← Product Code [ITI](/productcode/ITI) · K243961

# Electrically Powered Wheelchair (K243961)

_Kunshan Hi-Fortune Health Products Co., Ltd. · ITI · Feb 12, 2025 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243961

## Device Facts

- **Applicant:** Kunshan Hi-Fortune Health Products Co., Ltd.
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Feb 12, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

## Device Story

Electrically powered wheelchair (HP206) provides mobility for disabled or elderly users. Device consists of aluminum alloy frame, wheels, seat, armrests, lithium battery, DC brushless motors, and controller. User operates device via joystick controller; inputs processed by controller to drive rear wheels for forward, backward, and turning movements. Features automatic electromagnetic braking. Designed for indoor/outdoor use; foldable for transport. Benefits include increased mobility for users with limited walking ability. Healthcare providers do not operate device; it is for patient self-use.

## Clinical Evidence

Bench testing only. Performance verified against ISO 7176 series (stability, braking, energy consumption, dimensions, speed, obstacle climbing, fatigue strength, climatic tests, EMC, and power/control systems). Biocompatibility testing conducted per ISO 10993-5, -10, and -23. Flame-retardant testing per ISO 16840-10.

## Technological Characteristics

Aluminum alloy frame; solid PU tires; DC brushless motor (24Vdc, 170W); lithium battery (25.2VDC, 10.4Ah); joystick controller (M7084). Automatic electromagnetic brake system. Foldable design. Compliance with ISO 7176 series for power/control systems, stability, and EMC (IEC 60601-1-2).

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Electrically Powered Wheelchair HP358EA ([K211647](/device/K211647.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square.

February 12, 2025

Kunshan Hi-Fortune Health Products Co., Ltd % Ariel Xiang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China

Re: K243961

Trade/Device Name: Electrically Powered Wheelchair (HP206) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 23, 2024 Received: December 23, 2024

Dear Ariel Xiang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K243961

Device Name

Electrically Powered Wheelchair (HP206)

Indications for Use (Describe)

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Kunshan Hi-Fortune Health Products Co., Ltd. No.625 Juxiang Road zhangpu town , Kunshan City Jiangsu Province, China

### 510(K) Summary K243961

### Document Prepared Date: 2025/2/11

#### A. Applicant

Kunshan Hi-Fortune Health Products Co.. Ltd Address:NO.625 Juxiang Road, Zhangpu Town,Kunshan City, Jiangsu, 215321, CHINA Contact Person: zhanli.Tian Email:zhanli.tian@hi-fortune.net

Submission Correspondent: Primary contact: Ms.Ariel Xiang Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: shouqiu.xiang@sungoglobal.com

Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com

#### B. Device

Trade Name: Electrically powered wheelchair Common Name: Powered wheelchair Models: HP206

Regulatory Information Classification Name: Powered wheelchair Classification: Class II. Product code: ITI Regulation Number: 890.3860 Review Panel: Physical Medicine

#### C. Predicate device

510Knumber: K211647 Device Name:Electrically Powered Wheelchair Modesl:HP358EA Kunshan Hi-Fortune Health Products Co., Ltd.

#### D. Indications for use of the device

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

{5}------------------------------------------------

#### ய் Device Description:

This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded and it can be easily carried or rolled after folding. The seat cushion is detachable. The armrest can be flipped upside down, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device.

The wheelchair uses lithium batteries as its power source. The controls the drive leftright motor to realize the wheelchair forward, backward and turn functions.

The frame of the device is aluminum alloy. The front wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires.

When in use, the operator drives the motor of the rear wheel by operating the controller joystick to achieve the rear wheels movement.

The DC brushless motor and brake system are fixed on the rear wheels.

The max loading of the device is 120KG. Only for one person sit.

#### F. Comparison with predicate Device

Compared to the predicate devices, the subject device has the same intended use, similar product design, similar performance, same safety as the predicate device, the summarized comparison information is listed in the following table:

| changes                    | Proposed Device                        | Predicate Device (K211647)         |
|----------------------------|----------------------------------------|------------------------------------|
| 1.Change Motor type        | DC Brushless Motor,<br>24Vdc 170W;2pcs | DC Brush Motor,<br>24Vdc 250W;2pcs |
| 2. Change controller model | M7084                                  | PG D51492                          |

#### a) There are two changes to the proposed device:

{6}------------------------------------------------

Kunshan Hi-Fortune Health Products Co., Ltd. No.625 Juxiang Road zhangpu town , Kunshan City Jiangsu Province, China

# b) Technical comparison

# Table 1 General Comparison

| Elements of<br>Comparison                                     | Proposed Device                                                                                                                                                                                                                                                                              | Predicate Device (K211647)                                                                                                                                                                                                                                                              | Remark                                                                                                                                                                                     |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                  | Kunshan Hi-Fortune Health<br>Products Co., Ltd                                                                                                                                                                                                                                               | Kunshan Hi-Fortune Health<br>Products Co., Ltd                                                                                                                                                                                                                                          | --                                                                                                                                                                                         |
| Common or Usual name                                          | Electrically powered<br>wheelchair                                                                                                                                                                                                                                                           | Electrically Powered Wheelchair                                                                                                                                                                                                                                                         | --                                                                                                                                                                                         |
| Model(s)                                                      | HP206                                                                                                                                                                                                                                                                                        | HP358EA                                                                                                                                                                                                                                                                                 | --                                                                                                                                                                                         |
| Classification                                                | Class II                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                | S.E.                                                                                                                                                                                       |
| Classification regulation                                     | 21 CFR890.3860                                                                                                                                                                                                                                                                               | 21 CFR890.3860                                                                                                                                                                                                                                                                          | S.E.                                                                                                                                                                                       |
| Product code                                                  | ITI                                                                                                                                                                                                                                                                                          | ITI                                                                                                                                                                                                                                                                                     | S.E.                                                                                                                                                                                       |
| Indications for use                                           | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.                                                                                                                      | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.                                                                                                                 | S.E.                                                                                                                                                                                       |
| Use condition                                                 | indoor and outdoor use                                                                                                                                                                                                                                                                       | indoor and outdoor use                                                                                                                                                                                                                                                                  | S.E.                                                                                                                                                                                       |
| Number of wheels                                              | 6, including two front wheels and two rear wheels, two anti-tip wheels                                                                                                                                                                                                                       | 6, including two front wheels and two rear wheels, two anti-tip wheels                                                                                                                                                                                                                  | S.E.                                                                                                                                                                                       |
|                                                               |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                            |
| Function of wheels                                            | Front wheels: driven<br>wheels suitable for<br>rotation, accelerati<br>on, retrograde<br>Rear wheels: driving wheels<br>to control the speed and<br>direction<br>Anti-tip wheels: preventing<br>the wheelchair from<br>tipping turning over when<br>driving on the slope. Non-<br>adjustable | Front wheels: driven wheels<br>suitable for<br>rotation, acceleration,<br>retrograde<br>Rear wheels: driving wheels<br>to control the speed and<br>direction<br>Anti-tip wheels: preventing the<br>wheelchair from tipping turning<br>over when driving on the slope.<br>Non-adjustable | S.E.                                                                                                                                                                                       |
| Movement control<br>method                                    | By Joystick control                                                                                                                                                                                                                                                                          | By Joystick control                                                                                                                                                                                                                                                                     | S.E.                                                                                                                                                                                       |
| Driving system                                                | Direct drive on the rear<br>wheels                                                                                                                                                                                                                                                           | Direct drive on the rear<br>wheels                                                                                                                                                                                                                                                      | S.E.                                                                                                                                                                                       |
| Brake system                                                  | Automatic electromagnetic<br>brake system                                                                                                                                                                                                                                                    | Automatic electromagnetic<br>brake system                                                                                                                                                                                                                                               | S.E.                                                                                                                                                                                       |
| Armrest                                                       | PU                                                                                                                                                                                                                                                                                           | PU                                                                                                                                                                                                                                                                                      | S.E.                                                                                                                                                                                       |
| Back cushion                                                  | Terylene                                                                                                                                                                                                                                                                                     | Terylene                                                                                                                                                                                                                                                                                | S.E.                                                                                                                                                                                       |
| Seat cushion                                                  | Terylene                                                                                                                                                                                                                                                                                     | Terylene                                                                                                                                                                                                                                                                                | S.E.                                                                                                                                                                                       |
| Main frame material                                           | Aluminum alloy                                                                                                                                                                                                                                                                               | Aluminum alloy                                                                                                                                                                                                                                                                          | S.E.                                                                                                                                                                                       |
| Max loading weight                                            | 120KG                                                                                                                                                                                                                                                                                        | 120KG                                                                                                                                                                                                                                                                                   | S.E.                                                                                                                                                                                       |
| Maximum obstacle<br>climbing                                  | 35mm                                                                                                                                                                                                                                                                                         | 30mm                                                                                                                                                                                                                                                                                    | Similar,<br>Minor difference will not cause different<br>performance.                                                                                                                      |
| Total mass                                                    | 22KG                                                                                                                                                                                                                                                                                         | 22KG                                                                                                                                                                                                                                                                                    | S.E.                                                                                                                                                                                       |
| Overall Dimension<br>(length*width*height)                    | 950mmX660mmX940mm                                                                                                                                                                                                                                                                            | 990mmX620mmX940mm                                                                                                                                                                                                                                                                       | Similar,<br>Minor difference on wheelchair dimension<br>will not cause different performance.<br>All safety and performance have been<br>validated with the maximum rated weight<br>dummy. |
| Folded Dimension<br>(length*width*height)                     | 660mmX420mmX720mm                                                                                                                                                                                                                                                                            | 620mmX550mmX850mm                                                                                                                                                                                                                                                                       | S.E.                                                                                                                                                                                       |
| Front wheel size/type                                         | 8" PU tire                                                                                                                                                                                                                                                                                   | 8" PU tire                                                                                                                                                                                                                                                                              | S.E.                                                                                                                                                                                       |
| Rear wheel size/type                                          | 12" PU tires                                                                                                                                                                                                                                                                                 | 12" Pneumatic tires                                                                                                                                                                                                                                                                     | Similar,Minor difference will not cause<br>different performance.                                                                                                                          |
| Max speed forward                                             | 1.5 m/s                                                                                                                                                                                                                                                                                      | 1.4 m/s                                                                                                                                                                                                                                                                                 | Similar                                                                                                                                                                                    |
| Max Speed backward                                            | 0.9m/s                                                                                                                                                                                                                                                                                       | 0.8m/s                                                                                                                                                                                                                                                                                  | S.E.                                                                                                                                                                                       |
| Maximum safe<br>operational incline<br>Degree                 | 10°                                                                                                                                                                                                                                                                                          | 10°                                                                                                                                                                                                                                                                                     | S.E.                                                                                                                                                                                       |
| Maximum Braking<br>distance                                   | Forwards 1.2m;Reverse 0.6m                                                                                                                                                                                                                                                                   | Forwards 0.8m;Reverse 0.6m                                                                                                                                                                                                                                                              | Similar                                                                                                                                                                                    |
| Maximum distance of<br>travel on the fully<br>charged battery | 10.6km                                                                                                                                                                                                                                                                                       | 10km                                                                                                                                                                                                                                                                                    | Minor difference will not cause different<br>performance.                                                                                                                                  |
| Turning Radius                                                | 750mm                                                                                                                                                                                                                                                                                        | 500mm                                                                                                                                                                                                                                                                                   | Minor difference will not cause different<br>performance.                                                                                                                                  |
| Battery                                                       | li-ion battery pack;<br>rechargeable,<br>25.2VDC,10.4Ah                                                                                                                                                                                                                                      | li-ion battery;<br>rechargeable, 25.2VDC, 10.4Ah                                                                                                                                                                                                                                        | S.E                                                                                                                                                                                        |
| Battery charger                                               | Off-board charger,<br>Input :100-240VAC,                                                                                                                                                                                                                                                     | Off-board charger,<br>Input : 100-240VAC, 50-<br>60Hz, 1.2-0.5A                                                                                                                                                                                                                         | S.E                                                                                                                                                                                        |
|                                                               | 50-60Hz, 1.2-0.5A<br>DC output: 24Vdc 2A                                                                                                                                                                                                                                                     | DC output: 24Vdc2A                                                                                                                                                                                                                                                                      |                                                                                                                                                                                            |
| Motor                                                         | DC Brushless Motor,<br>24Vdc 170W;2pcs                                                                                                                                                                                                                                                       | DC Brush Motor,<br>24Vdc 250W;2pcs                                                                                                                                                                                                                                                      | Analysis, The design principles of the<br>controller and Driving system are the                                                                                                            |
| controller                                                    | M7084,<br>24V 35A                                                                                                                                                                                                                                                                            | PG D51492,<br>24V 35A                                                                                                                                                                                                                                                                   | same, and both meet the requirements of<br>the ISO 7176-14:2008. Software validation<br>is carried out on both control systems.                                                            |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

# Table 2 safety comparison

| Item               | Proposed Device                                                                                                           | Predicate Device                                                                                          | Results |
|--------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------|
| Biocompatibility   | All user directly contacting materials are<br>compliance with ISO10993-5、<br>ISO10993-10 and ISO10993-23<br>requirements. | All user directly contacting materials are<br>compliance with ISO10993-5 and<br>ISO10993-10 requirements. | SE      |
| EMC                | IEC 60601-1-2&ISO7176-21                                                                                                  | IEC 60601-1-2&ISO7176-21                                                                                  | SE      |
| Performance        | ISO7176 series                                                                                                            | ISO7176 series                                                                                            | SE      |
| Label and labeling | Conforms to FDA Regulatory                                                                                                | Conforms to FDA Regulatory                                                                                | SE      |

#### G. Substantial Equivalence Discussion

The proposed device and predicate device are complying to the same ISO standards, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-22,ISO 7176-25, IEC 60601-1-2 and FDA guidance Submission for Power Wheelchair.

The indications for use, design, structure, composition and energy source of proposed device and predicate device are identical.The dimensions and performance parameters are similar.The main differences are motor and controller.

The design principles of the controller and Driving system are the requirements of the ISO 7176-

{10}------------------------------------------------

14. Software validation is carried out on both control systems. Brake system and speed control are same way as well, and both meet the ISO 7176-3. Turning radius, Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the ISO 7176-2 and ISO 7176-10.

The materials in contact with human tissue are identical for both proposed device and predicate device. The biocompatibility of the proposed device and predicate device meet the ISO 10993-5:2009 & ISO 10993-10:2021 & ISO 10993-23: 2021.

The flame-retardant test of the seat cyshion/ backrest of both subject device is carried out acording to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level.

In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

#### H. Product Performance

- c) Product Material Safety
To ensure the safety of the wheelchair material, the bio-compatibility testing was done to the testing was conducted according to the following standards:

ISO 10993-5 Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity

ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization

ISO 10993-23 Biological evaluation of medical devices - Part 23: Tests for irritation

From the test result, we can find the material are safety and can meet the requirements.

- d) Performance of the products
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.

Risk Analysis developed in accordance with ISO 14971: 2019.

Software validation

ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability

ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs

ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes

{11}------------------------------------------------

ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range

ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space

ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration of electric wheelchairs

ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions

ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength

ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs

ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs

ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies

ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.

ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters --Requirements and test methods

ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information and labeling.

ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method

ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters

ISO 7176-22: 2014 Wheelchairs-Part 22: Set-up procedures

ISO 7176-25: 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs.

Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020 8.11.

#### 1. Conclusion

The proposed device and predicate have the same technological characteristics, features, specifications, materials, mode of operation, and intended use of the device. To ensure all the key characteristic of the products can meet the requirements, the testing was done and all the results indicate the positive conclusion.

Based on the analysis above, we confirm these two products are essentially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K243961](https://fda-staging.innolitics.com/device/K243961)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).