K244061 · Imperative Care, Inc. · MOF · Aug 26, 2025 · Cardiovascular
Device Facts
Record ID
K244061
Device Name
X-Wire Guidewire
Applicant
Imperative Care, Inc.
Product Code
MOF · Cardiovascular
Decision Date
Aug 26, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
Device Story
X-Wire Guidewire is a shapeable-tip guidewire used to access peripheral and neuro vasculature; facilitates introduction and positioning of catheters and interventional devices. Device features include 200cm-300cm lengths, varying stiffness profiles (Standard/Support), and 0.014"-0.024" diameters. Distal tip includes radiopaque marker for fluoroscopic visualization; hydrophilic coating on distal segment and PTFE coating on proximal segment reduce friction. Supplied with shaping mandrel, introducer, and torque device. Used by clinicians in interventional procedures; provides navigation through tortuous anatomy. Benefits include improved access to target vascular locations and delivery of interventional devices. Device is sterile, single-use, and EO-sterilized.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including dimensional verification, torqueability, torque strength, tip flexibility, tensile strength, kink resistance, coating integrity/lubricity, particulate evaluation, corrosion resistance, radiopacity, and simulated use. Biocompatibility testing per ISO 10993-1, -4, -5, -10, -11, -12, -23, and USP <151> confirmed non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, and hemocompatible status.
Indicated for general intravascular use in peripheral and neuro vasculature to introduce and position catheters and interventional devices. Not for use in coronary vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 26, 2025
Imperative Care, Inc.
Shivani Patel
Principal Regulatory Affairs Specialist
1359 Dell Avenue
Campbell, California 95008
Re: K244061
Trade/Device Name: X-Wire Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: July 23, 2025
Received: July 24, 2025
Dear Shivani Patel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K244061 - Shivani Patel
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K244061 - Shivani Patel
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K244061 | |
| Device Name X-Wire Guidewire | |
| Indications for Use (Describe) The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K244061
# 510(k) Summary
K244061
## A. Submitter Information:
Submitter's Name: Imperative Care, Inc.
Address: 1359 Dell Avenue
Campbell, CA 95008
Contact Person: Shivani Patel
Telephone: 510-468-8862
Email: spatel@imperativecare.com
Date of Preparation: August 21, 2025
## B. Subject Device:
Proprietary Name: X-Wire Guidewire
Common/Usual Name: Guidewire
Classification Name: Catheter guide wire
Product Codes: MOF, DQX
Regulation: 21 CFR 870.1330
Manufacturer: Imperative Care, Inc.
## C. Predicate Devices:
Proprietary Names: Aristotle 18 Guidewire; Aristotle 24 Guidewire (K231954)
Aristotle 14 Guidewire (K220398)
Common/Usual Name: Guidewire
Classification Name: Catheter guide wire
Product Codes: MOF, DQX
Regulation: 21 CFR. 870.1330
Manufacturer: Scientia Vascular, Inc.
## D. Device Description:
The Imperative Care’s X-Wire Guidewires are guidewires with shapeable tips to aid in accessing the peripheral and neuro vasculature. The guidewires are available in 200cm - 300cm lengths, standard (S) and support (T) stiffness profiles, and 0.014”, 0.018” and 0.024” diameters. The distal portion of the guidewire tip includes a radiopaque marker to facilitate fluoroscopic visualization. A hydrophilic coating on the distal segment and PTFE coating on the proximal segment serve to reduce friction during manipulation in vessels. The X-Wire Guidewire is supplied with a shaping mandrel, introducer, and torque device.
Page 1 of 8
{5}
K244061
# E. Indications for Use:
The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
# F. Comparison with the Predicate Devices:
The predicate devices are the Aristotle 18, 24, and 14 Guidewires cleared under K231954 and K220398. The predicate and subject devices share the same intended use and basic technological characteristics as shown in Table 1, below.
Table 1: Comparison of Subject and Predicate Devices
| | Predicate Aristotle 18 and 24 Guidewires (K231954) | Predicate Aristotle 14 Guidewire (K220398) | Subject Device X-Wire Guidewire (K244061) |
| --- | --- | --- | --- |
| Regulation Number | 21 CFR 870.1330 | Same | Same |
| Regulation Name | Catheter Guide Wire | Same | Same |
| Regulatory Class | Class II | Same | Same |
| Product Code | MOF, DQX | Same | Same |
| Manufacturer | Scientia Vascular, Inc. | Scientia Vascular, Inc. | Imperative Care., Inc |
| Anatomical Location | Peripheral and Neuro Vasculature | Same | Same |
| Indications for Use | The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | The Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. | The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature. |
{6}
K244061
| | Predicate Aristotle 18 and 24 Guidewires (K231954) | Predicate Aristotle 14 Guidewire (K220398) | Subject Device X-Wire Guidewire (K244061) |
| --- | --- | --- | --- |
| | in the coronary vasculature. | | |
| **Condition Supplied** | Sterile, single-use | Same | Same |
| **Accessory Devices** | Shaping Mandrel, Introducer, Torque Device | Same | Same |
| **Wire Outer Diameter** | 0.018” (0.46mm)
0.024” (0.61mm) | 0.014” (0.36mm) | 0.014” (0.36mm)
0.018” (0.46mm)
0.024” (0.61mm) |
| **Device Length** | 200 cm, 300 cm | 200cm, 300cm | X-Wire 14: Same as Aristotle 14
X-Wire 18: 200cm, 260cm, 300cm
X-Wire 24: Same as Aristotle 24 |
| **Tip Length (Hypotube)** | 35cm | 35cm | X-Wire 14S: 29cm
X-Wire 14T: 19cm
X-Wire 18S: 35cm
X-Wire 18T: 25cm
X-Wire 24S: 43.5cm
X-Wire 24T: 50cm |
| **Tip Type and Shape** | Core-to-tip Shapeable | Same | Same |
| **Flexible Tip Material (Distal Tip)** | Nitinol | Same | Same |
| **Core Wire Material** | Stainless Steel | Same | Same |
| **Stiffness Profile** | Support, Standard, Soft | Soft, Standard | Support, Standard |
| **Coatings** | Distal End: Hydrophilic
Proximal End: PTFE | Same | Same |
| **Hydrophilic Coating Length** | 46cm | 46cm | X-Wire 14S: 41cm
X-Wire 14T: 27.5cm
X-Wire 18S: 50cm
X-Wire 18T: 41cm
X-Wire 24S: 60cm
X-Wire 24T: 60cm |
Page 3 of 8
{7}
K244061
| | Predicate Aristotle 18 and 24 Guidewires (K231954) | Predicate Aristotle 14 Guidewire (K220398) | Subject Device X-Wire Guidewire (K244061) |
| --- | --- | --- | --- |
| Radiopaque Marker | 1 radiopaque marker at distal tip | Same | Same |
| Radiopaque Length | 10cm | 10cm | 85mm (8.5cm) |
| Bushing | Aristotle 24: Nitinol Bushing
Aristotle 18: None | None | None |
| Centering Coil | Aristotle 18: 1 centering coil
Aristotle 24: 2 centering coils | 1 centering coil | X-Wire 14S: 1 centering coil
X-Wire 14T: 1 centering coil
X-Wire 18S: 2 centering coils
X-Wire 18T: 1 centering coil
X-Wire 24S: 3 centering coils
X-Wire 24T: 2 centering coils |
| Sterilization Method | EO | Same | Same |
| Sterility Assurance Level | SAL of 10^{-6} | Same | Same |
| Shelf Life | 36 months | 4 months | 8 months |
G. Performance Data - Bench:
Bench testing was completed to evaluate the differences between the subject devices and predicate devices and to ensure the subject device meets the specifications and performs as intended. The test methods were based primarily on the FDA Guidance, "Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling, Guidance for Industry and Food and Drug Administration Staff," issued in October 2019. A summary of the evaluated performance specifications is presented in Table 2.
The test results demonstrated that any differences between the subject devices and predicate devices do not significantly impact performance parameters that would negatively affect the safety or effectiveness of the subject devices.
Page 4 of 8
{8}
K244061
Table 2: Summary of Bench Tests and Performance Specifications
| Test Attribute | Specification | Results |
| --- | --- | --- |
| Dimensional Verification | All defined guidewire dimensions are within the specified tolerances. | Pass |
| Visual Inspection | The guidewire shall be free of visual defects when removed from packaging. | Pass |
| Torqueability | The guidewire shall transmit rotation from the proximal end to the distal tip to allow users to select branches of the vasculature and reach the target location. | Pass |
| Torque Strength | In the event where the distal end of the guidewire is unable to move/rotate during the procedure, the device shall not fail under expected torsional input. | Pass |
| Tip Flexibility | The tip of the guidewire shall not cause vessel damage. Tip flexibility is defined by the ability of the tip to buckle when contacting vessel walls. In doing so it reduces the contact pressure to prevent vessel perforation. | Pass |
| Tensile Strength and Tip Pull | The guidewire shall withstand tensile forces expected in clinical use without breaking. | Pass |
| Kink Resistance | The guidewire shall be able to traverse through clinically relevant bends without kinking. | Pass |
| Coating Integrity | The coating shall remain intact during clinical use. | Pass |
| Coating Lubricity | The coating shall be lubricious to reduce frictional forces to allow for navigation through tortuous anatomy. | Pass |
| Particulate Evaluation | The guidewire shall not generate particles at a level that is greater than the range generated by the predicate. | Pass |
| Corrosion Resistance | The guidewire shall not corrode from the time of manufacture through its shelf life. | Pass |
| Radiopacity | Opacity to x-ray shall allow physicians to visualize guidewire under fluoroscopy. | Pass |
| Simulated Use | The guidewire must be able to reach anatomical locations and deliver catheters and other interventional devices used in common neurovascular procedures. | Pass |
{9}
K244061
# H. Biocompatibility Testing:
Biocompatibility testing was performed on a representative final, finished guidewire to evaluate all direct patient contacting components of the device. This testing was conducted to ensure that the components and raw materials used in the subject X-Wire Guidewires and the manufacturing processes and sterilization processes result in a biocompatible product. Test results and an expert evaluation confirmed that the subject X-Wire Guidewires met biological safety requirements per ISO 10993-1 for externally communicating medical devices with direct circulating blood contact for $\leq 24$ hours.
Table 3: Biocompatibility Test Summary - X-Wire Guidewire
| Test | Test Method | Extraction Methods/Conditions | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| Cytotoxicity | ISO 10993-5 ISO 10993-12 | Test device extracted with MEM with 5% serum at 37°C for 72 hours. | Test article extracts must yield grade 2 or lower. | Pass, Non-cytotoxic |
| Sensitization | ISO 10993-10 ISO 10993-12 | Test device extracted with both SC and SO at 50°C for 72 hours. | Overall pattern, intensity, duration, character of reactions compared to control conditions. | Pass, Non-sensitizer |
| Irritation or Intracutaneous Reactivity | ISO 10993-23 ISO 10993-12 | Test device extracted with both SC and SO at 50°C for 3 hours. | Difference between test extract mean score and corresponding control mean score ≤1. | Pass, Non-irritating |
| Acute Systemic Toxicity | ISO 10993-11 ISO 10993-12 | Test device extracted with both SC and SO at 50°C for 72 hours. | None of the animals treated with the test article extracts must show a significantly greater biological reactivity than animals treated with the control article. | Pass, Non-toxic (acute systemic) |
| Material Mediated Pyrogenicity | ISO 10993-11 USP <151> ISO 10993-12 | Test device extracted with SNPS at 50°C for 72 hours. | Test article extract must yield <0.5°C rise of | Pass, Non-pyrogenic |
| | | | concentration of <0.5% increase of the amount of the material added to the test article. | reaction rate of <0.5% increase of the amount of the material added to the test article. |
{10}
K244061
| Test | Test Method | Extraction Methods/Conditions | Acceptance Criteria | Results |
| --- | --- | --- | --- | --- |
| | | | individual animal baseline temperature. | |
| Hemocompatibility – Hemolysis (Direct and Indirect Contact) | ISO 10993-4 ASTM F756 ISO 10993-12 | Direct Contact: Test device incubated directly in diluted blood at 37°C for 72 hours.
Indirect Contact: Test device extracted in CMF-PBS at 50°C for 72 hours and then extracts incubated with diluted blood at 37°C for 3 hours. | Test article extract must result in hemolytic index of <2%. | Pass, Non-hemolytic for both Direct and Extract Methods |
| Hemocompatibility - SC5b-9 Complement Activation | ISO 10993-4 | Test device exposed to Normal Human Serum at 37°C for 60 minutes. | SC5b-9 concentration of the test article is similar to both the activated NHS and negative controls. | Pass, Hemocompatible |
| Hemocompatibility – In vivo thromboresistance | ISO 10993-4 | N/A, non-anticoagulated venous implant study. | Test article performs similarly to the predicate. | Pass, Hemocompatible |
I. Sterilization:
The X-Wire Guidewires are sterilized using a validated EO process with a sterility assurance level of $1 \times 10^{-6}$ validated per the overkill method in accordance with ISO 11135, "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices".
J. Shelf Life and Packaging:
Accelerated aging testing based on ASTM F1980 was conducted to verify device, accessory, and packaging performance. An accelerated aging equivalent of 8 months was used to support 8-month shelf life. Device performance was verified by functional and performance testing.
Page 7 of 8
{11}
K244061
Packaging and sterile barrier integrity through transportation challenge has been verified. Aging testing has also been performed that supports the sterile barrier integrity following 8 months of accelerated aging.
**K. Animal Study:**
Animal study was not deemed necessary to demonstrate substantial equivalence.
**L. Clinical Study:**
Clinical study was not deemed necessary to demonstrate substantial equivalence.
**M. Conclusions:**
The X-Wire Guidewire has the same intended use and similar technological characteristics as the predicate devices. The performance testing was conducted to ensure the subject device performs as intended, meets all design specifications, and performs consistently throughout its labeled shelf life. The differences between subject and predicate devices do not raise any new or different questions of safety and effectiveness. Therefore, the X-Wire Guidewire is substantially equivalent to the predicate devices.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.