissue spillage compared to uncontained procedures.
Clinical Evidence
Bench testing only. No clinical data. Evidence includes bacterial immersion (Brevundimonas Diminuta), dimensional verification, pressure relief valve testing, dye leak testing, puncture force comparative testing, bond strength, burst pressure, and usability studies (n=31, n=6, n=15) in simulator/porcine models. All tests met prespecified acceptance criteria.
Indicated for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery for isolation and containment of benign tissue resected during power morcellation. Compatible with electromechanical power morcellators (15-18 mm shaft diameter, 135-180 mm working length). For women with fibroids, indicated only for pre-menopausal women under age 50. Contraindicated for tissue known or suspected to contain malignancy, or for patients who are post-menopausal, over 50, or candidates for en-bloc removal.
Regulatory Classification
Identification
A gynecologic laparoscopic power morcellation containment system is a prescription device consisting of an instrument port and tissue containment method that creates a working space allowing for direct visualization during a power morcellation procedure following a laparoscopic procedure for the excision of benign gynecologic tissue that is not suspected to contain malignancy.
Special Controls
In combination with the general controls of the FD&C Act, the Gynecologic Laparoscopic Power Morcellation Containment System is subject to the following special controls:
- 1. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 2. Device components that are labeled sterile must be validated to a sterility assurance level of 10-°.
- 3. Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life.
- 4. Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
- a. Demonstration of the device impermeability to tissue, cells and fluids.
- b. Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum.
- c. Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera.
- d. Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system.
- Demonstration that intended users can adequately deploy the device, morcellate a e. specimen without compromising the integrity of the device and remove the device without spillage of contents.
- 5. Training must be developed and validated to ensure users can follow the instructions for use.
- 6. Labeling must include:
- . Contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy.
- Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: peri- or post-menopausal; or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision.
- . The following boxed warning: "Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk."
- Statement limiting use of device to physicians who have completed the training ● program.
- An expiration date or shelf life. ●
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible;
(2) Device components that are labeled sterile must be validated to a sterility assurance level of 10
−6 ;(3) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the intended shelf life;
(4) Non-clinical performance data must demonstrate that the device meets all design specifications and performance requirements. The following performance characteristics must be tested:
(i) Demonstration of the device impermeability to tissue, cells, and fluids;
(ii) Demonstration that the device allows for the insertion and withdrawal of laparoscopic instruments while maintaining pneumoperitoneum;
(iii) Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera;
(iv) Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system; and
(v) Demonstration that intended users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device, and remove the device without spillage of contents;
(5) Training must be developed and validated to ensure users can follow the instructions for use; and
(6) Labeling must include the following:
(i) A contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy;
(ii) Unless clinical performance data demonstrates that it can be removed or modified, a contraindication for removal of uterine tissue containing suspected fibroids in patients who are: Peri- or postmenopausal, or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision;
(iii) The following boxed warning: “Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk.”
(iv) A statement limiting use of device to physicians who have completed the training program; and
(v) An expiration date or shelf life.
LapBox Tissue Containment Removal System (K221365)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 15, 2025
ARK Surgical
% Bosmat Friedman-Cox
Regulatory Consultant
ProMedoss, Inc.
6026 Beech Cove Ln.
Charlotte, North Carolina 28269
Re: K250212
Trade/Device Name: LapBox Power Tissue Containment System
Regulation Number: 21 CFR 884.4050
Regulation Name: Gynecologic Laparoscopic Power Morcellation
Containment System
Regulatory Class: II
Product Code: PMU
Dated: January 23, 2025
Received: January 24, 2025
Dear Bosmat Friedman-Cox:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250212 - Bosmat Friedman-Cox
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250212 - Bosmat Friedman-Cox
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250212
Device Name
LapBox Power Tissue Containment System
Indications for Use (Describe)
The LapBox Power Tissue Containment System is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
ARK
SURGICAL
LapBox Power Tissue Containment System
K250212
Traditional 510(k)
510(k) Summary
Page 1 of 7
510(k) Summary [Traditional 510(k)]
LapBox Power Tissue Containment System
510(k) Number K250212
1 SUBMITTER
Applicant's Name:
ARK Surgical, Ltd.
Stav Tori, CEO
13 Wadi El Hadj
Nazareth, Israel
stav@ark-surgical.com
Primary Contact:
Bosmat Friedman
Regulatory Affairs Consultant
6026 Beech Cove Ln.
Charlotte, NC 28269
Phone: 980-308-1636
bosmat.f@promedoss.com
Date Prepared: April 10, 2025
Trade Name:
LapBox Power Tissue Containment System
Classification Code:
Device: Containment System, Laparoscopic Power
Morcellation, With Instrument Port
Product Code: PMU
Regulation No: 884.4050
Class: 2
Review Panel: Obstetrics/Gynecology
2 PREDICATE DEVICE
Primary Predicate:
- PneumoLiner, manufactured by Advanced Surgical Concepts Ltd., K192898; Product
Code: PMU. The predicate device has not been subject to a design-related recall.
Reference device:
- LapBox Tissue Containment Removal System, manufactured by ARK Surgical, cleared
under K221365; Product Code: GCJ.
3 DEVICE DESCRIPTION
The LapBox Power Tissue Containment System is a single use sterile device. It is comprised of a
double wall inflatable polyurethane chamber which is mounted on an insertion shaft and is
provided with two port sizes. Once the shaft is inserted to the abdominal cavity, the chamber is
deployed and the organ to be morcellated is placed within the chamber. The chamber is then
{5}
ARK SURGICAL
LapBox Power Tissue Containment System K250212
Traditional 510(k) 510(k) Summary Page 2 of 7
inflated using an external handpump and the sleeve of the chamber is exteriorized. The selected port is then placed over the sleeve in the incision site and the organ can be morcellated using a power morcellator. Once morcellation is complete, the port is removed, and the chamber is deflated and removed from the patient.
# 4 INDICATIONS FOR USE
The LapBox Power Tissue Containment System is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50.
# 5 SUBSTANTIAL EQUIVALENCE
The subject and predicate devices are both single use tissue bags intended to retrieve and contain specimen during power morcellation.
The following table provides a comparison with the predicate:
| Feature | LapBox Power Tissue Containment System | PneumoLiner (K192898) |
| --- | --- | --- |
| Reg. Number | 884.4050 | 884.4050 |
| Product Code | PMU | PMU |
| Indication for Use | The LapBox device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. The LapBox is compatible with electromechanical laparoscopic power morcellators that are between 15 mm and 18 mm in shaft outer diameter and 135 mm and 180 mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. When used in women with fibroids, the LapBox Power Tissue Containment System is for women who are pre-menopausal and under age 50. | The PneumoLiner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi-site laparoscopic surgery during power morcellation and removal. When used in women with fibroids, the PneumoLiner is for women who are pre-menopausal and under age 50. The PneumoLiner is compatible with bipolar or electromechanical laparoscopic power morcellators that are between 15mm and 18mm in shaft outer diameter and 135mm and 180mm in shaft working length and which have an external component that allows for the proper orientation of the laparoscope to perform a contained morcellation. |
{6}
ARK SURGICAL
LapBox Power Tissue Containment System K250212
Traditional 510(k) 510(k) Summary Page 3 of 7
| Feature | LapBox Power Tissue Containment System | PneumoLiner (K192898) |
| --- | --- | --- |
| Principle of Operation | Inserted through umbilical incision to the abdominal cavity, organ placed in chamber, hand pump is connected, and chamber is inflated, , incision size is increased (if needed), and chamber sleeve is exteriorized, port is placed, power morcellation is performed and at completion chamber is deflated, port is removed and chamber extracted. | Incision is made and retractor is placed over incision; boot assembly placed over retractor; bag inserted through boot assembly; organ placed in bag and the boot assembly removed; bag exteriorized and boot assembly put back in place; bag inflated and organ is morcellated; once morcellation is complete, all instruments including boot assembly are removed, and the bag is removed. |
| Component allowing tools and morcellator access | Use of either a medium or small size designated port | Boot assembly |
| Pneumoperitoneum/insufflation | Pneumoperitoneum is maintained in the abdomen with LapBox during the morcellation procedure | Pneumoperitoneum is maintained inside the PneumoLiner during the procedure |
| Umbilical incision length | 25-30 mm | 20-25 |
| Chamber/bag and Port/Guard Materials | TPU (Polyurethane) film, Nylon reinforced fabric and polyurethane port | TPU (Polyurethane) film, Nylon reinforced fabric and polyurethane and plastic port |
| Wall material thickness (TPU thickness) | 80 μm | 42 μm |
| Chamber size | 140mm X 320 mm | 350mm X 500mm |
| Chamber/bag pattern | None | Grid |
| Single use | Yes | Yes |
| Sterility | EtO | Gamma |
| Shelf-Life | 18 months | 36 months |
The differences in the indications for use do not represent a new intended use.
The differences in technological characteristics do not raise different questions of safety or effectiveness.
{7}
ARK SURGICAL
| LapBox Power Tissue Containment System | | K250212 |
| --- | --- | --- |
| Traditional 510(k) | 510(k) Summary | Page 4 of 7 |
## 6 SUMMARY OF NON-CLINICAL TESTING
The following non-clinical testing is provided in support of our substantial equivalence claim and to demonstrate compliance with the special controls associated with this device type:
## Sterilization and Shelf-Life:
Sterilization validation was performed in accordance with ISO 11135, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
Shelf-life of 18 months was validated on device samples following accelerated aging and simulated transit conditioning. Testing evaluated packaging integrity and device functionality
## Biocompatibility:
The biocompatibility evaluation for the LapBox Power Tissue Containment System device was conducted in accordance with the FDA biocompatibility guidance “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process” issued September 2023 and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The LapBox is considered an externally communicating device, tissue contacting with limited duration.
The following biocompatibility tests were conducted:
- Cytotoxicity per ISO 10993-5
- Sensitization per ISO 10993-10
- Intracutaneous irritation per ISO 10993-10
- Acute Systemic Toxicity per ISO 10993-11
- Pyrogenicity testing per USP <151>
Testing demonstrated that the device was non cytotoxic, non-sensitizer, non-irritant, not systemically toxic and non-pyrogenic.
## Non-Clinical Performance Testing (Bench):
Non-clinical performance testing on LapBox Power Tissue Containment System was conducted in accordance with the FDA guidance document “Non-Clinical Performance Assessment of Tissue Containment Systems Used During Power Morcellation Procedures” issued May 26, 2023.
Testing included:
- Bacterial Immersion Test: testing included 32 device samples aged beyond shelf life and previously exposed to simulated use; both positive and negative controls were used. Samples were then disinfected and sterilized prior to immersion. The tested chambers were inflated and immersed to assess for impermeability to Brevundimonas Diminuta. After the Immersion Test the entire volume of the TSB within each chamber was membrane filter plated to enumerate the presence of a single colony forming unit of the bacterial species. The test samples did not show any evidence of growth of Brevundimonas Diminuta. The positive and negative controls performed as intended.
- Dimensional Verification Tests: supplemental test to the one provided in K221365 (referenced below); a total of 30 non aged units were tested for the dimensions of the following: proximal handle outer diameter; inner tube inner diameter: outer tube inner diameter; and inflation chamber inner diameter. All samples met the prespecified
{8}
ARK SURGICAL
LapBox Power Tissue Containment System K250212
Traditional 510(k) 510(k) Summary Page 5 of 7
acceptance criteria.
- Pressure Relief Valve Testing: a total of 30 aged and 30 non aged pressure relief valves were tested to ensure they would relieve pressure above the set target. All samples met the prespecified acceptance criteria.
- Design and Performance Validation Test: a total of 30 aged and 27 no-aged units were tested by 2 trained surgeons using two different morcellators in a simulator with bovine tongue. All samples passed a dye leak test. Design changes were implemented following this test which were validated in supplemental tests.
- Puncturing Force Comparative Test: testing was conducted on double layers and a single layer of the LapBox chamber material with a variety of morcellator and graspers and compared to the results obtained with the predicate. Testing utilized 72 predicate material samples and 72 aged and 72 non aged LapBox material samples. The results obtained with LapBox supported non-inferiority when compared to the predicate.
- Bond Strength Tests: supplemental test to the one provided in K221365 (referenced below); 30 aged units were tested to evaluate the air tube bond connection as well as the handpump assembly. All samples met the prespecified acceptance criteria.
- Burst Pressure Evaluation: supplemental test to the one provided in K221365 (referenced below); 10 non-aged units with the modified new air tube were tested. All samples met the prespecified acceptance criteria.
- Inflation Pressure Range: a customized jig was developed to evaluate the pressure range required for the LapBox chamber to maintain its shape when subjected to insufflation. One LapBox unit was used. The testing demonstrated that the chamber will maintain its shape at a pressure below that relieved by the pressure relief valve.
- Comparison Force Test: a customized jig was developed to evaluate and compare the outward force exerted by the LapBox and by the predicate. One predicate and one LapBox were utilized in this test. The results demonstrate that when used in accordance with device labeling, a uniform distribution of force by LapBox does not pose a greater risk of compression on adjacent organs when compared to the predicate.
The following additional bench tests previously submitted under K221365 for the reference device were also considered:
- Corrosion Resistance: 30 units of the 2 different metal components were used; samples underwent double sterilization and 18 months accelerated aging prior to testing. No corrosion was noted on any samples.
- Burst Pressure Evaluation: a total of 30 aged and 30 non-aged units were tested to evaluate the burst pressure of the LapBox chamber. All tested units met the prespecified acceptance criteria.
- Bond Strength Tests: 30 non-aged units and 30 aged units were tested. Tensile strength of the following bonds were assessed: chamber assembly; air tube-chamber connection; delivery system-chamber connection; and the handpump assembly. All samples met the prespecified acceptance criteria.
- Dimensional Verification Tests: a total of 36 non aged units were tested for the dimensions for the following: shaft effective length; inner sleeve height; chamber double wall height; chamber outer diameter; and chamber opening. All samples met the prespecified acceptance criteria.
- Viral Penetration ASTM Method F 1671: 30 material samples that have undergone
{9}
ARK SURGICAL
| LapBox Power Tissue Containment System | | K250212 |
| --- | --- | --- |
| Traditional 510(k) | 510(k) Summary | Page 6 of 7 |
accelerated aging equivalent to 12 months were tested. All samples met the prespecified acceptance criteria.
- Maximum organ size testing: 2 aged devices used to evaluate the maximum organ size compatible with the LapBox.
## Training Validation and Usability Studies:
Multiple studies were conducted to evaluate the usability of the LapBox System and the training program developed by ARK Surgical.
- An initial usability study was conducted which included 31 physicians of varying levels of experience performing a total of 40 procedures in a simulator with bovine tongue. Following training, participants performed the procedure. No device samples had an observed leakage of dye following use. There were minor device issues noted that were addressed via manufacturing and labeling modifications.
- A supplementary usability study in a simulator was conducted, in which 6 physicians conducted a total of 33 procedures. These units were used in the bacterial immersion testing discussed above. There was no evidence of any device leak.
- Training validation was conducted in a porcine model with bovine tongue. A total of 16 physicians of varying levels of experience completed 32 morcellation procedures. The participants underwent the training and then had a 1 hour break prior to performing the procedure. Following the procedure, device samples were subject to a dye leak test. No samples leaked. Minor issues were noted with respect to adherence to the instructions for use. The training and instructions for use (IFU) were revised to address the noted issues and a follow-up usability study was conducted which included 15 physicians who performed 30 power morcellation procedures in a simulator. This study demonstrated that the revisions made to the training and IFU were effective, ensuring the device's safe and effective use in clinical settings.
## Additional Animal Studies:
- Performance Characterization Animal In-Vivo Study (Part 1 and 2): The purpose of these studies was to define the performance characteristics of the LapBox System in a porcine model. The first study utilized one animal and evaluated 3 devices, the second study utilized one animal and evaluated 5 devices.
## 7 CONFORMANCE TO SPECIAL CONTROLS
LapBox Power Tissue Containment System conforms to the following special controls:
1. The patient-contacting components of the device have been demonstrated to be biocompatible.
2. Device components that are labeled sterile have been validated to a sterility assurance level of $10^{-6}$.
3. Performance data support shelf-life by demonstrating continued sterility of the device and sterile components, package integrity, and device functionality over the intended shelf life.
{10}
ARK SURGICAL
| LapBox Power Tissue Containment System | | K250212 |
| --- | --- | --- |
| Traditional 510(k) | 510(k) Summary | Page 7 of 7 |
4. Non-clinical performance data demonstrate that the device meets design specifications and performance requirements. The following performance characteristics have been tested.
a. Demonstration of the device impermeability to tissue, cells and fluids.
b. Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum.
c. Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera.
d. Demonstration that intended laparoscopic instruments and morcellators do not compromise the integrity of the containment system.
e. Demonstration that intended users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device and remove the device without spillage of contents.
5. Training has been developed and validated to ensure users can follow the instructions for use.
6. Labeling includes:
- Contraindication for use in gynecologic surgery in which the tissue to be morcellated is known or suspected to contain malignancy.
- Contraindication for removal of uterine tissue containing suspected fibroids in patients who are: peri- or post-menopausal; or candidates for en bloc tissue removal, for example, through the vagina or via a mini-laparotomy incision.
Note: The language of this contraindication was revised during the review of the predicate device for clarity. The labeling for the LapBox Power Tissue Containment System is consistent with the predicate device and identifies the contraindication as follows: "Do not use for removal of uterine tissue containing suspected fibroids in patients who are: post- menopausal or over 50 years of age; or candidates for en-bloc tissue removal through the vagina or via a mini-laparotomy incision."
- The following boxed warning: "Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. Uterine tissue may contain unsuspected cancer. The use of laparoscopic power morcellators during fibroid surgery may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk."
- Statement limiting use of device to physicians who have completed the training program.
- An expiration date or shelf life.
# 8 CONCLUSION
The LapBox Power Tissue Containment System's performance testing support the substantial equivalence determination as well as compliance with the special controls. The LapBox Power Tissue Containment System is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.