Medline UNITE® REFLEX® Hybrid Nitinol Implant System
Applicant
Medline Industries, LP
Product Code
JDR · Orthopedic
Decision Date
Apr 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are intended to provide fixation for fractures, fusions or osteotomies of the bones of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle such as: First metatarsalcuneiform arthrodesis, First metatarsophalangeal arthrodesis, Talo-Navicular Fusion, Lis-Franc arthrodesis, Scarf and Chevron osteotomies. Implants are intended for single use only. The Medline UNITE® Locking and Non-Locking Cortical Screws are indicated for use with the Medline UNITE® REFLEX® Hybrid Nitinol Implant. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Device Story
Nitinol (nickel-titanium alloy) bone fixation implant system; includes implants with polyaxial locking holes and compression slots; accommodates Ø2.7mm, Ø3.5mm, and Ø4.0mm locking/non-locking cortical screws. Used in orthopedic foot and ankle surgery; implanted by surgeons. Device provides mechanical fixation for fractures, fusions, or osteotomies. Implants utilize chemical etching and passivation for surface oxidation. Provided non-sterile; requires steam sterilization at healthcare facility prior to use. System includes reusable and disposable instrumentation for implantation. Output is physical stabilization of bone segments to facilitate healing.
Clinical Evidence
No clinical data. Bench testing only. Testing included corrosion susceptibility (ASTM F2129), nickel-ion release (ASTM F3306), and galvanic corrosion (ASTM F3044). Results demonstrated nickel-ion release levels significantly below FDA tolerable intake (TI) of 35 µg/day. Engineering analysis confirmed mechanical equivalence for static bending and axial pullout strength.
Technological Characteristics
Material: Nickel titanium alloy (Nitinol) per ASTM F2063. Surface treatment: Chemical etching and passivation. Design: Polyaxial locking holes and compression slots for Ø2.7mm, Ø3.5mm, and Ø4.0mm cortical screws. Sterilization: Steam sterilization (non-sterile delivery).
Indications for Use
Indicated for patients requiring bone fixation for fractures, fusions, or osteotomies of the foot and ankle (tarsals, metatarsals, phalanges, distal tibia, fibula). Includes specific procedures: 1st metatarsal-cuneiform arthrodesis, 1st metatarsophalangeal arthrodesis, Talo-Navicular fusion, Lis-Franc arthrodesis, Scarf and Chevron osteotomies. Screws indicated for bone reconstruction, joint fusion, and fracture repair.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Medline UNITE® REFLEX® Nitinol Staple System (K231885)
Reference Devices
Medline UNITE® Ancillary Foot Recon Plating System (K234031)
Medline UNITE® MIS Foot Recon Screw System (K241359)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 11, 2025
Medline Industries, LP
Jennifer Mason
Regulatory Affairs Principal
Three Lakes Drive
Northfield, Illinois 60030
Re: K243888
Trade/Device Name: Medline UNITE® REFLEX® Hybrid Nitinol Implant System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: JDR, HWC
Dated: March 12, 2025
Received: March 12, 2025
Dear Jennifer Mason:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243888 - Jennifer Mason
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243888 - Jennifer Mason
Page 3
Sincerely,
# RYAN TROMBETTA -S
For: Limin Sun, Ph.D.
Assistant Director
DHT6A: Division of Joint
Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
K243888 Page 1 of 1
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243888
Device Name
Medline UNITE® REFLEX® Hybrid Nitinol Implant System
Indications for Use (Describe)
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are intended to provide fixation for fractures, fusions or osteotomies of the bones of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle such as: First metatarsalcuneiform arthrodesis, First metatarsophalangeal arthrodesis, Talo-Navicular Fusion, Lis-Franc arthrodesis, Scarf and Chevron osteotomies. Implants are intended for single use only.
The Medline UNITE® Locking and Non-Locking Cortical Screws are indicated for use with the Medline UNITE® REFLEX® Hybrid Nitinol Implant. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
MEDLINE
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
K243888 Page 1 of 5
# 510(k) SUMMARY
[AS REQUIRED BY 21CFR 807.92]
**Submitter / 510(k) Sponsor**
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
Registration Number: 1417592
**Contact Person**
Contact Person: Jennifer Mason, Regulatory Affairs Principal
Phone: 847-643-3652
Email: jamason@medline.com
**Summary Preparation Date**
April 10, 2025
**Type of 510(k) Submission**
Traditional
**Device Name / Classification**
Trade Name: Medline UNITE® REFLEX® Hybrid Nitinol Implant System
Common Name: Staple, Fixation, Bone
Screw, Fixation, Bone
Classification Name: Single/multiple component metallic bone fixation appliances and accessories
Smooth or threaded metallic bone fixation fastener
Product Code: JDR, HWC
Classification Panel: Orthopedic
Regulatory Class: Class II
Regulation Number: 21 CFR 888.3030, 21 CFR 888.3040
**Primary Predicate Device**
Medline UNITE® REFLEX® Nitinol Staple System
K231885
{5}
MEDLINE
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
K243888 Page 2 of 5
# Predicate Devices
Medline UNITE® Ancillary Foot Recon Plating System
K234031
Medline UNITE® MIS Foot Recon Screw System
K241359
Medline UNITE® REFLEX® Dynamic Discs
K211612
# Device Description
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are manufactured from nickel titanium alloy (nitinol). The implants utilize chemical etching and passivation to form a protective oxidation layer on the outer surface. Chemical etching and passivation are common processes to create a uniform oxidation layer on the surface of the implant. The system includes implants offered in various styles, sizes, and options; each designed for specific anatomy and procedures. The implants can accommodate Ø2.7mm, Ø3.5mm, and Ø4.0mm locking and non-locking cortical screws to be used with the polyaxial locking holes and compression slots. The system also includes reusable and disposable instrumentation necessary for implantation of the REFLEX® Hybrid Nitinol Implant.
# Indications for Use
The Medline UNITE® REFLEX® Hybrid Nitinol Implants are intended to provide fixation for fractures, fusions or osteotomies of the bones of short (tarsals) and long (metatarsals, phalanges, distal tibia and fibula) bones comprising the foot and ankle such as: First metatarsal-cuneiform arthrodesis, First metatarsophalangeal arthrodesis, Talo-Navicular Fusion, LisFranc arthrodesis, Scarf and Chevron osteotomies. Implants are intended for single use only.
The Medline UNITE® Locking and Non-Locking Cortical Screws are indicated for use with the Medline UNITE REFLEX® Hybrid Nitinol Implant. The Non-Locking Cortical Screws are also indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
# Summary of Technological Characteristics
The proposed Medline UNITE® REFLEX® Hybrid Nitinol Implant System and the primary predicate device have the same intended use.
{6}
MEDLINE
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
K243888 Page 3 of 5
The design features of the Medline UNITE® REFLEX® Hybrid Nitinol Implant System are compared to the predicate, Medline UNITE® REFLEX® Nitinol Staple System (K231885).
- Intended Use – same. Both the subject device and the primary predicate device are intended to provide fixation for fractures, fusions or osteotomies of the bones of the feet.
- Indications for Use – similar. The indications for use for the new proposed Medline UNITE® REFLEX® Hybrid Nitinol Implant System are identical to the original indications for use cleared in the primary predicate K231885. This new 510(k) includes additional indications covering locking and non-locking cortical screws used with the polyaxial locking holes and compression slots. These screws were cleared under K234031.
- Materials – same. Both the subject device and the primary predicate device are made from nickel titanium alloy (Nitinol) and meets the material specifications outlined in ASTM F2063.
- Leg Lengths – different. The primary predicate and subject devices are both offered in identical leg lengths of 16mm, 18mm, and 20mm. The new proposed device is also available in a 22mm leg length that is within the cleared range of lengths of the predicate device.
- Bridge Width – different. The primary predicate and subject devices are offered different bridge widths. The predicate device is offered in 1.5mm, 4mm, and 5mm widths, while the new proposed device is offered in 7.3mm and 10mm widths. The difference in bridge widths does not affect safety or effectiveness.
- Design features – Similar. Both the subject and primary predicate devices are offered in straight top configurations. This new proposed device also include polyaxial locking holes and compression slots that accept Ø2.7mm, Ø3.5mm, and Ø4.0mm locking and non-locking cortical screws. These additional design features do not affect safety or effectiveness.
- Sterility – same. The primary predicate and subject Medline UNITE® REFLEX® Hybrid Nitinol Implant System will be provided non-sterile and are intended to be steam sterilized at the healthcare facility prior to use.
## Summary of Non-Clinical Testing
The subject Medline UNITE® REFLEX® Hybrid Nitinol Implants does not represent a new worst-case when compared to the previously cleared Medline UNITE® REFLEX® Nitinol Staples (K231885).
{7}
MEDLINE
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
K243888 Page 4 of 5
An engineering analysis was performed to determine the subject Medline UNITE® REFLEX® Hybrid Nitinol Implants do not represent a new worst-case when compared to the previously cleared Medline UNITE® REFLEX® Nitinol Staples (K231885) for elastic static bending and axial pullout strength. The results of this analysis demonstrate the subject Medline UNITE® REFLEX® Hybrid Nitinol Implants are substantially equivalent to the predicate, Medline UNITE® REFLEX® Nitinol Staples (K231885).
To determine the worst-case staple for corrosion susceptibility representing the subject Medline UNITE® REFLEX® Hybrid Nitinol Implant, an additional engineering analysis was conducted. The results of this analysis and subsequent testing demonstrate the subject Medline UNITE® REFLEX® Hybrid Nitinol Implants are substantially equivalent to the predicate Medline UNITE® REFLEX® Nitinol Staples (K231885).
## Performance Testing (Bench)
The following tests were performed to demonstrate substantial equivalence between the proposed Medline UNITE® REFLEX® Hybrid Nitinol Implants and the predicate Medline UNITE® REFLEX® Nitinol Staples (K231885).
## Corrosion Susceptibility
Corrosion susceptibility testing was conducted per ASTM F2129 and the FDA guidance document *Technical Considerations for Non-Clinical Assessment of Medical Devices Containing Nitinol*. Testing was conducted to ensure that the proposed Medline UNITE® REFLEX® Hybrid Implants meet the predefined acceptance criteria. During testing, not all samples met the minimum acceptance criteria for electrostatic breakdown potential, so Nickel ion release testing was conducted.
## Nickel-ion Release
Nickel ion release testing was conducted per ASTM F3306 and the FDA guidance document *Technical Considerations for Non-Clinical Assessment of Medical Devices Containing Nitinol*. Testing was conducted to ensure that the proposed Medline UNITE® REFLEX® Hybrid Implants meet the acceptance criteria. The total Ni-ion release and maximum daily Ni-ion release determined for any of the three (3) samples tested over 60-days was significantly lower than the FDA guidance for the Tolerable Intake (TI) value for parenteral (non-oral) exposure to nickel of $0.5\ \mu\mathrm{g}/\mathrm{kg}/\mathrm{day}$, or $35\ \mu\mathrm{g}/\mathrm{day}$ for a $70\ \mathrm{kg}$ adult.
## Galvanic Corrosion
The Medline UNITE® REFLEX® Hybrid Nitinol Implants were tested for galvanic corrosion per ASTM F3044 “Standard Test Method for Evaluating the Potential for Galvanic Corrosion for Medical Implants”. The hybrid implants and screws representing the largest surface area ratio of the expected cathode material
{8}
MEDLINE
Medline Industries, LP
Three Lakes Drive
Northfield, IL 60093
K243888 Page 5 of 5
(titanium alloy) to the expected anode material (nitinol) were chosen for testing. Microscopic examination of all three (3) test samples revealed no pitting or indications of corrosion. The average calculated material release from mass loss, which includes both titanium alloy and nitinol materials, was significantly less than the FDA guidance for the Tolerable Intake (TI) value for parenteral (non-oral) exposure to nickel of 0.5 µg/kg/day, or 35 µg/day for a 70 kg adult.
## Performance Testing (Animal)
This section does not apply. No animal testing was performed.
## Performance Testing (Clinical)
This section does not apply. No clinical testing was performed.
## Summary of Clinical Testing
Not applicable.
## Conclusion
In accordance with 21 CFR Part 807 and based on the information provided in this premarket notification, Medline Industries, LP concludes that the Medline UNITE® REFLEX® Hybrid Nitinol Implants are as safe and as effective for their intended use as the predicate device, Medline UNITE® REFLEX® Nitinol Staple System (K231885).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.