Linshom Continuous Predictive Respiratory Monitoring System (CPRMS)

K250093 · Linshom Medical, Inc. · BZQ · Sep 26, 2025 · Anesthesiology

Device Facts

Record IDK250093
Device NameLinshom Continuous Predictive Respiratory Monitoring System (CPRMS)
ApplicantLinshom Medical, Inc.
Product CodeBZQ · Anesthesiology
Decision DateSep 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2375
Device ClassClass 2

Indications for Use

Linshom Continuous Predictive Respiratory Monitor System (CPRMS) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult, (at least 22 years of age) patients. The CPRMS is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume. CPRMS measurements are used as an adjunct to other clinical information sources.

Device Story

Linshom CPRMS is a portable, non-invasive respiratory monitor for adult patients in clinical settings. It utilizes a thermistor-based sensor assembly to detect spontaneous respiration by measuring temperature changes against time. The system autonomously adapts to the local thermal environment to provide a signal resistant to motion artifacts. It processes these thermal inputs to calculate and display respiratory rate, seconds since last breath, and a trend of tidal volume via a GUI interface. Healthcare professionals use these outputs as an adjunct to other clinical information to monitor patient breathing status. The device is intended to assist in clinical decision-making by providing continuous, reliable respiratory data.

Clinical Evidence

Bench testing with 12 human volunteers (open mask) and 6 volunteers (standard mask), with 5 overlapping subjects. Testing covered respiratory rates of 10–60 BPM under metronome guidance. Results showed high correlation (r ≥0.98) between measured rates and metronome reference, with minimal bias and narrow limits of agreement in Bland-Altman analysis. Raw signal overlays confirmed equivalent waveform patterns. No clinical patient data reported.

Technological Characteristics

Thermistor-based sensing; 2-piece sensor assembly; compression connection; GUI interface. Operates in 65°F–85°F range. Respiration range 5-60 BPM. Biocompatibility per ISO 10993-5:2009, ISO 18562-2:2017, ISO 18562-3:2017. Safety/EMC per IEC 60601-1 & 60601-1-2.

Indications for Use

Indicated for adult patients (at least 22 years of age) in healthcare facilities (procedural areas, recovery rooms) for non-invasive respiratory monitoring.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 26, 2025 Linshom Medical, Inc. William Stoll VP, Quality & Regulatory 2922 Excelsior Springs Court Ellicott City, Maryland 21042 Re: K250093 Trade/Device Name: Linshom Continuous Predictive Respiratory Monitoring System (CPRMS) Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZQ Dated: August 22, 2025 Received: August 22, 2025 Dear William Stoll: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250093 - William Stoll Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250093 - William Stoll Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Binoy J. Mathews -S Digitally signed by Binoy J. Mathews -S Date: 2025.09.26 14:00:08 -04'00' For Rachana Visaria Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250093 | | | Device Name Linshom Continuous Predictive Respiratory Monitoring System (CPRMS) | | | Indications for Use (Describe) Linshom Continuous Predictive Respiratory Monitor System (CPRMS) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult, (at least 22 years of age) patients. The CPRMS is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rate and seconds since last breath, along with a trend of tidal volume. CPRMS measurements are used as an adjunct to other clinical information sources. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Linshom Every Breath Counts™ K250093 # 510(k) Summary (per 21 CFR 807.92) # K250093 ## Submitter Information Name: Linshom Medical, Inc. Address: 2922 Excelsior Springs Court Ellicott City, MD 21042 Phone Number: (410) 480-2700 Contact Person: William A. Stoll VP, Quality & Regulatory Date Prepared: September 26, 2025 ## Device Information Trade Name: Linshom Continuous Predictive Respiratory Monitoring System (CPRMS) Common Name: Respiratory Monitor Classification: Breathing frequency monitor 21 CFR 868.2375 (Product Code BZQ) ## Predicate Device Information Device Name: Linshom Continuous Predictive Respiratory Monitoring System (CPRMS) Classification: Breathing frequency monitor 21 CFR 868.2375 (Product Code BZQ) 510(k) Number: K240271 ## Device Description The Linshom CPRMS (Continuous Predictive Respiratory Monitoring System) is a portable and reliable system for detection of spontaneous respiration. It's non-invasive and is not corrupted by motion artifacts. The system autonomously adapts to the local thermal environment to deliver a usable signal without complicated hardware and firmware processing. Linshom Medical, Inc. (CPRMS) – 510(k) Summary {5} Linshom Every Breath Counts K250093 # Indications for Use Linshom Continuous Predictive Respiratory Monitoring System (CPRMS) is indicated for use by healthcare professionals in healthcare facilities, such as procedural areas and recovery rooms, to monitor breathing in adult, (at least 22 years of age) patients. CPRMS is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rates and seconds since last breath, along with a trend of tidal volume. CPRMS measurements are used as an adjunct to other clinical information sources. # Technological Characteristics – Comparison Table 1: Comparison of Technological Characteristics | Characteristic | Subject Device Linshom CPRMS (K250093) | Predicate Device Linshom CPRMS (K240271) | Comparison Outcome | | --- | --- | --- | --- | | Trade Name | Continuous Predictive Respiratory Monitoring System (CPRMS) | Continuous Predictive Respiratory Monitoring System (CPRMS) | Same | | Common Name | Respiratory Monitor | Respiratory Monitor | Same | | 510(K) Number | K250093 | K240271 | | | Regulation Classification (Product Code) | 21 CFR 868.2375 (BZQ) | 21 CFR 868.2375 (BZQ) | Same | | Intended Use | Non-invasive monitoring of respiration and tidal volume trends for adults in healthcare settings. | Non-invasive monitoring of respiration and tidal volume trends for adults in healthcare settings. | Same | Linshom Medical, Inc. (CPRMS) – 510(k) Summary {6} Linshom Every Breath Counts K250093 | Characteristic | Subject Device Linshom CPRMS (K250093) | Predicate Device Linshom CPRMS (K240271) | Comparison Outcome | | --- | --- | --- | --- | | Indications for Use | Linshom Respiratory Monitoring Device (CPRMS) is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult, (at least 22 years of age) patients. CPRMS is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rates and seconds since last breath, along with a trend of tidal volume. CPRMS measurements are used as an adjunct to other clinical information sources. | Linshom Respiratory Monitoring Device (CPRMS) is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in adult, (at least 22 years of age) patients. CPRMS is a non-invasive system that graphically displays temperature changes against time and reports values of respiratory rates and seconds since last breath, along with a trend of tidal volume. CPRMS measurements are used as an adjunct to other clinical information sources. | Same | | Mechanism (General) | Thermistor | Thermistor | Same | | Assembly (Specifications) | 2 pieces | 2 pieces | Same | | Connection (Specifications) | Compression | Compression | Same | | Measurements | Respiratory Rate Seconds Since Last Breath Tidal Volume Trend | Respiratory Rate Seconds Since Last Breath Tidal Volume Trend | Same | | Communication Method | GUI Interface | GUI Interface | Same | | Software / Firmware | Proprietary Algorithm | Proprietary Algorithm | Same | | Safety / EMC Testing | IEC 60601-1 & 60601-1-2 2-piece sensor and ILM | IEC 60601-1 & 60601-1-2 2-piece sensor and ILM | Same | | Material Biocompatibility | ISO 18562-2:2017 ISO 18562-3:2017 ISO 10993-5:2009 | ISO 18562-2:2017 ISO 18562-3:2017 ISO 10993-5:2009 | Same | | Mask Type | Face mask (oxygen) - Adult open | Face mask (oxygen) - Adult standard | Same Function | Linshom Medical, Inc. (CPRMS) – 510(k) Summary {7} Linshom Every Breath Counts K250093 | Characteristic | Subject Device Linshom CPRMS (K250093) | Predicate Device Linshom CPRMS (K240271) | Comparison Outcome | | --- | --- | --- | --- | | Mounting Design | Sensor attached to separate electronics box | Sensor attached to separate electronics box | Same | | Ambient Operating Temperature | 65°F – 85°F (18.3°C – 29.4°C) | 65°F – 85°F (18.3°C – 29.4°C) | Same | | Working Range | Respiration: 5-60 BPM Tidal Volume Trend | Respiration: 5-60 BPM Tidal Volume Trend: | Same | | Accuracy | Respiration: ± 1 BPM Tidal Volume Trend: ~0.97 (r2 correlation) | Respiration: ± 1 BPM Tidal Volume Trend: ~0.97 (r2 correlation) | Same | | Weight (at point of measurement) | ~15g (Thermistor Sensor Assembly only) | ~15g (Thermistor Sensor Assembly only) | Same | | Dimensions (of unit as point of measurement) | 60mm H 254mm L 158mm W (ILM Core) | 60mm H 254mm L 158mm W (ILM Core) | Same | As described above in Table 1 the technological differences between the CPRMS open oxygen mask and CPRMS standard oxygen mask devices do not raise any/new questions of safety or effectiveness. Performance testing further confirms that these differences do not affect safety and/or efficacy. ## Non-Clinical and/or Clinical Tests Summary Clinical bench testing was conducted to evaluate the performance of the subject device, the Open Oxygen Mask (K250093), compared to the predicate Standard Oxygen Mask (K240271), using the same FDA-cleared Linshom two-piece sensor and electronics. Human volunteer testing was performed across the full physiologic range of respiratory rates (10–60 breaths per minute) under metronome guidance. A total of 12 volunteers were tested with the open mask and 6 volunteers with the standard mask, with 5 volunteers participating in both arms to allow direct within-subject comparison. Data collection included raw respiratory signals and derived respiratory rate values, which were statistically analyzed using Pearson correlation and Bland-Altman methods. Results demonstrated very high correlation (r ≥0.98 across all cases) between the measured respiratory rates and the metronome reference, with minimal bias and narrow limits of agreement. Overlays of raw respiratory signals confirmed that both mask configurations produced consistent waveform patterns and equivalent respiratory rate outputs. Together, these bench performance tests confirm that the change in mask design does not adversely affect sensor function and support a determination of substantial equivalence. Linshom Medical, Inc. (CPRMS) – 510(k) Summary {8} Linshom Every Breath Counts™ K250093 ## Biocompatibility The biocompatibility evaluation for the Linshom CPRMS [K250093] was conducted in accordance with ISO 10993-1:2018, ISO 18562-1:2024, and FDA’s 2023 Biocompatibility Guidance document and was based on materials-based rationale in conjunction with biological testing. ## Conclusions Clinical and non-clinical testing demonstrates substantial equivalence of the subject device CPRMS being as safe, as effective, and performing as well as the CPRMS predicate device. Linshom Medical, Inc. (CPRMS) – 510(k) Summary
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...