Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses; Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses
Applicant
Pegavision Corporation
Product Code
LPL · Ophthalmic
Decision Date
Jul 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia) in aphakic or non-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit refractive astigmatism of 2.00 diopters (D) or less where the astigmatism does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D. Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic or non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D and astigmatic corrections from -0.25D to -3.50D. Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia or hyperopia) with presbyopia in aphakic or non-aphakic person with non-diseased eyes who may have +0.25D to +3.00D of ADD powers or less. The lens may be prescribed in spherical powers ranging from +6.00D to -12.25D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear. Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
Device Story
Soft hydrophilic contact lenses; Toufilcon B copolymer material; UV-blocking; visibility tint. Available in spherical, aspherical, toric, and multifocal designs. Daily disposable or frequent/planned replacement wear. Prescribed by eye care professionals for vision correction of refractive ametropia, astigmatism, and presbyopia. Cast-molded; moist heat sterilized. Patient wears lens on eye; provides refractive correction; improves visual acuity. Benefits include vision correction and UV protection.
Clinical Evidence
Prospective, randomized, double-blind, parallel, active-controlled clinical trial (n=56 enrolled, 54 completed). Compared subject lenses to Biofinity (comfilcon A). Primary endpoint: corrected visual acuity of 1.0 decimal (0.0 logMAR) or better. Results showed no difference between test and control groups. No serious adverse events reported; no slit lamp findings > Grade 2.
Technological Characteristics
Material: Toufilcon B (2-HEMA, NVP, DMA, MAA, SiGMA, PDMS macromer). Cross-linked with EGDMA and TAIC. 50% water content. UV-blocking (benzotriazole). Cast-molded. Moist heat sterilization. Oxygen permeability 91 x 10^-11. Refractive index 1.405. Standards: ISO 18369-4 (physicochemical), ISO 18369-3 (transmittance), ASTM D1708-13 (mechanical), ISO 10993 (biocompatibility).
Indications for Use
Indicated for daily wear vision correction of refractive ametropia (myopia/hyperopia) in aphakic or non-aphakic persons with non-diseased eyes. Includes toric designs for astigmatism up to 2.00D (or -0.25D to -3.50D cylinder) and multifocal designs for presbyopia (+0.25D to +3.00D ADD).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Biofinity (Comfilcon A) Soft Contact Lens (K052560)
Reference Devices
Pegavision (Hioxifilcon A) Daily Disposable Soft Contact Lenses (K232649)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2025
Pegavision Corporation
Estela Lin
Official Correspondent
2F-1, No. 5, Shing Yeh St., Guishan Dist
Taoyuan City, 33341
Taiwan
Re: K243868
Trade/Device Name: Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses; Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses
Regulation Number: 21 CFR 886.5925
Regulation Name: Soft (Hydrophilic) Contact Lens
Regulatory Class: Class II
Product Code: LPL, MVN
Dated: November 25, 2024
Received: December 17, 2024
Dear Estela Lin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243868 - Estela Lin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243868 - Estela Lin
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ELISSA
WONG-S
Digitally signed
by ELISSA
WONG-S
for Angelo Green
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243868
Device Name
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses;
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses
Indications for Use (Describe)
## Sphere/ Asphere
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia) in aphakic or non-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit refractive astigmatism of 2.00 diopters (D) or less where the astigmatism does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D.
## Toric
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic or non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D and astigmatic corrections from -0.25D to -3.50D.
## Multifocal
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia or hyperopia) with presbyopia in aphakic or non-aphakic person with non-diseased eyes who may have +0.25D to +3.00D of ADD powers or less. The lens may be prescribed in spherical powers ranging from +6.00D to -12.25D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear.
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
K243868
# 510(k) SUMMARY
The following 510(K) Summary is being submitted as required by 21CFR 807.92(a).
## Submitter Information
| Company: | PEGAVISION CORPORATION
2F-1 No.5, Shing Yeh St., Guishan Dist., Taoyuan City 333, Taiwan |
| --- | --- |
| Contact Person: | Estela Lin, Regulatory Affair Engineer Supervisor |
| Phone: | 886-3-329-8808 |
| Fax: | 886-3-329-8897 |
| E-Mail: | EstelaLin@pegavision.com |
| Date Prepared: | June 6, 2025 |
## Identification of Device
| Trade Name: | Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic)
Contact Lenses
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses |
| --- | --- |
| Common Name: | Soft (hydrophilic) Contact Lenses (daily wear) |
| Classification Name: | Lenses, Soft Contact, Daily Wear 21CFR. 886.5925, Product Code LPL
Lens. Soft Contact (Disposable). 21CFR. 886.5925, Product Code MVN |
| FDA Classification: | Class II |
| Predicate Device: | K052560, Biofinity (Comfilcon A) Soft Contact Lens |
## Description of Device
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are clear and visibility tint with UV blocker are available as a spherical lens. The lens material, Toufilcon B is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA), N-Vinyl-2-Pyrrolidinone (NVP), N,N-Dimethylcarylamide (DMA), Methacrylic Acid (MAA), (3-Methacryloxy-2-hydroxypropoxy)propyl-bis(trimethylsiloxy)methylsilane (SiGMA) and Polydimethylsiloxane macromer (monofunctional Polydimethylsiloxane) (PDMS macromer), cross-linked with Ethylene glycol dimethacrylate (EGDMA) and
Page 1 of 7
{6}
Triallyl isocyanurate (TAIC) via photo-polymerization. The copolymer consists 50% Toufilcon B and 50% water by weight when immersed in buffered borate solution. The lens further contains a benzotriazole UV absorbing monomer and thus is able to block UV radiation. The lens is visibly tined with "Reactive Blue19" color additive, 21 CFR part 73.3121. The UV Blocking averages 95% in the UVB range of 280 nm to 315 nm and 50% in the UVA range of 315 nm to 380 nm. The Toufilcon B name has been adopted by the United States Adopted Names Council (USAN).
## Indications for use
### Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses
#### Sphere and Asphere
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia) in aphakic or non-aphakic persons with non-diseased eyes. The lenses may be worn by person who exhibit refractive astigmatism of 2.00 diopters (D) or less where the astigmatism does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D.
#### Toric
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Toric designs are indicated for daily wear for vision correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic or non-aphakic persons with non-diseased eyes. The lens may be prescribed in spherical powers ranging from +6.00D to -12.00D and astigmatic corrections from -0.25D to -3.50D.
#### Multifocal
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses with Multifocal designs are indicated for daily wear for the correction of refractive ametropia (myopia or hyperopia) with presbyopia in aphakic or non-aphakic person with non-diseased eyes who may have +0.25D to +3.00D of ADD powers or less. The
Page 2 of 7
{7}
lens may be prescribed in spherical powers ranging from +6.00D to -12.25D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear.
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfecting and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
## Technological characteristics studies
The technological characteristics of the Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses, as compared to the predicate devices, are summarized in the following table.
| Device Name/ Subgroups | Subject Device | Predicate Device |
| --- | --- | --- |
| | PEGAVISION (Toufilcon B) Soft (Hydrophilic) Contact Lenses
PEGAVISION (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses | Biofinity (Comfilcon A) Soft Contact Lens |
| 510(k) Number | New Device | K052560 |
| USAN name | Toufilcon B | Comfilcon A |
| FDA Category (Group) | Group 5 | Group 5 |
| Intended use | Spherical and Aspherical, Toric, Multifocal | Spherical and Aspherical, Toric, Multifocal |
| Expected clinical performance | Eye for vision correction | Eye for vision correction |
| Indications | Daily wear | Daily wear |
Page 3 of 7
{8}
| Device Name/ Subgroups | Subject Device | Predicate Device |
| --- | --- | --- |
| | PEGAVISION (Toufilcon B) Soft (Hydrophilic) Contact Lenses
PEGAVISION (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses | Biofinity (Comfilcon A) Soft Contact Lens |
| Replacement schedule | Daily/Bi-weekly/Monthly | Monthly |
| Production Method | Cast-molding | Cast-molding |
| Sterilization Method | Moist Heat | Moist Heat |
| Lens color | Light blue | Light blue |
| UV-blocking | Class II | NA |
| Water Content | 50%±2% | 48%±2% |
| Oxygen Permeability x10-11 (cm2/s) [ mlO2/(ml × mmHg) ] | 91 | 128 |
| Light Transmittance | 95±5% | >97% |
| Powers | +6.00D ~ -12.25D | +20.00D ~ -20.00D |
| Cylinder Power | -0.25D ~ -3.50D | -0.25D ~ -10.00D |
| ADD Power | +0.25D ~ +3.00D | +0.50D ~ +3.00D |
| Refractive Index | 1.405 | 1.400 |
## Summary of Clinical Study
A prospective, randomized, double-blind, parallel, active-controlled clinical trial with 3 months of treatment follow-up was conducted in Taiwan. The study screened 75 subjects, 10 subjects failed to be included and 9 subjects randomized but unsuccessful lens fitting. A total of 56 subjects were enrolled in this study (39 subjects wore the test lenses and 17 subjects wore control lenses "CooperVison Biofinity® (comfilcon A) Soft Contact Lens". A total 54 subjects completed the study, while 2 subjects in test group were discontinued from the study. The mean treatment compliance was high (98.81% ~ 98.77%) during the study period. Regarding the primary efficacy endpoint, the subjects with corrected contact lens visual acuity of 1.0 decimal (0.0 logMAR) or better at final visit shows no difference in control group and test group. The test and control groups are comparable in corrected contact lens visual acuity. In safety endpoints, there was no serious and significant adverse device event occurred during
Page 4 of 7
{9}
the conduct of the study in both control and test group. No subject with adverse device effects accompanying Slit Lamp Findings > Grade 2 was reported in the control group and test group. Therefore, this clinical investigation supports the claim of substantial equivalence between the test product and control product with regard to clinical effectiveness and safety.
## Non-clinical Study
All tests were conducted in accordance with the May 1994 FDA guidance title Premarket Notification 510(K) Guidance Document for Class II Contact Lenses. The non-clinical performance tests had been performed to demonstrate the safety and effectiveness of Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses.
Non-Clinical testing performed includes:
- Physicochemical Properties
- Refractive Index (ISO 18369-4)
- Oxygen Permeability (ISO 18369-4)
- Water content (ISO 18369-4)
- Extractable (ISO 18369-4)
- Visible & UV Light Transmittance (ISO 18369-3)
- Mechanical Property (ASTM D1708-13)
- Solution Compatibility test (ISO 11981)
- Preservative Uptake and Release (ISO 11986)
- Shelf Life Test and Sterility Test (ISO 11987)
The results demonstrated that the lens met all established specifications and requirements for physical, optical, and chemical properties.
- Biocompatibility Test
The biocompatibility evaluation of this product was conducted in accordance with the ISO 10993 series and ISO 9394 standards for the biological evaluation of medical devices. The evaluation included relevant tests for both the contact lens materials and the packaging components:
Page 5 of 7
{10}
| Contact Lenses (Toufilcon B) |
| --- |
| • Cytotoxicity Test (ISO 10993-5) • Ocular Irritation Test (ISO 10993-10) • Acute Systemic Toxicity Test (ISO 10993-11) • Skin Sensitization Study (Maximization Test) (ISO 10993-10) • 22-Day Ocular Irritation Study (ISO 9394) |
| PP blister and aluminum foil (identical to K232649) |
| • Cytotoxicity Test (ISO 10993-5) • Ocular Irritation Test (ISO 10993-10) • Acute Systemic Toxicity Test (ISO 10993-11) |
Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses use the same primary packaging materials as the K232649 Pegavision (Hioxifilcon A) Daily Disposable Soft Contact Lenses. The raw materials and formulation of the primary packaging materials (Aluminum Foil 4-Layer and PP blister) are identical. The safety of the these materials has been verified through biocompatibility testing performed for the referenced device (K232649) in accordance with the ISO 10993 series of standards; therefore, additional testing for primary packaging was not required.
The test results demonstrated that the contact lens exhibit no cytotoxicity, no ocular irritation, no skin sensitization, and no acute systemic toxicity. All biocompatibility tests yielded passing results, confirming that the product does not contain any toxic or harmful substances that may pose a risk to biological systems.
## Substantial Equivalence Statement
Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses are substantially equivalent to the predicate device BIOFINITY (comfilcon A) Soft Contact Lens.
All of these lenses are categorized as Group V silicone hydrogel lenses and are manufactured using the same processes: cast molding and steam sterilization. Designed for vision correction, they are available in various optical designs, including spherical/aspherical, toric, and multifocal. The Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision Soft (Hydrophilic)
Page 6 of 7
{11}
Contact Lenses are considered substantially equivalent to the predicate device, CooperVision Biofinity® (comfilcon A) Soft Contact Lens.
## Conclusion
A series of pre-clinical tests and clinical trial were performed to demonstrate the safety and effectiveness of the Pegavision (Toufilcon B) Daily Disposable Soft (Hydrophilic) Contact Lenses and Pegavision (Toufilcon B) Soft (Hydrophilic) Contact Lenses. It is concluded that the lenses are as safe, as effective and perform as well as the predicate devices.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.