K250035 · Mim Software, Inc. · QKB · Feb 3, 2025 · Radiology
Device Facts
Record ID
K250035
Device Name
Contour ProtégéAI+
Applicant
Mim Software, Inc.
Product Code
QKB · Radiology
Decision Date
Feb 3, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K250035 · Feb 3, 2025
Contour ProtégéAI+
Mim Software, Inc.
Clinical treatment plan images (CT) from 32 institutions
The device models were trained and validated using a dataset of 770 clinical images sourced from routine radiotherapy treatment plans. The performance of the subject device was compared against a reference device (MIM Maestro) using these clinical datasets to demonstrate substantial equivalence.
Model Training and Validation Dataset; Retrospective analysis of clinical images
Patients prescribed external beam or molecular radiotherapy; Sample Size: 770 images; Number of Sites: 32 institutions
MIM Maestro (K071964) reference device
Dice score, Mean Distance to Agreement (MDA), and cumulative Added Path (APL)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Anatomical structure segmentation
Neural network models
Statistical non-inferiority of Dice score and MDA compared to reference; average user evaluation score >= 2
Dice and MDA scores equivalent or superior to reference device
770 images from 32 institutions; re-segmented by consultants and reviewed by MIM Software staff.
>1 (consultants (physicians and dosimetrists)) + >1 (qualified staff (MD or licensed dosimetrists))
Validation set of images from institutions disjoint from training datasets.
>1 (trained medical professionals)
Indications for Use
Contour ProtégéAl is an accessory to MIM software used for the contouring of anatomical structures in imaging data using machine-learning-based algorithms automatically. Appropriate image visualization software must be used to review and, if necessary, edit results automatically generated by Contour ProtégéAI. Contour ProtégéAl is not intended to detect or contour lesions.
Device Story
Contour ProtégéAI+ is a software accessory for MIM software that automatically segments anatomical structures on CT and MR images. It utilizes machine-learning neural network models to generate contours, which are then reviewed and edited by clinicians using visualization software. The device is used in clinical settings (e.g., radiation oncology) to assist in quantitative analysis, adaptive therapy, and treatment planning. It operates on Windows, Mac, or Linux workstations, or via the MIMcloud service. By automating the time-consuming manual contouring process, the device aims to improve efficiency in radiation therapy workflows. Clinicians use the output to inform treatment planning and patient management. The device is not intended for lesion detection.
Clinical Evidence
Bench testing only. Evaluated 770 images from 32 institutions. Performance measured against MIM Maestro reference device using Dice coefficient, Mean Distance to Agreement (MDA), and cumulative Added Path Length (APL). Results demonstrated non-inferiority (p=0.05) for most structures. User evaluation on a three-point scale confirmed time-saving functionality. Training data included diverse scanner vendors (GE, Siemens, Philips, Toshiba) and patient demographics (53.4% female, 31.3% male, 15.3% unknown; 50% over age 60).
Technological Characteristics
Software-based medical image processing system. Utilizes machine-learning neural network models for automated segmentation. Operates on Windows, Mac, and Linux platforms; supports local or cloud-based (MIMcloud) deployment. Compatible with ACR/NEMA DICOM 3.0 standard images. No physical materials or energy sources.
Indications for Use
Indicated for trained medical professionals to assist in automated processing of digital CT and MR images (ACR/NEMA DICOM 3.0). Supports creation of contours via machine-learning for quantitative analysis, adaptive therapy, transfer to radiation therapy planning systems, and archiving. Enables segmentation of anatomical structures across CT locations and prostate, seminal vesicles, and urethra in T2-weighted MR images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 3, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal of the Department of Health & Human Services - USA, which features an abstract image of an eagle. To the right of the seal, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue color. The word "FDA" is in a square box.
MIM Software Inc. Sydney Lindner Regulatory Affairs Engineer 25800 Science Park Drive, Suite 180 Cleveland, Ohio 44122
Re: K250035
Trade/Device Name: Contour ProtégéAI+ Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB, Dated: January 7, 2025 Received: January 7, 2025
Dear Sydney Lindner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K250035
Device Name
Contour ProtégéAl+
Indications for Use (Describe)
Trained medical professionals use Contour ProtégéAI as a tool to assist in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. In addition, Contour ProtégéAl supports the following indications:
· Creation of contours using machine-learning algorithms for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
· Segmenting anatomical structures across a variety of CT anatomical locations.
· And segmenting the prostate, the seminal vesicles, and the urethra within T2-weighted MR images.
Appropriate image visualization software must be used to review and, if necessary, edit results automatically generated by Contour ProtégéAI.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a simple color palette.
#### 510(k) Summary (The following information is in conformance with 21 CFR 807.92)
## Submitter
K250035
MIM Software Inc. 25800 Science Park Drive - Suite 180 Cleveland, OH 44122
| Phone:<br>Fax: | 216-455-0600<br>216-455-0601 |
|------------------------|------------------------------|
| Contact Person: | Sydney Lindner |
| Date Summary Prepared: | January 7, 2025 |
## Device Name
Trade Name:
Common Name:
Regulation Number / Product Code:
Classification Name:
Contour ProtégéAl+
Medical Imaging Software
21 CFR 892.2050 Product Code QKB
System, Imaging Processing, Radiological
## Predicate and Reference Devices
| Predicate: | K231765 | Contour ProtégéAl | MIM Software Inc. |
|------------|---------|------------------------------------------|-------------------|
| Reference: | K071964 | MIM 4.1 (SEASTAR)<br>[i.e., MIM Maestro] | MIMvista Corp. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, to the left of the text "mim SOFTWARE." The text is in a sans-serif font, with "mim" in a larger size than "SOFTWARE."
## Intended Use
Contour ProtégéAl is an accessory to MIM software used for the contouring of anatomical structures in imaging data using machine-learning-based algorithms automatically.
Appropriate image visualization software must be used to review and, if necessary, edit results automatically generated by Contour ProtégéAI.
Contour ProtégéAl is not intended to detect or contour lesions.
#### Indications for Use
Trained medical professionals use Contour ProtégéAl as a tool to assist in the automated processing of digital medical images of modalities CT and MR, as supported by ACR/NEMA DICOM 3.0. In addition, Contour ProtégéAl supports the following indications:
- Creation of contours using machine-learning algorithms for applications ● including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
- · Segmenting anatomical structures across a variety of CT anatomical locations.
- And segmenting the prostate, the seminal vesicles, and the urethra within ● T2-weighted MR images.
Appropriate image visualization software must be used to review and, if necessary, edit results automatically generated by Contour ProtégéAl.
## Device Description
Contour ProtégéAl+ is an accessory to MIM software that automatically creates contours on medical images through the use of machine-learning algorithms. It is designed for use in the processing of medical images and operates on Windows, Mac, and Linux computer systems. Contour ProtégéAl+ is deployed on a remote server using the MIMcloud service for data management and transfer; or locally on the workstation or server running MIM software.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it.
Indications for use have not been modified. The intended use is the same as the last 510(k) clearance for Contour ProtégéAl (K231765).
In the upcoming 1.4.0 release, one new neural network model is being added and three existing neural network models are being updated by retraining with additional or updated contours which prompted this premarket submission of the subject device.
| ITEM | Subject Device:<br>Contour ProtégéAl+<br>(K250035) | Predicate Device:<br>Contour ProtégéAl<br>(K231765) | Substantial Equivalence<br>Discussion |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clearance<br>Date | TBD | November 8, 2023 | N/A |
| Operating<br>Platform | Server-based application<br>supporting:<br>• Linux-based OS<br>• Local deployment on<br>Windows or Mac | Server-based application<br>supporting:<br>• Linux-based OS<br>• Local deployment on<br>Windows or Mac | No change |
| Modalities | CT and MR | CT and MR | No change |
| Atlas-based<br>Segmentation | No | No | No change |
| Automatically<br>contour<br>imaging data<br>using<br>machine-<br>learning | Yes | Yes | No change |
| Cloud-based<br>Deployment | Yes | Yes | No change |
| Local<br>Deployment<br>or Installation | Yes | Yes | No change |
| Neural | (1.0.0 models) | (1.0.0 models) | The subject device contains |
| | Subject Device: | Predicate Device: | |
| ITEM | Contour ProtégéAl+<br>(K250035) | Contour ProtégéAl<br>(K231765) | Substantial Equivalence<br>Discussion |
| Network<br>Models<br>Included | • Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Liver CT<br>• Prostate MR<br>(1.1.0 model)<br>• Prostate MR<br>(2.0.0 models)<br>• Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Abdomen CT<br>• Lungs and Liver CT<br>(3.1.0 models)*<br>• Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Abdomen CT<br>• Lungs and Liver CT<br>• MRT Additional Structures<br>CT<br>(4.0.0 models)<br>• Head and Neck CT | • Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Liver CT<br>• Prostate MR<br>(1.1.0 model)<br>• Prostate MR<br>(2.0.0 models)<br>• Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Abdomen CT<br>• Lungs and Liver CT<br>(3.1.0 models)*<br>• Head and Neck CT<br>• Prostate CT<br>• Thorax CT<br>• Abdomen CT<br>• Lungs and Liver CT<br>• MRT Additional Structures<br>CT<br>(4.0.0 models)<br>• Head and Neck CT | one new neural network<br>model and three updated<br>neural network models<br>compared to the predicate.<br>These 4 models were all<br>trained on the same<br>architecture as the<br>predicate without any<br>changes. These 4 models<br>cover all the existing<br>structures for contouring<br>and additional, new<br>structures, which are<br>denoted in bold font.<br>These 4 models were all<br>tested according to the<br>same procedures and<br>acceptance criteria as the<br>predicate. No unexpected<br>results were observed. The<br>changes do not raise new<br>questions for safety and<br>effectiveness. |
| | • Thorax CT<br>• Abdomen CT<br>• Pelvis CT<br>• SurePlan MRT CT | • Thorax CT<br>• Abdomen CT<br>• Pelvis CT<br>• SurePlan MRT CT | |
| | (4.1.0 models)<br>• Head and Neck CT<br>• Thorax CT<br>• Whole Body -<br>Physiological<br>• Uptake Organs CT | (4.1.0 models)<br>• Head and Neck CT<br>• Thorax CT<br>• Whole Body -<br>Physiological<br>• Uptake Organs CT | |
| | (4.2.0 models) | | |
| | Subject Device: | Predicate Device: | |
| ITEM | Contour ProtégéAl+<br>(K250035) | Contour ProtégéAl<br>(K231765) | Substantial Equivalence<br>Discussion |
| | • Thorax CT<br>- LN_IMN_L<br>- LN_IMN_R<br>- LN_Sclav_L<br>- LN_Sclav_R<br>- LN_Ax_L1_L<br>- LN_Ax_L1_R<br>- LN_Ax_L2_L<br>- LN_Ax_L2_R<br>- LN_Ax_L3_L<br>- LN_Ax_L3_R<br>- BrachialPlex_L***<br>- BrachialPlex_R***<br>- Breast_L***<br>- Breast_R***<br>- Breast_L_RTOG**<br>- Breast_R_RTOG**<br>- Bronchus<br>- Carina<br>- Cricoid<br>- Esophagus<br>- Glnd_Thyroid<br>- GreatVes***<br>- Heart***<br>- Humerus_Head_L<br>- Humerus_Head_R<br>- Kidney_L<br>- Kidney_R<br>- Larynx***<br>- Liver<br>- Lung_L<br>- Lung_R<br>- Musc_Constrict<br>- Pancreas<br>- SpinalCord<br>- Stomach***<br>- Trachea<br>- Ribs**<br>- Chestwall_L**<br>- Chestwall_R**<br>- A_Asc_Aorta**<br>- A_LAD**<br>• Abdomen CT<br>- Bladder<br>- Bowel*** | | |
| | Subject Device: | Predicate Device: | Substantial Equivalence |
| ITEM | Contour ProtégéAl+<br>(K250035) | Contour ProtégéAl<br>(K231765) | Discussion |
| | - Bowel Bag***<br>- Cauda Equina<br>- Left Kidney<br>- Right Kidney<br>- Liver<br>- Spinal Cord<br>- Stomach***<br>- Female Pelvis CT**<br>- Bowel Bag**<br>- Bowel**<br>- LN Pelvics**<br>- Bladder**<br>- Uterocervix**<br>- Colon Sigmoid**<br>- Cauda Equina**<br>- Sacral Plexus**<br>- Sacrum**<br>- Rectum**<br>- Femur Head L**<br>- Femur Head R**<br>- SurePlan MRT CT<br>- Bone***<br>- Glnd_Lacrimal_L<br>- Glnd_Lacrimal_R<br>- Glnd_Submand_L<br>- Glnd_Submand_R<br>- Glnd_Thyroid<br>- Kidney_L (Kidney_L w/o<br>Renal Pelvis)<br>- Kidney_R (Kidney_R<br>w/o Renal Pelvis)<br>- Liver<br>- Lung_L<br>- Lung_R<br>- Parotid_L<br>- Parotid_R<br>- Spleen | | |
#### Table 1 - Comparison to Predicate Device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font. Below the text "mim" is the word "SOFTWARE" in a smaller, sans-serif font.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, and the colors are eye-catching.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle in the red square. To the right of the squares is the text "mim" in a bold, sans-serif font. Below the text "mim" is the text "SOFTWARE" in a smaller, sans-serif font.
* The 3.1.0 models share the same training images and architecture as the 3.0.0 models. Some errors and style inconsistencies in the training segmentations were corrected before re-training, resulting in the 3.1.0 models. ** Indicates new contour/model.
*** Indicates updated contour.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is clean and modern, with a focus on simplicity and clarity.
## Testing and Performance Data
Changes within this submission compared to the predicate device, Contour ProtégéAl (K231765), include one new 4.2.0 neural network model and three updated 4.2.0 neural network models with additional or updated contours. Non-inferiority testing was used to compare the Contour ProtégéAl+ subject device to Atlases created from the MIM Maestro (K071964) reference device.
Verification and validation tests were performed for each of the core features of Contour ProtégéAl+:
- Contour creation time savings ●
- Clinical quality contour generation ●
The testing methods include both internal verification by MIM Software's own qualified testers and external validation by trained medical professionals.
The Contour ProtégéAl+ subject device was evaluated on the test subjects from a pool of 770 images gathered from 32 institutions.
The CT images for this training set were obtained from clinical treatment plans for patients prescribed external beam or molecular radiotherapy, but the original segmentations were not used. Instead, the images were re-segmented by consultants (physicians and dosimetrists) specifically for this purpose, outside of clinical practice. Detailed instructions derived from relevant published contouring guidelines were prepared for the dosimetrists. The initial seqmentations were then reviewed and corrected by radiation oncologists against the same standards and guidelines. Qualified staff at MIM Software (MD or licensed dosimetrists) then performed a final review and correction. All segmenters and reviewers were instructed to spend additional time to ensure the highest quality training data. In particular, the consultants were asked to contour all specified OAR structures on all images according to referenced standards, whether or not they were proximal to the treatment field. All patients were imaged on an indexed couch in the treatment position ("simulation CT"). Series that were no-axial, had slices thinner than 0.5mm, or had non-Fan Beam or MV acquisitions were excluded.
The verification data used for testing is from a set of institutions that are totally disjoint from the datasets used to train each model in the Contour ProtégéAl+ device. The MIM Maestro (K071964) reference device was configured with Atlases built from the same training data used to train the models. We tested Contour ProtégéAl+ against the reference device, and the goal of this testing is to show that the subject device is equivalent or superior to the reference device. The performance of both segmentation devices was measured by calculating both the Dice score and MDA of the novel segmentations with the original ground-truth contours. User beta testing was also used to evaluate the performance of Contour ProtégéAl+ in the context of time savings
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, on the left side. To the right of the squares is the text "mim" in a stylized, lowercase font, with the word "SOFTWARE" underneath in a smaller font.
compared to contouring from scratch. This user evaluation was made on a three-point scale for each contour, with one indicating negligible time savings, two indicating moderate, and three indicating significant time savings. Our acceptance criteria combine the statistical tests and the user evaluation - only structures that pass two or more of the following three tests could be included in the final models:
- Statistical non-inferiority of the Dice score compared with the reference predicate.
- Statistical non-inferiority of the MDA score compared with the reference predicate.
- Average user evaluation of 2 or higher, when measured on a three-point scale. ●
Further clinical validation was also conducted in-house to qualitatively evaluate contours compared to detailed criteria based on established clinical guidelines.
In addition, each model as a whole was also evaluated. In order to be included in the released product, the cumulative Added Path (APL) of the contours in the model was evaluated relative to the ground-truth. Cumulative APL has been found to correlate well with the total time spent in editing and correcting auto-segmented contours (Vaassen et al, 2020)1. Each model was required to have statistically non-inferior cumulative APL compared to the reference predicate.
Finally, the localization accuracy of each structure in all models was measured. We did not impose a passing criterion on localization accuracy; the results however are included in this summary document as well as in our User Guide and White Paper to allow the user to better understand the performance of the device.
Across the testing data images, 53.4% were female, 31.3% were male, and 15.3% were unknown for sex. The manufacturer was GE for 43.2%, Siemens for 21.7%, Phillips for 15.5%, Toshiba for 0.4%, and unknown or other scanner vendors for 19.2%. 0.3% were between the ages of 0-20, 4.7% were between the aqes of 20-40, 20.9% were between 40-60, and 50.0% were over the age of 60. 24.1% were unknown for age.
The mean and standard deviation Dice coefficients and MDA scores, along with the lower 95th percentile confidence bound, were calculated for both the proposed Contour ProtégéAl device and the MIM Maestro atlas segmentation reference device for each structure of each neural network model. Contour ProtégéAl+ results were equivalent or had better performance than the MIM Maestro atlas segmentation reference device. Equivalence is defined such that the lower 95th percentile confidence bound of the Contour ProtégéAl+ segmentation is greater than 0.1 Dice lower than the mean MIM atlas seqmentation reference device performance.
Results over the validation set compared to the reference device are presented below.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a lowercase, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, with a clean design.
| 4.2.0 CT<br>Model | Structure | Dice MIM<br>Atlas | Dice<br>Contour<br>ProtégéAl | MDA<br>MIM<br>Atlas | MDA<br>Contour<br>ProtégéAl | External<br>Evaluation<br>Score |
|-------------------|---------------------|-------------------|------------------------------|---------------------|-----------------------------|---------------------------------|
| Thorax | BrachialPlex_L | 0.30 ± 0.14 | 0.41 ± 0.15<br>(0.28) * | 3.01 ±<br>1.19 | 2.89 ± 1.07<br>(3.94) * | 2.43 |
| | BrachialPlex_R | 0.31 ± 0.12 | 0.39 ± 0.15<br>(0.26) * | 3.08 ±<br>1.54 | 3.11 ± 1.38<br>(4.55) * | 2.43 |
| | Breast_L | 0.74 ± 0.11 | 0.79 ± 0.07<br>(0.73) * | 5.73 ±<br>3.10 | 5.00 ± 2.53<br>(6.98) * | 2.38 |
| | Breast_R | 0.76 ± 0.12 | 0.80 ± 0.11<br>(0.71) * | 5.03 ±<br>2.72 | 4.49 ± 2.02<br>(6.24) * | 2.57 |
| | Breast_L_RTO_G | 0.74 ± 0.11 | 0.77 ± 0.11<br>(0.69) * | 5.73 ±<br>3.10 | 5.47 ± 3.72<br>(7.87) | 2.5 |
| | Breast_R_RTO_G | 0.76 ± 0.12 | 0.77 ± 0.15<br>(0.67) * | 5.03 ±<br>2.72 | 5.10 ± 3.17<br>(7.24) | 2.5 |
| | Bronchus | 0.57 ± 0.19 | 0.62 ± 0.14<br>(0.50) ** | 2.69 ±<br>2.22 | 1.86 ± 0.90<br>(3.19) * | 2.63 |
| | Carina | 0.37 ± 0.18 | 0.50 ± 0.12<br>(0.41) * | 2.67 ±<br>2.65 | 1.93 ± 0.83<br>(3.07) * | 2.43 |
| | Cricoid | 0.02 ± 0.04 | 0.06 ± 0.05<br>(0.02) * | 4.77 ±<br>1.53 | 5.01 ± 1.40<br>(6.27) * | 2.86 |
| | Esophagus | 0.47 ± 0.17 | 0.69 ± 0.16<br>(0.63) * | 2.73 ±<br>2.19 | 1.08 ± 1.53<br>(1.71) * | 2.5 |
| | Glnd_Thyroid | 0.46 ± 0.18 | 0.66 ± 0.18<br>(0.53) * | 2.84 ±<br>1.90 | 1.68 ± 1.44<br>(2.86) * | 2.86 |
| | GreatVes | 0.64 ± 0.17 | 0.70 ± 0.16<br>(0.53) ** | 4.59 ±<br>1.99 | 3.44 ± 1.99<br>(5.46) * | 2.63 |
| | Heart | 0.88 ± 0.08 | 0.90 ± 0.07<br>(0.87) * | 3.05 ±<br>2.03 | 2.49 ± 1.84<br>(3.08) * | 2.63 |
| | Humerus_Head_L | 0.60 ± 0.24 | 0.62 ± 0.24<br>(0.42) ** | 0.38 ±<br>0.23 | 0.33 ± 0.22<br>(0.52) * | 3 |
| | Humerus_Head_R | 0.57 ± 0.22 | 0.60 ± 0.22<br>(0.41) ** | 0.78 ±<br>2.00 | 0.33 ± 0.35<br>(1.60) * | 3 |
| 4.2.0 CT<br>Model | Structure | Dice MIM<br>Atlas | Dice<br>Contour<br>ProtégéAl | MDA<br>MIM<br>Atlas | MDA<br>Contour<br>ProtégéAl | External<br>Evaluation<br>Score |
| | Kidney_L | 0.73 ± 0.19 | 0.90 ± 0.09<br>(0.83) * | 3.75 ±<br>2.66 | 1.32 ± 1.13<br>(2.27) * | 2.75 |
| | Kidney_R | 0.73 ± 0.19 | 0.89 ± 0.09<br>(0.83) * | 3.97 ±<br>2.60 | 1.41 ± 1.06<br>(2.34) * | 2.75 |
| | Larynx | 0.47 ± 0.19 | 0.59 ± 0.14<br>(0.51) * | 3.42 ±<br>1.32 | 2.87 ± 1.34<br>(3.51) * | 2.63 |
| | Liver | 0.84 ± 0.12 | 0.90 ± 0.13<br>(0.85) * | 5.06 ±<br>4.18 | 3.56 ± 9.12<br>(6.39) * | 2.71 |
| | Lung_L | 0.95 ± 0.02 | 0.96 ± 0.02<br>(0.96) * | 1.12 ±<br>0.46 | 0.79 ± 0.40<br>(0.94) * | 2.75 |
| | Lung_R | 0.95 ± 0.03 | 0.97 ± 0.03<br>(0.96) * | 1.36 ±<br>0.67 | 0.85 ± 0.48<br>(1.04) * | 2.75 |
| | Musc_Constrict | 0.40 ± 0.17 | 0.50 ± 0.17<br>(0.39) * | 2.00 ±<br>1.79 | 1.67 ± 1.55<br>(2.71) * | 3 |
| | Pancreas | 0.17 ± 0.17 | 0.47 ± 0.21<br>(0.36) * | 16.43 ±<br>16.42 | 6.80 ± 8.89<br>(14.66) ** | 2.17 |
| | SpinalCord | 0.66 ± 0.16 | 0.64 ± 0.17<br>(0.59) * | 1.29 ±<br>0.91 | 1.28 ± 0.72<br>(1.50) * | 2.5 |
| | Stomach | 0.46 ± 0.22 | 0.73 ± 0.21<br>(0.64) * | 12.56 ±<br>13.48 | 6.89 ± 20.57<br>(14.36) ** | 2.13 |
| | Trachea | 0.68 ± 0.15 | 0.74 ± 0.17<br>(0.66) * | 1.43 ±<br>0.65 | 1.12 ± 0.64<br>(1.46) * | 2.63 |
| | A LAD | 0.07 ± 0.09 | 0.32 ± 0.12<br>(0.22) * | 8.77 ±<br>10.69 | 3.43 ± 4.46<br>(11.41) ** | 2.67 |
| | A_Aorta_Asc | 0.72 ± 0.16 | 0.83 ± 0.17<br>(0.68) ** | 2.90 ±<br>2.16 | 1.11 ± 0.60<br>(2.64) * | 2.14 |
| | Rib | 0.25 ± 0.09 | 0.28 ± 0.11<br>(0.21) * | 35.64 ±<br>12.65 | 39.04 ± 14.62<br>(49.68) ** | 2.63 |
| | Chestwall_L | 0.40 ± 0.16 | 0.39 ± 0.17<br>(0.19) ** | 4.35 ±<br>1.12 | 4.44 ± 1.24<br>(5.87) * | 2.43 |
| | Chestwall_R | 0.45 ± 0.18 | 0.42 ± 0.18<br>(0.22) ** | 4.66 ±<br>1.65 | 4.69 ± 1.56<br>(6.51) * | 2.43 |
| | LN_Ax_L1_L | 0.59 ± 0.09 | 0.63 ± 0.10<br>(0.75) * | 2.76 ±<br>0.89 | 2.25 ± 0.80<br>(3.32) * | 2.67 |
| 4.2.0 CT<br>Model | Structure | Dice MIM<br>Atlas | Dice<br>Contour<br>ProtégéAl | MDA<br>MIM<br>Atlas | MDA<br>Contour<br>ProtégéAl | External<br>Evaluation<br>Score |
| | LN_Ax_L1_R | 0.52 ± 0.12 | 0.58 ± 0.13<br>(0.69) * | 3.46 ±<br>1.04 | 2.90 ± 1.52<br>(4.09) * | 2.67 |
| | LN_Ax_L2_L | 0.57 ± 0.16 | 0.64 ± 0.13<br>(0.77) ** | 2.57 ±<br>1.25 | 1.93 ± 0.88<br>(2.88) * | 2.67 |
| | LN_Ax_L2_R | 0.53 ± 0.20 | 0.60 ± 0.18<br>(0.73) ** | 2.91 ±<br>1.29 | 2.44 ± 1.26<br>(3.31) * | 2.67 |
| | LN_Ax_L3_L | 0.55 ± 0.17 | 0.62 ± 0.19<br>(0.76) ** | 2.21 ±<br>1.11 | 1.90 ± 1.26<br>(2.87) * | 2.67 |
| | LN_Ax_L3_R | 0.46 ± 0.16 | 0.52 ± 0.17<br>(0.65) * | 3.30 ±<br>1.95 | 2.74 ± 1.73<br>(4.24) * | 2.67 |
| | LN_IMN_L | 0.17 ± 0.11 | 0.41 ± 0.17<br>(0.59) * | 4.75 ±<br>3.26 | 1.82 ± 0.85<br>(5.05) ** | 3 |
| | LN_IMN_R | 0.23 ± 0.16 | 0.48 ± 0.20<br>(0.63) * | 3.36 ±<br>1.95 | 1.69 ± 1.38<br>(3.09) * | 3 |
| | LN_Sclav_L | 0.58 ± 0.15 | 0.66 ± 0.13<br>(0.79) * | 2.84 ±<br>1.94 | 2.49 ± 1.70<br>(4.18) * | 2 |
| | LN_Sclav_R | 0.48 ± 0.09 | 0.55 ± 0.09<br>(0.64) * | 2.97 ±<br>0.75 | 2.67 ± 0.90<br>(3.56) * | 2.33 |
| Abdomen | Bladder | 0.72 ± 0.23 | 0.92 ± 0.16<br>(0.81) * | 3.97 ±<br>3.00 | 0.78 ± 0.73<br>(1.93) * | 2.6 |
| | Bowel | 0.34 ± 0.14 | 0.52 ± 0.19<br>(0.37) * | 9.26 ±<br>3.49 | 5.05 ± 3.17<br>(8.13) ** | 2.5 |
| | BowelBag | 0.30 ± 0.09 | 0.36 ± 0.11<br>(0.29) * | 14.13 ±<br>3.62 | 10.08 ± 3.36<br>(12.50) * | 2.89 |
| | CaudaEquina | 0.62 ± 0.15 | 0.69 ± 0.13<br>(0.59) * | 1.17 ±<br>0.51 | 0.95 ± 0.53<br>(1.33) * | 2.6 |
| | Kidney_L | 0.74 ± 0.17 | 0.94 ± 0.03<br>(0.85) * | 3.86 ±<br>2.48 | 0.82 ± 0.41<br>(2.12) * | 2.9 |
| | Kidney_R | 0.75 ± 0.18 | 0.92 ± 0.07<br>(0.83) * | 3.97 ±<br>3.35 | 0.96 ± 0.62<br>(2.74) * | 2.9 |
| | Liver | 0.84 ± 0.12 | 0.93 ± 0.08<br>(0.86) * | 5.23 ±<br>3.65 | 2.01 ± 2.14<br>(3.85) * | 2.8 |
| | SpinalCord | 0.60 ± 0.16 | 0.65 ± 0.14<br>(0.56) * | 1.13 ±<br>0.60 | 0.83 ± 0.32<br>(1.10) * | 2.9 |
| 4.2.0 CT<br>Model | Structure | Dice MIM<br>Atlas | Dice<br>Contour<br>ProtégéAl | MDA<br>MIM<br>Atlas | MDA<br>Contour<br>ProtégéAl | External<br>Evaluation<br>Score |
| | Stomach | 0.49 ± 0.21 | 0.82 ± 0.11<br>(0.72) * | 11.56 ±<br>14.68 | 2.68 ± 2.02<br>(9.37) ** | 2.5 |
| Female<br>Pelvis | Bladder | 0.61 ± 0.18 | 0.91 ± 0.06<br>(0.83) * | 5.53 ±<br>3.76 | 1.04 ± 0.94<br>(2.66) * | 2.75 |
| | Bag_Bowel | 0.45 ± 0.17 | 0.52 ± 0.20<br>(0.36) ** | 8.89 ±<br>2.79 | 6.89 ± 2.25<br>(9.06) * | 2.63 |
| | Bowel | 0.34 ± 0.10 | 0.55 ± 0.19<br>(0.42) * | 8.81 ±<br>3.05 | 4.84 ± 3.03<br>(7.44) * | 2.75 |
| | Colon_Sigmoid | 0.05 ± 0.05 | 0.47 ± 0.24<br>(0.36) * | 20.55 ±<br>12.23 | 14.72 ± 15.14<br>(23.56) ** | 2.5 |
| | Femur_Head_<br>L | 0.86 ± 0.07 | 0.91 ± 0.08<br>(0.84) * | 1.43 ±<br>0.78 | 0.95 ± 1.03<br>(1.75) * | 2.75 |
| | Femur_Head_<br>R | 0.87 ± 0.05 | 0.92 ± 0.02<br>(0.89) * | 1.22 ±<br>0.59 | 0.69 ± 0.28<br>(1.09) * | 2.75 |
| | UteroCervix | 0.16 ± 0.16 | 0.65 ± 0.27<br>(0.47) * | 10.65 ±<br>5.73 | 4.39 ± 13.21<br>(13.49) ** | 2 |
| | LN_Pelvics | 0.64 ± 0.05 | 0.74 ± 0.06<br>(0.69) * | 4.37 ±<br>1.11 | 4.12 ± 1.52<br>(5.28) * | 2.75 |
| | Rectum | 0.38 ± 0.16 | 0.75 ± 0.12<br>(0.67) * | 6.30 ±<br>3.30 | 1.52 ± 1.66<br>(3.04) * | 2.63 |
| | SacralPlex | 0.02 ± 0.01 | 0.03 ± 0.01<br>(0.02) * | 13.03 ±<br>1.73 | 12.81 ± 1.86<br>(14.38) * | 2.5 |
| | Sacrum | 0.84 ± 0.02 | 0.89 ± 0.01<br>(0.87) * | 1.58 ±<br>0.32 | 1.09 ± 0.17<br>(1.31) * | 3 |
| | CaudaEquina | 0.65 ± 0.11 | 0.66 ± 0.11<br>(0.58) * | 1.19 ±<br>0.58 | 0.94 ± 0.59<br>(1.39) * | 2.57 |
| SurePlan<br>MRT | Bone | 0.76 ± 0.08 | 0.83 ± 0.05<br>(0.71) * | 4.77 ±<br>1.97 | 4.57 ± 3.35<br>(9.50) ** | 3 |
| | Glnd_Lacrimal<br>_L | 0.23 ± 0.17 | 0.30 ± 0.21<br>(0.16) * | 1.99 ±<br>0.86 | 1.26 ± 0.57<br>(1.96) * | 2.67 |
| | Glnd_Lacrimal<br>_R | 0.23 ± 0.16 | 0.36 ± 0.23<br>(0.22) * | 1.53 ±<br>0.80 | 1.18 ± 0.87<br>(1.97) * | 2.67 |
| | Glnd_Subman | 0.58 ± 0.12 | 0.67 ± 0.29 | 2.19 ± | 1.00 ± 0.35 | 3 |
| 4.2.0 CT<br>Model | Structure | Dice MIM<br>Atlas | Dice<br>Contour<br>ProtégéAl | MDA<br>MIM<br>Atlas | MDA<br>Contour<br>ProtégéAl | External<br>Evaluation<br>Score |
| | d_L | | (0.52) * | 0.76 | (1.45) * | |
| | GInd_Subman<br>d_R | 0.56 ± 0.16 | 0.66 ± 0.31<br>(0.48) * | 2.41 ±<br>1.14 | 0.97 ± 0.33<br>(1.68) * | 3 |
| | Glnd_Thyroid | 0.47 ± 0.19 | 0.75 ± 0.12<br>(0.63) ** | 2.98 ±<br>1.95 | 1.29 ± 1.20<br>(2.46) * | 2.67 |
| | Kidney_L | 0.72 ± 0.18 | 0.91 ± 0.04<br>(0.82) * | 4.08 ±<br>2.52 | 1.56 ± 0.61<br>(2.89) * | 3 |
| | Kidney_R | 0.76 ± 0.17 | 0.91 ± 0.03<br>(0.82) * | 3.87 ±<br>3.14 | 1.45 ± 0.63<br>(3.12) * | 3 |
| | Liver | 0.85 ± 0.12 | 0.93 ± 0.07<br>(0.88) * | 4.82 ±<br>3.51 | 1.79 ± 1.50<br>(3.25) * | 2.67 |
| | Lung_L | 0.94 ± 0.03 | 0.96 ± 0.04<br>(0.93) * | 1.42 ±<br>0.51 | 0.90 ± 0.49<br>(1.32) * | 3 |
| | Lung_R | 0.94 ± 0.04 | 0.96 ± 0.05<br>(0.92) * | 1.66 ±<br>0.75 | 1.07 ± 0.86<br>(1.75) * | 3 |
| | Parotid L | 0.71 ± 0.09 | 0.81 ± 0.05<br>(0.78) * | 2.17 ±<br>0.77 | 1.42 ± 0.40<br>(1.73) * | 3 |
| | Parotid_R | 0.71 ± 0.09 | 0.82 ± 0.05<br>(0.78) * | 2.17 ±<br>0.73 | 1.34 ± 0.48<br>(1.67) * | 3 |
| | Spleen | 0.72 ± 0.10 | 0.95 ± 0.02<br>(0.87) * | 4.38 ±<br>1.87 | 0.62 ± 0.45<br>(2.08) * | 2.67 |
# Table 2 – Dice, MDA, and external evaluation results.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the MIM Software logo. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a stylized font, with the word "SOFTWARE" below it.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a stylized font, with the word "SOFTWARE" underneath in smaller letters. The logo is simple and modern, with a focus on the company's name.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for MIM Software. The logo consists of a gray square overlapping a red square with a white circle in the middle. To the right of the squares is the text "mim" in a sans-serif font, with the word "SOFTWARE" below it in a smaller font.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle where they overlap. To the right of the squares is the text "mim" in a lowercase, sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, with a clean design.
Mean ± Std MDA mm (upper 95th percentile confidence bound based on normal distribution in parentheses) * Equivalence demonstrated at p=0.05 significance level between Contour ProtégéAI and MIM Atlas (green) ** Equivalence was NOT demonstrated because the minimum sample size was not met for this contour (yellow) Cells highlighted in orange indicate that equivalence was NOT demonstrated.
Additionally, preliminary user evaluation conducted as part of testing demonstrated that Contour ProtégéAl+ yields comparable time-saving functionality when creating contours as other commercially available automatic segmentation products.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle in the intersection. To the right of the squares is the text "mim" in a sans-serif font, with the word "SOFTWARE" in a smaller font below it. The logo is simple and modern, and the colors are eye-catching.
## Table 3 – 4.2.0 CT Models cumulative APL.
| 4.2.0 CT Models | MIM Atlas | Contour ProtégéAl |
|-----------------|-----------------|----------------------------|
| Thorax | 220.33 ± 232.42 | 181.44 ± 219.43 (200.79) * |
| Abdomen | 354.53 ± 386.05 | 238.05 ± 309.40 (312.07) * |
| Female Pelvis | 433.24 ± 392.19 | 314.13 ± 349.77 (383.86) * |
| SurePlan MRT | 216.52 ± 207.87 | 133.01 ± 160.23 (167.60) * |
Mean ± Std APL mm (upper 95th percentile confidence bound based on normal distribution in parentheses). * Equivalence demonstrated at p=0.05 significance level between Contour ProtégéAl and MIM Atlas.
| 4…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.