K250165 · Wontech Co., Ltd. · GEX · Jul 3, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250165
Device Name
Pastelle
Applicant
Wontech Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pastelle laser system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus
Device Story
Pastelle is a Q-switched Nd:YAG laser system for aesthetic, cosmetic, and surgical procedures. Device comprises Nd:YAG laser head, power supply, cooling system, and delivery system. Laser cavity utilizes two Nd:YAG media and two high-intensity xenon flash lamps. Operates in 1064nm (nanosecond and long-pulse Genesis modes) and 532nm (nanosecond mode) wavelengths; optional dye handpieces provide 595nm and 660nm. Operated by clinicians in dermatology/surgical settings. Laser energy delivered via articulated arm to various handpieces (zoom, collimated, microbeam). Output affects soft tissue via photothermal/photomechanical interaction for incision, excision, ablation, vaporization, coagulation, and hemostasis. Benefits include targeted treatment of pigmented lesions, vascular lesions, tattoos, and skin resurfacing.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Nd:YAG laser medium; 1064nm and 532nm wavelengths (expandable to 595nm/660nm via dye handpieces). Energy source: xenon flash lamps. Cooling: water-cooled housing. Delivery: articulated arm with zoom/microbeam handpieces. Materials: Aluminum powder (handpiece tip). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, ablation, or vaporization for general dermatology and surgery. Specific applications include tattoo removal (various ink colors), removal of Nevus of Ota, hair removal, treatment of common nevi, acne scars, wrinkles, melasma, inflammatory acne vulgaris, epidermal pigmented lesions, minor vascular lesions (telangiectasias), lentigines, café-au-lait, seborrheic keratoses, post-inflammatory hyperpigmentation, Becker's nevi, freckles, and nevi spilus.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Build Correspondence
Convert to PDF
July 3, 2025
WONTECH Co., Ltd.
Hyun Yoon
Team Leader
64 Techono 8-ro, Yuseong-gu
Daejeon, 34028
Korea, South
Re: K250165
Trade/Device Name: Pastelle
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: June 5, 2025
Received: June 5, 2025
Dear Hyun Yoon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250165 - Hyun Yoon
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250165 - Hyun Yoon
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK
MACIOS -S
Digitally signed by
MARK MACIOS -S
Date: 2025.07.03
09:49:40 -04'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K250165
Device Name
Pastelle
Indications for Use (Describe)
The Pastelle laser system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
1064nm in nanosecond mode, including microbeam handpieces:
- Tattoo removal: dark ink (black, blue, and brown)
- Removal of Nevus of Ota
- Removal or lightening of unwanted hair with or without adjuvant preparation
- Treatment of Common Nevi
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of melasma
1064nm in Genesis (long-pulse) mode:
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris
532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces):
- Tattoo removal: light ink (red, tan, purple, orange, sky blue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of Lentigines
- Treatment of Café-au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyperpigmentation
- Treatment of Becker's Nevi, Freckles, and Nevi spilus
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (6/20)
Page 1 of 2
PISC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 2 of 2
{5}
W
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
# 510(k) Summary
[As required by 21 CFR 807.92]
# K250165
## 1. Date Prepared [21 CFR 807.92(a)(a)]
June 27, 2025
## 2. Submitter's Information & Contact Person [21 CFR 807.92(a)(1)]
- Name of Manufacturer: WON TECH Co., Ltd.
- Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea
- Contact Name: Hyun Sik Yoon
- Telephone No.: +82-10-6750-5346
- Fax No.: +82-70-7836-0110
- Email Address: yoonhs21@wtlaser.com
## 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Common name: Q-switched Nd:YAG Laser System
Trade name: Pastelle
| Classification Description | 21 CFR Section | Product Code |
| --- | --- | --- |
| Powered Laser Surgical Instrument | 878.4810 | GEX |
As stated in 21 CFR, parts 878.4810, this generic types of devices has been classified as Class II.
Page 1 / 6
{6}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow:
### #1 Predicate device
- 510(k) Number: K123293
- Applicant: WON TECH Co., Ltd.
- Classification Name: Powered Laser Surgical Instrument
- Trade Name: Pastelle
### #2 Predicate device
- 510(k) Number: K241529
- Applicant: WON TECH Co., Ltd.
- Classification Name: Powered Laser Surgical Instrument
- Trade Name: Pastelle Pro
### #3 Predicate device
- 510(k) Number: K213569
- Applicant: Lutronic Corporation
- Classification Name: Powered Laser Surgical Instrument
- Trade Name: HOLLYWOOD SPECTRA Laser System
## 5. Description of the Device [21 CFR 807.92(a)(4)]
The Pastelle laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed, in the special adjustable holders composed the laser cavity.
## 6. Indications for Use [21 CFR 807.92(a)(5)]
The Pastelle system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
1064nm in nanosecond mode, including microbeam handpieces:
- Tattoo removal: dark ink (black, blue, and brown)
- Removal of Nevus of Ota
- Removal or lightening of unwanted hair with or without adjuvant preparation
- Treatment of Common Nevi
{7}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of melasma
1064nm in Genesis (long-pulse) mode:
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris
532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces):
- Tattoo removal: light ink (red, tan, purple, orange, sky blue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of Lentigines
- Treatment of Café-au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyperpigmentation
- Treatment of Becker's Nevi, Freckles, and Nevi spilus
Page 3 / 5
{8}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
# 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]
There are no significant differences between the Pastelle and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 | SE Decision |
| --- | --- | --- | --- | --- | --- |
| K Number | - | K123293 | K241529 | K213569 | - |
| Manufacturer | WONTECH Co., Ltd. | WONTECH Co., Ltd. | WONTECH Co., Ltd. | Lutronic Corporation | - |
| Model | Pastelle | Pastelle | Pastelle Pro | HOLLYWOOD SPECTRA Laser System | - |
| Product Code | GEX | GEX | GEX | GEX | Same |
| Intended Use | The Pastelle system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | Pastelle Q-Switched Nd:YAG Laser System is indicated for the Incision, Excision, Ablation and Vaporization of Soft Tissue for General Dermatology, Dermatologic and General Surgical Procedures for Coagulation and Hemostasis. | The Pastelle Pro system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | The HOLLYWOOD SPECTRA System is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation | Same as Predicate Device #2 and 3 |
{9}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
| | - Treatment of Common Nevi
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of melasma
1064nm in Genesis (long-pulse) mode:
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris
532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces):
- Tattoo removal: light ink (red, tan, purple, orange, sky blue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of Lentigines
- Treatment of Café-au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyperpigmentation
- Treatment of Becker's Nevi, Freckles, and Nevi spilus | | - Treatment of Common Nevi
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of melasma
1064nm in Genesis (long-pulse) mode:
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris
532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces):
- Tattoo removal: light ink (red, tan, purple, orange, sky blue, green)
- Removal of Epidermal Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
- Treatment of Lentigines
- Treatment of Café-au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyperpigmentation
- Treatment of Becker's Nevi, Freckles, and Nevi spilus | | |
| --- | --- | --- | --- | --- | --- |
Page 2 / 5
{10}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
| | | | Freckles, and Nevi spilus | | |
| --- | --- | --- | --- | --- | --- |
| Principle/Method of Operation | The Pastelle laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The Pastelle laser system consists of an Nd:YAG Laser head, a Power supply and a cooling system. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The Pastelle Pro laser system consists of an Nd:YAG laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains two Nd:YAG laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed ,in the special adjustable holders composed the laser cavity. | The HOLLYWOOD SPECTRA Laser System contains a Nd:YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates Q-switched and/or pulsed laser sources at the nominal wavelength of 1064 nm and 532 nm using KTP. The outputs of each laser generator and the aiming beam (655 nm) are delivered by articulated arm to a fixed (collimated), or focusing variable (zoom) spot handpiece, or a dual focused dots microbeam handpiece, or a 585nm/650nm dye laser converter handpiece. The dye handpieces convert the KTP 532 nm wavelength beam into a 585 nm or 650 nm wavelengths, correspondingly. | Same as Predicate Device #1 and #2 (Q-switched, Nd:YAG) |
| Laser Material | Nd:YAG | Nd:YAG | Nd:YAG | Nd:YAG | Same |
| | Proposed Device | Predicate Device #1 | Predicate Device #2 | Predicate Device #3 | SE Decision |
| --- | --- | --- | --- | --- | --- |
| Radiation diameter | Zoom: 2mm ~ 10mm
DOE: 5mm x 5mm ~ 7mm x 7mm | Zoom: 2mm ~ 10mm | Zoom: 2mm ~ 10mm
Collimation: 7mm
MLA: 3mm ~ 8mm
DOE: 5mm x 5mm ~ 7mm x 7mm
Dye: 3mm | 1064 Zoom: 1 ~ 7mm (Q-Switch, Spectra) / 2 ~ 7mm (Q-PTP, Q-3, Q-4)
532 Zoom: 0.8 ~ 6mm
Collimation: 3 ~ 8mm
MDF: 1064(4~8mm) / 532(3.4~6.9mm) | Same
(The range of specification of submitted device is covered by all predicate |
{11}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
| | | | | Dye: 2~5mm | devices' feature) |
| --- | --- | --- | --- | --- | --- |
| Laser output power | Genesis mode: 100 ~ 3,500 mJ
1064 nm mode: 80 ~ 1,300 mJ
1064 nm PTP mode: 80 ~ 1,600 mJ
532 nm mode: 10 ~ 500 mJ
1064 nm DOE mode: 80 ~ 900 mJ | Genesis mode: 100 ~ 2,000 mJ
1064 nm mode: 80 ~ 1,300 mJ
532 nm mode: 10 ~ 500 mJ | Genesis mode: 100 ~ 5,000 mJ
1064 nm mode: 50 ~ 1,200 mJ
1064 nm PTP mode: 600 ~ 2,000 mJ
1064 nm Triple mode: 1,000 ~ 1400 mJ
1064 nm 2x PTP mode: 1,000 ~ 1400 mJ
532 nm mode: 30 ~ 500 mJ
1064 nm DOE mode: 60 ~ 900 mJ
595 nm Dye mode: 30 ~ 250 mJ | Spectra: Up to 1500mJ
1064nm: Up to 1200mJ
1064nm PTP: Up to 1400mJ
1064nm Q-3, Q-4: Up to 1400mJ
532nm: Up to 400mJ | Same
(The range of specification of submitted device is covered by all predicate devices' feature) |
| Pulse width | Genesis mode: 80~480μs
PTP mode: 4~48ns
1064, 532 mode: 4~48ns | Genesis mode: 250~350μs
1064 mode: 4~48ns
532 mode: 4~48ns | 532nm/1064nm mode: 5ns~12ns
1064 PTP/Triple/2X PTP: Up to 20ns
Genesis mode: 80~300μs | Normal: 5~10ns
1064 Q-3, Q-4: 10~20ns
Spectra mode: 190μs | Same
(The range of specification of submitted device is covered by all predicate devices' feature) |
| Pulse repetition rate | Genesis mode: 1 ~ 10 Hz
1064 nm mode: 1 ~ 10 Hz
1064 nm PTP mode: 1 ~ 10 Hz
532 nm mode: 1 ~ 10 Hz
1064 nm DOE mode: 1 ~ 10 Hz | Genesis mode: 1 ~ 10 Hz
1064 nm mode: 1 ~ 10 Hz
532 nm mode: 1 ~ 10 Hz | Genesis mode: 1 ~ 10 Hz
1064 nm mode: 1 ~ 10 Hz
1064 nm PTP mode: 1 ~ 10 Hz
1064 nm Tripple mode: 1 ~ 10 Hz
1064 nm 2x PTP mode: 1 ~ 10 Hz
532 nm mode: 1 ~ 10 Hz
1064 nm DOE mode: 1 ~ 10 Hz
595 nm Dye mode: 1 ~ 5 Hz
660 nm Dye mode: 1 ~ 2 Hz | 1064, 532nm: Up to 10Hz | Same
(The range of specification of submitted device is covered by all predicate devices' feature) |
Page 4 / 5
{12}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
# Non-Clinical Test Summary [21 CFR 807.92(b)(1)]
## 1) Electrical Safety, Electromagnetic Compatibility Testing
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
| Standard (Edition) | Standard Title |
| --- | --- |
| IEC 60601-1:2005/AMD2:2020 | Medical electrical equipment — Part 1: General requirements for basic safety and essential performance — Amendment 2 |
| IEC 60601-1-2:2014+A1:2020 | Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests |
| IEC 60601-1-6:2010/AMD2:2020 | Medical electrical equipment — Part 1-6: General requirements for basic safety and essential performance — Collateral Standard: Usability — Amendment 2 |
| IEC 60601-2-22:2019 | Medical electrical equipment — Part 2-22: Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment |
| IEC 60825-1:2014 (Third Edition) | Safety of laser products - Part 1: Equipment classification and requirements |
## 2) Software Validation
The PASTELLE contains BASIC documentation level software. Software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
## 3) Biocompatibility
| Part | Material | Patient Contact | Duration of Contact by ISO 10993-1 | Bio-compatibility |
| --- | --- | --- | --- | --- |
| Handpiece Tip | Aluminium Powder (Cas No. 7429-90-5) | Intact Skin | Limited (< 24 hours) | Yes |
{13}
WONTECH
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea
TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491
## 4) Performance Testing
The performance of the PASTELLE has been defined as follows.
### Treatment Laser
- Genesis mode (1064nm): Up to 3500mJ (80μs to 480μs)
- PTP mode (1064nm): Up to 1600mJ (4ns to 48ns)
- 1064nm mode: Up to 1300mJ (4ns to 48ns)
- 532nm mode: Up to 500mJ (4ns to 48ns)
### Aiming Beam
- Wavelength: 635nm
- Output power: < 5mW
## Clinical Test Summary [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
## Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that the PASTELLE is substantially equivalent to predicate devices as described herein.
Page 2 / 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.