The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 µg/mL

K243738 · Thermo Fisher Scientific · NGZ · Feb 27, 2025 · Microbiology

Device Facts

Record IDK243738
Device NameThe Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 µg/mL
ApplicantThermo Fisher Scientific
Product CodeNGZ · Microbiology
Decision DateFeb 27, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2
AttributesPCCP

Indications for Use

The Sensititre YeastOne Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of Candida spp. This 510(k) is for micafungin in the dilution range of 0.008-16 ug/mL for testing Candida spp. on The Sensitite YeastOne Susceptibility System. Micafungin has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label: Candida albicans Candida glabrata Candida guilliermondi Candida krusei Candida parapsilosis Candida tropicalis

Device Story

Sensititre YeastOne Susceptibility System is an in vitro diagnostic device used for antimicrobial susceptibility testing. It determines the susceptibility of Candida species to the antifungal agent micafungin. The system utilizes a dilution-based method to measure the minimum inhibitory concentration (MIC) of the drug against specific yeast pathogens. The device is intended for use in clinical laboratory settings by trained laboratory personnel. Results are interpreted by healthcare providers to guide antifungal therapy decisions, potentially improving patient outcomes by ensuring appropriate drug selection for fungal infections.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Micro-broth dilution system; 96-well plates with dried antifungal agents and alamarBlue indicator. Inoculation via Sensititre AIM autoinoculator; reading via Vizion digital viewing device. Incubation at 35°C for 24 hours. Colorimetric growth detection (blue to pink/purple).

Indications for Use

Indicated for clinical susceptibility testing of Candida spp. (C. albicans, C. glabrata, C. guilliermondi, C. krusei, C. parapsilosis, C. tropicalis) to micafungin in the dilution range of 0.008-16 µg/mL. For prescription use only.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K243738 B Applicant Thermo Fisher Scientific C Proprietary and Established Names The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 $\mu \mathrm{g} / \mathrm{mL}$ D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | NGZ | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain a substantial equivalence determination for The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of $0.008 - 16~\mu \mathrm{g / mL}$ . B Measurand: Micafungin in the dilution range 0.008 to $16~\mu \mathrm{g / mL}$ C Type of Test: Quantitative antifungal susceptibility test (AST) growth-based detection Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} III Intended Use/Indications for Use: A Intended Use(s): The Sensititre YeastOne Susceptibility System is an *in vitro* diagnostic product for clinical susceptibility testing of *Candida* spp. B Indication(s) for Use: The Sensititre YeastOne Susceptibility System is an *in vitro* diagnostic product for clinical susceptibility testing of *Candida* spp. This 510(k) is for micafungin in the dilution range of 0.008-16 µg/mL for testing *Candida* spp. on The Sensititre YeastOne Susceptibility System. Micafungin has been shown to be active both clinically and *in vitro* against the following organisms according to the FDA drug label: - *Candida albicans* - *Candida glabrata* - *Candida guilliermondii* - *Candida krusei* - *Candida parapsilosis* - *Candida tropicalis* C Special Conditions for Use Statement(s): Rx - For Prescription Use Only The following limitation was applied to micafungin testing in the device labeling: Studies of Voriconazole, Caspofungin, Fluconazole, Rezafungin, and Micafungin with *Candida* spp. were performed using the AIM autoinoculator inoculation method and the VIZION reading method only. The use of alternative inoculation methods or alternative reading methods when testing Voriconazole, Caspofungin, Fluconazole, Rezafungin, and Micafungin have not been evaluated. Due to the insufficient number of resistant *Candida* spp. evaluated, the following limitation statement was applied to micafungin testing in the device labeling: The ability of the Sensititre YeastOne Susceptibility System to detect resistance with the following combination(s) is unknown because resistant strains were either not available or an insufficient number were encountered at the time of comparative testing: Micafungin (0.008-16 µg/ml) and *Candida guilliermondii*, *Candida krusei*, and *Candida parapsilosis*. If a resistant isolate is encountered, it should be submitted to a reference laboratory for further testing. K243738 - Page 2 of 11 {2} D Special Instrument Requirements: Sensititre AIM for device inoculation Sensititre Vizion digital viewing device IV Device/System Characteristics: A Device Description: The Sensititre YeastOne Susceptibility System is a micro-version of the broth dilution susceptibility test performed in multi-well microtiter plates. Various antifungal agents are serially diluted to concentrations bridging the range of clinical interest in autoclaved diluent which contains a colorimetric growth indicating compound. A standardized organism suspension is prepared using the Sensititre Yeast Susceptibility inoculum broth and 100 µL of the suspension is inoculated into the dried antifungal containing wells. After inoculation with a standardized suspension of organisms in inoculum medium and incubation at 35°C for 24 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antifungal concentration preventing the development of a pink or purple color change (as evidenced by no color change). Yeast growth in the antifungal solutions will be evident as change in the colorimetric growth indicator from blue (negative – no growth) to pink/purple (positive – growth). Turbidity is not read, only color change is used as an indicator of growth. B Principle of Operation: Colorimetric test in which MICs are determined by determining the lowest concentration of antifungal agent that shows no color change indicating inhibition of growth of the organism. A color change from blue to pink indicates growth of the organism. The Sensititre YeastOne panels can be read only with the Vizion digital viewing device which allows the panel image to be displayed on a touch screen directly from a video camera and allows the user to visually determine MIC results. V Substantial Equivalence Information: A Predicate Device Name(s): The Sensititre YeastOne Susceptibility System with Rezafungin in the dilution range of 0.008-8 µg/mL B Predicate 510(k) Number(s): K231433 K243738 - Page 3 of 11 {3} C Comparison with Predicate(s): | Device & Predicate Device(s): | Device: K243738 | Predicate: K231433 | | --- | --- | --- | | Device Trade Name | The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 μg/mL | The Sensititre YeastOne Susceptibility System with Rezafungin in the dilution range of 0.008-8 μg/mL | | General Device Characteristic Similarities | | | | Intended Use | The Sensititre YeastOne Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of Candida spp. | Same | | Technology | Colorimetric test in which MICs are determined by determining the lowest concentration of antifungal agent that shows no color change indicating inhibition of growth of the organism. A color change from blue to pink indicates growth of the organism. | Same | | Test Panel | Each 96 well plate is precision dosed with selected antifungal agents, a stabilizer and the alamarBlue indicator then dried thereby stabilized. The fungal suspension in the appropriate broth is used to rehydrate the plate. | Same | | Incubation | 24 hours | Same | | Read Method | Vizion | Same | | General Device Characteristic Differences | | | | Antifungal and Dilution Range | Micafungin 0.008-16 μg/mL | Rezafungin 0.008-8 μg/mL | | Indicated Species | Candida albicans Candida glabrata | Candida albicans Candida glabrata | K243738 - Page 4 of 11 {4} | | Candida guilliermondii Candida krusei Candida parapsilosis Candida tropicalis | Candida parapsilosis | | --- | --- | --- | VI Standards/Guidance Documents Referenced: CLSI M27M44S, “Performance Standards for Antifungal Susceptibility Testing of Yeasts -3rd Edition”, (August 2022) CLSI M27, “Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts - 4th Edition”, (November 2017) FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA (Issued August 28, 2009) VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: A reproducibility study of The Sensititre YeastOne Susceptibility System with Micafungin was performed at four sites using a panel of twelve (12) Candida isolates from indicated species (2 Candida albicans, 2 Candida glabrata, 1 Candida guilliermondii, 3 Candida krusei, 2 Candida parapsilosis, and 2 Candida tropicalis). In addition, three Candida isolates (1 Candida lusitaniae, and 2 Candida auris) were tested that are not intended for testing with this device. All isolates were tested in triplicate over three days and read visually with the Vizion. The Sensititre AIM autoinoculator was used for Sensititre plate inoculation. The mode MIC value was determined, and the reproducibility of the twelve (12) isolates was calculated based on MIC values falling within ±1 or ±2 doubling dilution of the mode MIC value. The reproducibility studies for both calculations demonstrated an acceptable performance of ≥95%. 2. Linearity: Not applicable. 3. Analytical Specificity/Interference: Not applicable. 4. Assay Reportable Range: Not applicable. K243738 - Page 5 of 11 {5} 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The CLSI-recommended quality control (QC) strains *C. parapsilosis* ATCC 22019 and *C. krusei* ATCC 6258 were tested at four sites. The QC strains were tested a minimum of 20 times per site and read visually with the Vizion. The QC strains were also tested with the reference method. The results demonstrate that The Sensititre YeastOne Susceptibility System with Micafungin produced quality control results within the recommended range >95% of the time (Table 1). Table 1. Quality Control Results for *C. parapsilosis* and *C. krusei* with Micafungin with the Reference Method and Vizion | QC Organism | Expected Range (μg/mL) | Concentration (μg/mL) | Reference | Vizion | | --- | --- | --- | --- | --- | | Candida parapsilosis ATCC 22019 | 0.5-2 | ≤0.25 | | | | | | 0.5 | 32 | | | | | 1 | 34 | 79 | | | | 2 | 16 | 9 | | | | ≥4 | | | | Candida krusei ATCC 6258 | 0.06-0.25 | ≤0.03 | | | | | | 0.06 | 5 | 1 | | | | 0.12 | 49 | 86 | | | | 0.25 | 27 | 1 | | | | ≥0.5 | | | Inoculum Density: Inoculum density checks were performed for all QC, reproducibility, challenge, and clinical isolates tested. Only results from cultures with appropriate inoculum densities were reported. Purity Checks: Purity checks were performed for all QC, reproducibility, challenge, and clinical isolates tested. Only results from pure cultures were reported. Growth Failure: There were no growth failures on The Sensititre YeastOne Micafungin panels. 6. Detection Limit: Not applicable. 7. Assay Cut-Off: Not applicable. B Comparison Studies: 1. Method Comparison with Predicate Device: Testing of The Sensititre YeastOne Susceptibility System with Micafungin was performed at one internal and three external sites. Results were compared to those obtained with the CLSI K243738 - Page 6 of 11 {6} broth microdilution reference method. Sensititre panels were inoculated using the AIM Autoinoculator and results were read visually by the Vizion only. Reference panels were inoculated according to recommendations in the CLSI M27 document and results were interpreted manually using a mirrored reader. No inoculation system other than the AIM Autoinoculator and no read method other than Vizion were used in the comparative study. To address the inoculation method and read method limitation, the following limitation was updated in the device labeling, the updates are shown in **bold**: Studies of Voriconazole, Caspofungin, Fluconazole, Rezafungin, and Micafungin with Candida spp. were performed using the AIM autoinoculator inoculation method and the VIZION reading method only. The use of alternative inoculation methods or alternative reading methods when testing Voriconazole, Caspofungin, Fluconazole, Rezafungin, and Micafungin have not been evaluated. The testing conditions for the reference method consisted of the following: - Media: Roswell Park Memorial Institute (RPMI) 1640 Culture Medium plus 0.2% glucose - Inoculum: Inoculated per CLSI M27 guidelines - Incubation: 35°C in a non-CO₂ incubator for 24 hours Inoculation and incubation procedure for Candida spp. - Media: Sensititre Yeast Susceptibility inoculum Broth - Inoculum: A suspension approximating a 0.5 McFarland standard was prepared in 5 mL sterile water. Twenty (20) μL of the standardized suspension was transferred to 11 mL of Yeast Susceptibility Inoculum Broth. Susceptibility plates were inoculated with 100 μL of the final organism suspension using the Sensititre AIM Autoinoculator. - Incubation: 35°C in a non-CO₂ incubator for 24 hours A total of 528 clinical isolates comprised of 142 C. albicans isolates, 98 C. glabrata isolates, 28 C. guilliermondii isolates, 80 C. krusei isolates, 91 C. parapsilosis isolates, 89 C. tropicalis isolates as well as 84 challenge isolates comprised of 21 C. albicans isolates, 27 C. glabrata isolates, 1 C. guilliermondii isolate, 8 C. krusei isolates, 10 C. parapsilosis isolates, 17 C. tropicalis isolates were evaluated with the Vizion and the results were provided in **Table 2**. For C. albicans read using the Vizion, the combined clinical and challenge results (163 isolates) were acceptable at 99.4% and 98.2% for EA and CA, respectively. There were 3 minor errors, no major errors, and no very major errors. For C. glabrata read using the Vizion, the combined clinical and challenge results (125 isolates) were acceptable at 100% and 96.8% for EA and CA, respectively. There were 4 minor errors, no major errors, and no very major errors. For C. guilliermondii read using the Vizion, the combined clinical and challenge results (29 isolates) were acceptable at 96.6% and 100% for EA and CA, respectively. There were no minor errors, no major errors, and no very major errors. K243738 - Page 7 of 11 {7} For C. krusei read using the Vizion, the combined clinical and challenge results (88 isolates) were acceptable at 98.9% and 97.7% for EA and CA, respectively. There were 2 minor errors, no major errors, and no very major errors. For C. parapsilosis read using the Vizion, the combined clinical and challenge results (88 isolates) were acceptable at 99.0% and 99.0% for EA and CA, respectively. There was 1 minor error, no major errors, and no very major errors. For C. tropicalis read using the Vizion, the combined clinical and challenge results (88 isolates) were acceptable at 99.1% and 98.1% for EA and CA, respectively. There were 2 minor errors, no major errors, and no very major errors. Due to the insufficient number of resistant isolates evaluated for some Candida spp., the following limitation was applied to micafungin testing in the device labeling: The ability of the Sensititre YeastOne Susceptibility System to detect resistance with the following combination(s) is unknown because resistant strains were either not available or an insufficient number were encountered at the time of comparative testing: Micafungin (0.008-16 μg/ml) and Candida guilliermondii, Candida krusei, and Candida parapsilosis. If a resistant isolate is encountered, it should be submitted to a reference laboratory for further testing. Table 2. Micafungin Performance of Candida spp. Read by Vizion | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Candida albicans [≤ 0.25 (S), 0.5 (I), ≥1 (R)] | | | | | | | | | | | | | | | Clinical | 142 | 141 | 99.3 | 97 | 95 | 97.9 | 139 | 97.9 | 6 | 132 | 3 | 0 | 0 | | Challenge | 21 | 21 | 100 | 12 | 11 | 91.7 | 21 | 100 | 0 | 21 | 0 | 0 | 0 | | Combined | 163 | 162 | 99.4 | 109 | 106 | 97.3 | 160 | 98.2 | 6 | 153 | 3 | 0 | 0 | | Candida glabrata [≤ 0.06 (S), 0.12 (I), ≥0.25 (R)] | | | | | | | | | | | | | | | Clinical | 98 | 98 | 100 | 83 | 83 | 100 | 94 | 95.9 | 4 | 92 | 4 | 0 | 0 | | Challenge | 27 | 27 | 100 | 20 | 20 | 100 | 27 | 100 | 11 | 16 | 0 | 0 | 0 | | Combined | 125 | 125 | 100 | 103 | 103 | 100 | 121 | 96.8 | 15 | 108 | 4 | 0 | 0 | | Candida guilliermondii [≤ 2 (S), 4 (I), ≥8 (R)] | | | | | | | | | | | | | | | Clinical | 28 | 27 | 96.4 | 28 | 27 | 96.4 | 28 | 100 | 0 | 28 | 0 | 0 | 0 | | Challenge | 1 | 1 | 100 | 1 | 1 | 100 | 1 | 100 | 0 | 1 | 0 | 0 | 0 | | Combined | 29 | 28 | 96.6 | 29 | 28 | 96.6 | 29 | 100 | 0 | 29 | 0 | 0 | 0 | | Candida krusei [≤ 0.25 (S), 0.5 (I), ≥1 (R)] | | | | | | | | | | | | | | | Clinical | 80 | 79 | 98.8 | 80 | 79 | 98.8 | 78 | 97.5 | 1 | 78 | 2 | 0 | 0 | | Challenge | 8 | 8 | 100 | 8 | 8 | 100 | 8 | 100 | 0 | 8 | 0 | 0 | 0 | | Combined | 88 | 87 | 98.9 | 88 | 87 | 98.9 | 86 | 97.7 | 1 | 86 | 2 | 0 | 0 | | Candida parapsilosis [≤ 2 (S), 4 (I), ≥8 (R)] | | | | | | | | | | | | | | | Clinical | 91 | 90 | 98.9 | 91 | 90 | 98.9 | 90 | 98.9 | 0 | 91 | 1 | 0 | 0 | | Challenge | 10 | 10 | 100 | 10 | 10 | 100 | 10 | 100 | 0 | 10 | 0 | 0 | 0 | | Combined | 101 | 100 | 99.0 | 101 | 100 | 99.0 | 100 | 99.0 | 0 | 101 | 1 | 0 | 0 | | Candida tropicalis [≤ 0.25 (S), 0.5 (I), ≥1 (R)] | | | | | | | | | | | | | | | Clinical | 89 | 88 | 98.9 | 89 | 88 | 98.9 | 87 | 97.8 | 2 | 86 | 2 | 0 | 0 | K243738 - Page 8 of 11 {8} | Challenge | 17 | 17 | 100 | 15 | 15 | 100 | 17 | 100 | 0 | 17 | 0 | 0 | 0 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Combined | 106 | 105 | 99.1 | 104 | 103 | 99.0 | 104 | 98.1 | 2 | 103 | 2 | 0 | 0 | EA - Essential Agreement CA - Categorical Agreement S - Susceptible maj - Major Discrepancies EVAL - Evaluable MICs R - Resistant min - Minor Discrepancies vmj - Very Major Discrepancies Essential agreement (EA) occurs when the result of the reference method and that of the Sensititre panel are within plus or minus two serial two-fold dilution of the antifungal agent. Evaluable results are those that are on scale for both the reference method and the Sensititre panel or those in which an off-scale result is more than two doubling dilutions from the on-scale result. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the Sensititre panel. # MIC Trending A trending analysis was conducted using the combined data (clinical and challenge) obtained from the Vizion for Candida spp. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower or at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis. Species for which the difference between the percentage of isolates with higher vs. lower readings was $>30\%$ and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that shows higher or lower MIC values compared to the reference is addressed in the labeling. Evaluation of results for Candida spp. with micafungin using the Vizion are summarized in Table 3. No trending was observed. Table 3. Micafungin Trending Analysis for Candida spp. with Vizion | Organism | Total Evaluable for Trending | ≥1 Dilution lower No. (%) | Exact No. | ≥1 Dilution Higher No. (%) | Percent Difference (95% CI) | Trending Noted | | --- | --- | --- | --- | --- | --- | --- | | Candida albicans | 143 | 56 (39.2) | 54 | 33 (23.1) | -16% (-26% to -5%) | No | | Candida glabrata | 122 | 47 (38.5) | 51 | 24 (19.7) | -19% (-30% to -7%) | No | | Candida guilliermondii | 29 | 5 (17.2) | 19 | 5 (17.2) | 0% (-20% to 20%) | No | | Candida krusei | 88 | 8 (9.1) | 51 | 29 (33.0) | 24% (12% to 35%) | No | | Candida parapsilosis | 101 | 15 (14.9) | 47 | 39 (38.6) | 24% (12% to 35%) | No | | Candida tropicalis | 106 | 22 (20.8) | 38 | 46 (43.4) | 23% (10% to 34%) | No | K243738 - Page 9 of 11 {9} # Testing/Reporting MICs for Non-indicated Species. For this review, the interpretive criteria are applied to the organisms/organism groups according to the FDA STIC website. As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the Warnings and Precautions section of the device labeling to address testing and reporting of non-indicated species: The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved. 2. Matrix Comparison: Not applicable. # C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. # D Clinical Cut-Off: Not applicable. # Expected Values/Reference Range: Table 4. FDA-Recognized Interpretive Criteria for Micafungin | Organism | Minimum Inhibitory Concentrations (μg/mL)a | | | | --- | --- | --- | --- | | | Susceptible | Intermediate | Resistant | | C. albicans | ≤0.25 | 0.5 | ≥1 | | C. glabrata | ≤0.06 | 0.12 | ≥0.25 | | C. guilliermondii | ≤2 | 4 | ≥8 | | C. krusei | ≤0.25 | 0.5 | ≥1 | | C. parapsilosis | ≤2 | 4 | ≥8 | | C. tropicalis | ≤0.25 | 0.5 | ≥1 | a According to the FDA STIC Webpage K243738 - Page 10 of 11 {10} K243738 - Page 11 of 11 VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission incorporated by reference a breakpoint change protocol that was reviewed and accepted by FDA in submission K231994 cleared on August 25, 2023. This referenced protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria). The referenced protocol outlined the specific procedures and acceptance criteria that Thermo Fisher Scientific intends to use to evaluate The Sensititre YeastOne Susceptibility System with Micafungin in the dilution range of 0.008-16 µg/mL when revised breakpoints for micafungin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, Thermo Fisher Scientific will update the micafungin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...