MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL)
Applicant
Beckman Coulter, Inc.
Product Code
LTT · Microbiology
Decision Date
Jul 18, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Attributes
PCCP
Indications for Use
For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Device Story
MicroScan Dried Gram-Negative MIC/Combo Panels are single-use, miniaturized broth dilution susceptibility tests. Panels contain dehydrated antimicrobial agents (Aztreonam 0.5-64 µg/mL) in broth. Laboratory professionals inoculate panels with standardized suspensions of isolated colonies from solid media. Panels are rehydrated with water and incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator. The minimum inhibitory concentration (MIC) is determined by identifying the lowest antimicrobial concentration inhibiting growth, either visually or via MicroScan instrumentation (WalkAway or autoSCAN-4). Results assist clinicians in selecting appropriate antimicrobial therapy for gram-negative bacterial infections.
Clinical Evidence
Performance evaluated using 622 clinical and 148 challenge isolates. After excluding non-indicated species, 540 clinical and 108 challenge isolates were analyzed. Overall essential agreement (EA) and category agreement (CA) were generally >90%. Specific limitations were established for organism/method combinations showing unacceptable major or very major error rates (e.g., Citrobacter koseri, Proteus mirabilis, E. coli). No clinical data; performance based on bench-top comparative testing against reference broth microdilution.
Technological Characteristics
Miniaturized broth microdilution panel; dehydrated antimicrobial agent (Aztreonam). Incubation: 35°C ± 1°C, aerobic, 16-20 hours. Reading: Visual or automated (WalkAway/autoSCAN-4). Single-use. Complies with Class II Special Controls Guidance for AST Systems.
Indications for Use
Indicated for quantitative and qualitative antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic gram-negative bacilli (Enterobacterales and Pseudomonas aeruginosa) from solid media. Contraindicated for Enterobacter cloacae complex, Serratia marcescens, Serratia liquefaciens complex, Proteus vulgaris, Providencia stuartii, and Providencia rettgeri.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K250084
B Applicant
Beckman Coulter, Inc.
C Proprietary and Established Names
MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LTT | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
| JWY | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LRG | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTW | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for aztreonam at concentrations of 0.5-64 µg/mL with the MicroScan Dried Gram-Negative MIC/Combo Panels for susceptibility testing of non-fastidious Gram-negative organisms.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Measurand:
Aztreonam in the dilution range of 0.5-64 µg/mL
C Type of Test:
Quantitative antimicrobial susceptibility test (AST)
III Intended Use/Indications for Use:
A Intended Use(s):
For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
B Indication(s) for Use:
The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.
This particular submission is for the addition of the antimicrobial aztreonam at concentrations of 0.5-64 µg/mL to the test panel. Testing is indicated for Enterobacterales and Pseudomonas aeruginosa, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC) webpage.
The MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL) has demonstrated acceptable performance with the following organisms:
Enterobacterales (Citrobacter freundii complex, Citrobacter koseri, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Morganella morganii, Yersinia enterocolitica)
Pseudomonas aeruginosa
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Results obtained with the organism/antimicrobial agent combinations listed below have shown discrepant MIC's when compared with an overnight reference method. Test results for these organisms will not be reported, and an alternate method for testing should be used.
Aztreonam: Enterobacter cloacae complex, Serratia marcescens, Serratia liquefaciens complex, Proteus vulgaris, Providencia stuartii, and Providencia rettgeri.
Performance of aztreonam when testing Citrobacter koseri have shown discrepant MICs (unacceptable essential agreement and an elevated very major error rate) when compared with the reference method across all read methods with the Prompt inoculation system. An alternate testing method (e.g., autoSCAN-4/WalkAway/manual reads with turbidity inoculation) should be used for aztreonam with Citrobacter koseri.
K250084 - Page 2 of 15
{2}
Performance of aztreonam when testing M. morganii and P. mirabilis using the Prompt inoculation method with all read methods was outside of essential agreement with unacceptable major error rates compared to the reference method. M. morganii and P. mirabilis should only be tested with the turbidity inoculation method.
Due to the occurrence of major errors with aztreonam, P. mirabilis isolates with an MIC ≥ 16 µg/mL run on the WalkAway/Turbidity method should be read manually prior to reporting.
Due to the occurrence of a very major error involving a single E. coli isolate tested with aztreonam, isolates with an MIC ≤ 4 µg/mL run on the autoSCAN-4/Turbidity method should be read manually prior to reporting.
Performance of aztreonam when testing Yersinia enterocolitica using the Prompt inoculation method with the WalkAway read method were within categorical agreement, but outside of essential agreement due to a single isolate when compared to the reference method. Yersinia enterocolitica results with the Prompt/WalkAway should be read manually prior to reporting.
The ability of the MicroScan Dried Gram Negative Panels to detect resistance to aztreonam is unknown with Y. enterocolitica because an insufficient number of resistant strains were available at the time of comparative testing. Isolates yielding MIC results suggestive of a resistant interpretive category should be submitted to a reference laboratory.
## D Special Instrument Requirements:
MicroScan panels can be read either manually or automatically on the WalkAway or autoScan-4 instrument systems.
## IV Device/System Characteristics:
### A Device Description:
The MicroScan Dried Gram-Negative MIC/Combo panel with aztreonam is used to determine the quantitative and/or qualitative antimicrobial agent susceptibility of aerobic and facultatively anaerobic Gram-negative bacilli colonies grown on solid media. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1° in a non-CO2 incubator and read either visually or with MicroScan instrumentation according to the package insert.
Inoculation methods: Turbidity or Prompt Inoculation System
Read methods: Manual, MicroScan WalkAway System and MicroScan autoSCAN-4
### B Principle of Operation:
The antimicrobial susceptibility tests are dehydrated miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in Mueller Hinton broth supplemented with calcium and magnesium to concentrations spanning the range of clinical interest. Breakpoint Combo panels use concentrations equivalent to the categorical breakpoints
K250084 - Page 3 of 15
{3}
determined or recognized by FDA. After inoculation and rehydration with a standardized suspension of organism and incubation at 35°C for a minimum of 16 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth.
V Substantial Equivalence Information:
A Predicate Device Name(s):
MicroScan Dried Gram-Negative MIC/Combo Panels with Ceftazidime (Caz) (0.5-64 µg/mL)
B Predicate 510(k) Number(s):
K202343
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K250084 | Predicate: K202343 |
| --- | --- | --- |
| Device Trade Name | MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (Azt) (0.5-64 µg/mL) | MicroScan Dried Gram-Negative MIC/Combo Panels with Ceftazidime (Caz) (0.5-64 µg/mL) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Determination of susceptibility with Gram-negative bacilli | Same |
| Technology | Overnight microdilution MIC susceptibility test | Same |
| Specimen | Isolated colonies from culture | Same |
| Incubation Temperature | 35 °C ± 1°C | Same |
| Incubation Atmosphere | Aerobic | Same |
| Incubation Time | 16-20 hours | Same |
| Reading Method | Automated (WalkAway or autoSCAN-4) or Manual | Same |
| Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Same |
| General Device Characteristic Differences | | |
| Antimicrobial Agent | Dried Aztreonam 0.5-64 µg/mL | Dried Ceftazidime 0.5-64 µg/mL |
K250084 - Page 4 of 15
{4}
VI Standards/Guidance Documents Referenced:
CLSI M07, 11th ed., "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard" (January 2018)
CLSI M100, 35th ed., "Performance Standards for Antimicrobial Susceptibility Testing" (January 2025)
Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A reproducibility study was conducted at three clinical sites using 10 isolates of Gram-negative organisms that were consistent with the device intended use. The range of aztreonam dilutions tested was 0.5-64 µg/mL. Isolates were tested in triplicate over three days at three clinical sites (27 data points per isolate). The isolates tested in the reproducibility study included 4 isolates of Pseudomonas aeruginosa, 3 isolates of Klebsiella aerogenes, 2 isolates of Klebsiella oxytoca, and 1 isolate of Enterobacter cloacae complex.
Inocula were prepared using both the turbidity and Prompt methods and results were read manually (visually) and with the WalkAway and autoSCAN-4 instrument systems. The mode (or median for results without a mode) of MIC values was determined for each isolate and the reproducibility was calculated based on the number of MIC values that fell within ± one doubling dilution of the mode/median MIC value. The majority of data points were within ± one doubling dilution of the mode/median MIC value. The data was analyzed as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. For those read/inoculation combinations that included off-scale results, reproducibility was assessed as best-case and worst-case scenarios (Table 1).
Table 1. Reproducibility of Aztreonam with All Inoculation and Read Methods
| Read Method | Reproducibility
No. within ± one dilution of the mode/median MIC value (%) | | | |
| --- | --- | --- | --- | --- |
| | Prompt Inoculation | | Turbidity Inoculation | |
| | Best | Worst | Best | Worst |
| WalkAway | 266/270 (98.5) | 259/270 (95.5) | 270/270 (100) | 264/270 (97.8) |
| autoSCAN-4 | 265/270 (98.1) | 258/270 (95.6) | 266/270 (98.5) | 256/270 (95.0) |
| Manual | 267/270 (98.9) | 262/270 (97.0) | 266/270 (98.5) | 261/270 (96.7) |
Reproducibility performance was considered acceptable for all inoculation and read methods.
K250084 - Page 5 of 15
{5}
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Not applicable.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Inoculum Density Check. A spectrophotometric device, the MicroScan Turbidity Meter, was used to ensure the accuracy of the turbidity inoculation method. A zero check of the turbidity meter was performed daily. The inocula prepared using the turbidity method were standardized using a reading of 0.08 ± 0.02 (equivalent to a 0.5 McFarland barium sulfate turbidity standard). The digital reading was recorded for each isolate and was considered acceptable based on recommendations in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. Inoculum density colony counts were evaluated from suspensions of the QC strain E. coli ATCC 25922 and were found to be within the acceptable concentration range as recommended in the CLSI document M07, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically.
Inoculum density data was collected for the Prompt inoculum preparation of all reproducibility isolates and weekly testing of QC strains E. coli ATCC 25922 and P. aeruginosa ATCC 27853, as well as monthly QC testing with the turbidity inoculation method. The overall average colony count was within the acceptable range for all isolates.
Purity Check. Purity checks were performed on all isolates for each inoculum preparation; only results from pure cultures were included.
Growth Failure Rate. During the clinical study, no isolates failed to grow on the dried test panels and frozen reference panel which is acceptable (<10% growth failure).
Quality Control Testing. The CLSI-recommended QC organisms E. coli ATCC 25922 and P. aeruginosa ATCC 27853 were tested with all inoculation and read methods using eight dilutions of aztreonam (0.5 – 64 µg/mL). The reference panel was inoculated using the turbidity method only. In this submission, the QC ranges for P. aeruginosa ATCC 27853 reflect the current MIC ranges recommended in the CLSI document M100, Performance Standards for Antimicrobial Susceptibility Testing 35th ed. The results of QC testing are shown in Table 2. For P. aeruginosa ATCC 27853, quality control results were within the acceptable range for all inoculation and read methods for ≥95% of tests which is acceptable. The device does not include the full CLSI/FDA-recommended dilution range for E. coli ATCC25922. Results ≤ the lowest dilution on the panel (i.e., ≤0.5 µg/mL) were considered an acceptable result and ≥95% of tests were in that range for all inoculation and read methods.
K250084 - Page 6 of 15
{6}
Table 2. Quality Control Results for all Inoculation and Read Methods for Aztreonam
| Organism | Conc. (μg/mL) ^{1} | Reference ^{2} | Prompt Inoculation Method | | | Turbidity Inoculation Method | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | autoSCAN-4 | Manual | WalkAway | autoSCAN-4 | Manual | WalkAway |
| E. coli ATCC 25922
Expected Range 0.06-0.5 μg/mL | ≤ 0.5 | 186 | 179 | 180 | 180 | 183 | 184 | 185 |
| | 1 | 1 | 5 | 3 | 4 | 1 | 2 | 1 |
| | 2 | | 2 | 3 | 2 | 1 | | 1 |
| | 4 | | 1 | | | | | |
| | 8 | | | 1 | | 1 | 1 | 1 |
| | 16 | | | | | | | |
| | 32 | | | | 1 | 2 | 2 | 2 |
| | 64 | 2 | | | | | | |
| | > 64 | | | | | | | |
| P. aeruginosa ATCC 27853
Expected Range 2-8 μg/mL | ≤ 0.5 | | | | | | | |
| | 1 | | | | | | | |
| | 2 | 2 | 117 | 113 | 89 | 123 | 114 | 87 |
| | 4 | 162 | 60 | 67 | 75 | 48 | 59 | 80 |
| | 8 | 24 | 11 | 8 | 18 | 11 | 13 | 16 |
| | 16 | | | | 7 | 2 | | 2 |
| | 32 | | | | | 1 | | 1 |
| | 64 | 1 | | | | | | |
| | > 64 | | | | | | | |
1 Does not include the full CLSI/FDA-recommended dilution range for QC testing of E. coli ATCC 25922 with the reference panel or the MicroScan panel. For E. coli ATCC25922, an acceptable result will be ≤ the lowest dilution on the panel (i.e., ≤0.5 μg/mL)<br/>
2 Frozen reference panel inoculated using the turbidity method and interpreted manually.
6. Detection Limit:<br/>
Not applicable/
7. Assay Cut-Off:<br/>
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:<br/>
The results obtained with the MicroScan Dried Gram-Negative MIC/Combo Panel with Aztreonam (Azt) (0.5-64 μg/mL) were compared to results obtained using a frozen broth microdilution reference panel (dilution range 0.5-64 μg/mL). Clinical isolates were evaluated at three testing sites in the U.S.; challenge isolates were evaluated at one internal and one external clinical site.<br/>
The reference panel was prepared as described in CLSI document M07-A11 except for the use of Pluronic-F (wetting agent) in the inoculum water for the reference panel. A summary of historical data from previously cleared antimicrobial tests was provided in the submission
K250084 - Page 7 of 15
{7}
which demonstrated that inclusion of the wetting agent did not affect testing. In addition, QC testing conducted during the clinical study was acceptable.
For the reference method and MicroScan panels inoculated using the turbidity method, panels were inoculated using the same standardized suspension further diluted into 25 mL of inoculum water with Pluronic-D (for the MicroScan panels) or Pluronic-F (for the frozen reference panels). MicroScan panels were also inoculated using the Prompt inoculation method with isolates inoculated into the Prompt inoculation bottle. Reference panels were read manually (visually) after 16-20 hours. MicroScan panels inoculated with both inoculation methods were read using the WalkAway and autoSCAN-4 instruments and by manual read after 16-18 hours.
## Clinical Study
To determine the performance of MicroScan Dried Gram-Negative MIC/Combo Panel with Aztreonam, a total of 622 gram-negative clinical isolates were evaluated with the Prompt inoculation method and the Turbidity inoculation method and all read methods at three clinical sites. Isolates included *Pseudomonas aeruginosa* (94 isolates) and Enterobacterales (528 isolates including 3 isolates of *Citrobacter amalonaticus*, 15 isolates of *Citrobacter fruendii*, 3 isolates of *Citrobacter koseri*, 41 isolates of *Enterobacter cloacae*, 1 isolate of *Enterobacter gergoviae*, 214 isolates of *Escherichia coli*, 1 isolate of *Hafnia alvei*, 17 isolates of *Klebsiella aerogenes* (formerly *Enterobacter aerogenes*), 14 isolates of *Klebsiella oxytoca*, 71 isolates of *Klebsiella pneumoniae*, 1 isolate of *Klebsiella spp.*, 16 isolates of *Morganella morganii*, 1 isolate of *pantoea agglomerans*, 89 isolates of *Proteus mirabilis*, 1 isolate of *Proteus penneri*, 2 isolates of *Proteus vulgaris*, 1 isolate of *Providencia rettgeri*, 3 isolates of *Providencia stuartii*, and 34 isolates of *Serratia marcescens*).
Performance when testing *Enterobacter cloacae* complex, *Serratia marcescens*, *Serratia liquefaciens* complex, *Proteus vulgaris*, *Providencia stuartii*, and *Providencia rettgeri* was not acceptable. These species were excluded from the intended use of the MicroScan Dried Gram-Negative MIC/Combo Panel with Aztreonam leaving a total of 540 clinical isolates (446 Enterobacterales and 94 *P. aeruginosa*) in the performance evaluation. This is addressed with the following limitation in the device labeling:
Results obtained with the organism/antimicrobial agent combinations listed below have shown discrepant MIC's when compared with an overnight reference method. Test results for these organisms will not be reported, and an alternate method for testing should be used.
Aztreonam: Enterobacter cloacae complex, Serratia marcescens, Serratia liquefaciens complex, Proteus vulgaris, Providencia stuartii, and Providencia rettgeri
Of the 540 clinical isolates (446 Enterobacterales and 94 *P. aeruginosa*) with results included in the performance analysis, a total of 471 (87.2%) were fresh isolates, and 69 (12.7%) were stock isolates (isolated from clinical specimens and tested beyond six months of isolation).
## Challenge Study
A total of 148 gram-negative challenge isolates were evaluated at two sites including *Pseudomonas aeruginosa* (21 isolates) and Enterobacterales (127 isolates including 6 isolates
K250084 - Page 8 of 15
{8}
of Citrobacter fruendii, 3 isolates of Citrobacter koseri, 5 isolates of Enterobacter cloacae, 21 isolates of Escherichia coli, 11 isolates of Klebsiella aerogenes (formerly Enterobacter aerogenes), 6 isolates of Klebsiella oxytoca, 11 isolates of Klebsiella pneumoniae, 14 isolates of Morganella morganii, 6 isolates of Proteus mirabilis, 6 isolates of Proteus vulgaris, 8 isolates of Providencia rettgeri, 8 isolates of Providencia stuartii, 4 isolates of Serratia liquefaciens, 9 isolates of Serratia marcescens, and 9 isolates of Yersenia enterocolitica). Results obtained with Enterobacter spp., Serratia spp., Proteus vulgaris, Providencia stuartii, and Providencia rettgeri were removed from the device indications for use and were not included in analysis, leaving a total of 108 challenge isolates (87 Enterobacterales and 21 P. aeruginosa isolates) in the performance evaluation. One isolate of P. aeruginosa was not assessed in the Prompt inoculation method due to colony mucoid morphology leaving only 20 isolates for this evaluation.
Results for essential agreement, category agreement, and categorical errors for Enterobacterales and Pseudomonas aeruginosa for all inoculation and read methods are shown in Table 3 and Table 4. Essential agreement of evaluable results was calculated considering MIC results that were clearly identical to reference method results or clearly one doubling dilution higher or lower than the reference method results. Performance was evaluated separately for each organism group (i.e., Enterobacterales and $P$ aeruginosa) due to differences in susceptibility test interpretive criteria.
For Y. enterocolitica, no resistant isolates were available for evaluation during clinical or challenge testing. This is addressed with the following limitation in the device labeling:
The ability of the MicroScan Dried Gram Negative Panels to detect resistance to aztreonam is unknown with Y. enterocolitica because an insufficient number of resistant strains were available at the time of comparative testing. Isolates yielding MIC results suggestive of a resistant interpretive category should be submitted to a reference laboratory.
Table 3. Performance of MicroScan Dried Gram-Negative Panels with Aztreonam, Using Prompt Inoculation and all Read Methods
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S/SDD | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| autoSCAN-4 Read | | | | | | | | | | | | | |
| Overall Enterobacterales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 422 | 94.6 | 57 | 33 | 57.9 | 428 | 96.0 | 46 | 392 | 10 | 8 | 0 |
| Challenge | 87 | 73 | 83.9 | 50 | 36 | 72 | 74 | 85.1 | 47 | 37 | 9 | 2 | 2 |
| Combined | 533 | 495 | 92.9 | 107 | 69 | 64.5 | 502 | 94.2 | 93 | 429 | 19 | 10 | 2 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 89 | 94.7 | 83 | 78 | 94.0 | 88 | 93.6 | 14 | 70 | 6 | 0 | 0 |
| Challenge | 20 | 19 | 95.0 | 20 | 19 | 95 | 12 | 60.0 | 8 | 4 | 8 | 0 | 0 |
| Combined | 114 | 108 | 94.7 | 103 | 97 | 94.2 | 100 | 87.7 | 22 | 74 | 14 | 0 | 0 |
| Manual Read | | | | | | | | | | | | | |
| Overall Enterobacterales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 422 | 94.6 | 57 | 33 | 57.9 | 426 | 95.5 | 46 | 392 | 11 | 9 | 0 |
| Challenge | 87 | 73 | 83.9 | 51 | 37 | 72.6 | 74 | 85.1 | 47 | 37 | 9 | 2 | 2 |
| Combined | 533 | 495 | 92.9 | 108 | 70 | 64.8 | 500 | 93.8 | 93 | 429 | 20 | 11 | 2 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 90 | 95.7 | 84 | 80 | 95.2 | 86 | 91.5 | 14 | 70 | 7 | 1 | 0 |
K250084 - Page 9 of 15
{9}
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S/SDD | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Challenge | 20 | 19 | 95.0 | 20 | 19 | 95 | 12 | 60.0 | 8 | 4 | 8 | 0 | 0 |
| Combined | 114 | 109 | 95.6 | 104 | 99 | 95.2 | 98 | 86.0 | 22 | 74 | 15 | 1 | 0 |
| WalkAway Read | | | | | | | | | | | | | |
| Overall Enterobacteriales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 416 | 93.3 | 63 | 33 | 52.4 | 424 | 95.1 | 46 | 392 | 10 | 12 | 0 |
| Challenge | 87 | 69 | 79.3 | 52 | 34 | 65.4 | 72 | 82.8 | 47 | 37 | 9 | 4 | 2 |
| Combined | 533 | 485 | 91.0 | 115 | 67 | 58.3 | 496 | 93.1 | 93 | 429 | 19 | 16 | 2 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 87 | 92.6 | 83 | 76 | 91.6 | 83 | 88.3 | 14 | 70 | 9 | 2 | 0 |
| Challenge | 20 | 17 | 85.0 | 20 | 17 | 85 | 15 | 75.0 | 8 | 4 | 5 | 0 | 0 |
| Combined | 114 | 104 | 91.2 | 103 | 93 | 90.3 | 98 | 86.0 | 22 | 74 | 14 | 2 | 0 |
*Includes 6.4% (34/533) of non-indicated Enterobacteriales
EA - Essential agreement
CA - Category agreement
EVAL - Evaluable isolates
min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
S - Susceptible
SDD - Susceptible-Dose Dependent
R - Resistant
Essential agreement (EA) occurs when the result of the reference method and that of the MicroScan Dried Gram-Negative MIC/Combo Panel are within plus or minus one serial two-fold dilution of the antibiotic. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the MicroScan Dried Gram-Negative MIC/Combo Panel.
Table 4. Performance of MicroScan Dried Gram-Negative Panels with Aztreonam, Using Turbidity Inoculation and all Read Methods
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S/SDD | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| autoSCAN-4 Read | | | | | | | | | | | | | |
| Overall Enterobacteriales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 436 | 97.8 | 47 | 37 | 78.7 | 432 | 96.9 | 46 | 392 | 9 | 4 | 1 |
| Challenge | 87 | 80 | 92.0 | 49 | 42 | 85.7 | 77 | 88.5 | 47 | 37 | 9 | 0 | 1 |
| Combined | 533 | 516 | 96.8 | 96 | 79 | 82.3 | 509 | 95.5 | 93 | 429 | 18 | 4 | 2 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 92 | 97.9 | 83 | 81 | 97.6 | 87 | 92.6 | 14 | 70 | 7 | 0 | 0 |
| Challenge | 21 | 21 | 100 | 20 | 20 | 100 | 12 | 57.1 | 9 | 4 | 9 | 0 | 0 |
| Combined | 115 | 113 | 98.3 | 103 | 101 | 98.1 | 99 | 86.1 | 23 | 74 | 16 | 0 | 0 |
| Manual Read | | | | | | | | | | | | | |
| Overall Enterobacteriales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 435 | 97.5 | 48 | 37 | 77.1 | 432 | 96.9 | 46 | 392 | 10 | 4 | 0 |
| Challenge | 87 | 81 | 93.1 | 50 | 44 | 88.0 | 76 | 87.4 | 47 | 37 | 10 | 0 | 1 |
| Combined | 533 | 516 | 96.8 | 98 | 81 | 82.7 | 508 | 95.3 | 93 | 429 | 20 | 4 | 1 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 91 | 96.8 | 84 | 81 | 96.43 | 86 | 91.5 | 14 | 70 | 7 | 1 | 0 |
| Challenge | 21 | 21 | 100 | 20 | 20 | 100 | 12 | 57.1 | 9 | 4 | 9 | 0 | 0 |
| Combined | 115 | 112 | 97.4 | 104 | 101 | 97.12 | 98 | 85.2 | 23 | 74 | 16 | 1 | 0 |
| WalkAway Read | | | | | | | | | | | | | |
| Overall Enterobacteriales*, ≤4 (S), 8 (I), ≥16 (R) | | | | | | | | | | | | | |
| Clinical | 446 | 436 | 97.8 | 48 | 38 | 79.2 | 431 | 96.7 | 46 | 392 | 10 | 5 | 0 |
| Challenge | 87 | 80 | 92.0 | 52 | 45 | 86.5 | 77 | 88.5 | 47 | 37 | 9 | 0 | 1 |
| Combined | 533 | 516 | 96.8 | 100 | 83 | 83.0 | 508 | 95.3 | 93 | 429 | 19 | 5 | 1 |
| Pseudomonas aeruginosa, ≤8 (S), 16 (I), ≥32 (R) | | | | | | | | | | | | | |
| Clinical | 94 | 90 | 95.7 | 84 | 80 | 95.24 | 86 | 91.5 | 14 | 70 | 6 | 2 | 0 |
| Challenge | 21 | 21 | 100 | 20 | 20 | 100 | 15 | 71.4 | 9 | 4 | 6 | 0 | 0 |
K250084 - Page 10 of 15
{10}
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S/SDD | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Combined | 115 | 111 | 96.5 | 104 | 100 | 96.15 | 101 | 87.8 | 23 | 74 | 12 | 2 | 0 |
*Includes 6.4% (34/533) of non-indicated Enterobacteriales
EA - Essential agreement
CA - Category agreement
EVAL - Evaluable isolates
min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
S - Susceptible
SDD - Susceptible-Dose Dependent
R - Resistant
Essential agreement (EA) occurs when the result of the reference method and that of the MicroScan Dried Gram-Negative MIC/Combo Panel are within plus or minus one serial two-fold dilution of the antibiotic. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the MicroScan Dried Gram-Negative MIC/Combo Panel.
The overall EA and CA performance for Enterobacteriales for the WalkAway, autoSCAN-4, and manual read methods were acceptable ( $>90\%$ ) for each inoculation method. A range of MAJ rates were observed for the different read and inoculation methods: $0.9\%$ (4/429) to $3.7\%$ (16/429), which is not acceptable. A range of VMJ rates were observed for the different read and inoculation methods: $1.1\%$ (1/93) to $2.1\%$ (2/93), which is acceptable. The unacceptable MAJ error rate observed for Enterobacteriales is addressed by the limitations described below.
When the performance for Enterobacteriales was evaluated individually by species, the following performance issues were observed:
When evaluating performance of $E$ coli, the overall EA and CA performance for $E$ coli for the WalkAway, autoSCAN-4, and manual read methods were acceptable with the prompt and turbidity inoculation methods. One very major error was observed, resulting in a VMJ rate of $3.2\%$ (1/31), which was unacceptable. This is addressed with the following limitation in the device labeling:
Due to the occurrence of a very major error involving a single $E$ coli isolate tested with aztreonam, isolates with an $MIC \leq 4 \mu \mathrm{g} / \mathrm{mL}$ run on the autoSCAN-4/Turbidity method should be read manually prior to reporting.
When evaluating performance of $P.$ mirabilis, the overall EA and CA performance for the manual read method was acceptable for the prompt inoculation method. The overall EA and CA performance for $P.$ mirabilis for the WalkAway, AutoSCAN-4, and manual read methods were acceptable with the turbidity inoculation method. However, four major errors were observed with the AutoSCAN-4 read method resulting in a MAJ rate of $4.4\%$ (4/92), six major errors were observed with the manual read method resulting in a MAJ rate of $6.5\%$ (6/92), and eight major errors were observed with the WalkAway read method resulting in a MAJ rate of $8.7\%$ (8/92) all using the prompt inoculation method, which was unacceptable. In addition, three major errors were observed with the WalkAway and turbidity inoculation method resulting in a MAJ rate of $3.3\%$ (3/92), which is not acceptable. This is addressed with the following limitations in the device labeling:
Performance of aztreonam when testing M. morganii and P. mirabilis using the Prompt inoculation method with all read methods was outside of essential agreement with
K250084 - Page 11 of 15
{11}
unacceptable major error rates compared to the reference method. M. morganii and P. mirabilis should only be tested with the turbidity inoculation method.
Due to the occurrence of major errors with aztreonam, P. mirabilis isolates with an MIC ≥ 16 µg/mL run on the WalkAway/Turbidity method should be read manually prior to reporting.
When evaluating performance of M. morgannii, the overall EA and CA performance for the WalkAway, AutoSCAN-4, and manual read methods were acceptable with the turbidity inoculation method. One major error was observed with the AutoSCAN-4 read method and manual read method resulting in a MAJ rate of 5.9% (1/17), and two major errors were observed with the WalkAway read method all using the prompt inoculation method resulting in a MAJ rate of 11.8% (2/17) which is not acceptable. This is addressed with a limitation in the device labeling (described above) to only test M. morganii with the turbidity inoculation method.
When evaluating performance of Citrobacter koseri, one very major error was observed with all read methods for the prompt inoculation method which resulted in a VMJ rate of 11.1% (1/9), which is not acceptable. This is addressed with the following limitation in the device labeling:
Performance of aztreonam when testing Citrobacter koseri have shown discrepant MICs (unacceptable essential agreement and an elevated very major error rate) when compared with the reference method across all read methods with the Prompt inoculation system. An alternate testing method (e.g. autoSCAN-4/WalkAway/manual reads with turbidity inoculation) should be used for aztreonam with Citrobacter koseri.
When evaluating performance of Citrobacter freundii complex, one very major error was observed with all read and inoculation methods resulting in a VMJ rate of 11.1% (1/9), which is not acceptable. This is addressed with the following footnote to the Performance Characteristics table in the device labeling:
For aztreonam, one resistant strain of Citrobacter freundii complex had a single very major error compared to the reference method across all read methods with turbidity inoculation.
When evaluating performance of Y. enterocolitica, one isolate was outside of essential agreement for the WalkAway read and Prompt inoculation method resulting in a EA rate of 88.9% (8/9), which is not acceptable. This is addressed with the following limitation in the device labeling:
Performance of aztreonam when testing Yersinia enterocolitica using the Prompt inoculation method with the WalkAway read method were within categorical agreement, but outside of essential agreement due to a single isolate when compared to the reference method. Yersinia enterocolitica results with the Prompt/WalkAway should be read manually prior to reporting.
The overall EA for P. aeruginosa for the WalkAway, autoSCAN-4, and manual read methods were acceptable (>90%) for both inoculation methods. In addition, a range of MAJ
K250084 - Page 12 of 15
{12}
rates was observed with all read methods when using the Prompt inoculation method: 0% (0/74) to 2.7% (2/74), which is acceptable. A range of MAJ rates was observed with all read methods when using the turbidity inoculation method: 0% (0/74) to 2.7% (2/74), which is acceptable. However, the CA was not acceptable across all read and inoculation methods. After further analysis, this was considered acceptable due to the acceptable performance of the evaluable isolates and the majority of the errors being minor. This is addressed with the following footnote to the Performance Characteristics table in the device labeling:
Categorical agreement for P. aeruginosa and aztreonam ranged from 85.2% (98/115) -- 87.8% (101/115) across all read and inoculation methods. Pseudomonas aeruginosa performance is acceptable due to the essential agreement of evaluable isolates ranging from 90.3% (93/103) -- 95.2% (99/104) and the majority of the categorical discrepancies were minor errors.
## Testing/Reporting MIC for Non-indicated Species
For this review, the interpretative criteria are applied to the organisms/organism groups according to the FDA STIC website. As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is added to the Warnings and Precautions section of the device labeling:
The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
## Trending
An analysis of trending was conducted using the combined clinical and challenge data for each organism group (Enterobacterales and P. aeruginosa) and for each inoculation and read method (Table 5). This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
Species for which the difference between the percentage of isolates with higher vs. lower readings was > 30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that shows higher or lower MIC values compared to the reference is addressed in the labeling.
No trending was observed for Enterobacterales or P. aeruginosa organism groups with all inoculation and read methods. At a species level, a trend toward higher MIC values was observed for P. mirabilis across both inoculation methods and all read methods. A trend toward lower MIC values was observed for K. oxytoca (turbidity inoculation/all read methods and Prompt inoculation/manual read method), M. morganii (turbidity inoculation/all read methods), and Y. enterocolitica (turbidity inoculation/autoSCAN-4 read method and turbidity inoculation/manual read method).
K250084 - Page 13 of 15
{13}
To address the MIC trending, the sponsor included the following footnote to the Performance Characteristics table in the device labeling:
Aztreonam MIC values for Enterobacterales were most frequently in exact agreement with the reference method. When not in agreement results tended to be one or more doubling dilutions higher for P. mirabilis with all read/all inoculation methods. When not in agreement results tended to be one doubling dilution lower for K. oxytoca and M. morganii with all read methods/turbidity inoculation, Y. enterocolitica with the autoSCAN-4/manual read methods/turbidity inoculation and K. oxytoca with the manual read /Prompt inoculation.
Table 5. Trending Observed for Aztreonam
| Inoculation/ Read Method | Organism Group | Total Evaluable for Trending | ≥1 Dilution Lower # (%) | Exact # (%) | ≥1 Dilution Higher # (%) | Percent Difference (95% CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Prompt/ autoSCAN-4 | Overall Enterobacterales | 130 | 47, (36.2) | 27 (20.7) | 56, (43.1) | 7%, (-5 to 18) | No |
| | Pseudomonas aeruginosa | 107 | 39, (36.4) | 60 (56.1) | 8, (7.5) | -29%, (-39 to -18) | No |
| Prompt/ Manual | Overall Enterobacterales | 126 | 45, (35.7) | 30 (23.8) | 51, (40.5) | 5%, (-7 to 16) | No |
| | Pseudomonas aeruginosa | 106 | 30, (28.3) | 62 (58.7) | 14, (13.2) | -15%, (-26, -4) | No |
| Prompt/ WalkAway | Overall Enterobacterales | 135 | 43, (31.9) | 29 (21.4) | 63, (46.7) | 15%, (3 to 26) | No |
| | Pseudomonas aeruginosa | 105 | 19, (18.1) | 65 (x) | 21, (20.0) | 2%, (-9 to 13) | No |
| Turbidity/ autoSCAN-4 | Overall Enterobacterales | 112 | 59, (52.7) | 26 (x) | 27, (24.1) | -29%, (-40 to -16) | No |
| | Pseudomonas aeruginosa | 106 | 42, (39.6) | 53 (x) | 11, (10.4) | -29%, (-40 to 18) | No |
| Turbidity/ Manual | Overall Enterobacterales | 112 | 55, (49.1) | 28 (x) | 29, (25.9) | -23%, (-35 to -11) | No |
| | Pseudomonas aeruginosa | 107 | 39, (36.5) | 55 (x) | 13, (12.2) | -24%, (-35 to -13) | No |
| Turbidity/ WalkAway | Overall Enterobacterales | 115 | 56, (48.7) | 30 (x) | 29, (25.2) | -23%, (-35 to -11) | No |
| | Pseudomonas aeruginosa | 106 | 30, (28.3) | 60 (x) | 16, (15.1) | -13%, (-24 to -2) | No |
## Resistance Mechanism Characterization
Challenge isolates of Enterobacterales and $P_{\cdot}$ aeruginosa harboring various molecular mechanisms of resistance were tested with aztreonam. Challenge isolates included strains from the CDC and FDA Antibiotic Resistance Isolate Bank for evaluation.
2. Matrix Comparison:
Not applicable
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
K250084 - Page 14 of 15
{14}
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
# D Clinical Cut-Off:
Not applicable.
# E Expected Values/Reference Range:
Table 6: FDA-Identified and Recognized Interpretive Criteria for Aztreonam (μg/mL)
| Organisms | Interpretive Criteriaa | | | |
| --- | --- | --- | --- | --- |
| | S | SDD | I | R |
| Enterobacterales | ≤4 | - | 8 | ≥16 |
| Pseudomonas aeruginosa | ≤8 | - | 16 | ≥32 |
S = Susceptible; SDD = Susceptible-dose dependent; I = Intermediate; R = Resistant
aFDA-Recognized Antimicrobial Susceptibility Test Interpretive Criteria Website https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria
# VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria). The protocol outlined the specific procedures and acceptance criteria that Beckman Coulter, Inc. intends to use to evaluate the MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) $(0.5 - 64\mu \mathrm{g} / \mathrm{mL})$ device when revised breakpoints for aztreonam are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, Beckman Coulter, Inc. will update the MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) $(0.5 - 64\mu \mathrm{g} / \mathrm{mL})$ device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K250084 - Page 15 of 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.