ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M)
K250126 · Livsmed, Inc. · GEI · Oct 21, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250126
Device Name
ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M)
Applicant
Livsmed, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ArtiSeal Vessel Sealing System - ArtiSeal Instruments are a bipolar electrosurgical instrument intended for use only with the ArtiSeal Generator in minimally invasive or open surgical procedures where ligation (sealing) and division (Cutting) of vessels and tissue bundles is desired. ArtiSeal Vessel Sealing System - ArtiSeal Instruments can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery.
Device Story
Bipolar electrosurgical instrument; used with ArtiSeal Generator; intended for minimally invasive or open surgery. Input: RF energy from generator applied to electrode jaw. Operation: User operates grip lever to open/close jaw; jaw articulates up to ±90° and rotates 360°. Output: Ligation and division of vessels/tissue bundles. Used by surgeons in OR. Benefits: Effective vessel sealing up to 7 mm; equivalent to predicate performance.
Clinical Evidence
Nonclinical program included bench testing (mechanical/electrical performance), ex vivo burst pressure testing (mean >240 mmHg), acute thermal spread study (<2 mm spread), and 21-day chronic survival study (equivalent tissue healing/seal integrity). Biocompatibility confirmed per ISO 10993-1. No clinical human data presented.
Technological Characteristics
Bipolar electrosurgical instrument; RF energy source; sterile, single-use. Materials identical to predicate. Dimensions: Ø8 mm shaft. Articulation: ±90° jaw bend, 360° rotation. Sterilization: Ethylene oxide (ISO 11135, SAL 10⁻⁶). Electrical safety/EMC: IEC 60601-1, 60601-1-2, 60601-2-2.
Indications for Use
Indicated for use in general surgery for ligation and division of vessels (arteries and veins) up to 7 mm and tissue bundles in minimally invasive or open surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ArtiSeal Vessel Sealing System - ArtiSeal Instruments, ArtiSeal Vessel Sealing System - ArtiSeal Generator (K241138)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 21, 2025
LivsMed Inc.
Dong Wook Lee
Quality Management Representative
304, D-dong, 700, Pangyo-ro, Bundang-gu
Seongnam-si, Gyeonggi-do 13516
Korea, South
Re: K250126
Trade/Device Name: ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: September 22, 2025
Received: September 22, 2025
Dear Dong Wook Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250126 - Dong Wook Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250126 - Dong Wook Lee
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.10.21
17:09:18 -04'00'
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250126
Device Name
ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M)
Indications for Use (Describe)
ArtiSeal Vessel Sealing System - ArtiSeal Instruments are a bipolar electrosurgical instrument intended for use only with the ArtiSeal Generator in minimally invasive or open surgical procedures where ligation (sealing) and division (Cutting) of vessels and tissue bundles is desired. ArtiSeal Vessel Sealing System - ArtiSeal Instruments can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
LIV'SMED
ArtiSeal Vessel Sealing System
# 510(k) Summary - K250126
October 20, 2025
## Contact Details
Applicant Name: LivsMed Inc.
Applicant Address: 304, D-dong, 700, Pangyo-ro, Bundang-gu, Seongnam-si, Gyenggi-do, 13516, Republic of Korea
Applicant Contact Telephone: +82 70 7708 1399
Applicant Contact: Dong Wook Lee
Applicant Contact Email: hera@livsmed.com
## Device Name
Device Trade Name: ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M);
Common Name: Electrosurgical cutting and coagulation device and accessories
Classification Name: Electrosurgical, Cutting & Coagulation & Accessories
Regulation Number: 878.4400
Product Code: GEI
## Legally Marketed Predicate Devices
Predicate device: K241138 / ArtiSeal Vessel Sealing System - ArtiSeal Instruments, ArtiSeal Vessel Sealing System - ArtiSeal Generator
## Device Description Summary
The ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M) is a sterile, single-use, invasive instrument which is used with electrosurgical generator (ArtiSeal Generator), intended to be used during laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels and tissue bundles is desired.
When activating the sealing button of ArtiSeal Vessel Sealing System-ArtiSeal instruments, the ArtiSeal Generator applies radiofrequency (RF) energy to the electrode portion of the jaw. Heat is generated from the electrode for ligation.
During surgical procedures, the user operates the grip lever to open and close the jaw. Furthermore, the jaw can be bent up, down, left, and right within a range of up to $\pm 90^{\circ}$ or more by adjusting the grip accordingly. The jaw also provides $360^{\circ}$ articulation (rotation) by turning the grip.
## Indications for Use
ArtiSeal Vessel Sealing System - ArtiSeal Instruments are a bipolar electrosurgical instrument intended for use only with the ArtiSeal Generator in minimally invasive or open surgical procedures where ligation (sealing) and division (Cutting) of vessels and tissue bundles is desired. ArtiSeal Vessel Sealing System - ArtiSeal Instruments can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery.
Page 1 / 4
{5}
LIV'SMED
ArtiSeal Vessel Sealing System
# Nonclinical Testing Summary
A nonclinical program determined that the subject and predicate devices are substantially equivalent in performance and safety.
- Bench Testing:
Evaluated mechanical and electrical performance (jaw force, jaw gap, jaw hyper-extension, grasping force, device cycle, and electrical properties including inductance, capacitance, and resistance). Results were consistent with the predicate device and met all acceptance criteria.
- Ex Vivo Burst Pressure Testing:
Mean burst pressure of sealed vessels exceeded 240 mmHg across all vessel sizes, with no statistically significant differences from the predicate.
- Acute Thermal Spread Study:
Thermal spread remained below 2 mm with seal quality scores ≤ 2, meeting predefined acceptance limits.
- 21-Day Chronic Survival Study:
Long-term tissue healing and seal integrity were equivalent to the predicate, with no rebleeding or adverse histological findings.
These results demonstrate equivalent mechanical performance, vessel sealing effectiveness, and tissue safety.
# Biocompatibility
All patient-contacting materials are identical to those used in the predicate device.
Testing in accordance with ISO 10993-1 and related standards, including cytotoxicity, hemolysis, irritation, sensitization, systemic toxicity, pyrogenicity, and material characterization, confirmed that all materials are biocompatible and substantially equivalent to the predicate device.
# Sterilization, Packaging, and Shelf Life
The device is sterilized by ethylene oxide (EO) in accordance with ISO 11135, achieving a Sterility Assurance Level (SAL) of 10⁻⁶.
Packaging validation in accordance with ISO 11607-1 and ISO 11607-2 confirmed seal integrity and sterile barrier performance, and accelerated and real-time aging verified a three-year shelf life consistent with the predicate device.
# Electromagnetic Compatibility and Electrical Safety
The device operates with the same ArtiSeal Generator as the predicate.
Testing per IEC 60601-1, 60601-1-2, and 60601-2-2 demonstrated compliance with applicable electrical safety and EMC standards.
Substantial Equivalence Comparison
| Characteristic | Proposed Subject Device (AS01M) | Predicate Device (AS01L, K241138) | Comparison |
| --- | --- | --- | --- |
| Manufacturer | LivsMed Inc. | LivsMed Inc. | Same |
{6}
LIV'SMED
ArtiSeal Vessel Sealing System
| Device Classification / Regulation | Class II / 21 CFR 878.4400 | Class II / 21 CFR 878.4400 | Same |
| --- | --- | --- | --- |
| Product Code | GEI | GEI | Same |
| Indications for Use | ArtiSeal Vessel Sealing System - ArtiSeal Instruments are a bipolar electrosurgical instrument intended for use only with the ArtiSeal Generator in minimally invasive or open surgical procedures where ligation (sealing) and division (Cutting) of vessels and tissue bundles is desired. ArtiSeal Vessel Sealing System - ArtiSeal Instruments can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery. | ArtiSeal Vessel Sealing System - ArtiSeal Instruments are a bipolar electrosurgical instrument intended for use only with the ArtiSeal Generator in minimally invasive or open surgical procedures where ligation (sealing) and division (Cutting) of vessels and tissue bundles is desired. ArtiSeal Vessel Sealing System - ArtiSeal Instruments can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery. | Same |
| Principle of Operation | Bipolar RF energy applied via electrodes to seal and divide tissue | Bipolar RF energy applied via electrodes to seal and divide tissue | Same |
| Energy Type | Bipolar Radiofrequency | Bipolar Radiofrequency | Same |
| Biocompatibility | Evaluated per ISO 10993-1 | Evaluated per ISO 10993-1 | Same |
| Sterilization Method | EO gas (SAL 10⁻⁶), single-use | EO gas (SAL 10⁻⁶), single-use | Same |
| Shaft Dimensions | Ø8 mm × 345 mm | Ø8 mm × 415 mm | Similar |
| Performance | Equivalent jaw force, sealing performance, and electrical characteristics verified through bench and animal testing | Equivalent jaw force, sealing performance, and electrical characteristics verified through bench and animal testing | Equivalent |
| Electrical Safety / EMC | Compliant with IEC 60601-1, -1-2, -2-2 | Compliant with IEC 60601-1, -1-2, -2-2 | Same |
# Conclusion
The ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M) and the predicate device (AS01L) are identical in intended use, operating principle, energy type, electrode configuration, and materials, with the only difference being shaft length.
Performance validation/verification, including bench testing, ex vivo burst pressure, acute and chronic in vivo studies, biocompatibility, sterilization, and electrical safety, demonstrates that the subject device meets all predefined acceptance criteria and performs equivalently to the predicate.
Page 3 / 4
{7}
LIV'SMED
ArtiSeal Vessel Sealing System
Therefore, the ArtiSeal Vessel Sealing System - ArtiSeal Instruments (AS01M) is substantially equivalent to the legally marketed predicate device (K241138) and is as safe and effective for its intended clinical use.
Page 4 / 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.