True Non-Surgical Spinal Decompression System (DRX9000-SL)
K243775 · Excite Medical of Tampa Bay, LLC · ITH · Jul 18, 2025 · Physical Medicine
Device Facts
Record ID
K243775
Device Name
True Non-Surgical Spinal Decompression System (DRX9000-SL)
Applicant
Excite Medical of Tampa Bay, LLC
Product Code
ITH · Physical Medicine
Decision Date
Jul 18, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DRX9000-SL True Non-Surgical Spinal Decompression System provides a primary treatment modality for the management of pain and disability for patients suffering with incapacitating low back pain and sciatica. It is designed to apply spinal decompressive forces to compressive and degenerative injuries of the spine. It has been found to provide relief of pain and symptoms associated with herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome and sciatica.
Device Story
Device provides non-surgical spinal decompression for lumbar spine; used in professional healthcare facilities/clinics by trained personnel. System consists of treatment computer, servo-amplifier, servo-motor, load-cell feedback, upper chest/lower pelvic harnesses, and free-floating lower body mattress. Patient reclined horizontally; therapist localizes pain and directs treatment. Device applies controlled, cyclic tensile forces to decompress intervertebral disc space; load-cell feedback continuously monitors and adjusts tension to ensure smooth application. Patient safety switch and emergency stop button allow immediate cessation of force; redundant systems reduce tension to zero and disable motor. Benefits include mobilization of disc segments and promotion of total patient relaxation to facilitate decompression.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), risk management (ISO 14971), biocompatibility (ISO 10993-1), and software lifecycle (IEC 62304).
Indicated for patients suffering from incapacitating low back pain and sciatica, including those with herniated discs, bulging/protruding intervertebral discs, degenerative disc disease, and posterior facet syndrome.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
Predicate Devices
DRX9000 True Non-Surgical Spinal Decompression System (K060735)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 18, 2025
Excite Medical of Tampa Bay, LLC
% Jon Werner
QA / RA Consultant
Cofactor Technologies, Inc.
2002 Blind Pond Ave
Lutz, Florida 33549
Re: K243775
Trade/Device Name: True Non-Surgical Spinal Decompression System (DRX9000-SL)
Regulation Number: 21 CFR 890.5900
Regulation Name: Power Traction Equipment
Regulatory Class: Class II
Product Code: ITH
Dated: June 27, 2025
Received: June 27, 2025
Dear Jon Werner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243775 - Jon Werner
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243775 - Jon Werner
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber T. Ballard -S
Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243775
Device Name
DRX9000-SL True Non-Surgical Spinal Decompression System
Indications for Use (Describe)
The DRX9000-SL True Non-Surgical Spinal Decompression System provides a primary treatment modality for the management of pain and disability for patients suffering with incapacitating low back pain and sciatica. It is designed to apply spinal decompressive forces to compressive and degenerative injuries of the spine. It has been found to provide relief of pain and symptoms associated with herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome and sciatica.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
EXCITE MEDICAL
SPECIAL INFORMATION
# 510(k) Summary
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Excite Medical of Tampa Bay is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K243775.
Date Prepared: 25 June 2025
## A. SUBMITTER
Excite Medical of Tampa Bay, LLC
4710 Eisenhower Blvd N Suite A-12
Tampa, FL 33634
Establishment Registration Number: 3008816697
**Company Contact:**
Saleem Musallam
President
Tel: (813) 210-1000
sm@excitemedical.com
## B. DEVICE NAME
**Proprietary Name:** DRX9000-SL True Non-Surgical Spinal Decompression System
**Model Numbers:**
DRX9000-SL True Non-Surgical Spinal Decompression System
Upper (Chest) Harnesses (E-UBH-XS/S/M/L/XL)
Lower (Pelvic) Harnesses (E-PH-XS/S/M/L/XL)
51-0007 Head Pillow
51-0008 Knee Pillow
**Common Name:** Equipment, Traction, Powered
**Classification Names:** Powered Traction Equipment
**Regulation Number:** 21 CFR Part 890.5900
**Classification:** Class II
**Product Code:** ITH
## C. PREDICATE / LEGALLY MARKETED DEVICES
**Device Name:** DRX9000 True Non-Surgical Spinal Decompression System
**Company Name:** Axiom Worldwide, Inc.
**510(k) #:** K060735
Page 1 of 5
{5}
EXCITE MEDICAL LICENSING INSTITUTE
## D. DEVICE DESCRIPTION
The DRX9000-SL True Non-Surgical Spinal Decompression System provides accurately controlled tensions designed to relax and confuse paraspinal muscles and allow distinctive forces to decompress intervertebral spinal disc space. The user interface provided by the treatment computer constantly updates a servo-amplifier controlling a servo-motor to immediately and safely apply forces as determined by qualified healthcare personnel. Load-cell feedback is utilized to further verify and adjust tensile forces, allowing for variations in patient posture and outside forces such that continuous and smooth tension is experienced by the patient.
An upper chest harness and a lower pelvic harness are used to help distribute the applied forces evenly. Once in place, the patient is slowly reclined to a horizontal position. Following the physician's orders, the therapist localizes the pain, makes any adjustments and directs the treatment to the proper area. The DRX9000-SL True Non-Surgical Spinal Decompression System helps to mobilize the troubled disc segment without introducing further damage to the spine.
The patient safety switch is held by the patient who at any time and for any reason may quickly pause any tensile forces. The patient safety switch is monitored and executed by two redundant systems, which will gradually reduce the tension to zero to stop the treatment and after a 5 second delay directly disable the treatment motor.
The emergency stop button is accessible to the operator who at any time and for any reason may quickly pause any tensile forces. The emergency stop button is monitored and executed by two redundant systems, which will gradually reduce the tension to zero to stop the treatment and after a 5 second delay directly disable the treatment motor.
Integral to effective spinal decompression and included in the device are continuous load-cell tensile feedback into the treatment computer, dedicated and matched servo-amplifier and servo-motor, smoothly modulated cyclic tension application (high and low tension plateaus transitioned into via non-linear tension change), two segment (upper and lower) textile patient harness, patient safety switch, and free-floating lower body mattress. The free-floating lower body mattress allows the interdiscal segments of the lumbar spine to decompress at their own rate. As tension is cycled, the lower body can extend independent of the upper body which is held in place via an upper body textile patient harness. The treatment bed and textile harness allow the patient to relax completely and require no conscious exertion on the part of the patient. Total patient relaxation encourages paraspinal muscle relaxation from both a physical and psychological standpoint and is a key to spinal decompression.
Page 2 of 5
{6}
EXCITE MEDICAL LICENSING INSTITUTE
# E. INTENDED USE / INDICATIONS FOR USE
The DRX9000-SL True Non-Surgical Spinal Decompression System provides a primary treatment modality for the management of pain and disability for patients suffering with incapacitating low back pain and sciatica. It is designed to apply spinal decompressive forces to compressive and degenerative injuries of the spine. It has been found to provide relief of pain and symptoms associated with herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome and sciatica.
# F. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS
The DRX9000-SL True Non-Surgical Spinal Decompression System indications for use is the same as the predicate device. The technical characteristics and operational parameters of the DRX9000-SL True Non-Surgical Spinal Decompression System are substantially equivalent. The table below shows a comparison of the primary technological characteristics and operational parameters of the DRX9000-SL True Non-Surgical Spinal Decompression System to the referenced predicate device.
| | Subject Device (K243775)
DRX9000-SL True Non-Surgical Spinal Decompression System | Predicate Device (K060735)
DRX9000 True Non-Surgical Spinal Decompression System |
| --- | --- | --- |
| Intended Use | Prescription use only | Prescription use only |
| Environment | Professional healthcare facilities, clinics and / or hospitals | Professional healthcare facilities, clinics and / or hospitals |
| Indications for Use | The DRX9000-SL True Non-Surgical Spinal Decompression System provides a primary treatment modality for the management of pain and disability for patients suffering with incapacitating low back pain and sciatica. It is designed to apply spinal decompressive forces to compressive and degenerative injuries of the spine. It has been found to provide relief of pain and symptoms associated with herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome and sciatica. | The DRX9000 True Non-Surgical Spinal Decompression System provides a primary treatment modality for the management of pain and disability for patients suffering with incapacitating low back pain and sciatica. It is designed to apply spinal decompressive forces to compressive and degenerative injuries of the spine. It has been found to provide relief of pain and symptoms associated with herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet syndrome and sciatica. |
| Treatment Areas | Lumbar | Lumbar |
| Electrical Classification and Type | Class I
Type BF | Not publicly available |
| Electrical Safety Consensus Standard | IEC 60601-1:2005 + A1:2012 + A2:2020 | Not publicly available |
{7}
EXCITE MEDICAL LICENSING INSTITUTE
MEDICAL GENEVA
MEDICAL LICENSE
MEDICAL LICENSE
Page 4 of 5
| | Subject Device (K243775)
DRX9000-SL True Non-Surgical
Spinal Decompression System | Predicate Device (K060735)
DRX9000 True Non-Surgical
Spinal Decompression System |
| --- | --- | --- |
| EMC Consensus Standard | IEC 60601-1-2:2014 + A1:2020 | Not publicly available |
| Usability Consensus Standard | IEC 60601-1-6:2010 + A1:2013 + A2:2020 | Not publicly available |
| Biocompatibility | In accordance with ISO 10993-1:2018 and FDA# G95-1 | Not publicly available |
| Traction Mechanism | Electromechanical | Electromechanical |
| Operational Components | Software controlled servo amplifier
and servo motor for traction and flexion | Not publicly available |
| Traction Force | Traction limits can be varied
between 0 – 150lbs pull | Traction limits can be varied
between 0 – 150lbs pull |
| Traction Speed | Continuous, Adjustable | Continuous, Adjustable |
| Treatment Time | Up to 30 minutes | Up to 30 minutes |
| Therapy Mode | Continuous | Continuous |
| Power Supply | 100-120VAC / 220-230 VAC, 50/60 Hz | Not publicly available |
| Battery Backup | Yes | Not publicly available |
| Environmental Conditions | Storage / Operation
+10 to +40 Celsius
30% to 75% non-condensing
700 to 1060 hPa | Not publicly available |
| Materials | Mattress / Pillow Surfaces:
REM114 ADF Remington | Not publicly available |
{8}
EXCITE MEDICAL LICENSING INSTITUTE
# G. SUMMARY OF NON-CLINICAL TESTING
Non-clinical tests were performed on the DRX9000-SL True Non-Surgical Spinal Decompression System in order to verify and validate the design and to assure conformance with the following voluntary design standards in connection with electrical safety, electromagnetic compatibility, usability, risk management, biocompatibility and software.
| Type | Description |
| --- | --- |
| Electrical safety | • IEC 60601-1:2005 + A1:2012 + A2:2020 [19-49] Medical electrical equipment – Part 1: General requirements for basic safety and essential performance |
| Electromagnetic compatibility | • IEC 60601-1-2:2014 + A1:2020 [19-36] Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests
• IEC TR 60601-4-2:2016 [19-19] Medical electrical equipment – Part 4-2: Guidance and interpretation – Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems |
| Risk management | • ISO 14971:2019 [5-125] Medical devices – Application of risk management to medical devices |
| Biocompatibility | • ISO 10993-1:2018 [5-258] Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
| Software | • IEC 62304:2006 + A1:2015 [13-79] Medical device software – Software life cycle processes |
# H. SUMMARY OF CLINICAL TESTING
Clinical testing is not required for this submission. The non-clinical performance testing described above is sufficient to support substantial equivalence of the subject device to the predicate device.
# I. CONCLUSION
The DRX9000-SL True Non-Surgical Spinal Decompression System has the same intended use, indications for use, treatment location, conditions of use, and fundamental scientific technology as the predicate device. The conclusions drawn from the nonclinical performance testing (discussed above) demonstrate that the device raises no new issues of safety or effectiveness and that the DRX9000-SL True Non-Surgical Spinal Decompression System is therefore substantially equivalent to the DRX9000 True Non-Surgical Spinal Decompression System cleared by K060735.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.