The DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.
Device Story
DCX Disposable Cassette provides sterile fluid pathway for kidney preservation; houses/protects kidneys during pulsatile hypothermic machine perfusion. System uses gravity flow; perfusate flows from organ chamber to pumphead, through heat exchanger, to bubble trap, then to cannulated kidney(s); returns to organ chamber via kidney vein. Used in healthcare professional environments by trained personnel. Device allows perfusion of one or two kidneys (individual or bloc). Direct oxygenation occurs in organ chamber; removable partition allows reduced perfusate volume (1L vs 2L) for single organ perfusion. Output is preserved kidney ready for transplantation. Benefits include improved storage conditions and reduced perfusate volume requirements.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993 (cytotoxicity, sensitization, irritation, hemolysis, acute systemic toxicity). Sterilization validated per ISO 11135 and ISO 10993-7. Stability testing confirmed 24-month shelf life at 2-25°C. Performance testing conducted with RM4 Control Unit confirmed device meets all functional requirements.
Technological Characteristics
Single-use disposable cassette; sterile barrier system; Ethylene Oxide sterilization (SAL 10^-6). Materials evaluated per ISO 10993. Features: luer connections, direct organ chamber oxygenation, removable partition for volume optimization, dual-lid design. Compatible with RM4 control unit. Shelf life: 24 months. Storage: 2°C to 25°C.
Indications for Use
Indicated for pulsatile hypothermic machine perfusion of kidneys for preservation and eventual transplantation into a recipient in a healthcare professional environment.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 25, 2025
Institut Georges Lopez (IGL)
Matthieu Prouteau
Director of Quality & Regulatory Affairs
Parc Tertiaire du Bois Dieu
1 Allée des Chevreuils
Lissieu, 69380
France
Re: K243998
Trade/Device Name: DCX Disposable Cassette (DCX)
Regulation Number: 21 CFR§ 876.5880
Regulation Name: Isolated kidney perfusion and transport system and accessories
Regulatory Class: II
Product Code: KDN
Dated: March 27, 2025
Received: March 27, 2025
Dear Matthieu Prouteau:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243998 - Matthieu Prouteau
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243998 - Matthieu Prouteau
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243998
Device Name
DCX Disposable Cassette (DCX)
Indications for Use (Describe)
The DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
IGL
Traditional 510(k) Premarket Notification – DCX Disposable Cassette
K243998
Page 1 of 4
# 510(k) Summary
[As required by 21 CFR 807.92]
Written on December, 17 2024
## I. Submitter
Institut Georges Lopez (IGL)
Parc Tertiaire du Bois Dieu
1 allée des Chevreuils
69380 Lissieu
France
Telephone: +33 437 646 332
Official Contact: Matthieu Prouteau – Director of Quality and Regulatory Affairs
## II. Device Information
Trade Name: DCX Disposable Cassette (DCX)
Common Name: System, Perfusion, Kidney
Regulation Description: Isolated kidney perfusion and transport system and accessories
Regulation Number: 21 CFR 876.5880
Class: Class II
Product Code: KDN
## III. Predicate Device Identification
| | K Number | Predicate Device Name | Manufacturer |
| --- | --- | --- | --- |
| Predicate | K211224 | RM4 control unit | Institut Georges Lopez - IGL (France) |
## IV. Device Description
The DCX Disposable Cassette, provides a sterile fluid pathway, houses and protects the kidneys during perfusion and it has a gravity flow system which allows circulation of perfusate through the kidneys.
The kidneys are placed in an organ chamber filled with perfusate. The perfusate flows from the organ chamber to the pumphead. Then, it is pumped through the heat exchanger to the bubble trap where it is delivered to the cannulated kidney(s). The perfusate then returns to the organ chamber via the kidney vein to repeat the perfusion cycle.
The cassette can provide circulation of perfusate to one or two kidneys, attached individually or in bloc.
## V. Indication for Use
The DCX Disposable Cassette, as part of the RM4 Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.
Confidential
Page 1 of 4
{5}
IGL
Traditional 510(k) Premarket Notification – DCX Disposable Cassette
K243998
Page 2 of 4
# VI. Comparison of Technological Characteristics
The DCX Disposable Cassette and the DCM-100 Disposable Cassette used with the Kidney Perfusion system have identical technical specifications.
DCX Disposable Cassette is the new version of the DCM-100 Disposable Cassette, compatible with the RM4 Perfusion system (control unit). In order to improve the system and the condition of kidney storage, IGL develops for the new cassette, following technical characteristics:
- connection luer for perfusion accessories
- in the DCX, the oxygenation is provided directly in the organ chamber, whereas on the DCM-100 the perfusate circulates through a membrane where O₂ exchange occurs (two different means to bring oxygen to the perfusate).
- the placement of a removable partition to allow the immersion of the organ using 1L of perfusate instead of 2L of perfusate for single organ perfusion in the DCX. In DCM-100, even if it for a single perfusion, 2L of solution have to be poured inside the system.
- the number of lids. Indeed, two lids are on the DCX whereas only one lid is on the DCM-100.
Both cassettes are packaged into a primary packaging which is considered as the sterile barrier system, then sterilized with the Ethylene Oxide sterilization method and intended for single-use only. Finally, both cassettes are labeled as sterile, single use and with a 24 months shelf life.
The difference on the labelling is on the storage conditions. Indeed, the DCX is labeled with storage conditions between 2°C and 25°C with temporary 40°C outing tolerated instead of 5°C to 50°C for DCM-100. Thus, even if the storage condition of both cassettes are different, the range of temperature is similar.
| Comparison with Predicate Device | | |
| --- | --- | --- |
| | Subject Device | Primary Predicate Device |
| | RM4 Kidney Perfusion System with DCX | RM4 Kidney Perfusion System with DCM-100 (K211224). |
| Regulatory Specifications | | |
| Intended Use | Kidney storage and preservation for transplantation | Identical |
| Indication for Use | The DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment. | The RM4 control unit, as part of the RM4 Kidney Perfusion System is intended to be used for the pulsatile hypothermic machine perfusion of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment. |
| Principle of Operation | Single use, disposable cassette used with a control unit which provides continuous pulsatile hypothermic physiologic perfusion | Identical |
| Manufacturer | Institut Geoges Lopez, France | Identical |
| Regulation number | 21 CFR 876.5880 | Identical |
Confidential
Page 2 of 4
{6}
IGL
Traditional 510(k) Premarket Notification – DCX Disposable Cassette
K243998
Page 3 of 4
| Comparison with Predicate Device | | |
| --- | --- | --- |
| | Subject Device | Primary Predicate Device |
| | RM4 Kidney Perfusion System with DCX | RM4 Kidney Perfusion System with DCM-100 (K211224). |
| Regulation description | Isolated kidney perfusion and transport system and accessories | Identical |
| Product Code | KDN | Identical |
| Class | Class II | Identical |
| Technical Specifications | | |
| Organ Type | Kidney | Identical |
| Organ Capacity | One or two kidneys | Identical |
| Solution use to preserve the organ | PERF-GEN or equivalent solution | Identical |
| Transportability | Short transportation within the healthcare facility by trained personnel. | Identical |
| Compatible Elements | RM4 control Unit | Identical |
| Disposable compatible elements | Cannulas and clamps | Identical |
| Cassette mounting | Top of control unit | Identical |
| Sterilization, Shelf Life, and Storage | | |
| Sterilization Method | Ethylene Oxide sterilization | Identical |
| Single use | Yes | Identical |
| Sterility Assurance level (SAL) | 10^{-6} | Identical |
| Shelf life | 24 months | Identical |
# VII. Summary of Non-Clinical Testing
No performance standard has been established by FDA for the DCX Disposable Cassette.
In order to evaluate the performance and safety of the subject device DCX. :
- The DCX is supplied sterile and non-pyrogenic in order to assure safety for transplant recipients. The validation of the sterilizing was carried out according to ISO 11135 and ISO 10993-7.
- The biomaterial safety of the DCX has been evaluated through ISO 10993 compliant testing, which included cytotoxicity test, skin sensitization test in guinea pigs, primary skin irritation, hemolysis test and acute systemic toxicity test in mice. Results of this testing, showed the DCX is safe for the intended biocontact.
- The stability testing has showed that aging of test articles at the recommended storage conditions of 2-25°C (35.6° - 77°F) does not affect the product specifications for the DCX labeled with 24 months shelf life.
- Performances tests have been performed also with perfusion accessories, and a RM4 Control Unit, in accordance with the recommendations outlined in the IFU of the subject device and in the instructions for use of the RM4 Control Units. For each test performed, the subject performs as expected and meets the requirements set for the device.
The results of these tests indicate that DCX Disposable Cassette is substantially equivalent to the
Confidential
Page 3 of 4
{7}
IGL
Traditional 510(k) Premarket Notification – DCX Disposable Cassette
K243998
Page 4 of 4
predicate devices.
## VIII. Conclusion
The technological differences between the subject and predicate devices were evaluated through non-clinical testing. The results of these tests demonstrated that the subject device does not raise new issues of safety and effectiveness compared to the predicate device. The indications for use, technological characteristics, and performance characteristics of DCX Disposable Cassette stored for up to 24 months assessed to be substantially equivalent to the predicate device.
Confidential
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.