← Product Code [EIH](/productcode/EIH) · K243940

# VITA YZ Multi Translucent (K243940)

_Vita Zahnfabrik H.Rauter GmbH & Co. · EIH · Feb 21, 2025 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243940

## Device Facts

- **Applicant:** Vita Zahnfabrik H.Rauter GmbH & Co.
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Feb 21, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

VITA YZ MULTI TRANSLUCENT is indicated for: • fully anatomical anterior and posterior crowns • fully anatomical 4 unit anterior and posterior bridges • fully and partially veneered single tooth and up to 4-unit bridge substructures in the anterior and posterior tooth regions • inlays • Onlays • veneers

## Device Story

VITA YZ MULTI TRANSLUCENT consists of yttria-stabilized zirconia blanks used by dental professionals to fabricate dental restorations; including crowns, bridges (up to 4 units), inlays, onlays, and veneers. The device is processed via CAD/CAM milling to create anatomical shapes. The material features optimized translucency through reduced aluminum oxide content (0-1%) compared to traditional zirconia. These restorations are placed in the oral cavity to restore tooth function and aesthetics. The device provides a durable, biocompatible, and chemically stable alternative to other dental ceramics. Benefits include high flexural strength and aesthetic shade reproduction.

## Clinical Evidence

Bench testing only. Testing performed per EN ISO 6872:2019 and DIN EN ISO 10993-1/7405. Key metrics: Flexural strength 922 ± 95 MPa (>800 MPa required); CTE 10.4×10⁻⁶ K⁻¹; Chemical solubility 0 µg/cm² (<100 µg/cm² required); Radioactivity ≤1.0 Bq/g U-238. Biocompatibility confirmed with no adverse effects.

## Technological Characteristics

Yttria-stabilized zirconia-based material; 0-1% aluminum oxide content; flexural strength >800 MPa; CTE 10.4×10⁻⁶ K⁻¹; biocompatible per DIN EN ISO 10993-1/7405; supplied as blanks for CAD/CAM milling; non-software device.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- VITA YZ ST ([K180703](/device/K180703.md))
- VITA Enamic ([K122269](/device/K122269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 21, 2025

Vita Zahnfabrik H.Rauter GmbH & Co. Lindsay Tilton Regulatory Affairs Consultant Spitelgasse 3 Bad Sackingen, : D-79713 GERMANY

Re: K243940

Trade/Device Name: VITA YZ Multi Translucent Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: June 30, 2024 Received: December 20, 2024

Dear Lindsay Tilton:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Submission Number (if known)

K243940

Device Name

#### VITA YZ MULTI TRANSLUCENT

Indications for Use (Describe)

VITA YZ MULTI TRANSLUCENT is indicated for:

- · fully anatomical anterior and posterior crowns
- · fully anatomical 4 unit anterior and posterior bridges
- · fully and partially veneered single tooth and up to 4-unit bridge substructures in the anterior and posterior tooth regions
- · inlays
- · Onlays
- · veneers

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #:                          | K243940                                                                                                                                                                                                                                                                                                                           |              |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
|                                    | 510(k) Summary                                                                                                                                                                                                                                                                                                                    |              |
|                                    | Prepared on: 2025-02-19                                                                                                                                                                                                                                                                                                           |              |
|                                    | 21 CFR 807.92(a)(1)                                                                                                                                                                                                                                                                                                               |              |
| Contact Details                    |                                                                                                                                                                                                                                                                                                                                   |              |
| Applicant Name                     | Vita Zahnfabrik H.Rauter GmbH & Co.                                                                                                                                                                                                                                                                                               |              |
| Applicant Address                  | Spitelgasse 3 Bad Sackingen : D-79713 Germany                                                                                                                                                                                                                                                                                     |              |
| Applicant Contact Telephone        | 9256999091                                                                                                                                                                                                                                                                                                                        |              |
| Applicant Contact                  | Mrs. Lindsay Tilton                                                                                                                                                                                                                                                                                                               |              |
| Applicant Contact Email            | Itilton@vitanorthamerica.com                                                                                                                                                                                                                                                                                                      |              |
| Device Name                        |                                                                                                                                                                                                                                                                                                                                   |              |
|                                    | 21 CFR 807.92(a)(2)                                                                                                                                                                                                                                                                                                               |              |
| Device Trade Name                  | VITA YZ MULTI TRANSLUCENT                                                                                                                                                                                                                                                                                                         |              |
| Common Name                        | Porcelain powder for clinical use                                                                                                                                                                                                                                                                                                 |              |
| Classification Name                | Powder, Porcelain                                                                                                                                                                                                                                                                                                                 |              |
| Regulation Number                  | 872.6660                                                                                                                                                                                                                                                                                                                          |              |
| Product Code(s)                    | EIH                                                                                                                                                                                                                                                                                                                               |              |
|                                    | 21 CFR 807.92(a)(3)                                                                                                                                                                                                                                                                                                               |              |
| Legally Marketed Predicate Devices |                                                                                                                                                                                                                                                                                                                                   |              |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first)                                                                                                                                                                                                                                                                          | Product Code |
| K180703                            | VITA YZ ST                                                                                                                                                                                                                                                                                                                        | EIH          |
| K122269                            | VITA Enamic                                                                                                                                                                                                                                                                                                                       | EIH          |
|                                    | 21 CFR 807.92(a)(4)                                                                                                                                                                                                                                                                                                               |              |
| Device Description Summary         | VITA YZ MULTI TRANSLUCENT are zirconia blanks for reliable shade reproduction. They can be used for the production of fully and partially veneered reconstructions and monolithic bridge restorations in the anterior and posterior tooth regions. VITA YZ MULTI TRANSLUCENT is part of the VITA YZ SOLUTIONS product group.      |              |
|                                    | 21 CFR 807.92(a)(5)                                                                                                                                                                                                                                                                                                               |              |
| Intended Use/Indications for Use   | VITA YZ MULTI TRANSLUCENT is indicated for:<br>• fully anatomical anterior and posterior crowns<br>• fully anatomical 4 unit anterior and posterior bridges<br>• fully and partially veneered single tooth and up to 4-unit bridge substructures in the anterior and posterior tooth regions<br>• inlays<br>• Onlays<br>• veneers |              |
|                                    | 21 CFR 807.92(a)(5)                                                                                                                                                                                                                                                                                                               |              |
| Indications for Use Comparison     | 1. Differences in Indications for Use<br>Device Indications for Use<br>Subject Device (VITA YZ MULTI TRANSLUCENT) - Fully anatomical anterior and posterior crowns                                                                                                                                                                |              |

{5}------------------------------------------------

- Fully anatomical 4-unit bridges (anterior and posterior)

- Fully & partially veneered single tooth and up to 4-unit bridge substructures

- Inlays, onlays, veneers

Primary Predicate (VITA YZ ST) - Fully anatomical anterior and posterior crowns

- Multi-unit fully anatomical anterior and posterior bridges (not limited to 4 units)

- Direct screw-retained restorations (crowns & bridges)

- Single-tooth and multi-unit bridge frameworks

Secondary Predicate (VITA Enamic) - Inlays, onlays, veneers, crowns

Key Differences in Indications:

· Subject device vs. primary predicate:

o Subject device is limited to 4-unit bridges, while the primary predicate allows multi-unit bridges.

o Primary predicate supports screw-retained restorations, which the subject device does not.

o Subject device includes inlays, onlays, and veneers, which the primary predicate does not.

· Subject device vs. secondary predicate:

o Subject device covers bridges and veneered structures, whereas Enamic does not.

o Enamic only covers inlays, onlays, veneers, and crowns, making it more similar to lithium disilicate-based materials rather than zirconia frameworks.

## Technological Comparison

## 21 CFR 807.92(a)(6)

Composition Differences:

- 1. Subject Device (VITA YZ Multi Translucent) vs. Primary Predicate (VITA YZ ST)
o Both are yttria-stabilized zirconia-based materials.

o The subject device has slightly lower aluminum oxide (AlgO ) content (0-1%) than the primary predicate (1-3%), improving translucency.

- o The primary predicate has higher Al2O3, making it stronger but less translucent.
2. Subject Device vs. Secondary Predicate (VITA Enamic)

- o The subject device is zirconia-based, while Enamic is silica-based (hybrid ceramic).
- o Enamic contains polymer-infiltrated silica, making it more flexible and shock-absorbent.

o Zirconia (subject & primary predicate) is stronger but brittle, while Enamic is more aesthetic and wear-friendly but lacks bridge strength.

Why Composition Matters:

1. Translucency & Aesthetic Use

o The subject device has improved translucency, making it ideal for anterior crowns, veneers, and esthetic restorations.

o The primary predicate is stronger but less translucent, better for large bridges and screw-retained restorations.

o Enamic has the highest translucency but lacks zirconia's strength.

2. Mechanical Strength & Bridge Use

o Subject device: Suitable for 4-unit bridges due to its translucent composition.

o Primary predicate: Supports multi-unit bridges and screw-retained restorations due to higher strength.

o Enamic: Not suitable for bridges due to lower strength.

3. Application Differences

o Subject device: Optimized for high-translucency esthetic restorations and smaller bridges.

o Primary predicate: Optimized for strength, making it suitable for large bridges and screw-retained crowns.

o Secondary predicate (Enamic): Best for veneers, inlays, and onlays due to its flexibility and aesthetics.

Testing Results for VITA YZ MT, YZ ST Color, and YZ ST Multicolor

1. Flexural Strength

o YZ MT (Multi Translucent): 922 ± 95 MPa (meets >800 MPa requirement).

o YZ ST Color: 1073 ± 116 MPa (highest strength).

o YZ ST Multicolor: 1054 ± 95 MPa.

o Conclusion: YZ ST Color had the highest strength, YZ MT had the lowest but still met standards.

2. Linear Coefficient of Thermal Expansion (CTE)

o All materials fall within acceptable ranges:

8 YZ MT: 10.4×10? K?1

& YZ ST Color: 10.2 ×10?6 K?1

1 YZ ST Multicolor: 10.3 ×10?6 K?1

#### 3. Chemical Solubility

o YZ MT: 0 µg/cm2 weight loss (best resistance, no solubility).

o YZ ST Color: 9 µg/cm² (below <100 µg/cm2 requirement).

o YZ ST Multicolor: 10 µg/cm2.

o Conclusion: YZ MT had the best chemical resistance.

4. Radioactivity & Biocompatibility

{6}------------------------------------------------

o All three met the ≤1.0 Bq/g Uranium 238 limit.

Final Conclusions

· YZ ST Color: Strongest material.

- · YZ MT (Multi Translucent): Best translucency and chemical resistance.
· All three materials met thermal expansion, solubility, radioactivity, and biocompatibility standards.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Summary of Non-Clinical Testing Outcomes for VITA YZ ST Multicolor The VITA YZ ST Multicolor material was tested according to EN ISO 6872 standards for dental ceramic materials. The following key nonclinical tests were performed:

1. Packaging, Marking, and Labeling

Visual inspection confirmed compliance with EN ISO 6872:2019. No deviations were found in labeling specifications such as brand name, color designation, and lot number. 2. Uniformity

Ensured even distribution of inorganic pigments and organic dyes. No visible segregation of pigments, meeting EN ISO 6872:2019 requirements. 3. Freedom from Extraneous Materials

Visual inspection confirmed the absence of foreign material. No contamination was detected in tested samples. 4. Flexural Strength

Measured strength: 1054 ± 95 MPa Acceptance criteria: Must be greater than 800 MPa Results exceeded the minimum requirement, confirming the material's durability. 5. Linear Coefficient of Thermal Expansion (CTE)

Measured value: 10.3 ± 0.1 × 10? 8 K? Acceptance criteria: Within manufacturer's specified range of 9.8 - 10.8 x 10?6 K? Results confirmed compliance with the standard. 6. Chemical Solubility

Measured weight loss: 10 µg/cm2 Acceptance criteria: Must be less than 100 ug/cm2 The material showed minimal solubility, confirming chemical stability. 7. Radioactivity

Tested for uranium-238 activity at Jülich Forschungszentrum. Measured value: ≤1.0 Bq/g, meeting EN ISO 6872:2019 requirements. Confirmed safe levels of radioactivity. 8. Biocompatibility

Evaluated according to DIN EN ISO 10993-1 and DIN EN ISO 7405. Testing verified no adverse biological effects, ensuring safety for clinical use. Conclusion The VTA YZ ST Multicolor material successfully met all requirements under ENISO 6872. It demonstrated high flexural strength, thermal stability, chemical resistance, and biocompatibility, making it suitable for dental applications.

---

**Source:** [https://fda-staging.innolitics.com/device/K243940](https://fda-staging.innolitics.com/device/K243940)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
