The AViTA Breast Pump is intended to express and collect milk from the breasts of lactating women, and to alleviate engorgement of the breast. The device is intended for a single user.
Device Story
AViTA Breast Pump (model BR10) is an electric, software-controlled device for expressing and collecting breast milk; alleviates engorgement. System comprises motor unit with touch screen interface and expression kit with milk collection container; connected via silicone tube. Motor unit supports single or double pumping. Device creates seal around nipple; applies/releases suction via microcontroller-driven pump. Silicone diaphragm in expression kit provides backflow protection, separating milk-contacting parts from vacuum system. Powered by rechargeable lithium-ion battery (3.7V, 2900mAh). User interface features four buttons (power/pause/play, mode selection, level adjustment) and LED display showing mode, level, timer, and battery status. Two operating modes (Stimulation, Expression) each offer five suction levels. Used by lactating women in home or professional settings. Output is expressed milk collected in container for storage. Device is non-sterile, reusable, and single-user.
Clinical Evidence
Bench testing only. Performance testing included vacuum pressure and cycle frequency verification, battery life, button durability, backflow protection, and use-life testing. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2 series. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Powered breast pump; microcontroller-based suction control. Materials: milk-contacting components compliant with 21 CFR 177. Power: 3.7V 2900mAh Li-ion battery, USB Type-C charging. Suction: -45 to -270 mmHg range. Modes: Stimulation (75-150 cycles/min) and Expression (38-60 cycles/min). Backflow protection via silicone diaphragm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, ISO 10993-5/10/23.
Indications for Use
Indicated for lactating women to express and collect breast milk and alleviate breast engorgement. Intended for single-user use in home and professional healthcare environments.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Philips Avent single/double electric breast pump Advanced (K201381)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 11, 2026
AViTA Corporation
Rex Hsieh
Regulatory Affairs Specialist
9F, No.78, Sec.1, Kwang-Fu Road, Sanchong Dist.,
New Taipei City, 24158
TAIWAN
Re: K250129
Trade/Device Name: AViTA Breast Pump
Regulation Number: 21 CFR 884.5160
Regulation Name: Powered Breast Pump
Regulatory Class: II
Product Code: HGX
Received: October 29, 2025
Dear Rex Hsieh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20903
www.fda.gov
{1}
K250129 - Rex Hsieh
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250129 - Rex Hsieh
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, PhD
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K250129
Device Name
AViTA Breast Pump
Indications for Use (Describe)
The AViTA Breast Pump is intended to express and collect milk from the breasts of lactating women, and to alleviate engorgement of the breast. The device is intended for a single user.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 441-6740 EF
{4}
K250129
Page 1 of 6
AViTA Corporation
AViTA Breast Pump
Traditional 510(k)
510(K) Summary
### 510(k) Summary - K250129
#### 1. Submitter Information
| Name: | AViTA Corporation |
| --- | --- |
| Address: | 9F, No.78, Sec.1, Kwang-Fu Road, Sanchong Dist., New Taipei City 24158, Taiwan (R.O.C.) |
| Phone: | +886-2-8512-1568 ext.5306 |
| Correspondent: | Rex Hsieh (rex_hsieh@avita.com.tw) |
#### 2. Date of Summary: June 10, 2026
#### 3. Device Information
| Trade Name: | AViTA Breast Pump |
| --- | --- |
| Common Name | Powered Breast Pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered breast pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Classification Panel | Obstetrics/Gynecology |
| Device Class: | Class II |
#### 4. Predicate Device Information
| 510(k) Number: | K201381 |
| --- | --- |
| Predicate Trade Name: | Philips Avent single/double electric breast pump Advanced |
The predicate device has not been subject to a design-related recall.
1 / 6
{5}
K250129
Page 2 of 6
AViTA Corporation
AViTA Breast Pump
Traditional 510(k)
510(K) Summary
## 5. Device Description
The AViTA Breast Pump is an electric breast pump intended to express milk and collect milk from the breasts of lactating women, and to alleviate engorgement of the breast. It is provided non-sterile and is a reusable device intended for a single user in home and professional healthcare environments.
The device includes a single model, the AViTA BR10, and consists of two key components: a motor unit with a touch screen user interface and expression kit with milk collection container. The motor unit is connected to the expression kit via a silicone tube. The motor unit can be connected to two expression kits simultaneously.
It is an electrically powered, software-controlled, digital breast pump. It is powered by a rechargeable lithium-ion battery (3.7 V, 2900mAh) that can be charged using an external charger. The device is designed not to be used during charging. The device operates with embedded software and does not incorporate any off-the-shelf (OTS) software.
The device includes two (2) operating modes – Simulation mode and Expression mode, and each mode has five (5) suction levels. The user interface of the device consists of four buttons and an LED display. The buttons include power button (used to turn on/off the device and play/pause the pumping program), mode selection button to choose between the two modes and the level increase and decrease buttons to select one of five levels. The device LED display shows the current mode, level, countdown timer, and battery level.
The device extracts milk from the breasts by creating a seal around the nipple and applying and releasing suction to the nipple. The milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, the expression kit includes a silicone diaphragm that physically separates the milk-contacting pathway from the vacuum system. All milk contacting components of the device are compliant with 21 CFR 177.
## 6. Indications for Use / Intended Use
The AViTA Breast Pump is intended to express and collect milk from the breasts of lactating women, and to alleviate engorgement of the breast. The device is intended for a single user.
2 / 6
{6}
K250129
Page 3 of 6
AViTA Corporation
AViTA Breast Pump
Traditional 510(k)
510(K) Summary
### 7. Comparison of Intended Use and Technological Characteristics of the Subject Device and Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Item | Subject Device AViTA Breast Pump (K250129) | Predicate Device Philips Avent single/double electric breast pump Advanced (K201381) | Comparison |
| --- | --- | --- | --- |
| Classification | Pump, Breast, Powered | Pump, Breast, Powered | Same |
| Regulation | Class II, 21 CFR 884.5160 | Class II, 21 CFR 884.5160 | Same |
| Product code | HGX | HGX | Same |
| Intended use | The AViTA Breast Pump is intended to express and collect milk from the breasts of lactating women, and to alleviate engorgement of the breast. The device is intended for a single user. | The Philips Avent single/double electric breast pump Advanced is intended to express and collect milk from the breast of a lactating woman, and to alleviate engorgement of the breast. The device is intended for a single user. | Same |
| User population | Lactating woman | Lactating woman | Same |
| Single user | Yes | Yes | Same |
| Pump options | Single or double pumping | Single or double pumping | Same |
| Media separation (Backflow protection) | Yes (Silicone diaphragm on the expression kit, which separates the vacuum system of the motor unit from milk-contacting parts) | Yes (Silicone diaphragm on the expression kit, which separates the vacuum system of the motor unit from milk-contacting parts) | Same |
| Cycling control mechanism | Microcontroller | Microcontroller | Same |
| Suction modes | Stimulation Mode Expression Mode | Stimulation Mode Expression Mode | Same |
| Adjustable suction levels | Yes | Yes | Same |
| Vacuum range | Stimulation mode: | Stimulation mode: | Different |
3 / 6
{7}
K250129
Page 4 of 6
AViTA Corporation
AViTA Breast Pump
Traditional 510(k)
510(K) Summary
| Item | Subject Device AViTA Breast Pump (K250129) | Predicate Device Philips Avent single/double electric breast pump Advanced (K201381) | Comparison |
| --- | --- | --- | --- |
| | -45 to -190 mmHg (5 Levels) Expression mode: -145 to -270 mmHg (5 Levels) | -45 to -150 mmHg Expression mode: -45 to -270 mmHg | |
| Cycle speed | Stimulation mode: 75 to 150 cycles/min (5 Levels) Expression mode: 38 to 60 cycles/min (5 Levels) | Stimulation mode: 105 to 120 cycles/min Expression mode: 53 to 85 cycles/min | Different. |
| Controls (User interface) | 4 Buttons (On-Off-Pause-Play, Stimulation or Expression mode, Decrease vacuum level, Increase vacuum level) | 4 Buttons (On-Off-Pause-Resume, Stimulation or Expression mode, Decrease vacuum level, Increase vacuum level) | Same |
| Indicators | LED indicators: Level/mode indicator, Timer, Battery level indicator | LED indicators: Level/mode indicator, Timer, Battery level indicator | Same |
| Power connection | USB Type-C (power only) | Micro-USB (power only) | Different. |
| Power supply | Lithium-ion battery (3.7 V 2900 mAh) | AC-Adapter (100-240 V primary and 5 V secondary) or Lithium-ion battery (3.6V 2600mAh) for battery-operated devices | Different |
The indications for use of the subject and predicate devices are similar, and both devices have the same intended use (i.e., for the collection of breast milk from the breasts of lactating women).
The subject and predicate devices have similar technological features, including pumping options, pumping options, backflow protection and same maximum suction pressure (270mmHg). While there are differences in technological characteristics between the subject and predicate devices, including vacuum pressure, cycle frequency, battery, and charger, these differences do not raise different questions of safety and effectiveness.
4 / 6
{8}
AViTA Corporation
AViTA Breast Pump
K250129
Page 5 of 6
Traditional 510(k)
510(K) Summary
### 8. Summary of Non-Clinical Performance Testing
The following performance tests were conducted on the AViTA Breast Pump to support a substantial equivalence determination:
# ➤ Biocompatibility Testing
The biocompatibility evaluation for the patient contacting components of the subject device was conducted in accordance with the 2023 FDA Biocompatibility Guidance Use of International Standard ISO10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”) as follows:
- ISO 10993-5:2009, Biological evaluation of medical devices–Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021, Biological evaluation of medical devices–Part 10: Tests for skin sensitization
- ISO 10993-23:2021, Biological evaluation of medical devices–Part 23: Tests for skin irritation
# ➤ Electrical Safety and EMC Testing
The electrical safety, and electromagnetic compatibility testing was conducted per the following standards:
- IEC 60601-1-2 :2014/AMD:2020 Medical electrical equipment–Part 1-2: General requirements for basic safety and essential performance–Collateral standard: electromagnetic compatibility
- IEC TR 60601-4-2 :2016 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1 :2005/AMD2:2020 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance with US national differences,
- IEC 60601-1-11 :2005/AMD2:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 62133-2:2017/AMD1:2021 – Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems
5 / 6
{9}
K250129
Page 6 of 6
AViTA Corporation
AViTA Breast Pump
Traditional 510(k)
510(K) Summary
# ➤ **Software Verification and Validation Testing**
Software verification and validation were conducted in compliance with 2023 FDA Guidance “Content of Premarket Submissions for Device Software Functions”.
# ➤ **Non-Clinical Performance Testing**
A series of performance tests were conducted on the subject device to demonstrate that the device meets its design requirements and performs as intended. The performance tests include:
- Vacuum pressure and cycle frequency testing at all modes and levels
- Battery operating and charging time testing
- Button life test of breast pump
- Backflow protection testing
- Use life testing
These tests were conducted in single and double pumping mode, and the device met the pre-defined acceptance criteria.
# 9. Conclusion
The results of the performance testing described above demonstrate that AViTA Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.