TEMP MASTER, PMMA-based dental resin developed by Taiwan Dental Materials CO., LTD. is a self-cured resin indicated for the fabrication of short-term restorations including temporary inlays, crowns, and bridges, which can protect or restore damaged teeth while permanent restorations are being prepared. It is a traditional auto-cured PMMA-based resin composed of powder and liguid. Polymerization occurs when the powder and liguid are mixed at room temperature.
Device Story
TEMP MASTER is a self-curing (auto-polymerizing) PMMA-based dental resin consisting of powder and liquid components. Used by dental professionals in a clinical setting to fabricate temporary inlays, crowns, and bridges. Polymerization occurs at room temperature upon mixing the two components. The device protects or restores damaged teeth while permanent restorations are being prepared. It functions as a temporary restorative material; clinical benefit is the maintenance of tooth function and protection during the interim period before final restoration placement.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 10477 (flexural strength, water sorption, water solubility) and biocompatibility per ISO 10993. Results demonstrate compliance with standards and performance comparable to the predicate device.
Technological Characteristics
Self-curing PMMA-based acrylic resin (Type 1 per ISO 10477). Two-component system (powder/liquid). Polymerization via room-temperature chemical reaction. Flexural strength >50 MPa. Non-sterile.
Indications for Use
Indicated for fabrication of short-term dental restorations (inlays, crowns, bridges) to protect or restore damaged teeth during preparation of permanent restorations.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 21, 2025
Taiwan Dental Materials CO., LTD. Pey-Fen Tang Senior R&D Engineer 11F., No. 136, Sec. 3, Zhongxiao E. Rd., Da'an Dist. Taipei, 106671 TAIWAN
Re: K243951
Trade/Device Name: TEMP MASTER, PMMA-based dental resin Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: December 23, 2024 Received: December 23, 2024
Dear Pey-Fen Tang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K243951
Device Name
TEMP MASTER, PMMA-based dental resin
Indications for Use (Describe)
TEMP MASTER, PMMA-based dental resin developed by Taiwan Dental Materials CO., LTD. is a self-cured resin indicated for the fabrication of short-term restorations including temporary inlays, crowns, and bridges, which can protect or restore damaged teeth while permanent restorations are being prepared. It is a traditional auto-cured PMMA-based resin composed of powder and liguid. Polymerization occurs when the powder and liguid are mixed at room temperature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# [TEMP MASTER, PMMA-based dental resin]
[2024/08/15]
{5}------------------------------------------------
### K243951 510(k) SUMMARY
### [TEMP MASTER, PMMA-based dental resin]
#### 1. Contact Person
Company name: Taiwan Dental Materials CO., LTD.
Name: Pey Fen, Tang
Title: Senior R&D Engineer
E-mail: tangzs012304@gmail.com
Office address: 7F., No. 161, Sec. 1, Jianguo S. Rd., Da'an Dist., Taipei City 106084,
Taiwan (R.O.C.)
Tel: (O) +886 2 8772 4023
#### 2. Device Name and Classification
| Product Name: | TEMP MASTER, PMMA-based dental resin |
|-----------------------|--------------------------------------|
| Classification Name: | Temporary crown and bridge resin. |
| Common or Usual Name: | Dental Material |
| Classification Panel: | Dental Device |
| Regulation Number: | 872.3770 |
| Device Class: | Class 2 |
| Product Code: | EBG |
#### 3. Predicate Device(s)
| Product Name: | K073106<br>Unifast III |
|-----------------------|-----------------------------------------------------------------------------------|
| Classification Name: | Denture Relining Repairing or Rebasing Resin<br>Temporary crown and bridge resin. |
| Common or Usual Name: | Dental Material |
| Classification Panel: | Dental Device |
| Regulation Number: | 872.3760, 872.3770 |
| Device Class: | Class 2 |
| Product Code: | EBI, EBG |
{6}------------------------------------------------
#### 4. Device Description
TEMP MASTER, PMMA-based dental resin, is a self-curing (cold-cure) material designed to fabricate temporary crowns and bridges. It consists of two components: powder and liquid. Polymerization occurs when the powder and liquid are mixed at room temperature. Classified as a Type 1 (autopolymerizable) acrylic resin according to ISO 10477 standards, TEMP MASTER, PMMA-based dental resin provides reliable performance for temporary restorations.
#### 5. Indications for Use
TEMP MASTER, PMMA-based dental resin developed by Taiwan Dental Materials CO., LTD., is a self-cured resin indicated for the fabrication of shortterm restorations including temporary inlays, crowns, and bridges, which can protect or restore damaged teeth while permanent restorations are being prepared. It is a traditional auto-cured PMMA-based resin composed of powder and liquid. Polymerization occurs when the powder and liquid are mixed at room temperature.
#### 6. Substantial Equivalence
TEMP MASTER, PMMA-based dental resin, developed by Taiwan Dental Materials, is substantially equivalent to the predicate device, Unifast III (K073106).
The subject device, TEMP MASTER, PMMA-based dental resin and the predicate device, Unifast III (K073106), are both indicated for the fabrication of temporary inlays, crowns, and bridges. However, the predicate device also includes an additional indication for the repair of fractured dentures, which is not included in the indications for the subject device.
The substantially equivalent comparison focuses on the indications and properties under the product code EBG (Crown and Bridge, Temporary, Resin). The results of this comparison demonstrate that TEMP MASTER, PMMA-based dental resin is substantially equivalent to the predicate device, Unifast III (K073106). The expanded indication of the predicate device is not relevant to the substantial equivalence determination in this document.
{7}------------------------------------------------
| Device Description | | | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Specifications | Subject Device<br>TEMP MASTER,<br>PMMA-based dental<br>resin | Predicate Device<br>Unifast III | Differences |
| K number | NA | K073106 | NA |
| Classification<br>Product code | EBG | EBI, EBG | |
| Regulation no. | 872.3770 | 872.3770, 872.3760 | |
| Indications for use | TEMP MASTER,<br>PMMA-based dental<br>resin is a self-cured<br>resin indicated for the<br>fabrication of short-term<br>restorations including<br>temporary inlays,<br>crowns, and bridges,<br>which can protect or<br>restore damaged teeth<br>while permanent<br>restorations are being<br>prepared. It is a<br>traditional auto-cured<br>PMMA-based resin<br>composed of powder<br>and liquid.<br>Polymerization occurs<br>when the powder and<br>liquid are mixed at room<br>temperature. | Unifast III is a self-<br>curing acrylic resin used<br>for a variety of dental<br>applications, including<br>construction of<br>temporary inlays, crowns<br>and bridges and repair of<br>fractured dentures in<br>either pressure or paint<br>on techniques. | The predicate device<br>also includes an<br>additional indication<br>for the repair of<br>fractured dentures,<br>which is not included<br>in the indications for<br>the subject device.<br>The comparison only<br>focuses on the<br>properties under the<br>product code EBG<br>(Crown and Bridge,<br>Temporary, Resin). |
| Acrylic Resin | self-curing Resin | self-curing Resin | Identical |
| Chemical<br>Characterization | PMMA-based resin | PMMA-based resin | Identical |
Table 1 The substantially equivalent comparison
| Physical Properties | | | | |
|---------------------|-----------------------------------------------------------|--------------------------------------------|--|--|
| ISO 10477 | Subject Device<br>TEMP MASTER,<br>PMMA-based dental resin | Predicate Device<br>Unifast III<br>K073106 | | |
| Flexural Strength | >50 MPa | >50 MPa | | |
| Water Absorption | Complies with ISO 10477 | Complies with ISO 10477 | | |
| Water Solubility | Complies with ISO 10477 | Complies with ISO 10477 | | |
{8}------------------------------------------------
#### 7. Performance Data
#### Biocompatibility Testing
The biocompatibility evaluation for TEMP MASTER, PMMA-based dental resin was conducted in accordance with the FDA Blue Book Memorandum #G95 and international Standard ISO 10993.
#### Bench Testing
ISO 10477. Dentistry-Polymer-based crown and bridge materials. 2018. -The nonclinical tests conducted in accordance with ISO 10477 assessed the flexural strength, water sorption, and water solubility of the temporary crown and bridge dental resin. The performance characteristics of TEMP MASTER, PMMA-based dental resin, were compared to those of the predicate device, and both materials demonstrated results within the acceptable ranges defined by the standard.
#### Sterility and Shelf-life
The device is provided non- sterile.
From the shelf-life testing, TEMP MASTER, PMMA-based dental resin has a shelf life of 2.5 years. The shelf testing has been conducted with the bench test fromthe ISO Standard 10477 and ISO 868.
#### 8. Similarity and differences
TEMP MASTER, PMMA-based dental resin is similar to the predicate device in the method of processing. The main difference is the slight differences in the percentage of chemical composition. The chemical composition might influence the biocompatibility of TEMP MASTER, as well as its performance (mechanical properties). Thus, both biocompatibility and performance has been demonstrated for TEMP MASTER. The same tests according to the same standards ISO 10993 and ISO 10477 were used to show biocompatibility and performance of the predicate.
#### 9. Conclusion
After analyzing bench test, device description and intended use/ indication for use, it can be concluded that TEMP MASTER, PMMA-based dental resin is as safe and effective as the predicate device, Unifast III (K073106).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.