Clever One is dental imaging software that is intended to provide tools for supporting diagnosis and treatment. These tools enable end users to view and interpret a series of DICOM compliant medical images and are intended for use by trained medical professionals. Clever One allows users to load, view, and save DICOM images from CT, panoramic, cephalometric, intraoral, and other imaging equipment. It also provides functionalities such as 2D viewing, 2D analysis, 3D visualization, 3D analysis.
Device Story
Dental imaging software; processes DICOM-compliant 2D/3D images from CT, panoramic, cephalometric, and intraoral X-ray systems. Transforms raw image data into visualizations including multi-planar reconstruction (MPR) and volume rendering; provides tools for implant simulation, canal drawing, and bone density analysis. Used in clinical network environments by dentists and radiologists; operates on Windows-based PCs. Output consists of processed images and analysis metrics displayed on-screen to assist clinicians in diagnosis and preoperative planning. Benefits include enhanced diagnostic accuracy and efficient patient management through unified 2D/3D imaging workflows.
Clinical Evidence
Bench testing only. Software verification and validation testing, including measurement accuracy tests, were performed against pre-determined pass/fail criteria. Cybersecurity verification and validation were conducted per FDA guidance.
Technological Characteristics
Software-only device; Windows-based platform (Windows 10 or higher). Supports DICOM, JPG, and BMP formats. Features include 2D/3D image processing, MPR, volume rendering, and measurement tools (length, angle, area, volume). Connectivity includes network and cloud-based data transfer. No patient-contacting components. Software architecture integrates existing validated modules from predicate devices.
Indications for Use
Indicated for trained medical professionals to view and interpret DICOM-compliant dental images (CT, panoramic, cephalometric, intraoral) for diagnosis and treatment planning. 3D tooth segmentation is limited to adult subjects with permanent teeth.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 23, 2025
Ewoosoft Co., Ltd
% Hyeonguk Kang
Regulatory Affairs Team Manager
801-Ho, Vatechnetworks Bldg., 13, Samsung 1-Ro 2-Gil
HWASEONG-SI, GYEONGGIDO 18449
SOUTH KOREA
Re: K250005
Trade/Device Name: Clever One
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: April 18, 2025
Received: April 21, 2025
Dear Hyeonguk Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250005 - Hyeonguk Kang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250005 - Hyeonguk Kang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250005 | |
| Device Name Clever One | |
| Indications for Use (Describe) Clever One is dental imaging software that is intended to provide tools for supporting diagnosis and treatment. These tools enable end users to view and interpret a series of DICOM compliant medical images and are intended for use by trained medical professionals. Clever One allows users to load, view, and save DICOM images from CT, panoramic, cephalometric, intraoral, and other imaging equipment. It also provides functionalities such as 2D viewing, 2D analysis, 3D visualization, 3D analysis. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250005
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
1. Date: May 23, 2025
2. Applicant / Submitter
Ewoosoft Co., Ltd.
801-ho, Vatechnetworks Bldg., 13, Samsung 1-ro 2-gil,
Hwaseong-si, Gyeonggido, Republic of Korea
Tel: +82 31 8015 6172 Fax: +82 31 8015 6196
Contact person: Hyeonguk Kang
Email: paul.kang@ewoosoft.com
3. Subject Device:
- Trade/Device Name: Clever One
- Regulation Number: 21 CFR 892.2050
- Regulation Name: Medical Image Management and Processing System
- Regulatory Class: Class II
- Product Code: QIH
4. Predicate Device:
- Manufacturer: Ewoosoft Co., Ltd.
- Trade/Device Name: EzDent-i
- 510k Number: K241114
- Regulation Number: 21 CFR 892.2050
- Regulation Name: Medical Image Management and Processing System
- Regulatory Class: Class II
- Product Code: LLZ
- Manufacturer: Ewoosoft Co., Ltd.
- Trade/Device Name: Ez3D-i
- 510k Number: K231757
- Regulation Number: 21 CFR 892.2050
- Regulation Name: Automated Radiological Image Processing Software
- Regulatory Class: Class II
- Product Code: QIH
Page 1 of 8
{5}
Page 2 of 8
6. Device Description:
Clever One is a dental imaging software designed to acquire, process, view, edit, and analyze medical images for supporting diagnostic and preoperative planning purposes. It supports standard DICOM formats for 2D and 3D image files and enables advanced imaging functionalities for enhanced diagnostic accuracy.
The software provides a range of features, including:
- 2D Imaging: Loading, editing, and viewing 2D dental images in standard formats (e.g., DICOM, JPG, BMP).
- 3D Imaging: Visualization and reconstruction of 3D CT images, including multi-planar views (MPR) and volume rendering.
- Preoperative Planning: Implant position simulation, canal drawing, and bone density analysis to support treatment planning.
- Data Interoperability: Facilitates data transfer and storage using DICOM-compliant workflows, ensuring compatibility with third-party systems and imaging devices.
The software interfaces with dental imaging equipment, such as CT, panoramic, cephalometric, and intraoral X-ray systems, as well as intraoral cameras, for image acquisition. It is designed for use in network environments, allowing users to upload and download clinical diagnostic images and patient data for enhanced collaboration and efficient patient management.
7. Indication for use:
Clever One is dental imaging software that is intended to provide tools for supporting diagnosis and treatment.
These tools enable end users to view and interpret a series of DICOM compliant medical images and are intended for use by trained medical professionals. Clever One allows users to load, view, and save DICOM images from CT, panoramic, cephalometric, intraoral, and other imaging equipment. It also provides functionalities such as 2D viewing, 2D analysis, 3D visualization, 3D analysis.
8. Intended Patient Population
Our product has no clinical diagnostic or treatment functions and its main purpose is visualizing data. All contents shown in the software are visualized information of numerical values acquired from acquisition equipment and visualization results only assist end-users in patient counseling, diagnosis, and treatment planning. Users can adjust the visualization settings according to content viewed by the user. All diagnosis and treatment decisions made are solely up to the user. The 3D tooth segmentation function is limited to adult subjects with permanent teeth.
{6}
9. Substantial Equivalence:
| | Subject Device | Equivalent Device | Equivalent Device |
| --- | --- | --- | --- |
| Device name | Clever One v1.0 | EzDent-i v3.4 | Ez3D-i v5.5 |
| 510K number | - | K241114 | K231757 |
| Manufacturer | Ewoosoft | Ewoosoft | Ewoosoft |
| Indications for use | Clever One is dental imaging software that is intended to provide tools for supporting diagnosis and treatment. These tools enable end users to view and interpret a series of DICOM compliant medical images and are intended for use by trained medical professionals. Clever One allows users to load, view, and save DICOM images from CT, panoramic, cephalometric, intraoral, and other imaging equipment. It also provides functionalities such as 2D viewing, 2D analysis, 3D visualization, 3D analysis. | EzDent-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist. EzDent-i is intended for use as software to acquire, view and save 2D image files, load DICOM project files from panorama, cephalometric, and intra-oral imaging equipment. | Ez3D-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist. Ez3D-i is intended for use as software to load, view and save DICOM images from CT, panorama, cephalometric and intraoral imaging equipment and to provide 3D visualization, 2D analysis, in various MPR (Multi-Planar Reconstruction) functions. |
| Technology/Principle of Operation | Clever One is a dental imaging software designed to acquire, process, view, edit, and analyze medical images for supporting diagnostic and preoperative planning | EzDent-i is a device that provides various features to acquire, transfer, edit, display, store, and perform digital processing of medical images. EzDent-i is a patient & | Ez3D-i v5.5 is 3D viewing software for dental CT images in DICOM format with a host of useful functions including MPR, 2-dimensional analysis and 3- |
Page 3 of 8
{7}
| | purposes. It supports standard DICOM formats for 2D and 3D image files and enables advanced imaging functionalities for enhanced diagnostic accuracy.
The software provides a range of features, including:
- 2D Imaging: Loading, editing, and viewing 2D dental images in standard formats (e.g., DICOM, JPG, BMP).
- 3D Imaging: Visualization and reconstruction of 3D CT images, including multi-planar views (MPR) and volume rendering.
- Preoperative Planning: Implant position simulation, canal drawing, and bone density analysis to support treatment planning.
- Data Interoperability: Facilitates data transfer and storage using DICOM-compliant workflows, ensuring compatibility with third-party systems and imaging devices. | image management software specifically for digital dental radiography. It also provides server/client model so that the users upload and download clinical diagnostic images and patient information from any workstations in the network environment.
EzDent-i supports general image formats such as JPG and BMP for 2D image viewing as well as DICOM format.
EzDent-i supports the acquisition of dental images by interfacing with OpenCV library to import the intra-oral camera images. It also supports the acquisition of CT/Panoramic/Cephal o/Intra-Oral Sensor images by interfacing with X-ray capture software. | dimensional image reformation. It provides advanced simulation functions such as Implant Simulation, Drawing Canal, and Implant Environ Bone Density, etc. for the benefit of effective doctor and patient communication and precise treatment planning. |
| --- | --- | --- | --- |
Page 4 of 8
{8}
| The software interfaces with dental imaging equipment, such as CT, panoramic, cephalometric, and intraoral X-ray systems, as well as intraoral cameras, for image acquisition. It is designed for use in network environments, allowing users to upload and download clinical diagnostic images and patient data for enhanced collaboration and efficient patient management. | | | |
| --- | --- | --- | --- |
| Platform | IBM-compatible PC or PC network | IBM-compatible PC or PC network | IBM-compatible PC or PC network |
| Operating System | Microsoft Window 10 or higher | Microsoft Windows 10,11 | Microsoft Window 10 or higher |
| User Interface | Mouse, Keyboard | Mouse, Keyboard | Mouse, Keyboard |
| Image Input Sources | Images can be scanned, loaded from digital cameras or card readers, or imported from a radiographic imaging device | Images can be scanned, loaded from digital cameras or card readers, or imported from a radiographic imaging device | Images can be scanned, loaded from digital cameras or card readers, or imported from a radiographic imaging device |
| 32 bit / 64 bit | 64 bit | 32 / 64 bit | 64 bit |
| Image format | DICOM | DICOM | DICOM |
| Patient Database Compatibility | SQL | SQL | SQL |
| Includes Image Measurement tools | Length, Multi Length, Angle, Multi Angle, Circle, ROI/Area, Volume, Profile | Linear distance, angle | Length, Multi Length, Angle, Multi Angle, Circle, ROI/Area, Volume, Profile |
| Image viewing | Full, side by side, gallery, thumbnail | Full, side by side, gallery, thumbnail | Full, side by side, gallery, thumbnail |
Page 5 of 8
{9}
| Image manipulation | Grayscale, invert, emboss, brightness, contrast, gamma, sharpen, median, despeckle, hue, saturation, equalize, flip, mirror, masking, rotate, magnify, annotation, cephalometric tracing, ceph growth, projections, implant simulations, film view, zooming, whitening, nerve canal tracing, memo | Brightness, contrast, sharpness, inverse, film view, rotate, zooming, whitening, nerve canal tracing, memo | Grayscale, invert, emboss, brightness, contrast, gamma, sharpen, median, despeckle, hue, saturation, equalize, flip, mirror, masking, rotate, magnify, annotation, cephalometric tracing, ceph growth, projections, implant simulations |
| --- | --- | --- | --- |
| Implant module | Generic implant libraries | Generic implant libraries | Generic implant libraries |
| 3D imaging capability | Clever One can view, transfer and process 3D radiographs. Furthermore, it supports Smart Click, Smart Clipping, Implant Simulation and Canal Draw. | Includes interface to 3D imaging software, Ez3D-i. EzDent-i imaging software does not view, transfer or process 3D radiographs. | Ez3D-I can view, transfer and process 3D radiographs. Furthermore, it supports Smart Click, Smart Clipping, Implant Simulation and Canal Draw. |
| Image annotation | Test, paint, ellipse, pointer, select, draw, magnify, line, rectangle, polygon, ruler, protractor, smile library, smudge, brush, redeye reduction, select region, copy / paste | Text, paint, ellipse, pointer, select, draw, magnify, line, rectangle, polygon, ruler, protractor, smile library, smudge, brush, redeye reduction, select region, copy / paste | Test, paint, ellipse, pointer, select, draw, magnify, line, rectangle, polygon, ruler, protractor, smile library, smudge, brush, redeye reduction, select region, copy / paste |
| Distribution of Installation File for Upgrade | USB, EzUpdater | EzUpdater | USB, EzUpdater |
| Customer Support | Manufacturer website, phone number, and e-mail information provided. | Manufacturer website, phone number, and e-mail information provided. | Manufacturer website, phone number, and e-mail information provided. |
Page 6 of 8
{10}
Page 7 of 8
| Dose Information Display | File information, Dose indicator (if applicable) | File information, Dose indicator (if applicable) | File information, Dose indicator (if applicable) |
| --- | --- | --- | --- |
| Report Management | Create, open, view, edit, delete | Create, open, view, edit, delete | Create, open, view, edit, delete |
| Pre-integrated PMS | Clever Dent, Weclever | Clever Dent, Weclever | Clever Dent, Weclever |
| Send E-mail | Send e-mail, attachment, compress to zip file, signature, convert report to, converted image to, patient information anonymization | Send e-mail, attachment, compress to zip file, signature, convert report to, converted image to, patient information anonymization | Send e-mail, attachment, compress to zip file, signature, convert report to, converted image to, patient information anonymization |
The subject device, Clever One, is a software product designed to combine the functionality of two predicate devices, EzDent-i and Ez3D-i. EzDent-i provides 2D imaging capabilities, while Ez3D-i offers 3D imaging capabilities. Clever One integrates the technologies of these two devices to support both 2D and 3D imaging, resulting in a unified platform for medical imaging.
1. Technological Comparison:
Clever One adopts the same technological principles as EzDent-i and Ez3D-i, utilizing DICOM image formats and digital image input sources. The combined technology enables Clever One to provide both 2D and 3D imaging functionalities, which align directly with the features offered by the predicate devices. All major functionalities—including bone density analysis and implant simulation—are identical to those of the predicate devices EzDent-I and Ez3D-I.
2. Principle of Operation:
The operation of Clever One is substantially equivalent to EzDent-i and Ez3D-i. Clever One inherits the 2D imaging capabilities from EzDent-i and the 3D imaging capabilities from Ez3D-i without introducing new principles of operation.
3. Software and Platform:
All devices, including Clever One, operate on a Windows-based platform using Windows 10 as the operating system. The integration of functionalities does not alter the fundamental platform compatibility.
4. User Interface and Functional Enhancements:
The user interface of Clever One integrates and enhances the graphical interfaces of EzDent-i and Ez3D-i to support 2D and 3D imaging in a seamless manner. This enhancement does not introduce new risks or functionalities beyond those already established by the predicates.
{11}
Page 8 of 8
5. Safety and Effectiveness:
The combined functionalities of Clever One do not raise new questions regarding safety or effectiveness. The system builds upon validated features of EzDent-i and Ez3D-i, ensuring that the overall safety profile remains consistent with the predicates.
Conclusion:
Clever One is substantially equivalent to the predicate devices EzDent-i (K241114) and Ez3D-i (K231757) in terms of technology, principle of operation, platform, and functional capabilities. The integration of 2D and 3D imaging does not introduce new technological characteristics that would require additional validation beyond that conducted for the predicate devices.
10. Technological Characteristics:
Clever One is a software device that does not contact the patient, nor does it control any life sustaining devices. Results produced by the software's diagnostic, treatment planning and simulation tools are dependent on the interpretation of trained and licensed radiologists, clinicians and referring physicians as an adjunctive to standard radiology practices for diagnosis.
11. Performance Data:
SW verification/validation and the measurement accuracy test were conducted to establish the performance, functionality and reliability characteristics of the subject devices. The device passed all of the tests based on pre-determined Pass/Fail criteria.
The subject device incorporates a new platform that features network and cloud connectivity. In accordance with the FDA's guidance document, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions," comprehensive cybersecurity risk management and verification and validation activities were conducted. The results of these cybersecurity assessments, the supporting documentation, and the cybersecurity management plan are included in this premarket submission.
12. Conclusion:
The subject device is substantially equivalent in the areas of technical characteristics, general function, application, and indications for use. The new device does not introduce a fundamentally new scientific technology, and the device has been validated through system level test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.