eufy Wearable Breast Pump S1 (T8D02); eufy Wearable Breast Pump E10 (T8D03)
Applicant
Anker Innovations Limited
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Aug 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eufy Wearable Breast Pump S1 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user. The eufy Wearable Breast Pump E10 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
Device Story
The eufy Wearable Breast Pump (models S1 and E10) is an electrically powered, wearable, single-user breast pump for home use. It features a closed-system design with an anti-backflow diaphragm to separate the air-pumping system from the milk-flow system. The device operates in two modes (stimulation and expression) with 7 adjustable vacuum levels (-75 to -292.5 mmHg) and varying cycle speeds (39-110 cycles/minute). The S1 model includes a heating function and magnetic charging, while the E10 uses Type-C charging. Users operate the device via onboard controls or an optional smartphone app, which allows for mode/vacuum selection, session history tracking, and milk volume display. The pump is powered by an internal 3.7V 1460mAh lithium-ion polymer battery. By providing controlled suction, the device assists lactating women in expressing and collecting milk, facilitating breastfeeding support.
Clinical Evidence
Bench testing only. Testing included electrical safety (ANSI AAMI ES60601-1, IEC 60601-1-2, HA60601-1-11), battery safety (IEC 62133-2), biocompatibility (ISO 10993-5, -10, -23), and performance testing (vacuum pressure, cycle speed, backflow protection, battery capacity, and use life). Packaging was validated per ISTA 3A:2018. All specifications were met.
Technological Characteristics
Powered by 3.7V 1460mAh Li-ion battery. Materials: Polypropylene (milk container), Silicone (flange), ABS plastic (housing). Closed-system with anti-backflow diaphragm. Vacuum range: -75 to -292.5 mmHg. Cycle speeds: 39-110 cycles/min. Connectivity: Optional smartphone app via Bluetooth. S1 includes heating pins and magnetic charging; E10 uses Type-C charging. Embedded software controls pump functions.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
Anker Innovations Limited
% Tracy Chen
Manager
Shenzhen Joyantech Consulting Co., Ltd.
1713A, Block A, Zhongguan Times Square,
Liuxian Avenue Xili Town
Shenzhen, Guangdong 518000
CHINA
Re: K250207
Trade/Device Name: eufy Wearable Breast Pump S1 (T8DO2); eufy Wearable Breast Pump E10 (T8DO3)
Regulation Number: 21 CFR 884.5160
Regulation Name: Powered Breast Pump
Regulatory Class: II
Product Code: HGX
Received: July 11, 2025
Dear Tracy Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250207 - Tracy Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250207 - Tracy Chen
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250207
Device Name
eufy Wearable Breast Pump S1 (T8D02)
eufy Wearable Breast Pump E10 (T8D03)
Indications for Use (Describe)
The eufy Wearable Breast Pump S1 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
The eufy Wearable Breast Pump E10 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
K250207
Page 1 of 5
# 510(k) Summary- K250207
## 1. Submitter Information
### 1.1 Applicant information
| Applicant Name | Anker Innovations Limited |
| --- | --- |
| Address | Unit 56, 8th Floor, Tower 2, Admiralty Centre, 18 Harcourt Road, Hong Kong |
| Tel. | +86 166 2079 6266 |
| Contact person | Darren Li |
| E-mail | darren.li@anker.com |
| Date Prepared | August 9, 2025 |
### 1.2 Contact Information
| | Shenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square,
Liuxian Avenue, Xili Town, Nanshan District, Shenzhen,
Guangdong Province, China
+86 755-86069197 |
| --- | --- |
| Tel. | Tracy |
| Contact person | tracy@cefda.com; grace@cefda.com |
| E-mail | http://www.cefda.com |
| Website | |
## 2. Device information
| Trade name | eufy Wearable Breast Pump S1 (T8DO2)
eufy Wearable Breast Pump E10 (T8DO3) |
| --- | --- |
| Common Name | Powered breast pump |
| Regulation Number | 21 CFR 884.5160 |
| Regulation Name | Powered breast pump |
| Classification Panel | Obstetrics/Gynecology |
| Product code | HGX (Pump, Breast, Powered) |
| Regulatory Class | Class II |
## 3. Predicate Device Information
| Trade Name | Perifit Pump |
| --- | --- |
| 510(k) Number | K231785 |
| Product Code | HGX |
| Manufacturer | X6 Innovations |
{5}
K250207
Page 2 of 5
The predicate device has not been subject to a design-related recall.
## 4. Device Description
The eufy Wearable Breast Pump is an electrically powered wearable single breast pump. There are two models, eufy Wearable Breast Pump S1 (T8DO2) and eufy Wearable Breast Pump E10 (T8DO3). S1 consists of a Pump Hub, Diaphragm, Milk Container, Pouring Spout, Heating Pins, Flange, Duckbill Valve and charging pins. E10 consists of Pump Hub, Diaphragm, Milk Container, Pouring Spout, Flange and Duckbill Valve.
The differences between S1 and E10 are as follows:
1) E10 has no heating function (S1 has a heating function);
2) The charging method is different (S1 for magnetic charging, E10 for Type-C charging).
The eufy Wearable Breast Pump is powered by an internal rechargeable lithium-ion polymer battery (3.7V 1460mAh). The eufy Wearable Breast Pump cannot be operated while it is being charged.
The eufy Wearable Breast Pump operates on embedded software, and software updates by end-users are supported. The subject device is intended for use in a home healthcare environment. An optional smartphone app may be used with the Breast Pump. The app may be used to control the operation of The Breast Pump (select mode and vacuum level, start and stop pumping), save pumping session history and display the quantity (in ml) of milk in the bottle on the user's smartphone. Use of the app is not mandatory to operate the Breast Pump.
The eufy Wearable Breast Pump is a closed-system breast pump with an anti-backflow diaphragm which physically separates the air-pumping system from the milk-flow system.
The eufy Wearable Breast Pump allows the user to power the breast pump on and off, and to operate the breast pump in two modes (stimulation, expression) and to control the vacuum settings within these modes. Each mode consists of 7 vacuum levels, capable of providing suction from -75 to -292.5mmHg, with cycle speeds ranging from 71 to 110 cycles/minute in stimulation mode and 39 to 60 cycles/minute in expression mode.
## 5. Indication for Use
The eufy Wearable Breast Pump S1 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
The eufy Wearable Breast Pump E10 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
## 6. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Items | Proposed Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | K250207 | K231785 | N/A |
| Device Name | eufy Wearable Breast Pump S1
eufy Wearable Breast Pump E10 | Perifit Pump | N/A |
| Model | T8D02: eufy Wearable Breast | / | N/A |
{6}
K250207
Page 3 of 5
| | Pump S1
T8D03: eufy Wearable
Breast Pump E10 | | |
| --- | --- | --- | --- |
| Indications for Use | The eufy Wearable Breast Pump S1 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user.
The eufy Wearable Breast Pump E10 is a powered breast pump intended to be used by lactating women to express and collect milk from their breasts. It is intended for a single user. | The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user. | Similar |
| Patient Population | Lactating Women | Lactating Women | Identical |
| Single/double pump | Single (or Double by using two units simultaneously) | Single (or Double by using two units simultaneously) | Identical |
| Cycling control Mechanism | Microcontroller | Microcontroller | Identical |
| Backflow Protection | Yes | Yes | Identical |
| Suction Modes | Stimulation Mode and Expression Mode | Stimulation Mode and Expression Mode | Identical |
| Suction levels | 7 | 8 | Similar |
| Adjustable suction levels | Yes | Yes | Identical |
| Flange Size | 15, 17, 19, 21, 24, 27mm | 18mm, 21 mm, 24 mm, and 27mm | Similar |
| Vacuum range | -75 to -292.5mmHg | -30 to -300 mmHg | Similar |
| Cycle speed: Stimulation | 71 to 110 cycles/minute | 68 to 120 cycles per minute | Similar |
| Cycle speed: Expression | 39 to 60 cycles/minute | 28 to 80 cycles per minute | Similar |
| Controls | Power button, switch between stimulation and expression modes (S1 only), stimulation mode (E10 only), expression mode (E10 only), heating on/off (S1 only), +/- (adjust the suction, heat (S1 only) level, and LR paring (S1 only), LR paring (E10 only) | Power button, +/- (adjust suction intensity), pumping mode and intensity level indicator, switch between stimulation and expression modes | Similar |
| Power Supply | Li-Ion Battery | Li-ion Battery | Identical |
| Indicators | Yes, LED | Yes, LED | Identical |
| Materials | Milk Container:
Polypropylene
Flange: Silicone
Pump Outer Housing:
Acrylonitrile Butadiene | Milk-contacting parts are made of silicone, polypropylene, and Tritan copolyester (bottle only)
Pump Outer Housing: | Similar |
{7}
K250207
Page 4 of 5
| | Styrene (ABS) plastic | PC/ABS plastic | |
| --- | --- | --- | --- |
| Software | Pump operates on embedded software. A smartphone app (optional) may also be used to control the pump | Pump operates on embedded software. A smartphone app (optional) may also be used to control the pump | Identical Similar: the smartphone apps for both devices provide similar functionalities and any differences do not raise different questions of safety or effectiveness. |
In comparison to the predicate device, the subject device, eufy Wearable Breast Pump E10 and eufy Wearable Breast Pump S1, has the same intended use and similar indications for use – the expression and collection of breast milk. As seen in the comparison table, the subject and predicate devices have different technological features, including the user interface, vacuum range, cycle speed range, suction levels, materials, and flange size. These technological differences do not raise different questions of safety and effectiveness.
## 7. Summary of Non-clinical Performance Testing
### 1) Electrical Safety, Electromagnetic Compatibility
The subject device has passed safety testing in according to following standards.
- ANSI AAMI ES60601-1: 2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- ANSI AAMI HA60601-1-11:2015 [Including AMD1:2021]: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 62133-2 Edition1.0 2017-02: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
### 2) Performance testing
Performance testing was conducted on the subject device:
- Vacuum Suction pressure & Cycle speed
- Backflow protection
- Battery capacity
- Operating time/charging time
- Use life testing
Device specifications were met for all tests conducted.
### 3) Packaging simulation testing
{8}
K250207
Page 5 of 5
The packaging simulation testing was conducted in accordance with ISTA 3A: 2018.
## 4) Biocompatibility
Biocompatibility testing was conducted in accordance with the 2023 FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
The subject device has passed biocompatibility testing in according to following standards.
- ANSI AAMI ISO 10993-5:2009/(R)2014: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Fourth edition 2021-11: Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23 First edition 2021-01: Biological evaluation of medical devices - Part 23: Tests for irritation
The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
## 5) Software
The subject device contains an embedded software and a mobile application. Software was evaluated at the Basic Documentation level as recommended in the 2023 FDA guidance document “Content of Premarket Submissions for Device Software Functions.”
## 6) Cybersecurity
The cybersecurity was evaluated according to the 2025 FDA’s Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.”
## 8. Conclusions
The results of the performance testing described above demonstrate that the subject device, eufy Wearable Breast Pump S1, eufy Wearable Breast Pump E10 is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.