K243860 · AtriCure, Inc. · PZX · Jan 15, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243860
Device Name
AtriClip PRO-Mini LAA Exclusion System (PROM)
Applicant
AtriCure, Inc.
Product Code
PZX · General, Plastic Surgery
Decision Date
Jan 15, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AtriClip LAA Exclusion System is indicated for the exclusion of the heat's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
Device Story
AtriClip PRO-Mini LAA Exclusion System is a sterile, single-use, implantable device for LAA exclusion. System comprises a self-closing AtriClip Mini preloaded on a manual clip applier. Surgeon operates device during cardiac surgery; uses handle with articulation knobs to position end effector; deploys clip to apply uniform pressure across LAA length. Clip remains as permanent implant to achieve electrical isolation of LAA. Device facilitates secure LAA exclusion; benefits patient by reducing risk of thromboembolic events associated with LAA. No electronic or software components.
Clinical Evidence
Bench testing only. Testing included mechanical, reliability, tissue, and dose audit testing. Results met predetermined acceptance criteria, demonstrating performance equivalent to predicate and reference devices.
Indicated for patients undergoing cardiac surgical procedures requiring exclusion of the left atrial appendage (LAA) under direct visualization.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
AtriClip FLEX-Mini LAA Exclusion System (ACHM) (K234125)
Reference Devices
AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (PRO2) (K160454, K172742)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 15, 2025
AtriCure, Inc. Diane Sung Senior Regulatory Affairs Specialist 7555 Innovation Wav Mason, Ohio 45040
Re: K243860
Trade/Device Name: AtriClip PRO-Mini LAA Exclusion System (PROM) Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: PZX Dated: December 16, 2024 Received: December 16, 2024
Dear Diane Sung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
# Sincerely, Kalkidan Molla -S
Kalkidan A. Molla, MS Acting Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243860
Device Name AtriClip PRO-Mini LAA Exclusion System (PROM)
Indications for Use (Describe)
The AtriClip LAA Exclusion System is indicated for the exclusion of the heat's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Applicant Information
| Manufacturer: | AtriCure, Inc.<br>7555 Innovation Way<br>Mason Ohio 45040 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Diane Sung<br>Senior Regulatory Affairs Specialist<br>P: 513-600-1804 |
| Alternate Contact: | Jim Taufen<br>Director, Regulatory Affairs<br>P: 952-314-1507 |
| Date Prepared: | December 16, 2024 |
| Device Information | |
| Proprietary Name: | AtriClip PRO-Mini LAA Exclusion System (PROM) |
| Common Name: | Implantable Clip and Clip Applier |
| Classification: | Implantable Clip and Clip Applier<br>Regulatory Class: Class II; per 21 CFR 878.4300<br>Product Code: PZX<br>Classification Panel: Cardiovascular |
| Predicate Device: | AtriClip FLEX-Mini LAA Exclusion System (ACHM) (K234125, July 29, 2024) |
| Reference Device: | AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip<br>(PRO2) (K160454, March 18, 2016, and K172742, October 11, 2017) |
## Device Description
The AtriClip PRO-Mini LAA Exclusion System consists of a single use, sterile, self-closing, implantable Clip (AtriClip Mini) preloaded on a Single Use Clip Applier along with a Selection Guide. When closed, the Clip applies uniform pressure over the length of the Clip to ensure consistent, reproducible, and secure exclusion of the left atrial appendage (LAA). The Clip is then deployed and is left as a permanent implant, and excludes the LAA, resulting in electrical isolation of the LAA. The Clip is available in the following lengths to accommodate different sizes of LAA: 35 mm, 40 mm, 45 mm, and 50 mm. The PRO-Mini device is a disposable device with a handle, leftright and up/down articulation knobs, articulation lock, deployment tab, shaft, suture anchors, and end effector containing the AtriClip Mini.
{5}------------------------------------------------
## Intended Use / Indications for Use
The AtriClip LAA Exclusion System is indicated for the heart's left atrial appendage, performed under direct visualization and in conjunction with other cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, with or without assistance from a camera, endoscope, etc., or other appropriate viewing technologies.
#### Comparison of Technological Characteristics
- The devices include the same intended use and indications for use, and; ●
- . No changes were made in operating principle, and;
- The results of the verification and validation testing: .
- o Demonstrated equivalency in performance
- O Did not raise any new issues of safety
| # | Feature | AtriClip FLEX-Mini<br>LAA Exclusion<br>System with AtriClip<br>Mini<br>(ACHM)<br>(Predicate)<br>K234125 | AtriClip LAA<br>Exclusion System<br>with Preloaded<br>Gillinov-Cosgrove<br>LAA Clip (PRO2)<br>(Reference)<br>K160454, K172742 | PRO-Mini<br>(PROM)<br>(Proposed) | Equivalence<br>Comparison |
|--------------|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| 1 | Proprietary<br>Name | AtriClip FLEX-Mini LAA<br>Exclusion System | AtriClip LAA Exclusion<br>System with Preloaded<br>Gillinov-Cosgrove Clip | AtriClip PRO-Mini LAA<br>Exclusion System | N/A |
| 2 | Product<br>Code(s) | ACHM35 ACHM40<br>ACHM45 ACHM50 | PRO235 PRO240<br>PRO245 PRO250 | PROM35 PROM40<br>PROM45 PROM50 | N/A |
| 3 | Indications for<br>Use | The AtriClip LAA<br>Exclusion System is<br>indicated for the<br>exclusion of the heart's<br>left atrial appendage,<br>performed under direct<br>visualization and in<br>conjunction with other<br>cardiac surgical<br>procedures.<br>Direct visualization, in<br>this context, requires<br>that the surgeon is able<br>to see the heart directly, | The AtriClip LAA<br>Exclusion System is<br>indicated for the<br>exclusion of the heart's<br>left atrial appendage,<br>performed under direct<br>visualization and in<br>conjunction with other<br>cardiac surgical<br>procedures.<br>Direct visualization, in<br>this context, requires<br>that the surgeon is able<br>to see the heart directly, | The AtriClip LAA<br>Exclusion System is<br>indicated for the<br>exclusion of the heart's<br>left atrial appendage,<br>performed under direct<br>visualization and in<br>conjunction with other<br>cardiac surgical<br>procedures.<br>Direct visualization, in<br>this context, requires<br>that the surgeon is able<br>to see the heart directly, | Same as predicate<br>and reference<br>devices. |
| | | with or without<br>assistance from a<br>camera, endoscope,<br>etc., or other<br>appropriate viewing<br>technologies. | with or without<br>assistance from a<br>camera, endoscope,<br>etc., or other<br>appropriate viewing<br>technologies. | with or without<br>assistance from a<br>camera, endoscope,<br>etc., or other<br>appropriate viewing<br>technologies. | |
| 4 | Packaging | Sterile, Single use,<br>disposable device | Sterile, Single use,<br>disposable device | Sterile, Single use,<br>disposable device | Same as predicate<br>and reference<br>devices. |
| 5 | Biocompatibility | Materials have been<br>assessed based on ISO<br>10993 and are<br>commonly employed in<br>tissue contacting<br>devices. | Materials have been<br>assessed based on ISO<br>10993-1 and are<br>commonly employed in<br>tissue contacting<br>devices. | Materials have been<br>assessed based on ISO<br>10993-1 and are<br>commonly employed in<br>tissue contacting<br>devices. | Same as predicate<br>and reference<br>devices. |
| 6 | Sterilization | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation | Same as predicate<br>and reference<br>devices. |
| 7 | Sterility<br>Assurance<br>Level | $10^{-6}$ SAL | $10^{-6}$ SAL | $10^{-6}$ SAL | Same as predicate<br>and reference<br>devices. |
| Clip | | | | | |
| 8 | Preloaded Clip | AtriClip Mini (AOD3) | N/A | AtriClip Mini (AOD3) | Same as predicate<br>device. |
| 9 | Operating<br>Mechanism | Uniform pressure along<br>length of the Clip | N/A | Uniform pressure along<br>length of the Clip | Same as predicate<br>device. |
| 10 | Contact<br>Surface | Knit Braided Polyester | N/A | Knit Braided Polyester | Same as predicate<br>device. |
| Clip Applier | | | | | |
| 11 | Deployment<br>End Effector | N/A | Stainless steel jaws that<br>open and close, as well<br>as provide the pathway<br>for stainless steel<br>cables to route and<br>attach to the Clip. | Stainless steel jaws that<br>open and close, as well<br>as provide the pathway<br>for stainless steel<br>cables to route and<br>attach to the Clip. | Same as<br>reference device. |
| 12 | Shaft | N/A | Stainless Steel | Stainless Steel | Same as<br>reference device. |
| 13 | Handle | N/A | Polycarbonate | Polycarbonate | Same as<br>reference device. |
{6}------------------------------------------------
{7}------------------------------------------------
## Performance Data
The following bench testing was conducted for design and performance elements deemed appropriate to demonstrate equivalence to the previously cleared AtriClip LAA Exclusion System devices (ACHM, PRO2). The AtriClip PRO-Mini LAA Exclusion System (PROM) met the predetermined acceptance criteria ensuring substantial equivalence to the previously cleared AtriClip LAA Exclusion System devices (ACHM, PRO2). No new safety or performance issues were raised during testing.
#### Non-clinical Bench Testing:
- Mechanical Testing
- . Reliability Testing
- Tissue Testing ●
- Dose Audit Testing .
## Conclusions
AtriCure has demonstrated that the AtriClip PRO-Mini LAA Exclusion System (PROM) is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use / indication for use as the previously cleared AtriClip FLEX-Mini LAA Exclusion System (ACHM) per K234125 and AtriClip LAA Exclusion System with Preloaded Gillinov-Cosgrove Clip (PRO2) per K160454 and K172742.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.