K243990 · Optmed, Inc. · MPN · Sep 19, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243990
Device Name
BondEase 2 Topical Skin Adhesive
Applicant
Optmed, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
Sep 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BondEase® 2 Topical Skin Adhesive is intended for topical use only, to hold together the skin edges of incisions and lacerations that are under minimal tension and easily approximated. Where significant tension exists on incisions or lacerations, BondEase® 2 Topical Skin Adhesive should be used in conjunction with deep dermal stitches.
Device Story
BondEase 2 is a sterile, liquid topical skin adhesive for closing incisions and lacerations. It consists of a methylidene malonate monomer formulation (Part A) and a polymerization activator (Part B) contained within a single-use applicator. The applicator features a plastic tube, a crushable glass ampoule, a porous filter, and a tip subassembly. To operate, the user mixes the two solutions to initiate polymerization. The device is applied topically by a clinician to approximate skin edges. It functions as a mechanical barrier and adhesive to hold tissue together during healing. It is intended for use in clinical settings where skin closure is required. The device benefits patients by providing a non-invasive or minimally invasive method for wound closure, potentially reducing the need for traditional sutures or staples in low-tension wounds.
Clinical Evidence
No human clinical data provided. Evidence includes biocompatibility testing (ISO 10993-5, -6, -10, -11; USP <151>), bench testing (ASTM F2458-05, F2258-05, F2255-05), and a porcine animal study evaluating healing and local biological response compared to a predicate device.
Technological Characteristics
Methylidene malonate monomer-based tissue adhesive. Two-part system (monomer and activator). Single-use applicator with crushable glass ampoule, porous filter, and tip subassembly. Sterile. Bench testing per ASTM F2458-05, F2258-05, and F2255-05. Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for topical approximation of skin edges of incisions and lacerations under minimal tension; may be used with deep dermal stitches for incisions/lacerations under significant tension.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
9/19/2025
OptMed, Inc.
% Lucie Dalet
Principal Consultant
Rqm+
2790 Mosside Blvd.
Suite 800
Monroeville, Pennsylvania 15146
Re: K243990
Trade/Device Name: BondEase 2 Topical Skin Adhesive
Regulation Number: 21 CFR 878.4010
Regulation Name: Tissue Adhesive
Regulatory Class: Class II
Product Code: MPN
Dated: August 25, 2025
Received: August 25, 2025
Dear Lucie Dalet:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243990 - Lucie Dalet
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243990 - Lucie Dalet
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Wilmarie Flores -S
Digitally signed by Wilmarie Flores -S
For Jodie Giordano, Ph.D.
Acting Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243990 | |
| Device Name BondEase 2 Topical Skin Adhesive | |
| Indications for Use (Describe) BondEase® 2 Topical Skin Adhesive is intended for topical use only, to hold together the skin edges of incisions and lacerations that are under minimal tension and easily approximated. Where significant tension exists on incisions or lacerations, BondEase® 2 Topical Skin Adhesive should be used in conjunction with deep dermal stitches. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K243990
# 510(k) SUMMARY
## DATE PREPARED
September 18, 2025
## MANUFACTURER AND 510(k) OWNER
OptMed, Inc.
745 Fifth Ave., Suite 500
New York, New York 10151
Telephone: +1 (833) 678-6331
Official Contact: Nitasha Yajnik, Associate Director, Quality and Regulatory Assurance
## REPRESENTATIVE/CONSULTANT
Lucie Dalet, Ph.D
Allison C. Komiyama, Ph.D., RAC
RQM+
Telephone: +1 (412) 816-8253
Email: ldalet@rqmplus.com, akomiyama@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: BondEase 2 Topical Skin Adhesive
510(k) Number: K243990
Common Name: Tissue Adhesive For The Topical Approximation Of Skin
Regulation Number: 21 CFR 878.4010
Class: II
Product Code: MPN
Premarket Review: Division of Health Technology 4 B (Plastic and Reconstructive Surgery Devices)
Review Panel: General & Plastic Surgery
## PREDICATE DEVICE IDENTIFICATION
BondEase 2 is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
| --- | --- | --- |
| K152096 | Dermabond Advanced Topical Skin Adhesive / Ethicon, Inc. | ✓ |
| K141157 | BondEase Topical Skin Adhesive / OptMed, Inc. | |
The predicate devices have not been subject to a design-related recall.
## DEVICE DESCRIPTION
BondEase 2 Topical Skin Adhesive is a sterile, liquid topical skin closure device composed of a methylidene malonate monomer formulation. The adhesive is composed of two solutions which must be mixed to polymerize: the methylidene malonate monomer solution (Part A) and the polymerization activator solution (Part B). BondEase 2 is provided as a single-use applicator, packaged in a foil pouch. The applicator comprises a plastic tube containing a crushable glass
Page 1 of 3
{5}
K243990
ampoule, a porous filter, and a tip subassembly, which is inserted into a plastic handle. With the exception of the plastic handle, the device is identical in design, formulation, and manufacturing, to the reference device TearRepair Liquid Skin Protectant cleared in K241916.
## INDICATIONS FOR USE
BondEase® 2 Topical Skin Adhesive is intended for topical use only, to hold together the skin edges of incisions and lacerations that are under minimal tension and easily approximated.
Where significant tension exists on incisions or lacerations, BondEase® 2 Topical Skin Adhesive should be used in conjunction with deep dermal stitches.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
OptMed, Inc. believes that BondEase 2 is substantially equivalent to the predicate devices based on the information summarized here:
Both the subject and predicate devices are intended for the topical approximation of skin edges of incisions and lacerations. While the subject device and the primary predicate device cleared in K152096 have different formulations and dispenser designs, the principle of operation, mechanism of action, and conditions of use are identical. The subject device has a similar design and uses the same monomer as the predicate device cleared in K141157 for the same intended use. Since the differences in formulation and dispenser design do not impact the indications for use, principle of operation, or conditions of use of the device, they do not raise different questions of safety and effectiveness compared to the primary predicate device. These technological characteristics have undergone testing to ensure the device is substantially equivalent to the primary predicate.
## SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate substantial equivalence based on current industry standards as recommended by FDA in the Special Controls Guidance Tissue Adhesive for the Topical Approximation of Skin:
### Biocompatibility:
Biocompatibility testing performed on the subject device included:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation (ISO 10993-10)
- Acute systemic toxicity (ISO 10993-11)
- Material-mediated pyrogenicity (ISO 10993-11, USP <151>)
- Subacute toxicity / implantation (ISO 10993-11 and ISO 10993-6)
### Bench Testing:
- Mechanical Testing:
- Wound Closure Strength per ASTM F2458-05 Standard Test Method for Wound Closure Strength of Tissue Adhesives and Sealants
Page 2 of 3
{6}
K243990
- Tensile Strength per ASTM F2258-05 Standard Test Method for Strength Properties of Tissue Adhesives in Tension
- Shear Strength per ASTM F2255-05 Standard Test Method for Strength Properties of Tissue Adhesives in Lap-Shear by Tension Loading
- Hydrolytic Degradation
- Heat of Polymerization
- Microbial Barrier
- Applicator Functionality
- Design Specifications Testing
## Animal Testing:
A porcine study was conducted to evaluate the healing process and local skin reactions to the subject device. The subject device and predicate device were used to treat full-thickness linear incisions in a porcine model and the healing and biological response of the treatment sites were determined.
## CONCLUSION
Based on the testing performed, including adhesive strength testing, hydrolytic degradation, heat of polymerization, microbial barrier, applicator functionality, and design specification testing, as well as animal testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed BondEase 2 Topical Skin Adhesive are assessed to be substantially equivalent to the predicate devices.
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.