← Product Code [GCJ](/productcode/GCJ) · K250124

# Viction Disposable Irrigation and suction catheter (VC-IRA0532, VC-IRA0523, VC-IRB0532, VC-IRB0523, VC-IRC0532, VC-IRC0523, VC-IRD0532, VC-IRD0532) (K250124)

_Hangzhou Viction Medical Technology Co.,Ltd · GCJ · Aug 6, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K250124

## Device Facts

- **Applicant:** Hangzhou Viction Medical Technology Co.,Ltd
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 6, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Viction disposable Irrigation and suction Catheter is indicated for use in conjunction with the handpiece, dual tubing, and probes to provide controlled powered irrigation during Laparoscopic surgery.

## Device Story

Device is a battery-powered, single-use, disposable suction and irrigation system for laparoscopic surgery. System comprises a handpiece, dual tubing (suction and irrigation), and probes. Operates via a mechanical pump powered by nine AA alkaline batteries; irrigation flow is controlled by a button on the handpiece. Suction function requires connection to hospital-provided negative pressure source. Used in clinical/hospital settings by healthcare professionals. Provides controlled fluid delivery to surgical sites and suction capability to clear fluids. Benefits include improved visualization and site management during minimally invasive procedures.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including fluid flow rate, negative pressure resistance, sealing performance, electrical safety, EMC, and biocompatibility testing (acute systemic toxicity, skin sensitization, intracutaneous reactivity, pyrogen, and in vitro cytotoxicity).

## Technological Characteristics

Materials: 304 Stainless Steel, ABS, PVC. Energy source: 9 AA alkaline batteries. Principle: Battery-powered mechanical pumping system. Form factor: Handpiece with dual tubing and probes. Connectivity: None. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- CORE E3 Suction/Irrigator ([K202303](/device/K202303.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

August 6, 2025

Hangzhou Viction Medical Technology Co.,Ltd
Zhicong Luo
General Manager
First Floor ,No.1 Building, No.95 Binwen Road,
Binjiang District
Hangzhou, Zhejiang 310051
China

Re: K250124
Trade/Device Name: Viction Disposable Irrigation and suction catheter (VC-IRA0532, VC-IRA0523, VC-IRB0532, VC-IRB0523, VC-IRC0532, VC-IRC0523, VC-IRD0532, VC-IRD0532)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ
Dated: January 17, 2025
Received: January 17, 2025

Dear Zhicong Luo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250124 - Zhicong Luo
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

{2}

K250124 - Zhicong Luo
Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Julie Morabito, Ph.D.
Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

k250124

Device Name

Viction Disposable Irrigation and suction catheter (VC-IRA0532、VC-IRA0523、VC-IRB0532、VC-IRB0523VC-IRC0532、VC-IRC0523、VC-IRD0532、VC-IRD0532)

Indications for Use (Describe)

The Viction disposable Irrigation and suction Catheter is indicated for use in conjunction with the handpiece, dual tubing, and probes to provide controlled powered irrigation during Laparoscopic surgery.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

510 (K) Summary

K250124

This summary of 510(K)safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.

Type of submission : Traditional

The date the summary was prepared: June8, 2025

I. SUBMITTER

Hangzhou Viction Medical Technology Co.,Ltd

First floor, No.1 Building,

No.95 Binwen Road, Binjiang District,

Hangzhou City, Zhejiang Province, China.MD310051

Contact Person: Yu Gu

Phone:+86-057188305089

Email:guyu@victionmed.com

Date Prepared: February 11, 2025

II. Device

Name of Device : Viction Disposable Suction Irrigation Catheter, (VC-IRA0532, VC-IRA0523, VC-IRB0532, VC-IRB0523, VC-IRC0532,VC-IRC0523,VC-IRD0532, VC-IRD0532)

Common or Usual Name: Suction Irrigation Catheter

Classification Name:Endoscope And Accessories( 21 CFR876.1500 )

Regulatory Class: 2

Product Code: GCJ

III. PREDICATE DEVICE

CORE E3 Suction/Irrigator (K202303), manufactured by ConMed Corporation. Located in 525 French Road Utica.

This predicate has not been subjected to a design-related recall.

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Viction Disposable Irrigation and suction Catheter is constitutes with Suction/irrigation handpiece, dual tubing, Irrigation device that is a battery-powered mechanical pumping system connect with dual tubing(suction tubing and irrigation tubing) to deliver sterile fluids to surgical sites. Control of the irrigation flow is generated by depressing the on-off Control Button on the handpiece. The mechanical pumping system is powered with nine(9) standard AAlkaline Battery.

{5}

Hand piece equipped with pistol style attach to Suction irrigation probes and suction tubing. And it should be used with negative pressure device of the hospital.

There are 8 models of the product, models VC-IRA0532, VC-IRA0523, VC-IR B0532, VC-IRB0523, VC-IRC0532, VC-IRC0523, VC-IRD0532, VC-IRD0532. All product's Structure are the same, the length of the irrigation suction tubes is the main difference. It is a single use, disposable device and is sold sterile.

## V. INDICATION FOR USE

The Viction Disposable Irrigation and Suction Catheter is indicated for use in conjunction with the handpiece, dual tubing, and probes to provide controlled powered irrigation during Laparoscopic surgery.

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

## WITH THE PREDICATED DEVICE

The Viction Disposable Suction Irrigation system is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

**Predicate device name**: CORE E3 Suction/Irrigator

**510(K) number**: K202303

**Common Name**: Suction Irrigation

**Manufacturer**: ConMed Corporation.

The following table shows similarities and differences of indications for use between our device and the predicate devices. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.

|  Item | Predicate device | Subject device | Comparison  |
| --- | --- | --- | --- |
|  Trade Name | CORE E3 Suction/Irrigator | Viction Disposable Irrigation and suction catheter | Similar product, different name  |
|  510(K) Number | K202303 | K250124 | /  |
|  Classification regulation | 21 CFR 876.1500 | 21 CFR 876.1500 | same  |
|  Classification and Code | ClassII, GCJ | ClassII, GCJ | same  |

{6}

|  Device Classification Name | Laparoscope, General & Plastic Surgery | Laparoscope, General & Plastic Surgery | same  |
| --- | --- | --- | --- |
|  Design | Battery-powered pump, dual tubing, trumpet handpiece, and optional probe. | Battery-powered pump, dual tubing, handpiece, and four types of probe. | Different; subject device’s probe has different shapes, but it can’t change, do not have optional Probes.  |
|  materials | 304 Stainless Steel, Polycarbonate, 420 Stainless Steel, EPDM Rubber, ABS, and PVC. Six (6) AAA alkaline batteries. | 304 Stainless Steel, ABS, and PVC. Nine (9) AA alkaline batteries. | Similar  |
|  Sterilization | Ethylene Oxide Sterility Assurance Level (SAL) of 10^{-6} | Ethylene Oxide Sterility Assurance Level (SAL) of 10^{-6} |   |
|  Single Use / Reusable | Single Use | Single Use | same  |
|  Principle of Operation | Battery powered mechanical pumping system | Battery powered mechanical pumping system | same  |
|  Environment Use | Healthcare facility/hospital | Healthcare facility/hospital | same  |

## Conclusion

Based on the comparison of indications of use and other product performances. The Viction Disposable Suction Irrigation system manufactured by “Hangzhou Viction Medical Technology Co., Ltd.” is substantial equivalent to its predicate devices.

## VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

### Electrical safety and electromagnetic compatibility

Electrical safety and EMC testing were conducted on the device.

Electromagnetic compatibility is evaluated following IEC 60601-1-2:2014+AMD

{7}

1:2020,IEC 60601-2-18:2009.

Electrical safety is evaluated following IEC 60601-1: 2005: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. Apply with standard IEC606011:2005+AMD1:2012+AMD2:2020

## Biocompatibility testing

The Biocompatibility evaluation for the CORE E3 device was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices – Part-1: Evaluation and testing within a risk management process and application of applicable principles detailed in FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and Testing within a Risk Management Process" (June 16, 2016). The following tests were completed:

- Acute Systemic Toxicity
- Skin Sensitization Test
- Intracutaneous Reactivity Test
- Pyrogen Test
- In Vitro Cytotoxicity Test

## Bench Testing

- Fluid flow rate of Irrigation tube
- Negative pressure resistance of suction
- Sealing of irrigation tube

## Sterilization

Performance qualification for ethylene oxide sterilization

## Clinical Testing

Not applicable to this submission

All the test results meets the predefined acceptance criteria and complied with the design specification of the subject device throughout the use life. The results of the non-clinical testing demonstrate that the victim disposable irrigation and suction catheter is as safe and effective as the predicate device.

## VIII. CONCLUSIONS

Based upon the intended use and known technical information provided in this pre-market notification the Viction Disposable Irrigation and Suction Catheter has been shown to be substantially equivalent to currently marketed predicate device K202303.

---

**Source:** [https://fda-staging.innolitics.com/device/K250124](https://fda-staging.innolitics.com/device/K250124)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
