← Product Code [HSB](/productcode/HSB) · K243907

# Affixus Tibial Nailing System - 4mm screws (K243907)

_Zimmer, Inc. · HSB · Feb 12, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243907

## Device Facts

- **Applicant:** Zimmer, Inc.
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Feb 12, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Affixus® Tibial Nailing System is indicated for temporary stabilization of tibial fractures and osteotomies including proximal, metaphyseal, and distal shaft fractures, closed fractures, open fractures, pathologic fractures, non-unions, malunions and deformity corrections.

## Device Story

Affixus Tibial Nailing System is an intramedullary fixation device for tibial fractures. System includes nails, end caps, and new 4mm cortical bone screws. Implants are manufactured from Ti-6Al-4V. Screws feature double-lead thread design and self-tapping tips; intended for insertion through distal locking holes of 8mm tibial nails. Used by orthopedic surgeons in clinical/surgical settings to restore bone shape and provide temporary stabilization. Device provides mechanical fixation to facilitate fracture healing. Benefits include versatile locking options and anatomical compatibility across various fracture types.

## Clinical Evidence

Bench testing only. Testing included construct bending fatigue strength, bone screw bending strength (ASTM F1264), screw torque to failure, breaking angle, insertion/removal torque, pullout strength, and self-tapping performance (ASTM F543). MRI compatibility evaluated per ASTM F2182, F2052, F2119, and F2213.

## Technological Characteristics

Material: Ti-6Al-4V. Design: Intramedullary fixation rod with 4mm cortical bone screws (double-lead thread, self-tapping). Standards: ASTM F1264 (bending strength), ASTM F543 (screw performance), ASTM F2182/F2052/F2119/F2213 (MRI). Connectivity: None. Sterilization: Validated methods consistent with predicate.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Affixus Tibial and Antegrade Femoral Nailing System ([K241651](/device/K241651.md))
- ZIMMER M/DN INTRAMEDULLARY FIXATION SYSTEM ([K142281](/device/K142281.md))
- BIOMET TIBIAL LOCKING NAIL SYSTEM ([K063570](/device/K063570.md))
- ZIMMER NATURAL NAIL SYSTEM TIBIAL NAIL ([K082770](/device/K082770.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 12, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zimmer Inc. Akash Savalia Regulatory Affairs Sr. Specialist 1800 W Center Street Warsaw, Indiana 46580

Re: K243907

Trade/Device Name: Affixus Tibial Nailing System - 4mm screws Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: December 19, 2024 Received: December 19, 2024

Dear Akash Savalia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Farzana Sharmin -S Digitally signed by Farzana Sharmin -S Date: 2025.02.12 12:56:40 -05'00'

Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K243907

Device Name

Affixus Tibial Nailing System - 4mm screws

Indications for Use (Describe)

The Affixus® Tibial Nailing System is indicated for temporary stabilization of tibial fractures and osteotomies including proximal, metaphyseal, and distal shaft fractures, closed fractures, open fractures, pathologic fractures, non-unions, malunions and deformity corrections.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #:                          | 510(k) Summary                                           | Prepared on: 2025-02-1                             |
|------------------------------------|----------------------------------------------------------|----------------------------------------------------|
| Contact Details                    |                                                          | 21 CFR 807.92(a)(1)                                |
| Applicant Name                     |                                                          | Zimmer Inc.                                        |
| Applicant Address                  |                                                          | 1800 W Center Street Warsaw IN 46580 United States |
| Applicant Contact Telephone        |                                                          | 1-978-908-0222                                     |
| Applicant Contact                  |                                                          | Mr. Akash Savalia                                  |
| Applicant Contact Email            |                                                          | akash.savalia@zimmerbiomet.com                     |
| Device Name                        |                                                          | 21 CFR 807.92(a)(2)                                |
| Device Trade Name                  |                                                          | Affixus Tibial Nailing System - 4mm screws         |
| Common Name                        |                                                          | Rod, Fixation, Intramedullary And Accessories      |
| Classification Name                |                                                          | Intramedullary fixation rod                        |
| Regulation Number                  |                                                          | 888.3020                                           |
| Product Code(s)                    |                                                          | HSB, HWC                                           |
| Legally Marketed Predicate Devices |                                                          | 21 CFR 807.92(a)(3)                                |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code                                       |
| K241651                            | Affixus Tibial and Antegrade Femoral Nailing System      | HSB                                                |
| K142281                            | ZIMMER M/DN INTRAMEDULLARY FIXATION SYSTEM               | HSB                                                |

K243907 - Page 1 of 2

нгв

HSB

21 CFR 807.92(a)(4)

21 CFR 807.92(a)(5)

BIOMET TIBIAL LOCKING NAIL SYSTEM K063570

# Device Description Summary

The Affixus Tibial Nalling System is a long bone nailing system that offer implants designed to treat a range of tibial fractures from simple to complex, with versatile locking options. They restore the shape of preinjured bone and are available in a variety of lengths and diameters to meet assorted anatomical needs. Screws and End Caps are included in the system. All implants are made of Ti-6Al-4V. The Affixus platform of instrumentation is designed to provide options and flexibility for many intraoperative approaches while maintaining ease of use and commonality.

ZIMMER NATURAL NAIL SYSTEM TIBIAL NAIL

This submission introduces 4mm Cortical Bone Screws are also made of Ti-6AL-4V and feature a double-lead thread design with a self-tapping tip. Screw options include

Ø4mm: 20mm - 60mm, increments of 2 mm

60mm - 90mm, increments of 5mm

K082770

Note: 4mm Screws are inserted through distal locking holes of 8mm Tibia nails only.

# Intended Use/Indications for Use

The Affixus® Tibial Nailing System is indicated for temporary stabilization of tibial fractures and osteotomies including

{5}------------------------------------------------

K243907 - Page proximal, metaphyseal, and distal shaft fractures, open fractures, pathologic fractures, non-unions, malunions and deformity corrections.

# Indications for Use Comparison

The indications for use for the subject device are identical to the primary predicate Affixus Tibial Nailing System.

# Technological Comparison

The proposed device has similar technological characteristics as the primary predicate system.

The following items are identical to the Affixus Tibial system cleared in K241651.

-Component Materials, raw and contact

-MRI conditional Status and MRI language

-Sterilization methods and validations

-Packaging materials, methods and validations

-Shelf life validations

-Biocompatibility

The rationale for substantial equivalence is based on consideration of the following characteristics:

· Materials: Identical to primary predicate device implants

- · Design Features: Similar to predicate device
- · Packaging: Similar to predicate device
- · Sterilization: Similar to primary predicate device

# Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Non-Clinical Tests:

- · Construct Bending Fatigue Strength
- · Bone Screw Bending Strength per ASTM F1264 Annex 4
- · Screw Maximum Torque to Failure and Breaking Angle per ASTM F543 Annex 1
- · Screw Insertion/ Removal Torque per ASTM F543 Annex 2
- · Screw Pullout Strength per ASTM F543 Annex 3
- · Screw Self Tapping Performance per ASTM F543 Annex 4
- · MR Compatibility Evaluation per ASTM F2182, ASTM F2052, F2119, ASTM 2213

The proposed device has a similar intended use as the predicates and similar technological characteristics. The information provided herein demonstrates that:

- · different technological characteristics of the proposed device do not raise new questions of safety and effectiveness; and
- · the proposed device is at least as safe and effective as the legally marketed predicate device.

## 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

---

**Source:** [https://fda-staging.innolitics.com/device/K243907](https://fda-staging.innolitics.com/device/K243907)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
