SMART PCFD software includes AI-powered algorithms and is intended to be used to support orthopedic healthcare professionals in the diagnosis and surgical planning of Progressive Collapsing Foot Deformity (PCFD) in a hospital or clinic environment. The medical image modality intended to be used in the software is weight-bearing CT (WBCT). SMART PCFD software provides for the user: - Visualization report of the three-dimensional (3D) mathematical models and measurements of the anatomical structures of foot and ankle and three-dimensional models of orthopedic fixation devices, - Measurement templates containing radiographic measures of foot and ankle, and - Surgical planning application for visualization of foot and ankle anatomical three-dimensional structures, radiographic measures, and surgical instrument parameters supporting the following common flatfoot procedures: Medial Displacement Calcaneal Osteotomy (MDCO), Lateral Column Lengthening (LCL), and Cotton Osteotomy (CO). The visualization report containing the measurements is intended to be used to support orthopedic healthcare professionals in the diagnosis of PCFD. The surgical planning application contains the visualizations of the three-dimensional structural models, orthopedic fixation device models and surgical instrument parameters combined with the measurements is intended to be used to support orthopedic healthcare professionals in surgical planning of PCFD.
Device Story
Software processes weight-bearing CT (WBCT) DICOM images to segment bone structures and generate 3D mathematical models of foot/ankle anatomy; includes AI-powered algorithms for bone and metal identification. Used by orthopedic healthcare professionals in hospitals/clinics for diagnosis and surgical planning of PCFD. Provides visualization reports, radiographic measurement templates, and surgical planning tools for specific procedures (MDCO, LCL, CO). Clinicians review 3D models, measurements, and simulated orthopedic fixation device/instrument parameters via a web-based GUI to inform surgical decision-making and planning. Benefits include improved visualization and standardized measurement for complex foot deformity correction.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated using 82 CT image series; bone identification accuracy 100%; metal identification sensitivity 100%, specificity 98%. Surgical planning component validated for angular (±1 degree) and distance (±1.0mm) measurement accuracy.
Technological Characteristics
Medical image management and processing software; web-based GUI. Uses machine learning algorithms for automated bone/metal segmentation from WBCT DICOM inputs. Provides 3D mathematical modeling and radiographic measurement tools. Software-only device; no hardware components. Operates on standard web browsers.
Indications for Use
Indicated for orthopedic healthcare professionals to support diagnosis and surgical planning of Progressive Collapsing Foot Deformity (PCFD) using weight-bearing CT (WBCT) images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 29, 2025
Disior Ltd
% Kelsey Gibson
Regulatory Affairs Specialist
Paragon 28, Inc.
14445 Grasslands Dr.
Englewood, CO 80134
Re: K250023
Trade/Device Name: Smart PCFD
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: August 29, 2025
Received: August 29, 2025
Dear Kelsey Gibson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250023 - Kelsey Gibson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250023 - Kelsey Gibson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250023
Device Name
SMART PCFD
Indications for Use (Describe)
SMART PCFD software includes AI-powered algorithms and is intended to be used to support orthopedic healthcare professionals in the diagnosis and surgical planning of Progressive Collapsing Foot Deformity (PCFD) in a hospital or clinic environment. The medical image modality intended to be used in the software is weight-bearing CT (WBCT).
SMART PCFD software provides for the user:
- Visualization report of the three-dimensional (3D) mathematical models and measurements of the anatomical structures of foot and ankle and three-dimensional models of orthopedic fixation devices,
- Measurement templates containing radiographic measures of foot and ankle, and
- Surgical planning application for visualization of foot and ankle anatomical three-dimensional structures, radiographic measures, and surgical instrument parameters supporting the following common flatfoot procedures: Medial Displacement Calcaneal Osteotomy (MDCO), Lateral Column Lengthening (LCL), and Cotton Osteotomy (CO).
The visualization report containing the measurements is intended to be used to support orthopedic healthcare professionals in the diagnosis of PCFD. The surgical planning application contains the visualizations of the three-dimensional structural models, orthopedic fixation device models and surgical instrument parameters combined with the measurements is intended to be used to support orthopedic healthcare professionals in surgical planning of PCFD.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250023
# 510(K) SUMMARY
510(k) Number: K250023
Manufacturer: Disior Ltd
HTC Helsinki, Building Pinta, Tammasaarenkatu 3
Helsinki Uusimaa, FL, 00180, Finland
Contact: Markuu Laitinen
Director of Enabling Technology
Phone: +358 405 430 673
Email: mlaitinen@paragon28.com
Prepared By: Kelsey R. Gibson
Regulatory Affairs Specialist II
Paragon 28, Inc.
14445 Grasslands Dr.,
Englewood, CO, 80134
Phone: 720-994-5458
Date Prepared: September 26, 2025
Device TradeName: SMART PCFD
Device Class and Common Name: Class II, Automated Radiological Image Processing Software
Classification: 21 CFR 892.2050: Medical image management and processing system
Product Codes: QIH
Indications for Use: SMART PCFD software includes AI-powered algorithms and is intended to be used to support orthopedic healthcare professionals in the diagnosis and surgical planning of Progressive Collapsing Foot Deformity (PCFD) in a hospital or clinic environment. The medical image modality intended to be used in the software is weight-bearing CT (WBCT).
SMART PCFD software provides for the user:
- Visualization report of the three-dimensional (3D) mathematical models and measurements of the anatomical structures of foot and ankle and three-dimensional models of orthopedic fixation devices,
- Measurement templates containing radiographic measures of foot and ankle, and
Page 1 of 5
{5}
K250023
- Surgical planning application for visualization of foot and ankle anatomical three-dimensional structures, radiographic measures, and surgical instrument parameters supporting the following common flatfoot procedures: Medial Displacement Calcaneal Osteotomy (MDCO), Lateral Column Lengthening (LCL), and Cotton Osteotomy (CO).
The visualization report containing the measurements is intended to be used to support orthopedic healthcare professionals in the diagnosis of PCFD. The surgical planning application contains the visualizations of the three-dimensional structural models, orthopedic fixation device models and surgical instrument parameters combined with the measurements is intended to be used to support orthopedic healthcare professionals in surgical planning of PCFD.
## Device Description:
The SMART PCFD software is intended to be used in reviewing and digitally processing computed tomography images for the purposes of interpretation by a specialized medical practitioner. The device segments the medical images and creates a 3D model of the bones of the foot and ankle. Measurements, including anatomical axes, are provided to the user and the device allows for presurgical planning.
The device includes the same machine learning derived outputs as the primary predicate SMART Bun-Yo-Matic CT (K240642) device and no new validations were conducted.
Details on the previously performed validation are summarized below. The testing for 82 CT image series presented 100% correctly identified bones of foot and ankle. The existence of metal was identified correctly for 98.8% of the images (specificity 98%, sensitivity 100%).
## Study Subjects
The AI algorithm for bone identification was developed using 145 CT image studies and metal identification was developed using 130 CT image studies. Testing was carried out using 82 CT image studies. Out of 357 CT image studies, 340 were from individual patients with few studies from same patient with different foot alignments. The CT image series' were collected from various sites across USA and Europe with a minimum of 50% of the images originating from the USA. The CT image studies were from patients with different ages and racial groups, with minimum of 35% male/female within each dataset, with mean age approximately 47 years (SD 15 years), and representatives from White, (Non-)Hispanic, African American, and Native racial groups. Each dataset was balanced in terms of subjects with different foot alignment, demographics, imaging devices and with subjects from clinical subgroups ranging from control/normal feet (44% with test data) to pre-/post-operative clinical conditions such as Hallux Valgus, Progressive Collapsing Foot Deformity, fractures, or with metal implants (40% of the test data).
## Imaging Systems
The 357 image studies were collected using CT imaging system made by five (5) manufacturers (7 different models in total). From the test data of 82 images, 61% of the images were acquired using Curvebeam PedCAT, 11% with Planmed Verify, and 26% with Carestream OnSight 3D Extremity. In addition, system test data contains images acquired with Toshiba Somatom. Typical imaging protocol is disclosed within the IFU, however, the test data contains wider range of parameters for generalization (tube
Page 2 of 5
{6}
K250023
voltages between 90-120 kV, tube currents 5-8 mA, and slice thickness/pixel spacing 0.37-1.5mm).
## Ground Truth
The ground truths for bone and metal identification were independently established by three (3) U.S. Orthopedic surgeons with a 3rd party software. Each clinicians reviewed each of the DICOM series through axial/sagittal/coronal views and/or 3D reconstruction and marked on a spreadsheet the presence of a bone and metal in the image series. Based on the majority vote of three, two same responses were required to establish a ground truth on each of the DICOM series.
## Training, Tuning, and Validation Data Independence
The SMART PCFD software machine learning algorithm training and tuning data used during the algorithm development, as well as test data used in the standalone software performance assessment study, were all independent data sets. Each CT image study was allowed to be allocated to only data set.
## Predicate:
SMART Bun-Yo-Matic CT (K240642)
## Substantial
The Indications for Use of the subject device and the predicate device are similar.
Differences do not constitute a different intended use because both devices are intended to provide 3D models, measurements, and presurgical planning generated from CT input to orthopaedic healthcare professionals.
## Equivalence:
The subject and predicate devices have similar technological characteristics. The main differences being in the surgical planning case report output for the device since the subject device is for Progressive Collapsing Foot Deformity (PCFD) while the predicate device is for Lapidus Arthrodesis. In support of the claim of substantial equivalence the comparison between the subject and predicate systems demonstrates a shared input, image processing, measuring and planning capabilities, and user interface.
| | Subject Device | Primary Predicate Device | Additional Predicate Device |
| --- | --- | --- | --- |
| Manufacturer | Disior Ltd | Disior Ltd | Disior Ltd |
| Trade Name | SMART PCFD | SMART Bun-Yo-Matic CT | Bonelogic |
| 510(k) | Subject Device | K240642 | K223757 |
| Indications for Use | SMART PCFD software includes AI-powered algorithms and is intended to be used to support orthopedic healthcare professionals in the diagnosis and surgical planning of Progressive Collapsing Foot Deformity (PCFD) in a hospital or clinic environment. The medical image modality intended to be used in the software is weight-bearing CT (WBCT). | SMART Bun-Yo-Matic CT software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment. The medical imaging type intended to be used as the input of the software is Computed Tomography (CT).
SMART Bun-Yo-Matic CT software provides:
• Visualization report of the three-dimensional | Bonelogic software is to be used by orthopaedic healthcare professionals for diagnosis and surgical planning in a hospital or clinic environment.
Bonelogic software provides:
• Semi-automatic segmentation with manual or assisted input of bony structure identification from CT imaging input, |
{7}
K250023
| | SMART PCFD software provides for the user: • Visualization report of the three-dimensional (3D) mathematical models and measurements of the anatomical structures of foot and ankle and three-dimensional models of orthopedic fixation devices, • Measurement templates containing radiographic measures of foot and ankle, and • Surgical planning application for visualization of foot and ankle anatomical three-dimensional structures, radiographic measures, and surgical instrument parameters supporting the following common flatfoot procedures: Medial Displacement Calcaneal Osteotomy (MDCO), Lateral Column Lengthening (LCL), and Cotton Osteotomy (CO). The visualization report containing the measurements is intended to be used to support orthopedic healthcare professionals in the diagnosis of PCFD. The surgical planning application contains the visualizations of the three-dimensional structural models, orthopedic fixation device models and | mathematical models of the anatomical structures of foot and ankle and three-dimensional models of orthopaedic fixation devices, • Measurement templates containing radiographic measures of foot and ankle, • Surgical planning application for visualization of foot and ankle anatomical three-dimensional structures, radiographic measures, and surgical instrument parameters. The visualization report containing the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgical planning application containing the visualizations of the three-dimensional structural models, orthopaedic fixation device models and surgical instrument parameters combined with the measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle. | • Three-dimensional mathematical models of the anatomical structures of foot and ankle, • Measurement templates containing radiographic measures of foot and ankle, and tools for manually obtaining lines and angular measurements, • Surgical planning application for foot and ankle using three-dimensional models of the anatomical structures and radiographic measures. The three-dimensional models of the anatomical structures combined with the measurements can be used for the diagnosis of orthopaedic healthcare conditions. The surgical planning application containing the three-dimensional structural models combined with the measurements can be used for the planning of treatments and operations to correct orthopaedic healthcare conditions of foot and ankle. |
| --- | --- | --- | --- |
Page 4 of 5
{8}
K250023
| | surgical instrument parameters combined with the measurements is intended to be used to support orthopedic healthcare professionals in surgical planning of PCFD. | | |
| --- | --- | --- | --- |
| Input | Weight Bearing CT DICOM Computed tomography | Computed tomography DICOM Computed tomography | Computed tomography DICOM Computed tomography |
| Image Processing | Segmentation of bone structures | Segmentation of bone structures | Segmentation of bone structures |
| Output | Automated case report of the 3D model of patient anatomy, surgical instrument parameters, and visualization of implant | Automated case report of the 3D model of patient anatomy, surgical instrument parameters, and visualization of implant | 3D model of patient anatomy and case report of the 3D model patient anatomy |
| Measuring and Planning | Perform measurements for presurgical planning | Perform measurements for presurgical planning | Perform measurements for presurgical planning |
| User Interface | Graphical user interface (GUI) to a web application used with a standard web browser. | Graphical user interface (GUI) to a web application used with a standard web browser. | Graphical user interface (GUI) built on the Unity development engine. |
Performance Testing:
Differences do not introduce new questions of safety and effectiveness.
All necessary testing has been performed on the SMART PCFD device to assure substantial equivalence to its predicate and demonstrate the subject device performs as intended.
Software Verification and Validation
Software verification and validation were carried out based on the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", at the unit, integration, and system levels to determine substantial equivalence to the predicate device.
Non-Clinical Bench Testing
Performance testing was conducted to evaluate the Surgical Planning Component of the device. The Surgical Planning Component was tested with multiple images and appropriate outputs for the subject device were evaluated by qualified truthers. Results showed the subject device performed as intended. Surgery planning executes mathematical operations for estimated correction $\pm 1$ degree for angular measurements and $\pm 1.0\mathrm{mm}$ for distance measurements.
Conclusions:
Clinical data are not needed to support the safety and effectiveness of the subject device. The SMART PCFD device subject to this submission possesses the same intended use and has similar technological characteristics as the predicate device. All performance testing conducted for the SMART PCFD device met the predetermined acceptance criteria or were otherwise considered acceptable. As such, the SMART PCFD device is substantially equivalent to the predicate device for the intended use.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.