Shenzhen Jermei Medical Device Technology Co., Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Sep 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The portable mesh nebulizer is intended to aerosolize medicine liquid for inhalation by the patient. The device is used with adult and pediatric (≥3 years old) both in hospital, home, and sub-acute settings. It is not intended for use with Pentamidine.
Device Story
Portable vibrating mesh nebulizer; utilizes piezoelectric transducer vibrating at nominal frequency to aerosolize liquid medications into fine particulates for inhalation; treats respiratory diseases. Powered by internal 3.7V DC lithium battery; rechargeable via adapter. Single-patient, multiple-use device; includes mask. Used in hospital, home, and sub-acute settings; operated by patient or caregiver. Healthcare provider prescribes medication; device transforms liquid into aerosol for patient inhalation; facilitates drug delivery to respiratory tract. Benefits include portable, efficient medication administration.
Clinical Evidence
Bench testing only. Aerosol characterization performed per FDA 1993 guidance using Levosalbutamol hydrochloride, Ipratropium bromide, and Budesonide. Compared MMAD, GSD, respirable mass, and respirable fraction against predicate. Results showed no statistically significant differences at 95% confidence level. Biocompatibility evaluated per ISO 10993 and ISO 18562. Electrical safety and EMC verified per IEC 60601-1, 60601-1-2, and 60601-1-11.
Technological Characteristics
Vibrating mesh nebulizer; piezoelectric transducer; 3.7V DC lithium battery. Shell: ABS; Medication cup: PCTG; Mask: TPU. Frequency: ~114KHz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 27427, ISO 10993, ISO 18562. Dimensions: 62.5x52.5x156mm; Weight: 176g. Connectivity: Type-C charging port.
Indications for Use
Indicated for aerosolizing liquid medication for inhalation by adult and pediatric patients (≥3 years old) in hospital, home, and sub-acute settings. Contraindicated for use with Pentamidine.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 19, 2025
Shenzhen Jermei Medical Device Technology Co., Ltd.
% James Tsai
Consultant
Shenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square
Nanshan District
Shenzhen, Guangdong 518000
China
Re: K244035
Trade/Device Name: Portable mesh nebulizer (JM821)
Regulation Number: 21 CFR 868.5630
Regulation Name: Nebulizer
Regulatory Class: Class II
Product Code: CAF
Dated: August 21, 2025
Received: August 21, 2025
Dear James Tsai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K244035 - James Tsai
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K244035 - James Tsai
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K244035 | |
| Device Name Portable mesh nebulizer | |
| Indications for Use (Describe) The portable mesh nebulizer is intended to aerosolize medicine liquid for inhalation by the patient. The device is used with adult and pediatric (≥3 years old) both in hospital, home, and sub-acute settings. It is not intended for use with Pentamidine. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
1. Administrative information
| Date of Summary prepared: | September 17th, 2025 |
| --- | --- |
| Manufacturer information: | Company: Shenzhen Jermei Medical Device Technology Co., Ltd.
Address: Room 201, 202, 4th Bldg XianYuXing Company, 4th Bldg XianYuXing industrial Zone, Yuhe Road, Gonghe Community, Shajing Street, Baoan District, Shenzhen, P. R. China
Contact person: Ms. Li Fang
Phone: +86-755-27800690
Fax: +86-755-27800690
E-mail: lifang@szjermei.com |
| Submission Correspondent: | Company: Shenzhen Joyantech Consulting Co., Ltd.
Address: 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District, Shenzhen
Contact person: James Tsai
Phone: +86-755-86069197
E-Mail: james_tsai@cefda.com |
2. Device Information
| Type of 510(k) submission: | Traditional |
| --- | --- |
| Trade Name: | Portable mesh nebulizer |
| Classification name: | Nebulizer (Direct patient interface) |
| Review Panel: | Anesthesiology |
| Product Code: | CAF |
| Device Class: | II |
| Regulation Number: | 868.5630 |
3. Predicate Device Information
| 510(K) Number: | K201397 |
| --- | --- |
| Trade Device: | Ultrasonic mesh nebulizer |
| Classification name: | Nebulizer (Direct patient interface) |
| Product Code: | CAF |
| Device Class: | II |
| Regulation Number: | 868.5630 |
| Review Panel: | Anesthesiology |
1 / 8
{5}
2 / 8
# 4. Device Descriptions
JM821 portable mesh nebulizer is a portable vibrating mesh nebulizer, which incorporates a piezoelectric transducer that vibrates at a nominal frequency when electrical current is applied. It is designed to nebulize liquid medications into fine particulates for inhalation by a patient, so as to therapy respiratory diseases.
The subject device is portable size, handheld design, it is powered by a 3.7V d.c. internally lithium battery, which is charged by an adapter.
The subject device does not contain any medicine. Only use physician-prescribed solutions that are approved for use with a general-purpose nebulizer. Consult drug manufacturer's instructions regarding suitability for nebulization.
The portable mesh nebulizer is single-patient, multiple-use. The mask can be can used for 30 sessions of aerosol treatment.
# 5. Intended Use/Indications for Use
The portable mesh nebulizer is intended to aerosolize medicine liquid for inhalation by the patient. The device is used with adult and pediatric (≥3 years old) both in hospital, home, and sub-acute settings.
It is not intended for use with Pentamidine.
# 6. Comparisons of technological characteristics with the predicate device
The subject nebulizer and Ivankaca nebulizer (K201397) are similar in intended purpose, structures, technical characteristics, and operation principle, the following table provides a comparison of the two devices.
| Items | Subject Device (JM821, K244035) | Predicate Device (Ivankaca, K201397) | Comments |
| --- | --- | --- | --- |
| Regulation number | 21 CFR 868.5630 | 21 CFR 868.5630 | Substantial equivalence |
| Regulatory class | Class II | Class II | Substantial equivalence |
| Product code | CAF | CAF | Substantial equivalence |
| Intended Use/Indication for Use | The portable mesh nebulizer is intended to aerosolize medicine liquid for inhalation by the patient. The device is used with adult and pediatric (≥3 years old) both in hospital, home, and sub-acute settings. It is not intended for use with Pentamidine. | The Ultrasonic Mesh Nebulizer designed to aerosolized liquid medications for inhalation by patient, the device may be used with pediatric (>4 years of age), defined by the prescribed medication, and adult patients in the home, hospital and sub-acute care settings. It is not intended for use with Pentamidine. | Substantial equivalence (Note 1) |
{6}
| Items | Subject Device (JM821, K244035) | Predicate Device (Ivankaca, K201397) | Comments |
| --- | --- | --- | --- |
| | | Pentamidine. | |
| Operation principle | Ultrasonic, vibrating mesh | Ultrasonic, vibrating mesh | Substantial equivalence |
| Environment of use | Clinic, home | Clinic, home | Substantial equivalence |
| OTC or Rx | Prescription | Prescription | Substantial equivalence |
| Lithium battery | d.c. 3.7 V | d.c. 3.7 V | Substantial equivalence |
| AC adapter/charger | Input: a.c. 100-240V, 50/60Hz, 0.16-0.07A; output: d.c. 5.0V, 1.0A Recharging cable/Type-C cable | Recharging cable/USB cable | Difference (Note 2) |
| Vibration frequency | Approx.114KHz | Approx.110KHz | Difference (Note 3) |
| Nebulization Rate/Aerosol Flowrate | ≥0.25ml/min | ≥0.2ml/min | Difference (Note 4) |
| Medicine Capacity | 25ml maximum 0.5ml minimum | 8ml maximum 0.5ml minimum | Difference (Note 5) |
| Hole size (nozzle) | 18mm | 16mm | Difference (Note 3) |
| Size of the nebulizer | 62.5*52.5*156mm | 42(L)x55(W)x109(H) mm (MY-123) | Difference (Note 6) |
| Weight of the nebulizer | 176 g | Model MY-123: 113±5g | Difference (Note 6) |
| Duration of use | Single patient, multi-use | Single patient, multi-use | Substantial equivalence |
| Materials | Shell: ABS; Medication cup: PCTG; Mask: TPU | PVC Others: Not publicly available | Difference (Note 7) |
| Patient Connector | Mouthpiece or mask | Mask | Substantial equivalence (Note 8) |
| Operating conditions | 5°C-40°C, ≤ 90%RH, 70kpa~106kpa | 5°C to 40°C, 15% to 90% RH | Substantial equivalence |
| Storage conditions | -25°C-70°C, ≤ 90%RH, 70kpa~106kpa | -25°C to 70°C, ≤90% RH | Substantial equivalence |
| Electrical safety & EMC | IEC60601-1 IEC60601-1-2 IEC60601-1-11 | IEC60601-1 IEC60601-1-2 IEC60601-1-11 | Substantial equivalence |
Note 1: Both the subject device and predicate device have the same indications for use, except the intended patient population, the subject device claims 3 years old
{7}
child, however, both 3-year-old and 4-year-old fall under the same category of children, which shall be used the subject device under the supervision of adults. In addition, other legally marketing devices also claims 3-year-olds or even younger patients, such as FOX MOBILE (K142059) claims patients of 3-year-old, InnoSpire Go (K170853) claims patients of 2-year-old. Therefore, the difference will not raise any new issues of safety and effectiveness for the subject device.
Note 2: The subject device provides the AC adapter and the charging port is type-C interface, which is different from the predicate device, the non-clinical tests show that the subject device conforms to the standards of IEC60601-1, IEC 60601-1-2, and IEC 60601-1-11. Therefore, the difference will not raise any new safe and effective issues for the subject device.
Note 3: The aerosol particles performance comparisons of the subject device and the predicate device are performed and analyzed due to the differences in the specifications of mesh sheet, hole size, and medication cup etc. Aerosol particulate performance comparisons are carried out in adult mode and child mode, the statistical analysis showed that differences in particulate characteristics are not statistically significant in the parameters of Total Mass, MMAD, GSD, Respirable fraction and Respirable mass between the subject device and the predicate device in a confidence level of 95%. Therefore, the difference will not raise any new issues of safety and effectiveness for the subject device.
Note 4: The nebulization rate claimed by the subject device are slightly larger than the predicate device, which means the subject device has a better nebulization efficiency, in the aerosol particle performance comparison testing, the "Durations from starting nebulization to ending nebulization" between the two devices are compared, and the results showed no statistical significance in a confidence level of 95%. Therefore, the difference will not raise any new safe and effective issues for the subject device.
Note 5: The medication volumes depend on the prescriptions by the doctors, the difference on the maximum fill volume will not influence the safety and performance of subject device, and will not raise any new issues of safety and effectiveness for the subject device.
Note 6: The differences of sizes and weights depend on designing variations among different manufacturers and differences in components. However, the subject device has undergone Electric safety and EMC tests, as well as performance tests and reliability tests. The difference will not raise any new issues of safety and effectiveness for the subject device.
Note 7: Materials of predicate device are not publicly available except for PVC, but the biocompatibility evaluations including the biocompatibility tests of the subject device are carried out according to FDA biological guidance, and the results of biocompatibility evaluation demonstrates a good biocompatibility performance of the subject device. Therefore, the differences will not raise any new safe and effective issues for the subject device.
Note 8: The accessories of predicate device include mask and mouthpiece, while the subject device contains only mask, the labeling of subject device is indicated
4 / 8
{8}
the accessory of mask clearly, and only mask is contained in the package. Therefore, the difference will not raise any new issues of safety and effectiveness for the subject device.
The subject device and the predicate device are similar in intended use/indication for use, structures, technical characteristics, and working principles, and there is no significant statistical difference in aerosol particle characteristics between subject device and predicate device. The minor differences will not raise any new issue on safety and effectiveness of the subject device. Therefore, the subject device is as safe and as effective to the legally marketed predicate device.
## 7. Brief discussions of the nonclinical tests
The following performance data were used in support of the substantial equivalence determination:
## 7.1 Electrical Safety and EMC of medical electrical device
- IEC 60601-1: 2005+A1:2012+ A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11: 2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 27427:2023 Respiratory therapy equipment - Part1: Nebulizing systems and their components Anaesthetic and respiratory equipment - Nebulizing systems and components
## 7.2 Biocompatibility evaluation
- ISO 10993-1: 2018 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process
- ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity
- ISO 10993-10: 2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation
- ISO 10993-17: 2023 Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents
- ISO 10993-18: 2020 Biological evaluation of medical devices - part 18: Chemical characterization of medical device materials within a risk management process
5 / 8
{9}
> ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
> ISO 18562-2:2017 Tests for emissions of volatile o Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
> ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds (VOCs)
# 7.3 Software verification and validation
FDA guidance: Content of Premarket Submissions for Device Software Functions
# 7.4 Cleaning and disinfection validation
FDA guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling
# 7.5 Performance data
Aerosol characterization testing for the subject devices and predicate devices were conducted in accordance with the relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH - 1993). Tests were performed at both low and high supplied air flow rates for the subject devices. For each device, three drugs: Levosalbutamol hydrochloride, Ipratropium bromide, Budesonide were tested respectively. The comparison items include Mass Median Aerodynamic Diameter (MMAD), Geometric Standard Deviation (GSD), Respirable mass, Respirable Fraction and Total Mass. Also, the variability of intra-sample and inter-sample was tested for the subject devices.
a) Adult mode: aerosol particle characteristics (the sampling rate is 28 L/min, adult mask)
| Parameters | Drugs | Subject nebulizer | Predicate nebulizer |
| --- | --- | --- | --- |
| Total Mass (μg) | Levosalbutamol hydrochloride | 278.15 ± 4.82 | 271.93 ± 6.80 |
| | Ipratropium bromide | 248.67 ± 25.80 | 219.33 ± 24.25 |
| | Budesonide | 223.40 ± 13.80 | 226.01 ± 4.45 |
| Particle Size-MMAD (μm) | Levosalbutamol hydrochloride | 2.76 ± 0.10 | 2.93 ± 0.09 |
| | Ipratropium bromide | 2.65 ± 0.08 | 2.79 ± 0.13 |
| | Budesonide | 2.75 ± 0.14 | 2.88 ± 0.23 |
| Geometric Standard Deviation | Albuterol sulfate | 2.81 ± 0.31 | 2.52 ± 0.08 |
| | Ipratropium bromide | 2.45 ± 0.18 | 2.74 ± 0.19 |
| | Budesonide | 2.60 ± 0.28 | 2.38 ± 0.17 |
{10}
(b) Pediatric mode: aerosol particle characteristics (the sampling rate is 12 L/min, child mask)
| Parameters | Drugs | Subject nebulizer | Predicate nebulizer |
| --- | --- | --- | --- |
| Total Mass (μg) | Levosalbutamol hydrochloride | 265.16 ± 12.09 | 264.15 ± 5.44 |
| | Ipratropium bromide | 227.43 ± 15.70 | 236.34 ± 19.76 |
| | Budesonide | 238.87 ± 22.57 | 227.13 ± 23.29 |
| Particle Size-MMAD (μm) | Levosalbutamol hydrochloride | 3.07 ± 0.05 | 3.15 ± 0.12 |
| | Ipratropium bromide | 3.17 ± 0.20 | 3.20 ± 0.16 |
| | Budesonide | 3.30 ± 0.08 | 3.31 ± 0.19 |
| Geometric Standard Deviation (GSD) | Albuterol sulfate | 2.31 ± 0.08 | 2.34 ± 0.03 |
| | Ipratropium bromide | 2.42 ± 0.11 | 2.38 ± 0.05 |
| | Budesonide | 2.47 ± 0.04 | 2.50 ± 0.12 |
| Respirable fraction (%, 0.5-5μm) | Levosalbutamol hydrochloride | 70.49% ± 0.01% | 69.14% ± 0.02% |
| | Ipratropium bromide | 67.97% ± 0.03% | 68.12% ± 0.02% |
| | Budesonide | 65.79% ± 0.01% | 65.49% ± 0.03% |
| Respirable Mass (μg, 0.5-5μm) | Levosalbutamol hydrochloride | 186.95 ± 10.21 | 182.65 ± 5.90 |
| | Ipratropium bromide | 154.28 ± 5.65 | 160.95 ± 14.02 |
| | Budesonide | 157.21 ± 15.78 | 148.24 ± 7.86 |
| Confidence level of testing | The test and number of samples tests provided a 95% confidence | | |
# 8. Conclusions
The results of the substantial equivalence assessment, taken together with safety and performance testing data, demonstrate that JM821-Portable mesh nebulizer is
{11}
substantially equivalent to the predicate device.
8 / 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.