The butterflyBVM™ when used in transport and non-clinical emergency settings (e.g., EMS, non-hospital) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates. The butterflyBVM™ when used in professional healthcare facilities (e.g., hospitals) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, and infants.
Device Story
Manual emergency bag-valve-mask resuscitator; provides pulmonary resuscitation via manual compression of side arms. Input: manual operator force. Transformation: bellows-based mechanism with adjustable tidal volume (Vt) and peak inspiratory pressure (PIP) dials. Output: controlled ventilation to patient airway. Used in EMS, transport, and hospital settings by professional healthcare providers. Device features patient connection port for masks/ET tubes and exhalation port for filters/PEEP valves. Adjustable dials allow single-device configuration for all patient populations (neonates to adults). Benefits include precise pressure limitation to prevent barotrauma and simplified inventory management compared to multi-size manual resuscitators.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 10651-4 and ISO 5356-1. Comparative study (Merrell et al. 2023) tested PIP dial precision against Ambu Spur II with disposable pressure manometer; butterflyBVM demonstrated higher accuracy in achieving target pressures (0 failures at 40cmH2O vs 21 failures for predicate). Biocompatibility testing performed per ISO 10993 and ISO 18562 series.
Technological Characteristics
Manual bellows-based resuscitator; adjustable Vt (30-715ml) and PIP (15-60cmH2O) dials. Materials evaluated per ISO 10993 and ISO 18562. Conical connectors per ISO 5356-1. Operating temp: -18°C to +50°C. MR Conditional. Single-use, non-sterile.
Indications for Use
Indicated for pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates in transport, non-clinical emergency, and professional healthcare settings.
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 25, 2025
Compact Medical, Inc.
% Allison Komiyama
VP, Medtech Innovations
Rqm+
2790 Mosside Blvd.
Suite 800
Monroeville, Pennsylvania 15146
Re: K243861
Trade/Device Name: butterflyBVM
Regulation Number: 21 CFR 868.5915
Regulation Name: Manual Emergency Ventilator
Regulatory Class: Class II
Product Code: BTM
Dated: December 12, 2024
Received: December 16, 2024
Dear Allison Komiyama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243861 - Allison Komiyama
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243861 - Allison Komiyama
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
John S. Bender -S
2025.04.25 15:14:01 -04'00'
for Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243861 | |
| Device Name butterflyBVM™ | |
| Indications for Use (Describe) | |
| The butterflyBVM™ when used in transport and non-clinical emergency settings (e.g., EMS, non-hospital) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates. | |
| The butterflyBVM™ when used in professional healthcare facilities (e.g., hospitals) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, and infants. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
butterflyBVM
# 510(k) SUMMARY K243861
## DATE PREPARED
April 25, 2025
## MANUFACTURER AND 510(k) OWNER
Compact Medical
525 South Meridian Street, Suite 2D2
Indianapolis, IN 46259
Telephone: +1 757-677-6422
Official Contact: Dr. Jonathan Merrell, M.D., CEO
Adam Scott, COO
## REPRESENTATIVE/CONSULTANT
Allison Komiyama, Ph.D., RAC
Joy Gutermuth
RQM+
Telephone: +1 (412) 816-8253
Email: akomiyama@rqmplus.com
jgutermuth@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: butterflyBVM™
Common Name: Manual Emergency Ventilator
Regulation Number: 21 CFR 868.5915
Class: II
Product Code: BTM
Premarket Review: Division of Anesthesia, Respiratory, and Sleep Devices (DHT1C)
Review Panel: Anesthesiology
## PREDICATE DEVICE IDENTIFICATION
The butterflyBVM™ is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
| --- | --- | --- |
| K152931 | Spur II/Ambu | ✓ |
## DEVICE DESCRIPTION
The butterflyBVM™ is a three-in-one bag-valve-mask capable of resuscitating adults, adolescents, children, infants, and neonates. The maximum tidal volumes (Vt) delivered by the butterflyBVM™ can be set via the tidal volume dial, as needed, to ranges that are generally appropriate for the size of patient receiving care. The butterflyBVM™ also has an adjustable peak inspiratory pressure (PIP) dial to help prevent barotrauma. The tidal volume and PIP are
{5}
butterflyBVM
selected by aligning the dials to the Vt/PIP indicator. Patient-facing accessories such as masks, laryngeal mask airways, endotracheal tubes, end-tidal $\mathrm{CO}_{2}$ samplers, and the like can be connected to the patient connection port. Exhalation accessories such as bio filters and PEEP valves can be connected to the exhalation port. Supplemental oxygen can be added via the oxygen inlet port. The device is operated by manually squeezing the side arms together with either one or two hands. A cover is provided over the PIP dial to provide instruction to the user via images and words to first set the tidal volume for the procedure based on the assessment of the patient.
## INDICATIONS FOR USE
The butterflyBVM™ when used in transport and non-clinical emergency settings (e.g., EMS, non-hospital) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates.
The butterflyBVM™ when used in professional healthcare facilities (e.g., hospitals) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, and infants.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Compact Medical believes that the butterflyBVM™ is substantially equivalent to the predicate devices based on the information summarized here:
{6}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Device Image | | | |
| Indications for Use | The butterflyBVM™ when used in transport and non-clinical emergency settings (e.g., EMS, non-hospital) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, infants, and neonates. The butterflyBVM™ when used in professional healthcare facilities (e.g., hospitals) is a single-use resuscitator that may be manipulated to provide pulmonary resuscitation of patients including adults, adolescents, children, and infants. | The Ambu SPUR II resuscitator is a single patient use resuscitator intended for pulmonary resuscitation. The range of the application for each version is: Adult: Adults and children with a body weight more than 30kg (66lb). Pediatric: Infant and children with a body weight up to 30kg (66lb). Infant: Neonates and infant with a body weight up to 10kg (22lb). | Similar. The subject device is offered in a single configuration which is manipulated for each patient while the predicate device comes in three sizes for each population. Both devices are intended to be used on the same spread of patient populations. |
| Product Codes / Regulation Number | BTM / 21 CFR 868.5915 | BTM / 21 CFR 868.5915 | Same |
| Regulation Description | Manual Emergency Ventilator | Manual Emergency Ventilator | Same |
{7}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Patient Population | Adults, adolescents, children, infants, and neonates. | Adults, adolescents, children, infants, and neonates. | Same |
| Environment of Use | Professional healthcare facilities and transport environments. | Professional healthcare facilities and transport environments. | Same |
| Single Patient Use | Single patient use | Single patient use | Same |
| Sterilization | Non-sterile | Non-sterile | Same |
| Prescription | Rx Only | Rx Only | Same |
| Magnetic Resonance Compatibility | MR Conditional | MR Conditional | Same |
| Performance Standards Utilized | • ISO 10651-4 Lung ventilators Part 4: Particular requirements for user-powered resuscitators • ISO 5356-1 Anaesthetic and respiratory equipment — Conical connectors | • ISO 10651-4 Lung ventilators Part 4: Particular requirements for user-powered resuscitators • ISO 5356-1 Anaesthetic and respiratory equipment — Conical connectors | Same |
{8}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Biocompatibility | • ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity • ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization • ISO 10993-11 Biological evaluation of medical devices - Part 11: Test for systems toxicity • ISO 10993-23 Biological evaluation of medical devices - Part 23: Tests for irritation • ISO 18562-1 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process • ISO 18562-2 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter • ISO 18562-3 Biocompatibility evaluation of breathing gas pathways in healthcare application - Part 3: Tests for emissions of volatile organic compounds • ISO 18562-4 Biocompatibility evaluation of breathing gas pathways in healthcare application - Part 4: Tests for leachables in condensate | • ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity • ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Similar. The subject device has been evaluated to demonstrate biocompatibility to more endpoints than the predicate. |
| Shelf Life | • 6 months | • Not stated | Similar. Difference in shelf life does not raise questions of safety or effectiveness. |
{9}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Principles of Operation | A manual emergency bag-valve-mask ventilator incorporating a bellows and valve, intended to provide emergency respiratory support to a patient's airway. A single device design is adjustable for each patient population. | A manual emergency bag-valve-mask ventilator incorporating a bag and valve, intended to provide emergency respiratory support to a patient's airway. Three sizes of device are available for three different patient populations. | Similar. The subject device is offered in a single configuration and is manipulated to be used on all intended patient populations while the predicate is provided in three separate sizes. |
| Pressure Limiting Valve (PIP) | Default: 40cmH2O Max: No pressure limiting (override) In a study of the butterflyBVM versus the Ambu Spur II [Merrell et al. 2023], the use of the PIP Dial on the butterflyBVM was tested against the Ambu Disposable Pressure Manometer in achieving certain target pressures. The default pressure mechanisms for each device were tested. Results: Target Max PIP: 40cmH2O butterflyBVM: 0 failures with 40cmH2O max PIP delivered | Infant and Pediatric: Default: 40cmH20 Max: No pressure limiting (override) Adult: Default: No pressure limiting (override) Results: Target Max PIP: 40cmH2O Ambu Spur II: 21 failures with 44.0cmH2O max PIP delivered | Similar. The subject device features a PIP dial rather than a pop-off valve. The subject device has the ability to be set to a lower minimum pressure; however, the default settings and override capability are the same between the subject and predicate, and the subject has demonstrated superiority to the predicate in venting pressures above the default setting. |
{10}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Delivered Volume One Hand | Min: 30 ml Max: 565 ml | Min (smallest size device): 150 ml Max (largest size device): 600 ml | Similar. The subject device is capable of delivering a smaller minimum value than the smallest size of the predicate. |
| Delivered Volume Two Hands | Min: 70 ml Max: 715 ml | Min (smallest size device): not reported Max (largest size device): 1000 ml | Similar. The predicate device does not state a two-hand delivered volume for the smallest device. The maximum volumes are similar. |
| Operating Temperature | -18°C to +50°C | -18°C to +50°C | Same |
| Storage Conditions | -40°C to +60°C | -40 °C to + 60 °C | Same |
{11}
butterflyBVM
| | Subject Device | | | | | Predicate Device | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | | | | | Ambu, Inc. Spur II K152931 | | | | | Substantial Equivalence Discussion |
| Delivered Oxygen Concentration | Setting | Inspiratory Rate (BPM) | Stroke Volume* (ml) | Nominal Oxygen Concentration (%) at a Given O2 Flow Rate | | O2 (l/min) | VT (ml) x f (pr. min.), l:E ratio = 1:2 | | 250 x 12 | 600 x 12 | 750 x 12 |
| | Adult Large | 10 | 565 | 81 | 78 | 90 | 2 | 74 | 43 | 38 | 34 |
| | Adult Med | 10 | 510 | 62 | 79 | 91 | 5 | 100 | 76 | 65 | 54 |
| | Adult Small | 10 | 415 | 66 | 83 | 93 | 10 | 100 | 100 | 100 | 87 |
| | 10-13 YRS | 20 | 380 | 62 | 79 | 90 | 15 | 100 | 100 | 100 | 100 |
| | 7-9 YRS | 20 | 275 | 72 | 87 | 95 | | | | | |
| | 5-6 YRS | 20 | 225 | 76 | 90 | 97 | | | | | |
| | 3-4 YRS | 20 | 185 | 81 | 93 | 99 | | | | | |
| | 1-24 MOS | 20 | 150 | 83 | 95 | 99 | | | | | |
| | NEO | 30 | 30 | 93 | 100 | 100 | | | | | |
| | Table 2: Expected oxygen concentration at a given O2 flow rate, inspiratory rate, and stroke volume. *Tested according to EN ISO 10651-4. | | | | | O2 (l/min) | 40 x 40 | 100 x 20 | 200 x 20 | 400 x 15 | |
| | | | | | | 1 | 70 | 60 | 40 | 34 | |
| | | | | | | 2 | 100 | 100 | 60 | 47 | |
| | | | | | | 4 | 100 | 100 | 100 | 73 | |
| | | | | | | 6 | 100 | 100 | 100 | 100 | |
| | | | | | | | | | | | |
| | | | | | | VT (ml) x f (pr. min.), l:E ratio = 1:1 | | | | | |
| | | | | | | O2 (l/min) | 40 x 40 | 100 x 20 | 150 x 20 | | |
| | | | | | | Reservoir Bag | 10" Tube | Reservoir Bag | 10" Tube | Reservoir Bag | 10" Tube |
| | | | | | | 1 | 70 | 70 | 60 | 60 | 47 |
| | | | | | 2 | 100 | 100 | 100 | 100 | 73 | |
| | | | | | 4 | 100 | 100 | 100 | 100 | 100 | |
| | | | | | 6 | 100 | 100 | 100 | 100 | 100 | |
| | | | | | Calculated delivered O2 %* | | | | | | |
{12}
butterflyBVM
| | Subject Device | Predicate Device | |
| --- | --- | --- | --- |
| | Compact Medical ButterflyBVM™ K243861 | Ambu, Inc. Spur II K152931 | Substantial Equivalence Discussion |
| Comparative Testing | butterflyBVM PIP Dial In a study of the butterflyBVM versus the Ambu Spur II [Merrell et al. 2023], the use of the PIP Dial on the butterflyBVM was tested against the Ambu Disposable Pressure Manometer in achieving certain target pressures. Results: Target Max PIP: ≤20cmH2O butterflyBVM: 1 failure with 20.3cmH2O max PIP delivered Target Max PIP: ≤30cmH2O butterflyBVM: 1 failure with 30.4cmH2O max PIP delivered | Ambu Disposable Pressure Manometer On Ambu Spur II pediatric and infant devices, the default pop off value is 40cmH2O. Results: Target Max PIP: ≤20cmH2O Ambu Spur II: 23 failures with 23.7cmH2O max PIP delivered Target Max PIP: ≤30cmH2O Ambu Spur II: 32 failures with 37.9cmH2O max PIP delivered | Similar. The Ambu Spur II Disposable Pressure Manometer is capable of measuring pressures between 5 and 60cmH2O whereas the butterflyBVM is capable of restricting the pressures between 15 and 60cmH2O. Comparative testing demonstrates that the butterflyBVM showcases repeatability and precision in achieving target pressures thereby demonstrating substantial equivalence. |
{13}
butterflyBVM
The butterflyBVM™ and the Spur II have the same intended use and similar indications, technological characteristics and similar principles of operation. The main technological differences between the subject device and its predicate are the design of the bellows, the variable tidal volume dial, the adjustable peak inspiratory pressure dial, and biocompatibility endpoints. The gas pathway of the subject device has been evaluated per the ISO 18562 series of standards. The minor differences do not present any new issues of safety or effectiveness because the materials of the subject device have passed biocompatibility testing according to rigorous endpoints and the bench testing has shown that despite design differences, the devices perform comparably and according to recognized consensus standards. Thus, the butterflyBVM™ is substantially equivalent to the Spur II.
## SUMMARY OF NON-CLINICAL TESTING
FDA-recognized performance standards have been established for the butterflyBVM™. The following tests were performed to demonstrate safety based on current industry standards:
- ISO 10651-4 Lung ventilators – Part 4: Particular requirements for operator-powered resuscitators
- ISO 5356-1 Anaesthetic and respiratory equipment – Conical connectors – Part 1: Cones and sockets
## Result: All tests passed.
## Connectors
- Patient connection port connector
- Expiratory port connector for breathing gases
- Oxygen tube connector
## Operation requirements
- Patient valve function after contamination with vomitus
- Mechanical strength /Drop test
- Immersion in water
- Minimum number of cycles
## Ventilatory requirements
- Supplementary oxygen and delivered oxygen concentration
- Expiratory resistance
- Inspiratory resistance
- Patient valve malfunction
- Resuscitator deadspace and rebreathing
## Ventilation performance
- Minimum delivered volume
- Pressure limitation
{14}
butterflyBVM
## Storage and Operation
- Inspiratory/Expiratory resistance test
- Stroke volume test
- Patient valve malfunction test
- Supplementary oxygen and delivery oxygen concentration
- Pressure limitation test
The results of these tests indicate that the butterflyBVM™ is substantially equivalent to the predicate device.
## CONCLUSION
Based on the testing, including performance testing per recognized consensus standards and biocompatibility, it can be concluded that the subject device does not raise new questions of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed butterflyBVM™ are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.