The Innovasis Endeavor™ Stand-Alone Cervical IBF System consists of a stand-alone interbody device indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The Endeavor device is intended to be used with the integrated fixation screws provided. The Endeavor device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach.
Device Story
Endeavor™ Stand-Alone Cervical IBF System is an intervertebral body fusion device with integrated fixation; used to stabilize cervical spinal segments to promote fusion, restrict motion, and decrease pain. Implanted via anterior cervical discectomy and fusion (ACDF) approach by surgeons. Device features PEEK body enhanced with HA and TCP; titanium faceplate; titanium alloy (Ti-6Al-4V ELI) fixation screws. Design includes tapered leading edge for insertion; convex profile; anti-migration teeth for stability; graft cavity for bone graft loading. Tantalum spheres provide radiographic visualization. Device restores vertebral height; radiolucent body allows fusion assessment. Supplied sterile with reusable instrumentation.
Clinical Evidence
Bench testing only. Performed static and dynamic torsion, axial compression, and compression shear testing per ASTM F2077; static subsidence testing per ASTM F2267; and static expulsion testing per FDA guidance. No clinical data were necessary.
Technological Characteristics
PEEK body with HA and TCP enhancement; titanium alloy (Ti-6Al-4V ELI) faceplate and screws per ASTM F136; tantalum markers per ASTM F560. Stand-alone interbody fusion device with integrated fixation. Sterile, single-use implant with reusable instrumentation.
Indications for Use
Indicated for anterior cervical interbody fusion in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1, following six weeks of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 25, 2025
Innovasis
Michael Thomas
Director Regulatory Affairs
614 E 3900 S
Salt Lake City, Utah 84107
Re: K250076
Trade/Device Name: Endeavor™ Stand-Alone Cervical IBF System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVE
Dated: August 26, 2025
Received: August 26, 2025
Dear Michael Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250076 - Michael Thomas
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250076 - Michael Thomas
Page 3
Sincerely,
EILEEN
CADEL-S
for
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250076
Device Name
Endeavor™ Stand-Alone Cervical IBF System
Indications for Use (Describe)
The Innovasis Endeavor™ Stand-Alone Cervical IBF System consists of a stand-alone interbody device indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.
This device is to be used in patients who have had six weeks of non-operative treatment. The Endeavor device is intended to be used with the integrated fixation screws provided. The Endeavor device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
510(k) #: K250076
510(k) Summary
Prepared on: 2025-09-23
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Innovasis |
| --- | --- |
| Applicant Address | 614 E 3900 S Salt Lake City UT 84107 United States |
| Applicant Contact Telephone | 801-261-2236 |
| Applicant Contact | Mr. Michael Thomas |
| Applicant Contact Email | mthomas@innovasis.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Endeavor™ Stand-Alone Cervical IBF System |
| --- | --- |
| Common Name | Intervertebral body fusion device |
| Classification Name | Intervertebral Fusion Device With Integrated Fixation, Cervical |
| Regulation Number | 888.3080 |
| Product Code(s) | OVE |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K181115 | CxHA PEEK Cervical IBF System | ODP |
| K162236 | AxHA Stand-Alone ALIF System | OVD |
| K220875 | HAcancellous PEEK-C Porous HA PEEK Cervical IBF System | ODP |
| K231899 | HAtetracell-C Titanium Cervical IBF System | ODP |
| K210497 | SeaSpine Shoreline ACS - Anterior Cervical Standalone, | OVE |
# Device Description Summary
21 CFR 807.92(a)(4)
The Endeavor™ Stand-Alone Cervical IBF System is an intervertebral body fusion (IBF) device with integrated fixation and associated instrumentation, used with bone graft material, that is intended to stabilize a cervical spinal segment to promote fusion which restricts motion and decreases pain. The Endeavor™ Stand-Alone Cervical IBF System is implanted via an Anterior Cervical Discectomy and Fusion (ACDF) surgical approach at one level from C2-T1 and is indicated for use in skeletally mature patients with degenerative disc disease (DDD). The Endeavor implants have a PEEK body enhanced with HA (Hydroxyapatite) and TCP (Tricalcium Phosphate) and feature a titanium faceplate and titanium fixation screws manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. The Endeavor implant features a tapered leading edge to aid in insertion due to limited anatomical space and features a slightly convex profile to match the anatomy and anti-migration teeth to ensure implant stability during the fusion process. The Endeavor implant features a graft cavity to provide volume for graft loading, and is radiolucent allowing assessment of the fusion process, while tantalum spheres per ASTM F560 enable implant visualization during and after the surgical procedure. The implant is available in multiple size options to match vertebral anatomy and is designed to restore height in the cervical spinal column during the fusion process. Implants are supplied sterile. Bone screws and reusable instruments to support the surgery are provided with the implants in sterilization sets.
Page 1 of 2
{5}
Page 2 of 2
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| --- | --- |
| The Innovasis Endeavor™ Stand-Alone Cervical IBF System consists of a stand-alone interbody device indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies.
This device is to be used in patients who have had six weeks of non-operative treatment. The Endeavor device is intended to be used with the integrated fixation screws provided. The Endeavor device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach. | |
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| The subject device has substantially equivalent indications for use as predicate device K210497 (SeaSpine Shoreline ACS). | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The subject device has substantially equivalent technological characteristics (i.e., design, material, chemical composition, principle of operation) as predicate and reference devices. While the material of the subject device is novel, additional characterization has shown the subject device to perform in a substantially equivalent manner as predicate devices. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The FDA Guidance for Cervical Intervertebral Body Fusion Devices, Class II Special Controls Guidance Document: Intervertebral Body Fusion Device (issued June 2007) recommends the following testing:
Static and dynamic torsion testing per ASTM F2077
Static and dynamic axial compression testing per ASTM F2077
Static and dynamic compression shear testing per ASTM F2077
Static subsidence testing per ASTM F2267
Static expulsion testing performed per recommendation in above guidance special controls document.
N/A - no clinical data were necessary.
The data submitted in this 510(k) submission demonstrates the substantial equivalence of the subject system as compared to the predicates identified and/or FDA published Cervical IBF data. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.