K250041 · Zhengzhou PZ Laser Slim Technology Co., Ltd. · GEX · May 14, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250041
Device Name
Medical Picosecond ND: YAG Laser System (PZ-6)
Applicant
Zhengzhou PZ Laser Slim Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medical Picosecond ND:YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows: 1064nm wavelength: - Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
Device Story
Medical Picosecond ND:YAG Laser System (PZ-6) is a pulsed laser device for dermatological and surgical applications. It utilizes a host unit with a laser power supply, Nd:YAG laser source, and water-cooling system to generate 1064nm and 532nm wavelengths. The laser energy is delivered via an articulated arm and handpiece to the target tissue. Operated by a clinician in a medical setting, the device is activated via a foot switch. The system features an LCD touch screen for parameter control. By delivering picosecond-range pulses, the device targets tattoo pigments and benign pigmented lesions, facilitating their removal or treatment. The device is intended for prescription use only.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Nd:YAG laser source; 1064nm/532nm wavelengths; 300-500ps pulse duration; 500mJ/250mJ max energy; articulated arm delivery; water circulation cooling; LCD touch screen interface; AC100-240V power. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, and ISO 10993 (biocompatibility).
Indications for Use
Indicated for tattoo removal (1064nm: Fitzpatrick I-VI, black/brown/green/blue/purple; 532nm: Fitzpatrick I-III, red/yellow/orange) and treatment of benign pigmented lesions (Fitzpatrick I-IV) in dermatology and plastic surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2025
Zhengzhou PZ Laser Slim Technology Co., Ltd.
% Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lujiazui Rd., Pudong
Shanghai, 200120
China
Re: K250041
Trade/Device Name: Medical Picosecond ND: YAG Laser System (PZ-6)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: April 15, 2025
Received: April 15, 2025
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250041 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250041 - Boyle Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.05.14
21:16:01 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250041 | |
| Device Name Medical Picosecond ND: YAG Laser System (PZ-6) | |
| Indications for Use (Describe) The Medical Picosecond ND:YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows: 1064nm wavelength: - Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
# 510(k) Summary
# K250041
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Zhengzhou PZ Laser Slim Technology Co., Ltd.
Address: Room 201, 2nd Floor, Building 2, No. 52 Hongsong Road, High-tech Development Zone, 450001 Zhengzhou City, Henan Province, P.R.China
Tel: 86-18403935976
Contact: Junmei Li
## Designated Submission Correspondent
Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China
Tel: +86-21-50313932
Email: Info@truthful.com.cn
Date of Preparation: Jan.03rd,2025
## 2.0 Device Information
Trade name: Medical Picosecond ND: YAG Laser System
Common name: Powered Laser Surgical Instrument
Classification name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Model: PZ-6
Production code: GEX
Regulation number: 21CFR 878.4810
Classification: Class II
Panel: General & Plastic Surgery
## 3.0 Predicate Devices and Reference Device
### Predicate Device:
{5}
510(k) Summary
Manufacturer: Shanghai Apolo Medical Technology Co., Ltd.
Trade/Device Name: PicoSecond Nd: YAG Laser System
510(k) number: K200116
## Reference Device 1:
Manufacturer: AMI Inc.
Trade/Device Name: PICO LEGEND Nd:YAG Laser System
510(k) number: K233007
## Reference Device 2:
Manufacturer: Beijing ADSS Development Co., Ltd.
Trade/Device Name: Picosecond Laser System
510(k) number: K220268
## 4.0 Device Description
The Medical Picosecond ND: YAG Laser System mainly consists of a host, a light guide arm, a treatment handle, a foot switch, a power cord, and other auxiliary treatment accessories. The host mainly includes laser power supply, laser, water circulation cooling system, control board, and display screen. Accessories include a foot switch, laser protective glasses, power cords, and a water filling funnel.
The Medical Picosecond ND: YAG Laser System is a multi-wavelength, pulsed laser system designed for the treatment of benign pigmented lesions. A key feature of the device is its ability to produce multiple laser wavelengths (1064 nm and 532 nm).
## 5.0 Indication for Use Statement
The Medical Picosecond ND: YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
532nm wavelength:
- Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
## 6.0 Comparison to the Predicate Device
{6}
510(k) Summary
| Item | Subject device K250041 | Predicate device K200116 | Reference device 1 K233007 | Reference device 2 K220268 |
| --- | --- | --- | --- | --- |
| Trade/Device Name | Medical Picosecond ND: YAG Laser System | PicoSecond Nd: YAG Laser System | PICO LEGEND Nd:YAG Laser System | Picosecond Laser System |
| Manufacturer | Zhengzhou PZ Laser Slim Technology Co., Ltd. | Shanghai Apolo Medical Technology Co., Ltd. | AMI Inc. | Beijing ADSS Development Co., Ltd |
| Class &Code | Class II GEX 878.4810 | Class II GEX 878.4810 | Class II GEX 878.4810 | Class II GEX 878.4810 |
| Intended Use/Indication for Use | The Medical Picosecond ND:YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | The PicoSecond Nd: YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | The PICO LEGEND Nd:YAG Laser System is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
-The 1064nm wavelength of the PICO LEGEND Nd:YAG Laser System is indicated for tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI).
-The 532nm wavelength of the | The Picosecond Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. |
3 / 8
{7}
510(k) Summary
| | 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. | PICO LEGEND Nd:YAG Laser System is indicated for tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III. | 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. |
| --- | --- | --- | --- | --- |
| Type of Use | Prescription Use | Prescription Use | Prescription Use | Prescription Use |
| Wavelength | 1064/532 nm | 1064/532 nm | 1064/532 nm (Accuracy ±10%) | 1064/532 nm |
| Pulse Duration (Pulse Width) | 300 – 500 ps | 300 – 500 ps | 300 ps - Both Modes | 500 ps – Both Modes |
| Pulse Energy | 500mJ (1064nm)
250mJ (532nm) | 500mJ (1064nm)
250mJ (532nm) | 1064nm mode: 100 - 500mJ (Adjustable by 25mJ)
532nm mode: 50 - 250mJ (Adjustable by 25mJ) | 500mJ (1064nm)
250mJ (532nm) |
| Aiming Beam | 635nm | Not Publicly Available | 635nm | 635 nm |
| Repetition Rate | 1Hz-10Hz,
with a step of 1Hz | Single, 1, 2, 3, 4, 5,
6, 7, 8, 9, 10 Hz | 1Hz-10Hz,
(Adjustable by 1Hz) | 1Hz-10Hz |
4 / 8
{8}
510(k) Summary
| Spot size | 2mm -10mm, ±20%, with a step of 1mm, - All treatment Modes | Adjustable spot size 2~10mm | 2mm - 10mm (Adjustable by 1mm) - All treatment Modes | 2mm - 10mm - All treatment Modes |
| --- | --- | --- | --- | --- |
| Maximum Average Fluence (J/cm²) | 1064 nm: 16 J/cm²
532 nm: 8 J/cm² | Not Publicly Available | 1064 nm: 15.92 J/cm²
532 nm: 7.96 J/cm² | 1064 nm: 15.5 J/cm²
532 nm: 8 J/cm² |
| Laser Type | Nd:YAG | Nd:YAG | Nd:YAG | Nd:YAG |
| Activation | foot-switch | foot-switch | foot-switch | foot-switch |
| Display | LCD Touch screen | LCD Touch screen | LCD Touch screen | LCD Touch screen |
| Cooling System | Water circulation cooling system | Cooling System | Cooling System | Internal water to air heat exchanger |
| Electrical Power | AC100-240V 50/60Hz | 120VAC 10A, 50/60Hz | Not Publicly Available | AC 230V, 50Hz |
| Beam Delivery System | Articulated Arm with Handpiece | Articulated Arm with Handpiece | Articulated arm with handpiece | Articulated arm with handpiece |
5 / 8
{9}
510(k) Summary
| System dimension | 98cm × 35 cm × 180 cm | 97cm H x 48cm W x 97cm D | Not Publicly Available | 93 cm × 40 cm × 93 cm |
| --- | --- | --- | --- | --- |
| System Weight (kg) | 186 kg | 130kg | Not Publicly Available | 128kg |
6 / 8
{10}
510(k) Summary
## 7.0 Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the subject devices met all design specifications as were Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-1:2005/AMD2:2020 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and essential performance
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC/TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
- IEC 60601-1-6:2013, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-2-22:2019, Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
- ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- ISO10993-10:2021 Biological evaluation of medical devices -- Part 10: Tests for skin sensitization
- ISO10993-23:2021 Biological evaluation of medical devices -- Part 23: Tests for irritation
Software Information:
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
## 8.0 Clinical Test Conclusion
No clinical study is included in this submission.
## 9.0 Conclusion
Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. And based on the performance
{11}
510(k) Summary
testing and compliance with acceptable voluntary standards, we believe the subject device is substantially equivalent to its predicate device.
8 / 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.