The uWS-Angio Pro is a post-processing software based on the uWS-Angio Basic for viewing, processing, evaluating and analyzing CT, MR, 3D XA images, and overlaying 2D medical images to support image guidance during interventional procedures. It has the following application: The uWS-Angio Pro uTarget view application is intended to provide tools for adding the trajectory to the lesion. It also supports overlay of 3D datasets on 2D fluoroscopy images. The uWS-Angio Pro uEmbo View application is intended to provide users with tools for planning and guiding interventional procedures. It offers a suite of features to mark targets, detect feeder vessels, and overlay the 3D image with the feeders and the corresponding anatomical X-ray image. When the overlay function is activated, it supports automatic 2D-3D image registration specifically for head neuroimaging, while the manual registration function is not restricted to any specific anatomical region. The uWS-Angio Pro uCardiac View application is intended to provide users tools to plan and guide the interventional procedures. This application can use CT angiography images for vascular access analysis, aortic root analysis and angle management, and overlay registration display with 2D XA images.
Device Story
Software for viewing, processing, and analyzing medical images (CT, MR, 3D XA) to support interventional procedure guidance. Operates on standalone workstations or networked systems; compatible with United Imaging Healthcare Angiography systems. Features include: uTarget View (trajectory planning, 3D/2D overlay); uEmbo View (target marking, feeder vessel detection, 2D-3D registration for neuroimaging); uCardiac View (vascular access/aortic root analysis, 2D/3D overlay). Clinicians use these tools to plan trajectories, identify vessels, and register 3D datasets to real-time 2D fluoroscopy. Output includes visual overlays, trajectory lines, and quantitative vascular measurements (area, diameter, perimeter). Benefits include improved procedural planning and guidance accuracy, potentially reducing clinical error and enhancing navigation during interventions.
Clinical Evidence
Bench testing only. Path generation validated on 40 clinical cases (112 vascular paths) via physician review. Neuro registration accuracy validated on 37 clinical cases (mean mTRE 0.24 mm). Cardiac segmentation validated on 39 CTA datasets using DICE coefficients (Aorta 0.987, LV 0.981, LCA 0.911, RCA 0.908). Parameter calculation validated against ground truth using ICC (>0.99) and Bland-Altman analysis.
Technological Characteristics
Standalone software; DICOM 3.0 compliant. Deterministic algorithms for segmentation and geometry-based parameter calculation. Supports 2D-3D registration and path generation. Deployed on standalone workstations or networked systems. Complies with NEMA PS 3.1-3.20, ISO 14971, and IEC 62304.
Indications for Use
Indicated for patients undergoing angiography and fluoroscopy-based procedures requiring image guidance. Supports viewing, processing, and analysis of CT, MR, and 3D XA images for interventional planning and guidance.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 24, 2025
Shanghai United Imaging Healthcare Co., Ltd.
% Xin Gao
RA Manager
No. 2258 Chengbei Rd. Jiading District
SHANGHAI, 201807
CHINA
Re: K250045
Trade/Device Name: uWS-Angio Pro
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: January 6, 2025
Received: August 25, 2025
Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250045 - Xin Gao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250045 - Xin Gao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250045 | |
| Device Name uWS-Angio Pro | |
| Indications for Use (Describe) The uWS-Angio Pro is a post-processing software based on the uWS-Angio Basic for viewing, processing, evaluating and analyzing CT, MR, 3D XA images, and overlaying 2D medical images to support image guidance during interventional procedures. It has the following application: The uWS-Angio Pro uTarget view application is intended to provide tools for adding the trajectory to the lesion. It also supports overlay of 3D datasets on 2D fluoroscopy images. The uWS-Angio Pro uEmbo View application is intended to provide users with tools for planning and guiding interventional procedures. It offers a suite of features to mark targets, detect feeder vessels, and overlay the 3D image with the feeders and the corresponding anatomical X-ray image. When the overlay function is activated, it supports automatic 2D-3D image registration specifically for head neuroimaging, while the manual registration function is not restricted to any specific anatomical region. The uWS-Angio Pro uCardiac View application is intended to provide users tools to plan and guide the interventional procedures. This application can use CT angiography images for vascular access analysis, aortic root analysis and angle management, and overlay registration display with 2D XA images. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250045
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
510 (k) SUMMARY
1. Date of Preparation:
September 24, 2025
2. Sponsor Identification
Shanghai United Imaging Healthcare Co., Ltd.
No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
Contact Person: Xin GAO
Position: Regulatory Affairs Specialist
Tel: +86-021-67076888-5386
Fax: +86-021-67076889
Email: xin.gao@united-imaging.com
3. Identification of Proposed Device
Trade Name: uWS-Angio Pro
Common Name: Angiography X-ray Image Processing Software
Model(s): uWS-Angio Pro
Regulatory Information
Classification Name: Medical image management and processing system
Classification: II
Product Code: LLZ,
Regulation Number: 21 CFR 892.2050
Review Panel: Radiology
4. Identification of Predicate Device(s)
Predicate Device
510(k) Number: K230955
Device Name: Syngo Application Software
Reference Device#1
510(k) Number: K142273
Device Name: EmboGuide
Page 1
{5}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
U
Reference Device#2
510(k) Number: K233209
Device Name: uOmnispace.CT
# 5. Device Description
uWS-Angio Pro is a software intended for viewing, processing, evaluating and analyzing medical images that comply with the DICOM 3.0 protocol. It supports interpretation and evaluation of examinations within healthcare institutions. It can be deployed on independent hardware such as a stand-alone diagnostic review and post-processing workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, it can be deployed on systems of the United Imaging Healthcare Angiography system family.
uWS-Angio Pro contains the following applications:
- uWS-Angio Pro uTarget View
- uWS-Angio Pro uEmbo View
- uWS-Angio Pro uCardiac View
# 6. Indications for use
The uWS-Angio Pro is a post-processing software based on the uWS-Angio Basic for viewing, processing, evaluating and analyzing CT, MR, 3D XA images, and overlaying 2D medical images to support image guidance during interventional procedures. It has the following application:
The uWS-Angio Pro uTarget view application is intended to provide tools for adding the trajectory to the lesion. It also supports overlay of 3D datasets on 2D fluoroscopy images.
The uWS-Angio Pro uEmbo View application is intended to provide users with tools for planning and guiding interventional procedures. It offers a suite of features to mark targets, detect feeder vessels, and overlay the 3D image with the feeders and the corresponding anatomical X-ray image. When the overlay function is activated, it supports automatic 2D-3D image registration specifically for head neuroimaging, while the manual registration function is not restricted to any specific anatomical region.
The uWS-Angio Pro uCardiac View application is intended to provide users tools to plan and guide the interventional procedures. This application can use CT angiography images for vascular access analysis, aortic root analysis and angle management, and overlay registration display with 2D XA images.
# 7. Summary of Technological Characteristics
The technology characteristics of the proposed device reflected in this 510(k)
Page 2
{6}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
submission are substantially equivalent to those of the predicate device and reference devices.
The following tables present a comparative analysis of the main features, principles of operation, fundamental scientific technology and intended use of the proposed device in relation to both predicate device and reference devices.
Page 3
{7}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
Table 1 General information comparison
| Item | Proposed Device uWS-Angio Pro | Predicate Device syngo Application Software (K230955) | Remark |
| --- | --- | --- | --- |
| General | | | |
| Device Classification Name | Medical Image Management and Processing System | Medical Image Management and Processing System | Same |
| Product Code | LLZ | LLZ | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Device Class | II | II | Same |
| Classification Panel | Radiology | Radiology | Same |
| Advanced Application | Yes | Yes | Same |
The proposed device uWS-Angio Pro and the predicate device syngo Application Software share similar indications for use: providing image viewing and processing to support device guidance during interventional procedures. Both devices are indicated for patients undergoing angiography and fluoroscopy-based procedures and are designed to support healthcare professionals in healthcare institution settings. The intended patient population and applicable clinical scenarios are the same for these two devices. The proposed device includes fewer applications compared to the predicate device. The difference will not impact the safety and effectiveness of the proposed device. Therefore, the proposed device can be considered substantially equivalent to the predicate device in terms of Indications for Use.
Table 2 Substantial equivalence discussion for uWS-Angio Pro applications
| Application | Function name | Proposed device uWS-Angio Pro | Predicate device syngo Application Software | Reference device#1 EmboGuide (K142273) | Reference device#2 uOmnispace.CT (K233209) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
{8}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
| | | | (K230955) | | | |
| --- | --- | --- | --- | --- | --- | --- |
| uTarget View | Trajectory planning | Yes | Yes | / | / | Note 1 |
| | Overlay | Yes | Yes | / | / | Same |
| | Sending position data | Yes | Yes | / | / | Note 2 |
| uEmbo View | Feeder Detection | Yes | / | Yes | / | Same |
| | Overlay | Yes | / | Yes | / | Same |
| uCardiac View | Vascular Access Analysis | Yes | / | / | Yes | Note 3 |
| | Aortic Root Analysis | Yes | / | / | Yes | Note 4 |
| | Overlay | Yes | Yes | / | / | Same |
| | Sending angle | Yes | Yes | / | / | Note 2 |
Note 1: Compared to the predicate device, uTarget view application is intended to provide tools for adding lesion markers (bounding box) and the trajectory to the lesion. The difference doesn't impact the safety and effectiveness of the subject device.
Note 2: Compared to the predicate device, position data of the proposed device are compatible to the Angiography Systems of United-Imaging Healthcare. This difference between the proposed device and the reference device doesn't impact the safety and effectiveness of the device.
Note 3: Compared to the reference device#2, the name of the function is different. The Vascular Access Analysis is the same as the Vessel analysis function of Cardiovascular Combined Analysis application in the reference device#2. The difference doesn't impact the safety and effectiveness of the subject device.
Note 4: Compared to the reference device#2, the Aortic Root Analysis is the same as the combination of TAVI evaluation function and Heart analysis function of Cardiovascular Combined Analysis application in the reference device#2. The difference doesn't impact the safety and effectiveness of the subject device.
Page 5
{9}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
# 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility
Not Applicable to the proposed device, because the device is stand-alone software.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Not Applicable to the proposed device, because the device is stand-alone software.
## Software
Software verification and validation testing was provided to demonstrate safety and efficacy of the proposed device. This includes a hazard analysis, and the potential hazards to satisfy the recommended documentation for basic documentation. These documentations include:
- Software Description
- Device Hazard Analysis
- Software Requirements Specification (SRS)
- Software Architecture Design Chart
- Software Development Environment Description
- Software Verification and Validation
- Cybersecurity Documents
## Animal Study
No animal study was required.
## Clinical Studies
No clinical study was required.
## Performance Verification
(1) uTarget View
> Workflow Test
a) Device Output
- Trajectory
- entry position data and lateral position data
- overlay planned trajectory with 3D image and 2D XA image
b) Test Data Description
- XA CBCT series and CT\MR data of the same patient
c) Test Method
- System Test
Page 6
{10}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
d) Results
- The uTarget View generated a trajectory line that is displayed on both the MPR and VR images and added to the trajectory list.
- Sent entry position data and lateral position data to the United Imaging Healthcare Angiography system.
- The uTarget View planned trajectory with 3D image and 2D XA image.
(2) uEmbo View
> Path Generation Algorithm
Based on XA 3D abdominal data or XA 3D head data, the paths to the target points from the catheter port are calculated based on user-marked locations.
e) Device Output
- Three-dimensional path to the target points from the catheter port, and it does not involve outputting quantitative parameters.
f) Test Data Description
- 40 cases of XA 3D data, including 62 hepatic artery branches from 23 patients and 50 intracranial artery branches from 17 patients.
g) Test Method
The path generation result was validated through a subjective evaluation conducted by U.S.-certified physicians.
h) Results
To verify the method's effectiveness, we evaluated it on 112 vascular paths from 40 clinical cases. Independent physicians reviewed the results and confirmed that all paths were successfully generated without any critical clinical errors, such as vessel wall penetration or incorrect branching. All paths received final clinical approval.
> Neuro Registration Algorithm
Based on 2D XA images and 3D XA images, perform 2D-3D registration for data in neural scenes.
a) Device Output
The result of 2D-3D registration, and it does not involve outputting quantitative parameters.
b) Test Data Description
The test dataset consisted of 37 real clinical cases of head XA data. The dataset was clinically diverse, including patients of varying ages, genders, and disease types (e.g., arterial stenosis, cerebral infarction), and encompassed challenging scenarios such as brain anatomy shift.
c) Test Method
Page 7
{11}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
The accuracy of the registration was evaluated using the mTRE between corresponding points in key anatomical regions, pre- and post-registration.
d) Results
All 37 cases across all predefined transformations passed the verification criterion. The overall mean mTRE was 0.24 mm (95% CI: 0.21 to 0.28 mm), significantly below the 1 mm;
The average registration time was 1.75 s (95% CI: 1.69 to 1.82 s).
(3) uCardiac View
> Aortic Root Vessel Extraction
Based on the 3D CTA images, we use deterministic algorithm to segment the aortic root.
a) Device Output
Segmentation results
♦ Aorta
♦ Left Ventricle
♦ Left coronary artery
♦ Right coronary artery
b) Test Data Description
A total of 39 CTA datasets were included, comprising 25 patients from the USA and 14 from China. Among these patients, 25 presented with calcification.
c) Test Method
Evaluation with the ground truth, with evaluation conducted using the DICE coefficient.
d) Results
| Volume Targets | Aorta | Left Ventricle | Left Coronary Artery | Right Coronary Artery |
| --- | --- | --- | --- | --- |
| DICE(Algorithm result) | 0.987 (95% CI 0.9862–0.9878) | 0.981 (95% CI 0.9786–0.9834) | 0.911 (95% CI 0.9047–0.9173) | 0.908 (95% CI 0.9002–0.9158) |
> Aortic Vessel Extraction Algorithm
Based on the 3D CTA images, we use deterministic algorithm to segment the aorta.
a) Device Output
Segmentation results
■ Aorta
Page 8
{12}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
b) Test Data Description
A total of 39 CTA datasets were included, comprising 13 patients from the USA and 26 from China. The patient cohort included 10 cases with calcification, 7 with aneurysm, 4 with aortic dissection, and 7 with hematoma.
c) Test Method
Evaluation with the ground truth, with evaluation conducted using the DICE coefficient.
d) Results
The DICE value is 0.968(95% CI 0.9652-0.9708).
> Parameter Calculation
Based on the 3D CTA images and the preceding image processing, we use simple geometry principle to calculate the parameters, including area, diameter, distance, perimeter and length.
a) Device Output and Quantitative Parameters
Area
Diameter
Perimeter
b) Test Data Description
Aortic Root Parameter Calculation: A total of 39 CTA datasets were included, comprising 25 patients from the USA and 14 from China. Among these patients, 25 presented with calcification.
Aortic Vessel Parameter Calculation: A total of 39 CTA datasets were included, comprising 13 patients from the USA and 26 from China. The patient cohort included 10 cases with calcification, 7 with aneurysm, 4 with aortic dissection, and 7 with hematoma.
c) Test Method
Evaluation with the ground truth, with evaluation conducted using the intraclass correlation coefficient (ICC) and Bland Altman plots.
d) Results
Aortic Root Parameter Calculation:
| Parameters | Analysis Results (including ICC and BA-plot) |
| --- | --- |
| Area | ICC: 0.9995 (95% CI: 0.9994 to 0.9995)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -15.5 to +19.0 mm²) |
Page 9
{13}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
| | |
| --- | --- |
| Maximum diameter | 0.9985 (95% CI: 0.9980 to 0.9989)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -0.64 to +0.63 mm)
Mean of uCardiacView_area and ROI_area |
| Minimum diameter | 0.9992 (95% CI: 0.9989 to 0.9994)
Bland-Altman analysis: 97.4% of points within limits of agreement (LoA: -0.56 to +0.56 mm)
Mean of uCardiacView_Dmax and ROI_Dmax |
| Perimeter | 0.9957 (95% CI: 0.9952 to 0.9960)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -2.2 to +2.6 mm) |
Page 10
{14}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING

## Aortic Vessel Parameter Calculation:
| Parameter | ICC results & Bland-Altman Analysis |
| --- | --- |
| Area | 0.9991 (95% CI: 0.9988 to 0.9993)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -13.2 to +12.3 mm²)
Mean of uCardiacView_area and ROI_area |
| Maximum diameter | 0.9983 (95% CI: 0.9969 to 0.9991)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -0.75 to +0.69 mm)
Mean of uCardiacView_Dmax and ROI_Dmax |
Page 11
{15}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
| Minimum diameter | 0.9979 (95% CI: 0.9960 to 0.9989)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -0.75 to +0.73 mm) |
| --- | --- |
| Perimeter | 0.9980 (95% CI: 0.9961 to 0.9989)
Bland-Altman analysis: 100% of points within limits of agreement (LoA: -2.3 to +2.6 mm) |
Note: features driven by user-input and the features permitting user editing as follow:
### Features driven by user input
| Application | Algorithm | Feature | User input |
| --- | --- | --- | --- |
| uTargetView | N.A. | Path Planning | Adding the entry spot and the target spot |
| | N.A. | Bounding Box ROI Placement | Holding down and dragging the mouse on the lesion to draw the long axis. |
| uEmboView | Path Generation | Feeder Detection | The user needs to mark the target points and catheter port |
### Features permitting user editing
Page 12
{16}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
UNITED IMAGING
| Application | Algorithm | Editing Content |
| --- | --- | --- |
| uEmboView | Path Generation | Path Results |
| | Neuro Registration | Registration Results |
| uCardiacView | Aortic Root Vessel Extraction | Segmentation Results |
| | Aortic Root Parameter Calculation | Vascular Contour |
| | Aortic Vessel Extraction | Segmentation Results |
| | Aortic Vessel Parameter Calculation | Vascular Contour |
## Other Standards and Guidance
- NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2023e).
- ISO 14971 Medical devices - Application of risk management to medical devices (Third Edition 2019-12).
- IEC 62304 Medical device software - Software life cycle processes (Edition 1.1, 2015).
- Cybersecurity in Medical Devices Quality System Considerations and Content of Premarket Submissions.
- Content of Premarket Submissions for Device Software Functions.
## Summary
The features described in this premarket submission are supported with the results of the testing mentioned above; the uWS-Angio Pro is found to have a safety and effectiveness profile that is similar to the predicate device and reference devices.
## 9. Substantially Equivalent (SE) Conclusion
The proposed device is equivalent to the predicate device and reference devices in regard to safety and efficacy. This conclusion is based upon a comparison of intended use, technological characteristics, performance specification, device hazards as well as verification and validation results.
In summary, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device and reference device.
Page 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.