For dental clinics treatment to irradiate polymer-based restorative materials to cure them. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
Device Story
Cordless, pen-style LED curing light for dental restorative procedures. Device emits light in 385-515nm range to polymerize dental resins or bond brackets; includes caries detection mode using violet light. Operated by qualified dental practitioners in clinical settings. User selects curing mode and time; device delivers light output via LED tip. Disposable PP protective sleeves prevent cross-contamination. Output aids in rapid solidification of restorative materials; caries detection mode uses fluorescence to identify dental decay. Benefits include efficient material curing and diagnostic assistance for caries.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 80601-2-60), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and performance testing per ISO 10650 and ISO 4049.
Indicated for dental clinic patients requiring polymerization of polymer-based restorative materials, including composite resins and orthodontic bracket bonding. Contraindicated for use in oxygen-rich environments.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
Predicate Devices
LED Curing Light (Model: DB686 NANO, DB686 SWIFT) (K220826)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 11, 2025
Foshan COXO Medical Instrument Co., Ltd.
Zhaorong Li
Certification Manager
No.17, Guangming Ave, New Light Source Industrial Base,
Nanha National High-tech Zone
Foshan, Guangdong 528226
China
Re: K243921
Trade/Device Name: LED Curing Lights (DB686 HALO)
Regulation Number: 21 CFR 872.6070
Regulation Name: Ultraviolet Activator For Polymerization
Regulatory Class: Class II
Product Code: EBZ, EAQ
Dated: February 25, 2025
Received: February 25, 2025
Dear Zhaorong Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K243921 - Zhaorong Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243921 - Zhaorong Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bobak
Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243921 | |
| Device Name LED Curing Lights (DB686 HALO) | |
| Indications for Use (Describe) For dental clinics treatment to irradiate polymer-based restorative materials to cure them. The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
510(k)#: K243921
Prepared on: 2025-04-11
1. Contact Details
Applicant Name: Foshan COXO Medical Instrument Co., Ltd.
Applicant Address: No.17,Guangming Ave.,New Light Source Industrial Base, Nanhai National High-tech Zone, Foshan 528226,Guangdong P.R. China
Applicant Contact Telephone: +86-13924548832
Applicant Contact: Mr. Zhaorong Li
Applicant Contact Email: lizhaorong@coxotec.cn
2. Device Name
Device Trade Name: LED Curing Lights
Model: DB686 HALO
Common Name: Ultraviolet activator for polymerization
Classification Name: Activator, Ultraviolet, For Polymerization
Device Class: Class II
Panel: Dental
Regulation Number: 21 CFR Part 872.6070
Product Code(s): EBZ, EAQ
Submission Type: Traditional 510(K)
3. Legally Marketed Predicate Devices
Primary Predicate Device
510(k) Number: K220826
Product Name: LED Curing Light (Model: DB686 NANO, DB686 SWIFT)
{5}
Product Code: EBZ
## Secondary Predicate Device
510(k) Number: K220471
Product Name: VALO™ X; VALO™ X Accessory Lenses
Product Code: EBZ, EAQ, PEQ
## Reference Device 1#
510(k) Number: K221194
Product Name: LEDEX WL-120
Product Code: EBZ
## Reference Device 2#
510(k) Number: K221152
Product Name: CuringPen Dental Curing Light
Product Code: EBZ
## 4. Device Description
The LED Curing Lights (Model: DB686 HALO) is an oral device for repairing teeth. It uses the principle of light curing to make the dental repair resin material solidify rapidly under the action of light wave in the specific wavelength range (385~515nm), so as to fill the tooth cavity or bond the bracket. It is a cordless pen-style device, and consists of handpiece, LED cure tip, charging base, power adapter, eyes protector, eye protector unit and disposable protective sleeve.
The LED Curing Lights protects the handpiece and LED cure tip from gross contamination and prevent cross infection between patients by applying the disposable protective sleeve. The disposable protective sleeve (patient contact part) is made of PP material.
## 5. Intended Use/Indications for Use
For dental clinics treatment to irradiate polymer-based restorative materials to cure them.
The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners.
2
{6}
6. Comparison to the Predicate Device
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALOTM X;
VALOTM X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental
Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Product Code | EBZ, EAQ | EBZ | EBZ, EAQ, PEQ | EBZ | EBZ | NOTE 1 |
| Regulation Number | 21 CFR 872.6070 | 21 CFR 872.6070 | 21 CFR 872.6070 | 21 CFR 872.6070 | 21 CFR 872.6070 | SE |
| Indications for Use | For dental clinics treatment to irradiate polymer-based restorative materials to cure them.
The instrument must only be used in hospital environments, clinics or dental offices, by qualified practitioners. | For light curing polymerization of dental composites, luting materials, cements and other light cured materials. | VALO X curing light is a source of illumination for curing photo-activated dental restorative materials and adhesives. It is also intended to provide illumination to aid in visualization during oral procedures. VALO X curing light accessory/diffusor lenses are not intended for | LEDEX WL-120 is a visible curing unit programmed for polymerization of dental light cured materials by dental professionals. | The CuringPen Dental Curing Light is intended to polymerize resinous dental materials, restorative composite materials, and orthodontic brackets, bonding and sealing materials that are photopolymerized in the 380~515nm waveband of visible light. This device must only be used in hospital | SE |
{7}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALOTM X;
VALOTM X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental
Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | | complete cure of photo-activated materials and adhesives. | | environments, clinics or dental offices by qualified dental personnel and not used in oxygenrich environment. | |
| Principles of operation | 1. Clean and disinfect all the surfaces of handpiece and tip before each use.
2. Cover the LED Cure Tip and handpiece with the Disposable Protective Sleeve.
3. Select curing mode and cure times/orthodonticcycle(s) .
4. Short press ON/OFF Button to start working. | DB686 NANO
1. Clean and disinfect all the surfaces of handpiece and tip before each use.
2. Cover the Cure Tip and handpiece with the Disposable Protective Sleeve.
4. Operate the rotary switch to Cure. Press the Start key to activate light. Press once or Double press to | / | / | Step-by-step:
1. Disinfect contaminated surfaces of the light source head, protective light shield, handpiece before each use.
2. Make sure that the stipulated light irradiance permits adequate polymerization. For that purpose, check the light source head for contamination | SE |
{8}
| Item | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device 1# | Reference Device 2# | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | LED Curing Lights | LED Curing Light (K220826) | VALO™ X; VALO™ X Accessory Lenses (K220471) | LEDEX WL-120 (K221194) | CuringPen Dental Curing Light (K221152) | |
| | 5. Once the selected curing time has elapsed, the curing program is automatically terminated. | select curing time .
5. Once the selected curing time has elapsed, the curing program is automatically terminated.
**DB686 SWIFT**
1. Clean and disinfect all the surfaces of handpiece and tip before each use.
2. Cover the Cure Tip and handpiece with the Disposable Protective Sleeve.
3. Select curing mode and cure times/orthodonticcycle(s) .
4. Short press | | | and damage, as well as the light irradiance at regular intervals.
3. Select curing program and time
4. Start: Once the selected curing time has elapsed, the curing program is automatically terminated. | |
{9}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALOTM X;
VALOTM X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | Power/Start Button to start working.
Once the selected curing time has elapsed, the curing program is automatically terminated. | | | | |
| Operational modes | A. Resin Curing: Cure the composite resin.
Low-temperature Mode: Variable illumination output with illumination values pulsed at 800/1300 mW/cm² cycles;
Normal Mode: Light output at a constant illumination of 1000mW/cm² for | DB686 NANO
1 programs:
Curing light Mode:10s, 20s(1600 ~ 1800mW/cm²)
DB686 SWIFT
3 programs:
- Soft up mode: 5s, 10s, 15s, 20s (1600mW/cm²)
- High power mode: | Operational Modes (Curing EBZ)
VALO X:
Standard Power Mode: 1,100 mW/cm²
Xtra Power Mode: 2,200mW/cm²
VALO X Accessory Lenses:
≥800 mW/cm² (PointCure – | Operational modes
Low, Ramp, Standard, High, Turbo, Plasma, White
Light intensity
MAX. 3200 mW/cm² | 1) Standard mode:
-2300 mW /cm²
-1500 mW/cm²
-1000 mW/cm²
2) RAMP mode:
1000 mW/cm²
3) PULSE mode:
1000 mW/cm²
4) Detect mode:
600 mW/cm² | NOTE 2 |
{10}
| Item | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device 1# | Reference Device 2# | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | LED Curing Lights | LED Curing Light (K220826) | VALO™ X; VALO™ X Accessory Lenses (K220471) | LEDEX WL-120 (K221194) | CuringPen Dental Curing Light (K221152) | |
| | conventional resin curing;
Turbo Mode: Light output at a constant illumination of 3000mW/cm² for fast resin curing;
B. Orthodontics Mode: mainly used for bracket bonding, the illumination value is about 3000mW/cm² ; work 3s stop 2s for a cycle, work cycle 1-8 times adjustable.
C. Caries detection Mode: using violet light to irradiate the teeth to produce a fluorescent reaction, checking dental caries or dental | 5s (1700mW/cm²)
- Orthodontic mode: 1-8Cycles (1800mW/cm²) | Recommended with High Power Mode) (ProxiCure – Recommended with mode suitable for material)
Operational Modes (Diagnostic EAQ, PEQ)
VALO X Accessory Lenses:
≥25 mW/cm², ≤420 nm wavelength (Diffuser Lens – Black Light Diagnostic Aid Mode)
≥1,000 lx luminescence, 3,800-6,500 K color temperature, ≥75 CRI (Diffuser Lens – | | | |
{11}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALOTM X;
VALOTM X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | calculus, a single working time of 60s. | | White Light
Diagnostic Aid Mode)
≥15,000 lx luminescence
(Interproximal Lens – Recommended with Standard Curing mode)
≥500 nm wavelength
(Translume Lens, any mode) | | | |
| Light source | 10W LED
blue and violet | 10W LED | LED light, blue and violet wavelengths (Curing mode)
LED light, violet or white wavelengths (Diagnostic mode)
12.5mm head size | 10W LED | 10W LED | SE |
{12}
| Item | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device 1# | Reference Device 2# | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Upper source | LED Curing Lights
Power adapter Input: 100-240V ~ 50/60Hz
Output: 5V 2A
Li-ion battery: 3.7V 800mAh | LED Curing Light (K220826)
Adapter: 100V - 240V~ 50 - 60 Hz
Output: 5V 1.5A
Batterycharging | VALO X curing light can be powered by an AC power supply or rechargeable batteries.
**AC Power Supply:**
Wall powered.
Output: 9VDC, 2.0A.
Input: 100VAC - 240VAC, 50-60 Hz with adapters for international capability. Ratings: Medical Grade, (CE, RoHS, REACH) Cord: 6 ft (1.8m), 2.5mm DC connector
**Battery Power:**
1IMR14/65 3.7V 900mAh 3.33 Wh Li Ion rechargeable | LEDEX WL-120 (K221194)
Power supply
Input: AC100~240V, 50/60Hz
Output: DC 5V/2A
**Power source**
3.7V Rechargeable Li-ion battery | 3.7V DC with Lithium ion battery
5V DC with charger power | NOTE 3 |
{13}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALO™ X;
VALO™ X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | | battery pack
**Power On Button:**
Located on the handle of the wand, back side and front side | | | |
| Accessories | -Handpiece
-Charging Base
-LED cure tip
-Battery (Included in the handpiece)
-Eyes protector
-Power adapter
-Disposable protective sleeve
-Eye protector unit
-User Manual | **DB686 NANO**
-Handpiece
-Charging Base
-Cure Tip
-Protection glasses
-Battery pack
-Adapter (with Power Cord)
-Protective Sleeve
-User Manual
**DB686 SWIFT**
-Handpiece | -Barrier Sleeve VALO™,
-Blue-Light Blocking Light Shield
-PointCure Lens,
ProxiCure Ball Lens,
Translume Lens,
Diffuser Lens, Inter-proximal Lens, | Handpiece,
Power supply,
Optical fiber light guide rod,
Filter,
Light guide sleeves,
Cradle,
Anti-glare shield | -Handpiece
-Charging base
-Light source head
-Protective light shield
-Adapter
-Disposable sleeves (1x100pcs)
-User Manual
Optional accessories:
-Light curing depth test board
-Protective light | NOTE 4 |
{14}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALO™ X;
VALO™ X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental
Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | -Charging Base
-Cure Tip
-Battery (Included in the handpiece)
-Adapter (with Power Cord)
-Protective Sleeve
-User Manual | | | glasses | |
| Use | Prescription / Hospital | Prescription / Hospital | Prescription / Hospital | Prescription / Hospital | Prescription / Hospital | SE |
| Wavelength range | 385~515nm | 385-515nm | 380~515 nm | 390nm-480nm | 380-515nm | SE |
| Peak wavelength (nm) | 402 & 460nm | 460nm | VALO X:
Nominal values: 380-420nm and 420-515nm
VALO X Accessory Lenses: | 405 & 460 nm | Dual peak:
400-410nm, 450-460nm | NOTE 5 |
{15}
| Item | Proposed Device LED Curing Lights | Primary Predicate Device LED Curing Light (K220826) | Secondary Predicate Device VALOTM X; VALOTM X Accessory Lenses (K220471) | Reference Device 1# LEDEX WL-120 (K221194) | Reference Device 2# CuringPen Dental Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | | All lenses match the curing lights peak wavelengths except: ≤420 nm wavelength (Diffuser Lens in Black Light Diagnostic Mode) ≥500 nm peak wavelength (Translume Lens, any mode) | | | |
| Composition of patient-contacting portions of device | Disposable protective sleeve, material: PP | Disposable protective sleeve, material: PP | Barrier Sleeve | Light guide sleeves | Disposable sleeve | SE |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile | SE |
| Infection Control | Use disposable protective sleeve and surface disinfection | Use disposable protective sleeve and surface disinfection | Barrier Sleeve | Light guide sleeves | Use disposable sleeve to prevent cross infection of the | SE |
{16}
| Item | Proposed Device
LED Curing Lights | Primary Predicate Device
LED Curing Light (K220826) | Secondary Predicate Device
VALO™ X;
VALO™ X
Accessory Lenses (K220471) | Reference Device 1#
LEDEX WL-120 (K221194) | Reference Device 2#
CuringPen Dental Curing Light (K221152) | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | to prevent cross infection of the patient. | to prevent cross infection of the patient. | | | patient | |
| Electrical Safety | IEC60601-1
IEC60601-1-2
ANSI AAMI ES60601-1
ANSI AAMI ES60601-1-2
IEC 80601-2-60 | IEC60601-1
IEC60601-1-2
IEC 80601-2-60 | IEC 60601-1
IEC 60601-1-2
IEC 80601-2-60 | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2
ANSI AAMI ES60601-1 | NOTE 6 |
| Photobiological Safety | IEC 62471 | IEC 62471 | IEC 62471 | / | IEC 62471 | SE |
| Biocompatibility | ISO 10993-5
ISO 10993-10
ISO 10993-11 | ISO 10993-5
ISO 10993-10
ISO 10993-11 | ISO 10993-1 | / | ISO 10993-5
ISO 10993-10 | SE |
{17}
14
Note:
Note 1 The curing function of proposed device is conformed with product code EBZ (K220826, K220471) and the detect function is conformed with product code EAQ (K220471), but PEQ is not applicable.
Note 2 The operational modes and light intensity of the proposed device is different with the predicate device. The proposed devices have several modes corresponding to the light output intensity and available times.
The maximum light intensity of the proposed device is higher than the predicate device which can achieve higher efficiency for irradiating polymer-based restorative materials, but it is still lower than a reference device K221194 (the claimed maximum light intensity is Max. $3200\mathrm{mW/cm^2}$).
The light output safety and performance test was conducted according to ISO 10650, IEC 62471, IEC 60601-1 and FDA guidance performance testing requirements.
The testing results show that these difference do not affect the safety and effectiveness.
Note 3 The proposed device is charged by the contact charging method as same as the primary predicate device. As a power source for the charger, the AC/DC adapter provided by the manufacturer is used. The input specifications of the adapter (100-240VAC, 50-60Hz) are the same, but the outputs differ from each other at $5\mathrm{V} \times 2\mathrm{A}$ and $5\mathrm{V} \times 1.5\mathrm{A}$ respectively. These difference do not affect safety and effectiveness.
Note 4 In order to block the blue light radiation outside the application part to protect the eyes of the operator and patient from blue light hazard, the proposed device provides two accessories: Eyes protector and Eye protector unit, and the predicate device only has Protection glasses which is same with Eyes protector. Eye protector unit provides an additional solution to protect the eyes of dentists and patients from hurting. This difference does not affect the safety and effectiveness.
Note 5 The proposed device and the secondary predicate device have dual peak wavelength. The peak wavelength of the proposed device is described as given values while the peak wavelength of reference device as a range. The values are similar. This difference does not raise issues of substantial equivalence.
{18}
Note 6 According to the requirements of US national differences, the proposed device has tested by ANSI AAMI ES60601-1 and ANSI AAMI ES60601-1-2.
Different but does not adversely impact safety and effectiveness of proposed device.
## Discussing:
The new device LED Curing Lights (Model: DB686 HALO) is substantial equivalence as the primary predicate device K220826, the secondary predicate device K220471 and the reference devices (K221194, K221152). The differences was in model and some minor parameters, it's because of they are different in design requirements between different devices, which not affect the safety performance of the device, and the new device passed Safety testing, so these differences will not cause any safety and effectiveness issues. Considering the same intended use/indications for use, principles of operation, and product performance etc., they are substantial equivalence.
## 7. Non-clinical Data
All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing demonstrate the compliance to the standards and matching the performance of new devices to the predicate devices.
- ISO 10993-5:2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- IEC 60601-1:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- AAMI ES60601-1:2021 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic
15
{19}
safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- ANSI/AAMI/IEC 60601-1-2:2021 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- IEC 80601-2-60:2019 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- IEC 62471: 2006 Photobiological safety of lamps and lamp systems
- IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them for use in portable applications - Part 2: Lithium systems
- ISO 10650:2018 Dentistry — Powered polymerization activators
- ISO 4049:2019 Dentistry - Polymer-based restorative materials
## 8. Clinical Data
There were no clinical tests performed for the LED Curing Lights (Model: DB686 HALO).
## 9. Conclusion
The LED Curing Lights (Model: DB686 HALO) is substantially equivalent to the primary predicate device (K220826), secondary predicate device (K220471) and the reference devices (K221194, K221152).
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.