This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. This product is not intended for apneic ventilation.
Device Story
F&P Optiflow Air/Oxygen Flow Source is a respiratory high flow therapy device; generates air/oxygen mixture for hospital use. Inputs: ambient air and external oxygen source (40–87 psi). Operation: electronic flow and FiO2 control; user interface via touch screen display. Outputs: high flow respiratory gases (up to 70 L/min; 21–100% oxygen concentration). Operated by qualified healthcare professionals in hospitals. Device monitors flow and FiO2 output; includes visual/audible alarm system for supply pressure, FiO2, power failure, disconnection, and obstruction. Output assists clinicians in managing respiratory gas delivery; benefits patients by providing controlled, high-flow respiratory support.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (ANSI/AAMI ES 60601-1), alarm systems (IEC 60601-1-8), software lifecycle (IEC 62304), usability (IEC 62366-1), electromagnetic immunity (AIM 7351731), respiratory high-flow equipment requirements (ISO 80601-2-90), and gas mixer performance (ISO 11195:2018).
Technological Characteristics
Mains-powered respiratory gas mixer. Electronic flow and FiO2 control. Flow range 0–70 L/min; FiO2 21–100%. Touch screen interface. Visual/audible alarm system. Complies with ISO 80601-2-90 and ISO 11195:2018. Software developed per IEC 62304.
Indications for Use
Indicated for delivery of high flow respiratory gases to patients in hospital settings by qualified healthcare professionals. Not for patients with bypassed upper airways, apneic ventilation, or life-sustaining/supporting use.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
Airvo 2 Humidifier and myAirvo 2 Humidifier (K131895)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 14, 2025
Fisher & Paykel Healthcare Ltd
Reena Daken
Regulatory Affairs Manager
15 Maurice Paykel Place, East Tamaki
Auckland, 2013
New Zealand
Re: K243917
Trade/Device Name: F&P Optiflow Air/Oxygen Flow Source
Regulation Number: 21 CFR 868.5330
Regulation Name: Breathing Gas Mixer
Regulatory Class: Class II
Product Code: BZR
Dated: June 12, 2025
Received: June 12, 2025
Dear Reena Daken:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243917 - Reena Daken
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K243917 - Reena Daken
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243917
Device Name
F&P Optiflow Air/Oxygen Flow Source
Indications for Use (Describe)
This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals.
This product is not intended to be used on patients with bypassed upper airways.
This product is not intended to be life-sustaining or life-supporting.
This product is not intended for apneic ventilation.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
F&P Optiflow Air/Oxygen Flow Source - K243917
# 510(k) Summary
As Required by 21 CFR 807.92
## I. SUBMITTER
| Company Name and Address | Fisher & Paykel Healthcare Limited
15 Maurice Paykel Place
East Tamaki
Auckland 2013, New Zealand
Telephone: +64 9 574 0100 |
| --- | --- |
| Prepared and Submitted by | Danica Tung
Regulatory Affairs Market Manager |
| Contact Person | Reena Daken
Regulatory Affairs Manager
Telephone: +64 9 574 0100
Email: reena.daken@fphcare.co.nz |
| Date Prepared | 14 July 2025 |
## II. DEVICE
| Name of Device | F&P Optiflow™ Air/Oxygen Flow Source |
| --- | --- |
| Common/Usual Name | Breathing Gas Mixer |
| Classification | 21 CFR §868.5330 |
| Regulatory Class | II |
| Product Code | BZR |
## III. PREDICATE DEVICES
Primary Predicate Device: MaxBlend 2 and MaxBlend Lite, K161718
Secondary Predicate Device: Airvo 2 Humidifier and myAirvo 2 Humidifier, K131895
{5}
F&P Optiflow Air/Oxygen Flow Source - K243917
## IV. DEVICE DESCRIPTION
The F&P Optiflow Air/Oxygen Flow Source is a respiratory high flow therapy device designed to generate a flow of air and/or oxygen. The device allows selectable flow rates up to 70 L/min with selectable oxygen concentrations ranging from 21% to 100%.
The subject device is a multi-patient use prescription only device, provided in a non-sterile state. It operates at flow ranges between 0 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals.
## V. INDICATIONS FOR USE
This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals.
This product is not intended to be used on patients with bypassed upper airways.
This product is not intended to be life-sustaining or life-supporting.
This product is not intended for apneic ventilation.
{6}
F&P Optiflow Flow Diverter - K243917
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| Classification | | | | |
| Device Classification | 21 CFR 868.5330 Class II | 21 CFR 868.5330 Class II | 21 CFR 868.5450 Class II | Identical |
| Classification Panel | Anesthesiology | Anesthesiology | Anesthesiology | Identical |
| Product Code | BZR | BZR | BTT | Identical |
| Indications for use and intended use | | | | |
| Indications for Use | This device is intended for the delivery of high flow respiratory gases (within the limits of its stated technical specifications) to patients in a hospital by appropriately qualified healthcare professionals. This product is not intended to be used on patients with bypassed upper airways. This product is not intended to be life-sustaining or life-supporting. | The MaxBlend 2 and MaxBlend Lite are designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, sub-acute, and nursing-care facilities where the delivery and | The Airvo 2 Humidifier is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2 - 60 L/min depending on the patient interface. The Airvo 2 Humidifier is for patients in hospitals and long-term care facilities. | Equivalent |
{7}
F&P Optiflow Flow Diverter - K243917
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| | This product is not intended for apneic ventilation. | monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device. | | |
| Availability | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Identical |
| Intended Use Environment | Hospital | Professional healthcare settings, i.e., hospital, sub-acute, and nursing-care facilities, where the delivery and monitoring of air/oxygen mixtures is required. Not for use in MRI environments | Hospitals and long-term care facilities | Equivalent |
| Users | Qualified healthcare professionals | Qualified trained personnel, under the direction of a physician | Healthcare professionals | Identical to secondary predicate. |
| Contraindications | No contraindications | No contraindications | No contraindications | Identical |
| Technological Characteristics | | | | |
| Power/Energy Source | Mains power | Battery (4 x AA Alkaline batteries) - Power source for the oxygen monitor only. | Mains power | Identical to secondary predicate |
| Control & Monitoring | Electronic flow and FiO2control. | Mechanical flow and FiO2control. Monitoring FiO2output. | Electronic flow control. | Equivalent |
{8}
F&P Optiflow Flow Diverter - K243917
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| | Monitoring of flow and FiO2 output. | | Monitoring of flow and FiO2 output. | |
| Device User Interface | Display with touch screen | Monitoring system has a display and push buttons for user input. Flow control is via rotary dial and a ball float for indicator | Display with push buttons | Different Similar to the primary and secondary predicates, the subject device has a display screen. Input and control for the subject device is via the touch screen, whereas both predicates utilise push buttons and a rotary dial (primary predicate only). The difference in device user interface does not raise new questions of safety and effectiveness. |
| Performance Specifications | | | | |
| Flow Range | 0 - 70 L/min | 0 - 100 L/min | 2 - 60 L/min | Equivalent |
| Flow Meter Accuracy | Accuracy of +/- 10% | Accuracy of +/- 10% | N/A | Identical |
| Operating Temperature Range | 64°F to 79°F (18 °C to 26 °C) | 59°F to 122°F (15°C to 50°C) | 64°F to 79°F (18 °C to 26 °C) | Equivalent |
{9}
F&P Optiflow Flow Diverter - K243917
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| % Oxygen Control | 21 – 100% | 21 – 100% | 21 – 100% | Identical |
| % Oxygen Control Accuracy | +/- 5% | +/- 3% | Not applicable | Different The subject device's % oxygen control accuracy limit is based on ISO 11195:2018, that specifies the requirements for the performance and safety of stand-alone gas mixers (blenders). The difference in oxygen control accuracy does not raise new questions of safety and effectiveness. |
| Alarms | | | | |
| Alarm method | Visual and audible alarm system. Mute button | Visual and audible alarm system. Mute button | Visual and audible alarm system. Mute button | Identical |
| Oxygen Supply Alarm | Issues alarm if the oxygen supply pressure is < 40psi | Air/oxygen mixer feature: Pressure supply differential alarm - Air/oxygen pressure must be < 20 psi an alarm sounds | Not Applicable | Equivalent |
{10}
F&P Optiflow Flow Diverter - K243917
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| FiO2 Alarm | Device alarms if it is unable to achieve set FiO2FiO2: 21 - 100% | Oxygen monitor features: High alarm range - 16 - 100% | Issues alarm if oxygen fraction as measured by oxygen sensor is above or below user adjustable high/low limit. | Equivalent |
| Power-fail Alarm | Power out audible alarm for time, t > 2 minutes or until power is restored, or capacitor charge is depleted. | Unknown | Power out audible alarm for time, t > 2 minutes or until paused by user pushing alarm mute button, power is restored (non-latching alarm), or capacitor charge is depleted. | Equivalent |
| Disconnection | Disconnection alarm activated in < 5 seconds when no connection/disconnection has been detected | Unknown | ‘Check tube’ and ‘Check for leaks’ alarms are activated in < 5 seconds if the breathing tube is not plugged in correctly | Identical to secondary predicate |
| Obstruction Detected/Flow Alarm | Obstruction alarm will be activated in < 5 seconds when an obstruction has been detected | Unknown | ‘Check for blockages’ alarm in <10 seconds when the device has detected a blockage | Equivalent |
| General | General alarm warning is activated when internal faults, such as hardware, visual alarm, audio alarm and alarm systems has been detected. | Unknown | Fault alarm is activated when the unit has detected an internal fault and has shut itself down. | Equivalent |
{11}
F&P Optiflow Flow Diverter - K243917
| Design/technological characteristics | Subject Device (F&P Optiflow Air/Oxygen Flow Source) | Primary Predicate Device (MaxBlend 2 and MaxBlend Lite - K161718) | Secondary Predicate Device (Airvo 2 Humidifier and myAirvo 2 Humidifier - K131895) | Comments |
| --- | --- | --- | --- | --- |
| Other | | | | |
| Gas Supply Type & Pressure | Room air: Ambient pressure External oxygen source : 40 – 87 psi | Air/Oxygen: 30 – 75 psi | Room air: Ambient pressure External oxygen source (optional) | Equivalent |
| Materials in Gas Pathway | Externally communicating – tissue contact, permanent duration. No exposure to humidified gases. VOC and PM2.5 | Externally communicating – tissue contact, permanent duration. No exposure to humidified gases. VOC and PM2.5 | Indirect contact - External communicating - tissue, bone, dentin | Identical |
{12}
F&P Optiflow Flow Diverter - K243917
# VII. PERFORMANCE DATA
# Summary of Non-Clinical Tests
Performance testing of the Flow Source was completed and confirms the subject device does not raise new questions of safety and effectiveness. The testing provided demonstrates substantial equivalence of the subject device to the predicate devices.
The Flow Source has been tested to the applicable requirements to the following standards:
- ANSI/AAMI ES 60601-1:2005/A2:2021 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety and Essential Performance - Amendments 2
- IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 62304:2006 + A1:2015 Medical device software - Software life cycle processes
- IEC60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ISO 17664-2:2021 Processing of health care products – Information to be provided by medical device manufacturer for the processing of medical devices. Part 2: Non-critical medical devices
- IEC 62366-1:2015 + A1:2020 Medical devices - Part 1: Application of usability engineering to medical devices
- AIM 7351731 Rev 3:2021 Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers
- ISO 80601-2-90:2021 Medical electrical equipment - Part 2-90: Particular requirements for basic safety and essential performance of respiratory high-flow therapy equipment
- ISO 11195:2018 Gas mixers for medical use - Stand-alone gas mixers
- ANSI/AAMI ST98:2022 Cleaning validation of health care products - Requirements for development and validation of a cleaning process for medical devices.
# VIII. CONCLUSION
The Flow Source is substantially equivalent to the predicates based on intended use, technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.