K243893 · Pearl, Inc. · MYN · May 5, 2025 · Radiology
Device Facts
Record ID
K243893
Device Name
Second Opinion® Pediatric
Applicant
Pearl, Inc.
Product Code
MYN · Radiology
Decision Date
May 5, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2070
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K243893 · May 5, 2025
Second Opinion® Pediatric
Pearl, Inc.
Retrospective clinical dental radiographs
A retrospective standalone study was conducted using 1,182 clinical radiographs to evaluate the sensitivity and accuracy of the CADe software in detecting caries in pediatric patients (ages 4-11).
Sensitivity of caries detection, FPPI, wAFROC FOM, HR-ROC FOM, and Dice Score
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Caries detection
Neural network-based computer vision
Sensitivity > 75%
Sensitivity 0.87 (95% CI: 0.84, 0.90)
—
—
Standalone retrospective study of 1182 radiographic images
—
Indications for Use
Second Opinion® Pediatric is a computer aided detection ("CADe") software to aid in the detection of caries in bitewing and periapical radiographs. The intended patient population of the device is patients aged 4 years and older that have primary or permanent teeth (primary or mixed dentition) and are indicated for dental radiographs.
Device Story
Second Opinion® Pediatric is a CADe software for dental clinicians; it processes bitewing and periapical radiographs to detect and segment caries. The device uses machine learning modules to analyze input images; it outputs polygonal overlays on the radiograph to highlight potential caries for clinical review. The system consists of an API, ML modules, and a client UI. Clinicians use the output as a second reader to assist in their assessment; it does not replace clinical judgment or dentist review. The device is used in dental clinics; it provides near-instantaneous results to aid in diagnostic decision-making, potentially improving the accuracy of caries detection in pediatric patients.
Clinical Evidence
Standalone retrospective study of 1,182 radiographic images (549 abnormal, 1,085 caries lesions). Primary endpoint: sensitivity > 75% for bitewing and periapical images. Results: lesion level sensitivity 0.87 (95% CI: 0.84, 0.90), p < 0.0001. Additional metrics: FPPI 1.22 (95% CI: 1.14, 1.30), wAFROC FOM 0.86 (95% CI: 0.84, 0.88), HR-ROC FOM 0.94 (95% CI: 0.93, 0.96), and LS mean Dice Score 0.76 (95% CI: 0.75, 0.77).
Technological Characteristics
Software-only device; runs on Windows OS. Uses neural network-based computer vision and machine learning algorithms for caries detection. Cloud-based processing environment. Outputs graphical overlays on radiographs via a client UI. Moderate level of software concern.
Indications for Use
Indicated for patients aged 4 years and older with primary or permanent teeth (primary or mixed dentition) who are indicated for dental radiographs to aid in the detection of caries in bitewing and periapical radiographs.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 5, 2025
Pearl, Inc.
% Ashley Brown
Director of Regulatory Affairs
2515 Benedict Canyon Dr.
BEVERLY HILLS, CA 90210
Re: K243893
Trade/Device Name: Second Opinion® Pediatric
Regulation Number: 21 CFR 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN
Dated: April 3, 2025
Received: April 3, 2025
Dear Ashley Brown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243893 - Ashley Brown
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243893 - Ashley Brown
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243893 | |
| Device Name | |
| Second Opinion® Pediatric | |
| Indications for Use (Describe) | |
| Second Opinion® Pediatric is a computer aided detection ("CADe") software to aid in the detection of caries in bitewing and periapical radiographs. The intended patient population of the device is patients aged 4 years and older that have primary or permanent teeth (primary or mixed dentition) and are indicated for dental radiographs. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Pearl
# K243893 510(K) SUMMARY
## 1. Submitter's Identification
Pearl Inc.
2515 Benedict Canyon Dr.
Beverly Hills, CA, 90210
USA
(239) 450-8829
compliance@hellopearl.com
Contact Person: Ashley Brown
Position: Director of Quality & Regulatory Affairs
Date Summary Prepared: May 2, 2025
## 2. Trade Name of the Device
Second Opinion® Pediatric
## 3. Common or Usual Name
Analyzer, Medical Image
## 4. Classification Name, Regulatory Classification & Product Code
Classification Name: Medical Image Analyzer
Regulatory Classification: 21CFR 892.2070, Class II
Product Code: MYN
## 5. Predicate Information
Predicate device: The primary proposed predicate device for clearance of Second Opinion® Pediatric is the original Second Opinion device, by Pearl Inc., cleared on March 04, 2022 (K210365) and classified as a Class II pursuant to 21 CFR §892.2070, under product code MYN (Analyzer, Medical Image).
The cleared device is a computer aided detection (CADe) software device, indicated for use by dental health professionals as an aid in their assessment of bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older, as second reader. The device utilizes computer vision technology, developed using machine learning techniques, to identify and mark anatomy & pathology in bitewing and periapical radiographs.
The subject device is substantially equivalent to the predicate as the overall intended
Hellopearl.com
{5}
Pearl
use and nature of the software remains the same. The difference with the subject device is the addition of the patient age group of 4 years of age and older.
# 6. Device Description
Second Opinion Pediatric is a radiological, automated, computer-assisted detection (CADe) software intended to aid in the detection and segmentation of caries on bitewing and periapical radiographs. The device is not intended as a replacement for a complete dentist's review or their clinical judgment which considers other relevant information from the image, patient history, or actual in vivo clinical assessment.
Second Opinion Pediatric consists of three parts:
- Application Programming Interface ("API")
- Machine Learning Modules ("ML Modules")
- Client User Interface (UI) ("Client")
The processing sequence for an image is as follows:
1. Images are sent for processing via the API
2. The API routes images to the ML modules
3. The ML modules produce detection output
4. The UI renders the detection output
The API serves as a conduit for passing imagery and metadata between the user interface and the machine learning modules. The API sends imagery to the machine learning modules for processing and subsequently receives metadata generated by the machine learning modules which is passed to the interface for rendering.
Second Opinion® Pediatric uses machine learning to detect caries. Images received by the ML modules are processed yielding detections which are represented as metadata. The final output is made accessible to the API for the purpose of sending to the UI for visualization. Detected caries are displayed as polygonal overlays atop the original radiograph which indicate to the practitioner which teeth contain detected caries that may require clinical review. The clinician can toggle over the image to highlight a potential condition for viewing. Further, the clinician can hover over the detected caries to show a hover information box containing the segmentation of the caries in the form of percentages.
# 7. Indications for Use
Hellopearl.com
{6}
Pearl
Second Opinion® Pediatric is a computer aided detection ("CADe") software to aid in the detection of caries in bitewing and periapical radiographs.
The intended patient population of the device is patients aged 4 years and older that have primary or permanent teeth (primary or mixed dentition) and are indicated for dental radiographs.
# 8. Summary of Substantial Equivalence:
The predicate devices and subject device are similar devices in the following ways:
1) Intended use: Both devices are intended to be used to aid dental clinicians in their review of caries lesions in bitewing and periapical radiographs of permanent teeth. The difference between the subject and predicate is the expanded age group to include 4+ in the subject.
2) Technology characteristics: Both devices employ computer vision and machine learning to output detections, use cloud-based environments to conduct processing, and display caries lesions within a user interface with a graphical overlay over radiographs.
3) Safety: As both the candidate and predicate devices are software systems, neither pose a direct safety hazard to the patient. The primary hazards for both devices, subject and predicate, are potential false positives and false negatives which could result in a temporary non-serious injury. In the case of each device, users are not meant to rely solely on detection output for clinical decision-making.
4) Clinical Performance: Both devices have undergone clinical studies which demonstrate statistically significant improvement in performance.
Table 1: Comparison of Second Opinion® Pediatric with the predicate devices.
| | Subject Device
Second Opinion® Pediatric | Primary Predicate
Second Opinion
K210365 |
| --- | --- | --- |
| Manufacturer | Pearl Inc. | Pearl Inc. |
| Classification | 892.2070 | 892.2070 |
| Product Code | MYN | MYN |
| Image Modality | Radiograph | Radiograph |
| Intended Use | Dental CADe to aid in dental radiograph review by HCP | Dental CADe to aid in dental radiograph review by HCP |
Hellopearl.com
{7}
Pearl
| Full IFU | Second Opinion® Pediatric is a computer aided detection (“CADe”) software to aid in the detection of caries in bitewing and periapical radiographs.
The intended patient population of the device is patients aged 4 years and older that have primary or permanent teeth (primary or mixed dentition) and are indicated for dental radiographs. | Second Opinion® is a computer aided detection (“CADe”) software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge and Implants.
It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a second reader. |
| --- | --- | --- |
| Intended body part | Dental | Dental |
| Technology | Automated, CADe software that utilizes machine learning | Automated, CADe software that utilizes machine learning |
| Device Description | Detection & display of caries lesions in intraoral radiographs | Detection and display of anatomy and pathologies in intraoral radiographs |
# 9. Technological Comparison to Predicate Devices
The fundamental technological principle for both the candidate and predicate device is the automatic detection and display of caries lesions using machine learning.
The candidate and predicate devices are technologically equivalent as follows:
- Both are software devices designed to run on Windows operating systems.
- Both devices are designed to process digital intraoral bitewing and periapical radiographs.
- Both devices use neural network-based computer vision algorithms for caries lesions display.
- Both devices display caries within the user interface with a graphical overlay on the radiograph.
- Both devices produce near-instantaneous detection results.
- Both devices are considered to be of “moderate” level of software concern.
- Both devices passed all verification and validation testing requirements.
Hellopearl.com
{8}
Pearl
The candidate and predicate devices technologically differ as follows: The candidate device includes only caries detection whereas the predicate includes the detection of several types of dental restorations as well. These differences in features available do not raise any concerns about the safety or efficacy of the device as compared to the predicate.
# 10. Assessment of Benefit-Risk, Safety and Effectiveness, and Substantial Equivalence to Predicate Device
Pearl demonstrated the benefits of the device through a standalone clinical study. The results of the study showed the system is substantially equivalent to the predicate device. When the probable benefits and probable risks of Second Opinion® Pediatric are weighed against one another, the weight of benefits significantly exceeds that of risks. This judgment can be made based on review of the submitted materials showing that Second Opinion® Pediatric meets the design verification and validation requirements. It is thus concluded that Second Opinion® Pediatric can be considered safe and effective, in comparison to the predicate, such that the device will aid users in the indicated user population in their radiographic review of caries lesions.
# 11. Cybersecurity
Pearl developed Security controls and processes in accordance with FDA Guidance - Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions dated September 2023. These processes are used in both the development of Second Opinion® Pediatric and in post-market surveillance to ensure the product upholds the highest standards of privacy and security.
# 12. Discussion of Non-Clinical Tests Performed
The device is a software-only device, so most testable characteristics common to other device types, including Biocompatibility/Materials, Shelf Life/Sterility, Electromagnetic Compatibility and Electrical Safety, Magnetic Resonance (MR) Compatibility, are not applicable to this device.
# 13. Discussion of Clinical Tests Performed
The effectiveness of Second Opinion® Pediatric was evaluated in a standalone performance assessment to validate the inclusion of caries lesion detection in pediatric patients 4-11 years of age.
The effectiveness of Second Opinion® Pediatric was evaluated in a standalone performance assessment to validate the CAD. The standalone retrospective study of 1182 radiographic images containing 1085 caries lesions on 549 abnormal images assessed the sensitivity of caries detection of Second Opinion® Pediatric compared to the Ground Truth.
Hellopearl.com
{9}
Pearl
The key results from this study are:
- The study met the primary endpoint, where Second Opinion® Pediatric sensitivity was > 75% for bitewing and periapical images.
- The secondary endpoints supported the conclusion of the primary endpoint.
In the 1182 images included in the standalone clinical study, there were 942 LL identified by Second Opinion® Pediatric, with a mean of 1.8 LL per image. The mean (range) Second Opinion® Pediatric rating of LLs was 0.96 (0.31, 1.00). There were 1445 NL identified by Second Opinion® Pediatric, with a mean of 2.1 NL per image. The mean (range) Second Opinion rating of NLs was 0.74 (0.31, 1.00). On normal images (no true caries identified on the image), there were 538 (1.8 per image) NL identified by Second Opinion® Pediatric for caries overall.
The z-score analysis of the lesion level sensitivity for caries detection showed the estimated lesion level sensitivity (95% CI) was 0.87 (0.84, 0.90). The test that the sensitivity was greater than 70% was statistically significant (p-value: <0.0001).
The FPPI (95% CI) was 1.22 (1.14, 1.30), wAFROC FOM (95% CI) was 0.86 (0.84, 0.88), HR-ROC FOM was 0.94 (0.93, 0.96) and the LS mean Dice Score (95% CI) was 0.76 (0.75, 0.77). All additional accuracy measures support the primary endpoint and demonstrate the high accuracy of Second Opinion® Pediatric, including the LS mean Dice score lower bound of the CI > 0.70.
The lesion level sensitivity, FPPI, HR-ROC FOM, wAFROC FOM and Dice score (segmentation accuracy) and their associated 95% CI.
The results of the standalone clinical testing of Second Opinion® Pediatric demonstrate that the caries detection performance of Second Opinion® Pediatric is comparable to that of primary predicate Second Opinion®.
# 14. Comparison to Predicate Clinical Outcomes
Predicate: Original Second Opinion
Standalone testing demonstrated that Second Opinion performed comparably to unaided readers in detecting pathologic and restoration features, based on Jaccard Index of ≥ 0.4 for lesion localization.
- Jaccard Index of 0.4 corresponds to an overlap area between the device's outputs and truth of 57% in theory. Using Jaccard Index of 0.4 leads to the device's standalone performance wAFROC-FOM 95% CI (0.73, 0.79), (0.71, 0.78), (0.78, 0.85) and (0.75, 0.84) for Caries, margin discrepancy (MD), Calculus, and periapical radiolucency (PR) with consensus truthing method, respectively.
Hellopearl.com
{10}
Pearl
- Jaccard Index of 0.5 corresponds to an overlap area between the device's outputs and truth of 67% in theory. Using Jaccard Index of 0.5 leads to the device's standalone performance wAFROC-FOM 95% CI (0.61, 0.68), (0.62, 0.68), (0.75, 0.81) and (0.69,0.78) for Caries, MD, Calculus, and PR with consensus truthing method, respectively.
The standalone sensitivity and false positive rate were also assessed:
- Sensitivity defined as the number of dental pathologies (of a given type) detected as a percentage of the same type of GT pathologies on a given slide. Range 76.39% – 89.77%
- False positive rate, defined as the number of false positive findings of a given type identified on a given slide and expressed in terms of FPPI (false positive per image).
Range 0.46 – 4.85
## 15. Conclusions
Conclusion: Based on the information presented above, Second Opinion® Pediatric and its predicate device, Second Opinion®, are deemed to have similar intended uses as devices which aid in the detection of caries that can appear in dental radiographic imagery. Second Opinion® Pediatrics' clinical trial results demonstrated that the device effectively performs as well as the predicate device.
Hellopearl.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.