Raptor Aspiration Catheter; Balt Aspiration Tubing Set
K243948 · Balt USA, LLC · NRY · Jun 3, 2025 · Cardiovascular
Device Facts
Record ID
K243948
Device Name
Raptor Aspiration Catheter; Balt Aspiration Tubing Set
Applicant
Balt USA, LLC
Product Code
NRY · Cardiovascular
Decision Date
Jun 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Raptor Aspiration Catheter: The Raptor Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. Balt Aspiration Tubing Set: The Balt Aspiration Tubing Set is intended to connect the Raptor Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow.
Device Story
Raptor Aspiration Catheter is a single-lumen, variable stiffness composite catheter used for thrombus removal from neurovasculature. Device inputs include vacuum pressure from a compatible aspiration pump, delivered via the Balt Aspiration Tubing Set. The catheter features a distal markerband, hydrophilic coating, and luer fitting for navigation and fluid infusion. Used in clinical settings by physicians, the device is navigated to the occlusion site; aspiration is applied to remove thrombus. Output is the physical removal of the clot, facilitating revascularization. Benefits include restoration of blood flow in acute ischemic stroke patients. The tubing set allows user control of fluid flow and vacuum connection to the pump canister.
Clinical Evidence
No clinical data. Substantial equivalence is based on non-clinical bench testing, including dimensional verification, tensile strength, kink resistance, leakage, burst pressure, torque strength, particulate matter, radiopacity, and clot retrieval in clinically relevant anatomical models.
Technological Characteristics
Single-lumen composite catheter; variable stiffness; hydrophilic coating. Materials: medical-grade plastics and metals. Dimensions: 0.038"-0.054" ID; 0.051"-0.074" OD. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 10555-1 (mechanical), ISO 10993-1 (biocompatibility), ISO 80369-7 (hub), ASTM F1980 (aging).
Indications for Use
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, and vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or who failed thrombolytic drug therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 3, 2025
Balt USA, LLC
Catherine Chiou
Senior Regulatory Affairs Specialist
29 Parker
Irvine, California 92618
Re: K243948
Trade/Device Name: Raptor Aspiration Catheter; Balt Aspiration Tubing Set
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY
Dated: May 2, 2025
Received: May 2, 2025
Dear Catherine Chiou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243948 - Catherine Chiou
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243948 - Catherine Chiou
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243948 | |
| Device Name Raptor Aspiration Catheter; Balt Aspiration Tubing Set | |
| Indications for Use (Describe) Raptor Aspiration Catheter: The Raptor Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. Balt Aspiration Tubing Set: The Balt Aspiration Tubing Set is intended to connect the Raptor Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
balt
# 510(k) Summary: K243948
| Applicant: | Balt USA, LLC
29 Parker
Irvine, CA 92618
Registration No.: 3014162263 |
| --- | --- |
| Contact Person: | Catherine Chiou
Senior Specialist, Regulatory Affairs
Email: Catherine.chiou@baltgroup.com |
| Date Summary Prepared: | May 30, 2025 |
| --- | --- |
| Trade Name: | Raptor Aspiration Catheter; Balt Aspiration Tubing Set |
| Common Name: | Catheter, Thrombus Retriever |
| Review Panel: | Neurology |
| Product Code: | NRY |
| Regulation Number: | 21 CFR 870.1250 |
| Regulation Name: | Percutaneous Catheter |
| Device Classification: | Class II |
| Predicate Device: | ZOOM (71, 55, 45, 35) Reperfusion Catheters; ZOOM Aspiration Tubing 510(k)#: K210996 |
| Reference Device: | Next Generation Aspiration Catheter; Balt Aspiration Tubing Set 510(k)#: K234083 |
# Device Description:
The Raptor Aspiration Catheter is a single-lumen, variable stiffness composite catheter offered in various sizes that facilitates removal of thrombus from the neurovasculature when connected to a compatible aspiration pump and the Balt Aspiration Tubing Set. The distal tip of the catheter shaft includes a markerband and the proximal end of the catheter has a luer fitting to allow for the attachment of ancillary devices for navigation, infusion of fluids, and aspiration through the catheter. The Raptor Aspiration Catheter distal shaft has an external hydrophilic coating which provides a lubricious surface during use. The Raptor Aspiration Catheters are compatible with $0.014'' - 0.018''$ guidewires.
A peelable split Introducer Sheath is provided in the package to provide support and facilitate the introduction of the distal tip of the Raptor Aspiration Catheter into an appropriate vascular sheath. The catheter and Introducer Sheath are provided sterile, non-pyrogenic, and intended for single use only.
The Balt Aspiration Tubing Set consists of one $100^{\prime \prime}$ HPF proximal tubing with a suction connector on one end and male rotator on the other, one 1-way Large Bore Stopcock, and one $10^{\prime \prime}$ HPF distal tubing with a female luer connector on one end and a male rotator on the other.
The Balt Aspiration Tubing Set facilitates the supply of vacuum from a compatible aspiration pump to the distal tip of the Raptor Aspiration Catheter. Balt Aspiration Tubing Set is provided sterile, non-pyrogenic and it is intended for single use only
Page 1 of 6
{5}
Page 2 of 6
balt
# Indications for Use:
## Raptor Aspiration Catheter:
The Raptor Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
## Balt Aspiration Tubing Set:
The Balt Aspiration Tubing Set is intended to connect the Raptor Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow.
## Comparison of Technological Characteristics:
| | Predicate Device | Reference Device | Subject Device |
| --- | --- | --- | --- |
| | ZOOM (71, 55, 45, 35)
Reperfusion Catheters;
ZOOM Aspiration Tubing
(K210996) | Next Generation Aspiration Catheter; Balt Aspiration Tubing Set
(K234083) | Raptor Aspiration Catheter; Balt Aspiration Tubing Set
(K243948) |
| Indications for Use | The ZOOM Reperfusion Catheters, with the ZOOM Aspiration Tubing and ZOOM Aspiration Pump (or equivalent vacuum pump), are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The ZOOM Aspiration Tubing is intended to connect the ZOOM Reperfusion Catheter to the ZOOM Canister of the ZOOM Aspiration Pump and to allow the user to control the fluid flow. | Next Generation Aspiration Catheter: The Next Generation Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
Patients who are ineligible for intravenous tissue plasminogen activator (IV-tPA) or who fail IV-tPA therapy are candidates for treatment.
Balt Aspiration Tubing Set: The Balt Aspiration Tubing Set is intended to connect the Next Generation Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. | Raptor Aspiration Catheter: The Raptor Aspiration Catheter with a compatible aspiration pump and Balt Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Balt Aspiration Tubing Set: The Balt Aspiration Tubing Set is intended to connect the Raptor Aspiration Catheter to the canister of a compatible aspiration pump and to allow the user to control the fluid flow. |
{6}
balt
| | Predicate Device | Reference Device | Subject Device |
| --- | --- | --- | --- |
| | ZOOM (71, 55, 45, 35)
Reperfusion Catheters;
ZOOM Aspiration Tubing
(K210996) | Next Generation Aspiration Catheter; Balt Aspiration Tubing Set
(K234083) | Raptor Aspiration Catheter; Balt Aspiration Tubing Set
(K243948) |
| Device Classification/
Product Code | Class II / NRY
(Percutaneous Catheter) | Same as K210996 | Same as K210996 |
| Intended Use | Revascularization of patients with acute ischemic stroke. | Same as K210996 | Same as K210996 |
| Dimensional Specifications | | | |
| Inner Diameter | 0.035” – 0.071” | 0.071” and 0.074” | 0.038” and 0.054” |
| Outer Diameter | Proximal: 0.061” – 0.083”
Distal: 0.051” – 0.083” | Proximal: 0.083” and 0.086”
Distal: 0.082” and 0.085” | Proximal: 0.062” and 0.074”
Distal: 0.051” and 0.067” |
| Effective Length | 137 cm – 160 cm | 128 cm, 132 cm and 120 cm, 132 cm | 145 – 160 cm and 125 – 150 cm |
| Device Attributes | | | |
| Materials | Commonly used medical grade plastics and metals with hydrophilic coating. | Same as K210996 | Same as K210996 |
| Coating | Hydrophilic | Same as K210996 | Same as K210996 |
| Packaging Configuration | The catheter is placed in a protective polyethylene tube and then mounted, along with the accessories, onto a polyethylene packaging card.
The packaging card is inserted into a Tyvek® pouch which is then sealed.
The sealed pouch and IFU are placed in a carton box. | Same as K210996 | Same as K210996 |
| Shelf Life | 12 months | Same as K210996 | Same as K210996 |
| Sterilization | | | |
| Condition Supplied | Sterile and Single Use | Same as K210996 | Same as K210996 |
| Method | Ethylene Oxide (EO), SAL 10^{-6} | Same as K210996 | Same as K210996 |
| Accessories | | | |
| Accessories | RHV | Introducer Sheath | Same as K234083 |
| Aspiration Tubing | ZOOM Aspiration Tubing | Balt Aspiration Tubing Set | Same as K234083 |
| Dimensional Specifications | Length: 104”
Inner Diameter: 0.110” Minimum | Length: 114”
Inner Diameter: 0.110” ± 0.003” | Same as K234083 |
| Condition Supplied | Sterile and Single Use | Same as K210996 | Same as K210996 |
| Sterilization Method | Ethylene Oxide (EO), SAL 10^{-6} | Same as K210996 | Same as K210996 |
| Flow Control Mechanism | Flow Control Clamp | Flow Control Valve | Same as K234083 |
| Aspiration Pump | • Capable of achieving pressure between -20 inHg to -29.9 inHg
• Airflow rating of 0 – 23 LPM
• IEC 60601-1 Compliant | • Capable of achieving pressure between -20 inHg and -25.6 inHg
• Airflow rating ≥ 16 LPM | Same as K234083 |
Page 3 of 6
{7}
balt
Performance Testing Summary:
Biocompatibility:
Biological safety evaluation was conducted for the Raptor Aspiration Catheter in accordance with the FDA biocompatibility guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", issued September 2023.
Additionally, the risk management process determined that the biocompatibility testing conducted on the reference device Next Generation Aspiration Catheter (K234083) and the Next Generation 6F+ Access Sheath (K242376), per ISO 10993-1 and applicable regulatory requirements, adequately assesses the biocompatibility profile of the subject Raptor Aspiration Catheters for endpoints recommended by the FDA biocompatibility guidance for an externally communicating device with limited (≤24 hours) duration of contact with circulating blood. Additionally, a comparative assessment of the Raptor Aspiration Catheter, Next Generation Aspiration Catheter, and Next Generation 6F+ Access Sheath did not identify any significant differences in the surface characteristics or geometry of these devices:
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Hemocompatibility – Comparative Surface Assessment | The test article and the comparison article(s) were visually inspected at a minimum 40x magnification. | Requirements have been met by the test article. |
The Balt Aspiration Tubing Set and Introducer Sheath are identical in materials and manufacturing to those cleared as part of the Next Generation Aspiration Catheter and Balt Aspiration Tubing Set (K234083) and the Ballast 088 Long Sheath (K182918), respectively.
Performance Data – Bench:
The following performance bench testing was conducted for the Raptor Aspiration Catheter:
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Dimensional Verification | The catheter outer diameter, inner diameter, usable length, tip length, and coating length were measured. | Pass |
| Surface Contamination | Visual inspection completed for surface defects. | Pass |
| Tensile Strength | The peak tensile force was evaluated per ISO 10555-1 after preconditioning in a simulated use model. | Pass |
| Kink Resistance | Kink resistance was evaluated after preconditioning in a simulated use model. | Pass |
| Liquid Leakage | The device was exposed to a liquid pressure for 30 seconds. The device was inspected for leakage per ISO 10555-1. | Pass |
| Air Leakage | The device was tested for air leakage into the hub per ISO 10555-1. | Pass |
| Dynamic Burst | Testing verified the device can withstand internal liquid pressure under dynamic flow conditions with the distal end open. | Pass |
| Torque Strength | The device was evaluated for torque strength by measuring the number of catheter rotations until failure after preconditioning in a simulated use model and compared to the predicate. | Pass |
| Particulate Matter | The catheter underwent simulated use testing and particulate testing was conducted including a reference device for comparison. | Pass, comparable to predicate device |
| Hub Testing | The device shall meet the established acceptance criteria per ISO 80369-7. | Pass |
Page 4 of 6
{8}
balt
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Corrosion | Corrosion tested per ISO 10555-1. | No evidence of corrosion and met requirements per ISO 10555-1. |
| Coating Integrity | The coating integrity was inspected before and after preconditioning through a simulated use model. | No evidence of surface damage or coating defects. |
| Contrast and Saline Exposure | After the device was used to deliver saline and contrast media, the device was inspected for damage, and dimensional attributes were measured. | No visual evidence of damage or dimensional changes. |
| Static Burst | The distal tip of the catheter was blocked, and fluid was injected into the lumen at increasing pressure until the catheter burst per ISO 10555-1. Static burst pressures were compared to pressures of manual syringe delivery of worst-case fluids through the catheter. | The static burst values met the acceptance criterion. |
| Tip Buckling | The maximum force to cause catheter tip buckling while constrained at varying distances was measured. | The tip stiffness was similar to the predicate and reference device. |
| Radiopacity (Visibility) | The device was tested to demonstrate acceptable radiopacity. | Marker radiopacity is comparable to the predicate. |
| Design Validation /Usability | The subject and predicate devices were prepared in accordance with their respective instructions for use and tested for device usability in a clinically relevant anatomical model. | Device preparation, introduction, and trackability were similar to the predicate. |
| Clot Retrieval | The subject and predicate devices were prepared in accordance with their respective instructions for use and tested for clot retrieval in a clinically relevant anatomical model. | Clot retrieval was comparable to the predicate. |
| Lumen Collapse | Tested to verify that the catheter lumen does not collapse at maximum vacuum pressure for a duration of 6 minutes. | Pass |
| Vacuum Pressure | Tested to verify the vacuum pressure at the aspiration pump and distal tip of the catheter. | Pass |
The following performance bench testing was conducted for the Balt Aspiration Tubing Set:
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Dimensional Verification | The total length, inner diameter, and outer diameter were measured. | Pass |
| Surface Contamination | Surface contamination testing was performed in alignment with ISO 10555-1. | Pass |
| Lumen Collapse | Tested to verify that the aspiration tubing lumen does not collapse at maximum vacuum pressure for a duration of 6 minutes. | Pass |
Page 5 of 6
{9}
Page 6 of 6
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Tensile Strength | Testing was adopted from ISO 10555-1. | Pass |
| Tubing to Pump | The tubing to pump pressure was measured. | Pass |
| On/Off Functionality | The On/Off functionality of the tubing set was measured. | Pass |
| Tubing Hub Compatibility | The tubing hub compatibility was evaluated. | Pass |
## Sterilization and Shelf Life:
The Raptor Aspiration Catheter and Balt Aspiration Tubing Set are sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of $1 \times 10^{-6}$ in accordance with ISO 11135. Accelerated aging testing for the Raptor Aspiration Catheter and Balt Aspiration Tubing Set based on ASTM F1980 have established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and device functionality were performed and met the acceptance criteria.
## Performance Data – Animal:
A determination of substantial equivalence is based upon successful completion of non-clinical bench testing.
## Performance Data – Clinical:
A determination of substantial equivalence is based upon successful completion of non-clinical bench testing.
## Conclusion:
Based on the same intended use, similar technological characteristics, and the results of the non-clinical testing, the Raptor Aspiration Catheter and Balt Aspiration Tubing Set is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness. Testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.