TSK Syringe

K250127 · Tsk Laboratory, Japan · QNQ · Jul 30, 2025 · General Hospital

Device Facts

Record IDK250127
Device NameTSK Syringe
ApplicantTsk Laboratory, Japan
Product CodeQNQ · General Hospital
Decision DateJul 30, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TSK Syringe is a low dead volume syringe intended for use in patients that need injection and withdrawal of substances for examination, treatment, diagnosis or prevention.

Device Story

TSK Syringe is a manual, low dead volume piston syringe used by healthcare professionals to inject or withdraw fluids. Device consists of a copolyester barrel, elastomer gasket, polypropylene plunger, and silicone lubricant. Operates via manual plunger depression within the barrel to move fluids through a Luer-Lock connection. Used in clinical settings for general medical applications. Low dead volume design minimizes fluid retention after full plunger depression. Benefits include precise delivery and reduced waste of substances. Device is sterile and single-use.

Clinical Evidence

Bench testing only. No clinical data. Performance validated against ISO 7886-1 (sterile hypodermic syringes) and ISO 80369-7 (small-bore connectors). Biocompatibility testing conducted per ISO 10993 series, including cytotoxicity, sensitization, irritation, systemic toxicity, and hemocompatibility. Sterilization validated per ISO 11137-1.

Technological Characteristics

Piston syringe; 0.7mL and 1.0mL capacities; Luer-Lock tip. Materials: Copolyester barrel, elastomer gasket, polypropylene plunger, silicone lubricant. Sterilization: Gamma radiation. Performance standards: ISO 7886-1, ISO 80369-7. Dead space specification: ≤ 0.023mL.

Indications for Use

Indicated for patients requiring injection or withdrawal of substances for examination, treatment, diagnosis, or prevention. Intended for use by healthcare professionals.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION July 30, 2025 TSK Laboratory, Japan % Anna Galea Principal Strategy Consultant NAMSA 400 Highway 169 South, Suite 500 Minneapolis, Minnesota 55426 Re: K250127 Trade/Device Name: TSK Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ, FMF Dated: June 23, 2025 Received: June 27, 2025 Dear Anna Galea: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250127 - Anna Galea Page 2 Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K250127 - Anna Galea Page 3 Sincerely, Kyran R. Gibson -S For Shruti Mistry Assistant Director, Injection Devices DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250127 | | | Device Name TSK Syringe | | | Indications for Use (Describe) | | | The TSK Syringe is a low dead volume syringe intended for use in patients that need injection and withdrawal of substances for examination, treatment, diagnosis or prevention. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) Summary TSK TSK LABORATORY, JAPAN # K250127 510(k) Summary # TSK Syringe ## Submitter: TSK Laboratory, Japan 1510-1, Soja-Machi, Tochigi-Shi Tochigi-Ken 328-0002 Japan +81-282-25-7575 Ms Yumi Ueda y.ueda@tsklab.co.jp Date Prepared: July 30, 2025 ## Correspondent: NAMSA 400 Highway 169 South, Suite 500, Minneapolis, MN 55426 United States +1 339-927-2945 Dr. Anna Galea agalea@namsa.com Device Name: TSK Syringe Device Trade Name: TSK Syringe Common Name: Piston Syringe Classification Name: Low Dead Space Piston Syringe, Piston Syringe Regulation Number: 880.5860 Product Codes: QNQ, FMF Regulation Class: II Predicate: PLPT LDV (Low Dead Volume) Sterile Syringe (K210443), Product Code QNQ ## Device Description Summary: The TSK Syringe is a syringe intended for use with a needle to inject or withdraw substances, for general use. The TSK Syringe leaves a low volume of fluid when the plunger is fully depressed. There are two model numbers for the TSK Syringe. The technological differences are as follows: | Number | Syringe Nominal Capacity (mL) | | --- | --- | Page 1 of 5 {5} 510(k) Summary TSK TSK LABORATORY, JAPAN TSY-0071P 0.7 TSY-0101Y 1.0 # Indications for Use: The TSK Syringe is a low dead volume syringe intended for use in patients that need injection and withdrawal of substances for examination, treatment, diagnosis or prevention. # Indications for Use Comparison: The indication for use is the equivalent to the predicate device with minor differences in wording. # Technological Comparison: The components and principle of operation are the same as the predicate. Device materials are similar with slight differences that do not impact safety or effectiveness, evaluated through testing. Syringe sizes are similar. The subject device also comes in a size smaller than $1\mathrm{mL}$ , which is not uncommon for other syringes in this classification. All features are documented in the table below. | Comparison of Subject Device to Predicate Device | Subject Device TSK Syringe | Predicate Device PLPT LDV (Low Dead Volume) Sterile Syringe (K210443) | Comparison | | --- | --- | --- | --- | | Manufacturer | TSK Japan | Poonglim Pharmatech Inc. | N/A | | Regulation | 21 CFR 880.5860 | 21 CFR 880.5860 | Same | | Product Code | QNQ FMF | QNQ | Same | | Class | II | II | Same | | Regulation Description | Piston syringe | Piston syringe | Same | | Indications for Use | The TSK Syringe is a low dead volume syringe intended for use in patients that need injection and withdrawal of substances for examination, treatment, diagnosis or prevention. | PLPT LDV (Low Dead Volume) Sterile Syringe is intended to be used for medical purpose to inject fluid into or withdraw fluid from body. | Same | | Intended Use | The TSK Syringe is a general purpose syringe intended for use with a needle to inject or withdraw substances. | It is intended for various medical applications and is not dedicated to medication administration. | Same | | Intended User and Environment | The TSK Syringe is intended for use by healthcare professionals | Healthcare professionals | Same | | Principal of Operation/Mechanism of Action | For manual use with a needle for injection or withdrawal of fluids. | A calibrated barrel (cylinder) with plunger intended to be used for | Same | {6} 510(k) Summary TSK TSK LABORATORY, JAPAN | | | injection/withdrawal of fluids/gas (e.g., medication) to/from a medical device or the body (i.e., capable of both); a needle is not included | | | --- | --- | --- | --- | | Device Components | Piston syringe consisting of barrel, gasket, plunger, and lubricant | Piston syringe consisting of barrel, plunger rod, and plunger | Different: The subject device contains lubricant. This is evaluated through performance and biocompatibility testing. It does not raise new questions of safety and effectiveness for the subject device when compared to the predicate. | | Materials | Barrel - Copolyester | Barrel - Polypropylene | Different: The differences in device material are not expected to impact safety or effectiveness. This is evaluated through biocompatibility testing. It does not raise new questions of safety and effectiveness for the subject device when compared to the predicate. | | | Gasket - Elastomer | Plunger - Rubber | | | | Plunger - Polypropylene | Plunger rod - Polypropylene | | | | Lubricant - Silicone | N/A | | | Syringe Size | 0.7 mL, 1 mL | 1 mL | Different: The subject device also comes in a size smaller than 1 mL, which is not uncommon for other syringes in this classification and does not raise safety concerns. | | Connection Type | Luer-Lock tip | Luer-lock type | Same | | Dead Space specification | ≤ 0.023mL | ≤ 0.023mL with 95% confidence/95% reliability | Same | | Sterilization and Shelf Life | Gamma Sterilization 5 years | EO gas 3 years | Different. The subject device has been validated in accordance with FDA recognized standards and therefore it does not raise new questions of safety and effectiveness for the subject device when compared to the predicate. | | Biocompatibility | Conforms to the requirements of ISO 10993 | Cytotoxicity Sensitization | Different: The subject device has undergone | Page 3 of 5 {7} 510(k) Summary TSK TSK LABORATORY, JAPAN | | series standards. Cytotoxicity Hemolysis Pyrogen test Intracutaneous reactivity test Skin sensitization test Acute systemic toxicity test Bacterial Endotoxins Test (USP39<85>) Sterility test (ISO 11737-2) USP<788> | Irritation Acute Systemic Toxicity Material-Mediated Pyrogenicity Hemocompatibility USP 39, <71> Sterility Test USP 39, <85> Bacterial Endotoxins Test | additional testing to support the biocompatibility of the device. This does not raise new questions of safety and effectiveness | | --- | --- | --- | --- | | Performance Data | ISO 7886-1:2017 ISO 80369-7:2021 | ISO 7886-1 ISO 80369-7:2016 | Same | # Non-Clinical Tests Summary: The TSK Syringe underwent biocompatibility testing according to the following: ISO 10993-1:2018/Cor1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process ISO 10993-4:2017 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity ISO 10993-12:2021 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials ISO 10993-18:2020/A1:2022 Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process AMENDMENT 1 ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation ISO 7886-1 Second edition 2017-05 Small-bore connectors for liquids and gases in healthcare applications - Part 7 Connectors for intravascular or hypodermic applications ISO 7886-1:2017 Second Edition 2017-05 Sterile hypodermic syringes for single use: Syringes for manual use, ISO 80369-7:2021 Second Edition 2021-05 Small bore connectors for liquids and gases in healthcare applications: Connectors for intravascular or hypodermic applications, USP&lt;788&gt; Particulate Matter in Injections Bench testing was conducted according to the following performance standards: ISO 7885-1: 2017 Sterile hypodermic syringes for single use Page 4 of 5 {8} 510(k) Summary TSK TSK LABORATORY, JAPAN ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications The TSK Syringe was sterilized in accordance with the following: ISO 11137-1:2006/Amd 2:2018 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices ISO 11607-1:2019 Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems ASTM F3039-15 Standard Test Method for Detecting Leaks in Nonporous Packaging or Flexible Barrier Materials by Dye Penetration ISO 4180:2019 Packaging - Complete, filled transport packages - General rules for the compilation of performance test schedules ASTM F 1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices ISO 7886-1 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use ISO 80369-7 Small bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications. USP39&lt;85&gt; Bacterial Endotoxins Test samples are representative of finished devices including sterilization. The TSK Syringe met all specified criteria and did not raise new safety or performance questions when compared to the predicate device. ## Conclusion: The results of non-clinical testing demonstrate that the TSK Syringe is substantially equivalent to, and is as safe and effective as, the predicate device for its intended use. Page 5 of 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...