K243819 · Apr 9, 2025
Luminopia
Luminopia, Inc.
Medical health records; Real-world registry
The registry study was used to evaluate the effectiveness and safety of the device in a broader pediatric population (aged 4 to <13) in a real-world setting, supporting the label expansion from the previously cleared age range (4-7).
Real-world registry; Medical health records; Amblyopia; Label expansion; Pediatric
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Real-World Registry study; Prospective registry collecting retrospective medical health record data; Follow-up/Duration: Average of approximately 8.2 months of treatment
Pediatric patients aged 4 to <13 with amblyopia associated with anisometropia and/or strabismus; Sample Size: 290
Not applicable for this study
Change in best-corrected visual acuity (BCVA) in the amblyopic eye; Adverse events
Indications for Use
Luminopia is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. Luminopia is indicated for improvement in visual acuity in amblyopia patients, aged 4 to <13, associated with anisometropia and/or with mild strabismus, having received treatment instructions (frequency and duration) as prescribed by a trained eye-care professional. Luminopia One is intended for both previously treated and untreated patients. Luminopia is intended to be used as an adjunct to full-time refractive correction, such as glasses, which should also be worn under the HMD during Luminopia therapy. Luminopia is intended for prescription use only, in an at-home environment.
Device Story
Luminopia is a software-only digital therapeutic (SaMD) for pediatric amblyopia. It operates on off-the-shelf head-mounted displays (HMDs) requiring Wi-Fi. The device consists of a Video Content Platform and Therapeutic Algorithms. Patients select TV shows or movies; the software modifies the video input to each eye via contrast reduction and dichoptic masks to break interocular suppression and encourage amblyopic eye usage. The HMD provides a stereoscopic viewing experience. Used at home under clinician prescription, the device aims to improve visual acuity. It functions as an adjunct to refractive correction (glasses).
Clinical Evidence
Prospective real-world registry study (N=290, aged 4 to <13) evaluated BCVA improvement. Patients used Luminopia for average 8.2 months. Mean BCVA improved 1.1 lines (95% CI: 0.92-1.3). Subgroup analysis: 4-7 years (N=186) improved 1.2 lines; 8-12 years (N=104) improved 0.95 lines. Adverse events were low (3% overall), consistent with prior clinical data (DEN210005).
Technological Characteristics
SaMD mobile application; two software units (Video Content Platform, Therapeutic Algorithms). Operates on off-the-shelf HMDs (minimum requirements: ≥60Hz refresh, ≥48 cd/m2 luminance, ≥90% Michelson contrast). Dichoptic presentation principle. No hardware modifications. Wi-Fi connectivity required.
Indications for Use
Indicated for improvement in visual acuity in amblyopia patients, aged 4 to <13, with anisometropia and/or mild strabismus. Intended as an adjunct to full-time refractive correction for both previously treated and untreated patients. Prescription use only in home environment.
Regulatory Classification
Identification
A digital therapy device for amblyopia is a device that incorporates dichoptic presentations on visual displays through therapeutic algorithms to treat amblyopia or to improve visual acuity of patients with amblyopia. Luminopia One is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. It is indicated for improvement in visual acuity in amblyopia patients, aged 4-7, associated with anisometropia and/or with mild strabismus, as an adjunct to full-time refractive correction.
Special Controls
In combination with the general controls of the FD&C Act, the digital therapy device for amblyopia is subject to the following special controls:
- Clinical performance testing must demonstrate that the device performs as intended (1) under anticipated conditions of use with labeled compatible visual display devices, including evaluation of all adverse events and device performance to improve measures of visual function.
- (2) Software verification, validation, and hazard analysis must be performed. Documentation must include characterizations of the technical specifications of the software.
- (3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. All visual displays intended for use must undergo compatibility testing to ensure adequate display resolution, luminance, contrast, field of view, image quality, appropriate optical image distance, and verify their compatibility with the software and intended user (such as appropriate interpupillary distance).
- (4) Labeling must include the following:
- (i) The minimum hardware and operating system requirements that support the software of the device:
- (ii) The models of the visual displays validated to be compatible with this device;
- The length of treatment and/or retreatment supported by clinical performance (iii) testing; and
- A summary of the clinical performance testing conducted with the device. (iv)
- Labeling comprehension testing with intended users must be performed. (૨)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 9, 2025
Luminopia, Inc.
Scott Xiao
CEO
955 Massachusetts Ave #335
Cambridge, Massachusetts 02139
Re: K243819
Trade/Device Name: Luminopia
Regulation Number: 21 CFR 886.5500
Regulation Name: Digital Therapy Device For Amblyopia
Regulatory Class: Class II
Product Code: QQU
Dated: December 9, 2024
Received: December 12, 2024
Dear Scott Xiao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243819 - Scott Xiao
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243819 - Scott Xiao
Page 3
Sincerely,
Elvin Y. Ng -S
Elvin Ng
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243819
Device Name
Luminopia
Indications for Use (Describe)
Luminopia is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. Luminopia is indicated for improvement in visual acuity in amblyopia patients, aged 4 to <13, associated with anisometropia and/or with mild strabismus, having received treatment instructions (frequency and duration) as prescribed by a trained eye-care professional. Luminopia One is intended for both previously treated and untreated patients. Luminopia is intended to be used as an adjunct to full-time refractive correction, such as glasses, which should also be worn under the HMD during Luminopia therapy. Luminopia is intended for prescription use only, in an at-home environment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K243819
# 510(K) SUMMARY
510(k) Owner: Luminopia, Inc.
Address: 955 Massachusetts Ave #335, Cambridge, MA 02139
Phone: 857-365-6636
Fax: 857-336-6605
Contact Person: Scott Xiao
Date: Dec 9, 2024
Trade Name: "Luminopia One" and "Luminopia"
Common Name: N/A
Classification Name: Digital Therapy Device for Amblyopia (21 CFR 886.5500, Product Code: QQU)
Predicate Device: Luminopia (Submission Number: K233720)
Indications for Use: Luminopia is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. Luminopia is indicated for improvement in visual acuity in amblyopia patients, aged 4 to <13, associated with anisometropia and/or with mild strabismus, having received treatment instructions (frequency and duration) as prescribed by a trained eye-care professional. Luminopia is intended for both previously treated and untreated patients. Luminopia is intended to be used as an adjunct to full-time refractive correction, such as glasses, which should also be worn under the HMD during Luminopia therapy. Luminopia is intended for prescription use only, in an at-home environment.
## Device Description:
Luminopia is Software as a Medical Device (SaMD) Mobile Application that is intended to improve vision in pediatric patients with amblyopia. The Mobile Application ("Mobile App") consists of two software units: the Video Content Platform and the Therapeutic Algorithms. The Video Content Platform allows the Patient to browse from a library of popular TV shows and movies and select videos to watch. The Video Content Platform, without the Therapeutic Algorithms, is analogous to consumer video applications (e.g., YouTube, Netflix). The Therapeutic Algorithms provide the actual treatment, by applying modifications to the videos shown by the Video Content Platform.
It is hypothesized that the Therapeutic Algorithms improve vision by breaking interocular suppression and encouraging amblyopic eye usage. The Therapeutic Algorithms are applied to patient-selected video content in the same manner for every video. The algorithms reduce contrast to the stronger eye's input to break interocular suppression and encourage amblyopic eye usage. Additionally, parts of each eye's input are occluded by dichoptic masks superimposed over the video content to promote binocular combination. The masks rotate through predefined pairs over the course of treatment.
Page 1 of 8
{5}
K243819
The Mobile App is designed to be used with commercially available, off-the-shelf head-mounted displays ("HMDs") and does not require any hardware modifications or customization. These HMDs can either consist of a headset combined with a display unit or consist of an all-in-one unit. The HMD serves two functions. Firstly, the HMD serves as the computing platform for the Mobile App, analogous to a smartphone for a mobile medical application. The HMD requires a Wi-Fi connection to run the Mobile App. Secondly, the HMD serves as a viewing device for dichoptic presentation of the content in the app, similar to a stereoscope device (FDA Product Code: HJR). During usage, the Mobile App displays a separate image to each of the Patient's eyes within the HMD, and the images appear as one when viewed together.
All compatible HMDs meet the following set of minimum requirements:
| Parameter | HMD |
| --- | --- |
| Luminance and luminance uniformity | ≥ 48 cd/m2 |
| Michelson contrast (low spatial frequency) | ≥ 90% across field of view |
| Michelson contrast (high spatial frequency grille pattern) | Baseline requirement |
| Resolution | ≥ 14.0 pixels/degree (vertical) ≥ 14.4 pixels/degree (horizontal) |
| IPD range support | Meets requirements ≥ 52mm IPD |
| Internet capability | Yes |
| Battery capacity | > 90 min |
| Weight | < 500 g (± 5%) |
| RF compliance | Yes |
| Audio support | Yes |
| Power button | Yes |
| Eye glasses compatibility | Yes |
| Processing capacity | > Snapdragon 821 |
| Refresh rate | ≥ 60 Hz |
| Field of view | ≥ 51.9 degrees (horizontal) ≥ 30.6 degrees (vertical) |
Substantial Equivalence:
| Feature | Subject Device | Cleared Device | Impact |
| --- | --- | --- | --- |
{6}
K243819
| | | (K233720) | |
| --- | --- | --- | --- |
| Class | II | II | No change |
| Classification | 886.5500 | 886.5500 | No change |
| Product code | QQU | QQU | No change |
| Labeling | | | |
| Indications for Use | Luminopia is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. Luminopia is indicated for improvement in visual acuity in amblyopia patients, aged 4 to <13, associated with anisometropia and/or with mild strabismus, having received treatment instructions (frequency and duration) as prescribed by a trained eye-care professional. Luminopia is intended for both previously treated and untreated patients. Luminopia is intended to be used as an adjunct to full-time refractive correction, such as glasses, which should also be worn under the HMD during Luminopia therapy. Luminopia is intended for prescription use only, in an at-home environment. | Luminopia is a software-only digital therapeutic designed to be used with commercially available Head-Mounted Displays (HMDs) which are compatible with the software application. Luminopia is indicated for improvement in visual acuity in amblyopia patients, aged 4-7, associated with anisometropia and/or with mild strabismus, having received treatment instructions (frequency and duration) as prescribed by a trained eye-care professional. Luminopia is intended for both previously treated and untreated patients; however, patients with more than 12 months of prior treatment (other than refractive correction) have not been studied. Luminopia is intended to be used as an adjunct to full-time refractive correction, such as glasses, which should also be worn under the HMD during Luminopia therapy. Luminopia is intended for prescription use only, in an at-home environment. | Similar |
| Prescription/ over-the-counter use | Prescription-use only | Prescription-use only | No change |
| Use environment | Home use | Home use | No change |
Page 3 of 8
{7}
K243819
| Compatible HMDs listed in labeling | - DPVR P1 Pro 4K
- Pico G2 4K | - DPVR P1 Pro 4K
- Pico G2 4K | No change |
| --- | --- | --- | --- |
| **Technological Characteristics** | | | |
| **Device type** | Software as a Medical Device (SaMD) | Software as a Medical Device (SaMD) | No change |
| **Device design** | 2 software-only components:
- Mobile Application
- Backend Service Layer | 2 software-only components:
- Mobile Application
- Backend Service Layer | No change |
| **Device materials** | N/A (device is SaMD) | N/A (device is SaMD) | No change |
| **Energy source** | N/A (device is SaMD) | N/A (device is SaMD) | No change |
| **Device feature:** Therapeutic mechanism | Modification of visual stimuli using:
contrast reduction + dichoptic masks | Modification of visual stimuli using:
contrast reduction + dichoptic masks | No change |
| **Device feature:** Visual stimuli | Video content | Video content | No change |
| **Hardware platform** | Off-the-shelf Head-Mounted Display (HMD) | Off-the-shelf Head-Mounted Display (HMD) | No change |
There were no changes to the Luminopia device in this submission from the predicate device, and therefore, there are no changes to the technological characteristics.
The change in the Indications for Use represent a label expansion for older children and is supported by clinical performance data.
## Performance Data:
### Clinical Performance Evaluation
The safety and efficacy of Luminopia for 4 to <8-year-old patients with amblyopia was established in DEN210005. This Clinical Performance Evaluation describes effectiveness of Luminopia in improving vision in children aged 4 to <13 in a real-world study in accordance with FDA Guidance: Use of Real-World Evidence to Support Regulatory Decision-Making for Medical Devices.
Luminopia was evaluated in a Real-World Registry study, designed prospectively, which collected retrospective data from medical health records of patients treated with Luminopia under usual care. The registry employed an all-comers design and included any patient with an
Page 4 of 8
{8}
K243819
amblyopia diagnosis and Luminopia use of at least 12 weeks, and only excluded patients who had taken part in past Luminopia clinical trials. Visual Acuity change from the time of Luminopia prescription to the last visit in the registry study was reported. The registry included 334 patients, of whom 290 were aged 4 to $< 13$ with a diagnosis of amblyopia associated with anisometropia and/or strabismus.
The registry study patients had an average of 2.8 years in refractive correction and 1.8 years using patching and/or atropine before starting Luminopia. Patients used Luminopia treatment for an average of approximately 8.2 months. Across all patients aged 4 to $< 13$ with a diagnosis of amblyopia associated with anisometropia and/or strabismus, the mean amblyopic eye best-corrected visual acuity (BCVA) improved 1.1 lines (95% CI: 0.92-1.3 lines, N=290). For the subgroup of patients aged 4-7, the mean amblyopic eye best-corrected visual acuity (BCVA) improved 1.2 lines (95% CI: 1.0-1.4 lines, N=186). For the subgroup of patients aged 8-12, the mean amblyopic eye best-corrected visual acuity (BCVA) improved 0.95 lines (95% CI: 0.66-1.3 lines, N=104).
The Adverse Event risk profile of Luminopia has been established through DEN210005. The risks of Luminopia reported through real-world data is lower than the risk reported within the RCT, with only 9 (3%, N=290) non-serious adverse events reported, although this is likely due to underreporting in the real-world environment. The risk of Luminopia in 4-12-year-olds is expected to be similar to the risk for 4-7-year-olds, as established in DEN210005.
| Amblyopic Eye Change in BCVA1 | | | |
| --- | --- | --- | --- |
| | Baseline | Last Visit | Improvement in BCVA (Lines)2 |
| Age 4-7 | 0.397 ± 0.211 (186) | 0.282 ± 0.219 (186) | 1.16 ± 1.4 (186) |
| | 0.4 (0.1, 1.18) | 0.18 (-0.12, 1.1) | 1.0 (-2.2, 5.2) |
| | [0.367, 0.428] | [0.250, 0.313] | [0.96, 1.36] |
| Age 8-12 | 0.432 ± 0.231 (104) | 0.337 ± 0.216 (104) | 0.95 ± 1.5 (104) |
| | 0.4 (0.1, 1.3) | 0.3 (0.0, 1.0) | 0.9 (-2.0, 7.0) |
| | [0.387, 0.477] | [0.285, 0.379] | [0.66, 1.25] |
| Age 4-12 | 0.410 ± 0.218 (290) | 0.301 ± 0.219 (290) | 1.08 ± 1.4 (290) |
| | 0.4 (0.1, 1.3) | 0.3 (-0.12, 1.1) | 1.0 (-2.2, 7.0) |
| | [0.384, 0.435] | [0.276, 0.327] | [0.92, 1.25] |
| 1. Based on participants with available data at baseline and one follow-up visit. Data presented as mean ± standard deviation (N) median (min, max). Change from baseline also includes [95% CI].2. Original visual acuity measurements captured using logMAR. Change presented in Lines of Visual Acuity. A 1-line improvement from baseline corresponds to a change of -0.10 logMAR. | | | |
{9}
K243819
| Amblyopic Eye Change in BCVA Over Time1 | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Time-point | N | Mean2 | Std Dev | Lower 95% CI | Upper 95% CI |
| 4 to 7 | 12 week - Baseline | 135 | 1 | 1.19 | 0.79 | 1.2 |
| | 24 week - Baseline | 91 | 1.12 | 1.3 | 0.85 | 1.39 |
| | 36 week - Baseline | 68 | 1.22 | 1.41 | 0.88 | 1.57 |
| | 48 week - Baseline | 44 | 1.28 | 1.47 | 0.84 | 1.73 |
| | 60 week - Baseline | 10 | 0.92 | 1.84 | -0.4 | 2.24 |
| | 72 week - Baseline | 5 | 0.6 | 1.85 | -1.7 | 2.9 |
| 8 to 12 | 12 week - Baseline | 61 | 0.94 | 1.44 | 0.57 | 1.31 |
| | 24 week - Baseline | 58 | 0.74 | 1.36 | 0.39 | 1.1 |
| | 36 week - Baseline | 39 | 1.47 | 1.41 | 1.01 | 1.93 |
| | 48 week - Baseline | 19 | 0.68 | 1.76 | -0.16 | 1.53 |
| | 60 week - Baseline | 7 | 2.26 | 2.44 | 0.01 | 4.51 |
| | 72 week - Baseline | 1 | 0 | | 0 | 0 |
| 4 to 12 | 12 week - Baseline | 196 | 0.98 | 1.27 | 0.8 | 1.16 |
| | 24 week - Baseline | 149 | 0.97 | 1.33 | 0.76 | 1.19 |
| | 36 week - Baseline | 107 | 1.31 | 1.41 | 1.04 | 1.58 |
| | 48 week - Baseline | 63 | 1.1 | 1.57 | 0.71 | 1.5 |
| | 60 week - Baseline | 17 | 1.47 | 2.14 | 0.37 | 2.57 |
| | 72 week - Baseline | 6 | 0.5 | 1.67 | -1.25 | 2.25 |
| 1. Based on participants with available data at baseline and at the timepoint with +/- 6 week window.2. Original visual acuity measurements captured using logMAR. Mean change presented in Lines of Visual Acuity. A 1-line improvement from baseline corresponds to a change of -0.10 logMAR. | | | | | | |
Page 6 of 8
{10}
K243819
| Amblyopic Eye Change in BCVA1 | | | |
| --- | --- | --- | --- |
| Number of Lines Change2(follow-up - baseline) | Age 4-7 | Age 8-12 | Age 4-12 |
| 6> to ≤7 line improvement | 0.0% (0/186) | 1.0% (1/104) | 0.3% (1/290) |
| 5> to ≤6 line improvement | 0.5% (1/186) | 1.0% (1/104) | 0.7% (2/290) |
| 4> to ≤5 line improvement | 3.2% (6/186) | 0.0% (0/104) | 2.1% (6/290) |
| 3> to ≤4 line improvement | 3.8% (7/186) | 5.8% (6/104) | 4.5% (13/290) |
| 2> to ≤3 line improvement | 13.4% (25/186) | 11.5% (12/104) | 12.8% (37/290) |
| 1> to ≤2 line improvement | 31.7% (59/186) | 25.0% (26/104) | 29.3% (85/290) |
| 0> to ≤1 line improvement | 15.6% (29/186) | 17.3% (18/104) | 16.3% (47/290) |
| No change | 22.6% (42/186) | 26.0% (27/104) | 23.8% (69/290) |
| 0> to ≤1 line decrease | 4.3% (8/186) | 3.8% (4/104) | 4.1% (12/290) |
| 1> to ≤2 line decrease | 4.3% (8/186) | 8.7% (9/104) | 5.9% (17/290) |
| 2> to ≤3 line decrease | 0.5% (1/186) | 0.0% (0/104) | 0.3% (1/290) |
| 1. Based on participants with available data at baseline and one follow-up visit. Categorical variables presented as n/N (%) where N is the number of participants with available data.2. Visual acuity measurements captured using logMAR. A 1-line improvement from baseline corresponds to a change of -0.10 logMAR. | | | |
| Adverse Events (AEs) | | | |
| --- | --- | --- | --- |
| AE Type | Age 4-7(N=186) | Age 8-12(N=104) | Age 4-12(N=290) |
| Eye Redness | 0 (0%) | 1 (1%) | 1 (<1%) |
| Headache | 4 (2%) | 1 (1%) | 5 (2%) |
| Dizziness | 0 (0%) | 1 (1%) | 1 (<1%) |
| Teary Eye | 1 (1%) | 0 (0%) | 1 (<1%) |
| Nightmare Event | 0 (0%) | 1 (1%) | 1 (<1%) |
| Overall | 5 (3%) | 4 (4%) | 9 (3%) |
| Includes events reported n (rate of n/N). | | | |
Page 7 of 8
{11}
K243819
## Conclusion:
As described in this 510(k) Summary, the subject and predicate devices have the same intended use and similar indications for use. There are no differences in technological characteristics between the subject and predicate devices. Clinical performance data from a real-world registry study in addition to the clinical evidence established in DEN210005 of the same device are sufficient to establish the safety and effectiveness of the subject device when used in accordance with its proposed labeling.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.