HEXIM Implant; Samwon General Abutments

K243731 · P&A USA · DZE · Jul 9, 2025 · Dental

Device Facts

Record IDK243731
Device NameHEXIM Implant; Samwon General Abutments
ApplicantP&A USA
Product CodeDZE · Dental
Decision DateJul 9, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HEXIM Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations. The HEXIM Implant System is for two stage surgical procedures. It is intended for delayed loading. Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.

Device Story

HEXIM Implant System comprises threaded, root-form endosseous dental implants and associated Samwon General Abutments (cover screws, healing abutments, couple abutments, and screws). Implants feature internal hexagon connections, self-tapping external threads, and a bottom cavity with five perpendicular channels for intra-bone fixation. Used by dental clinicians in clinical settings to provide stable foundations for prosthetic restorations (crowns, bridges, overdentures). Implants are surgically placed in the jawbone; after osseointegration, cover screws are removed and replaced with healing abutments to shape soft tissue, followed by final abutment and prosthetic attachment. System facilitates delayed loading protocols. Benefits include restoration of dental function and aesthetics for edentulous patients.

Clinical Evidence

Bench testing only. No clinical or animal data provided. Performance validated via fatigue testing (ISO 14801), biocompatibility (ISO 10993 series), sterilization validation (ISO 11137, ISO 17665), and MRI safety assessment.

Technological Characteristics

Materials: CP Ti Gr 4 (ASTM F67) for implants; Ti 6Al-4V ELI (ASTM F136) for abutments. Design: Root-form, threaded, internal hex connection. Surface: SLA blasted. Sterilization: Gamma irradiation (implants), steam (abutments). Connectivity: Mechanical, non-networked.

Indications for Use

Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations in the mandible or maxilla. Requires two-stage surgical procedure and delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION P&A USA July 09, 2025 % Chandler Thames Director of Quality, Consultant Rook Quality Systems 1155 Mount Vernon Hwy Suite 800 Dunwoody, Georgia 30338 Re: K243731 Trade/Device Name: HEXIM Implant; Samwon General Abutments Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 21, 2025 Received: June 5, 2025 Dear Chandler Thames: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243731 - Chandler Thames Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243731 - Chandler Thames Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Sherrill Lathrop Blitzer for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243731 Device Name HEXIM Implant; Samwon General Abutments Indications for Use (Describe) The Hexim Implant - The HEXIM Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations. The HEXIM Implant System is for two stage surgical procedures. It is intended for delayed loading. Samwon General Abutments - Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K243731 # 510(K) Summary ## I. SUBMITTER P&A USA, Inc. Mr. Jung Shin President 7535 Little River Turnpike Suite 310F Annadale, VA Phone: 503-314-9101 Email: pnausa7@gmail.com Date Prepared: 07/08/2025 ## II. DEVICE Trade Name: HEXIM Implant Classification Name: Implant, Endosseous, Root-Form Manufacturer Name: Uniance, Inc. Regulatory Class: II Primary Product Code: DZE Secondary Product Code: NHA Regulation Number: 872.3640 ## III. PREDICATE DEVICE a. HEXIM Implant Tapered Screw-Vent® M Implants/ Zimmer Dental Inc. (K111889 Primary Predicate) AR_N SLA Type Implant System/ Biotemp Co., Ltd. (K190641 Reference Device) b. Samwon General Abutment S-Plant Dental Implant System/ IDIS Co, Ltd. (K221866 Reference Device) SNUCONE Tissue Level Implant System / SNUCONE Co., LTD. (K222792 Reference Device) ## IV. DEVICE DESCRIPTION The HEXIM Implant with Samwon General Abutments is a Root-form Endosseous Dental Implant and Endosseous Dental Abutment system designed for use in dental implant surgery. These devices have special controls as described by the document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance for Industry and FDA Staff. The HEXIM Implant is a threaded root-form endosseous dental implant with an internal hexagon connection. The HEXIM implants are surgically inserted into the upper and/or lower jawbone to provide a stable foundation for restorations. Geometrically, the implant is screw type. It has external self-tapping threads with chip pocket vertical flutes to facilitate implantation. The HEXIM implant also includes a cavity which extends through the bottom of the implant and connects to 5 channels running perpendicular to the implant body long axis to aid in intra bone fixation. The bottom cavity is not directly interconnected to the internal hex connection nor to the internal threading feature used to fix abutments to the implant. {5} The Samwon General Abutments are connected to the implant through the implant's internally threaded hole and an internal hex connection. Samwon General Abutments come in several types: Cover Screw, Healing Abutment, Hex type Couple Abutment, and Abutment Screw. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. Healing abutments are used after osseointegration, once the cover screw is removed, to make a natural soft tissue shape before setting up prosthetics. The Hex-type Couple Abutment is designed to mate with the internal hex cavity of the endosseous implant and is fixed to the implant via an abutment screw. The Hex-type Couple Abutment is a transgingival component, which serves as the support for the artificial tooth or other prosthetic. ## System Dimensions a. HEXIM Implant | Fixture Body Diameter (mm) | 4.2,4.5, 5.0, 5.5, 6.0 | | --- | --- | | Fixture Length (mm) | Ø4.2: 7.0, 8.5,10,11.5,13 Ø4.5: 7.0, 8.5,10,11.5,13 Ø5.0: 7.0, 8.5,10,11.5,13 Ø5.5: 7.0, 8.5,10,11.5,13 Ø6.0: 7.0, 8.5,10,11.5,13 | b. Samwon General Abutment Screw | Fixture Body Diameter (mm) | Ø 3.35 | | --- | --- | c. Samwon General Abutment Healing Abutment | Fixture Body Diameter (mm) | Ø4.5, Ø4.8, Ø5.0, Ø5.5, Ø6.5 mm | | --- | --- | | Gingival Height | 3.5 – 7.5mm | | Total Length (mm) | 8.4 - 11.4 | d. Samwon General Abutment Screw {6} 3 | Samwon General Abutment Screw | | | --- | --- | | Diameter | Ø 2.32mm | | Length | 10.7mm | | General Abutment Hex Type Cover Abutment | | | --- | --- | | Body Diameter (mm) | Ø4.5, Ø4.8, Ø5.0, Ø5.5, Ø6.5 | | G/H (mm) | 0.9,1.9,2.9,3.9 | | Post Height (mm) | 4.0 | V. INDICATIONS FOR USE HEXIM Implant System: The HEXIM Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations. The HEXIM Implant System is for two stage surgical procedures. It is intended for delayed loading. Samwon General Abutments: Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover the screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. VI. TECHNOLOGICAL CHARACTERISTICS P&A USA submits the following information to demonstrate that the subject device the HEXIM Implant with Samwon General Abutments, is substantially equivalent to the following legally marketed predicate device: HEXIM Implant | 510(k) Number | Predicate Device Name/Manufacturer | Predicate/Reference | | --- | --- | --- | | K111889 | Tapered Screw-Vent® M Implants/ Zimmer Dental Inc. | Primary Predicate | | K190641 | AR_N SLA Type Implant System/ Biotemp Co., Ltd. | Reference | Samwon General Abutment {7} The HEXIM Implant has the same intended use, indications, principle of operation, similar technological and material characteristics as the predicate device, AR_N SLA Type Implant System (K190641). The Samwon Abutments have the same intended use, indications, principle of operation, and similar technological and material characteristics as the predicate device, S-Plant Dental Implant System (K221866) and reference device SNUCONE Tissue Level Implant System (K222792). The reference device was added for the healing abutment to give an example of gingival heights from cleared devices. Comparisons of the devices can be found in Tables 1-5: Table 1: Comparison of the Subject and Primary Predicate for HEXIM Implant | | Subject Device: HEXIM Implant System with Samwon General Abutments | Primary Predicate Zimmer Dental Inc. Tapered Screw-Vent® M Implants Predicate Device (K111889) | Reference Device AR_N SLA Type Implant System Reference Device (K190641) | Comparison | | --- | --- | --- | --- | --- | | Device Name | HEXIM Implant | Tapered Screw-Vent® M Implant | AR_N SLA Type Implant System | N/A | | Manufacturer | Uniance, Inc. | Zimmer Dental Inc. | Biotem Co., Ltd. | N/A | | Regulation Number | 872.3640 | 872.3640 | 872.3640 | Same | | Classification Name | Implant, Endosseous, Root-Form | Implant, Endosseous Root-Form | Implant, Endosseous, Root-Form | Same | | Product Code | DZE | DZE | DZE | Same | | Device Class | II | II | II | Same | | OTC or Rx | Rx | Rx | Rx | Same | | Indication for use | The HEXIM Implant is indicated for use in partially or fully edentulous mandibles and | The Tapered Screw-Vent® M Implants are designed for use in the maxilla or | AR_N SLA Type Implant System is indicated for use in partially or fully | Similar - The predicate and reference | {8} | | maxillae, in support of single or multiple-unit restorations. The HEXIM implant system is for two stage surgical procedures. It is intended for delayed loading. | mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. | edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. AR_N SLA Type Implant System is for two stage surgical procedures. It is intended for delayed loading. | device's Indications for Use do not include examples of single and multiple unit restorations . The predicate and reference device's Indications for Use do also include the option for immediate or delayed loading, whereas the subject device is indicated for delayed loading only. The difference does not impact substantial equivalence. | | --- | --- | --- | --- | --- | | Principle of Operation | This product is a root-type fixture which is inserted into the alveolar bone. It replaces the functions of the missing teeth as a dental implant fixture. | None given | None given | Different – The predicate and reference device did not explicitly describe a | 5 {9} | | | | | principle of operation but based on the special controls and the indications for use, we can infer they have equivalent principles of operation. | | --- | --- | --- | --- | --- | | **Material** | CP Ti Gr4 ASTM F67 | Titanium 6Al-4v | CP Ti Gr 4 ASTM F67 | Similar: The predicate device is made of a different titanium alloy than the reference and subject device. However, the material is the same metal which has similar mechanical properties, corrosion resistance, and purity. | | **Surface Treatment** | SLA | MTX Surface | SLA | Similar: The predicate device surface treatment is different than the | 6 {10} | | | | | predicate but is designed to perform a the same function. The reference device surface treatment is the same. This does not affect substantial equivalence. | | --- | --- | --- | --- | --- | | Sterilization | Gamma irradiation | Not specified on 510k summary | Gamma irradiation | Similar: The reference and subject device have the same sterilization method. The predicate device 510k summary did not list the sterilization method. | | Connection Type | Internal Hex | Internal Hex | Internal Hex | Same | | Fixture Body Diameter (mm) | 4.2,4.5, 5.0, 5.5, 6.0 | 3.7, 4.1, 4.7, 6.0 | 3.7, 4.2, 4.6, 5.1, 6.0 | Different: The predicate and reference device's fixture body | 7 {11} | | | | | diameters are slightly different from the subject device’s. The subject device’s diameters do fall within the range of the predicate and reference device, however. These differences do not impact substantial equivalence. | | --- | --- | --- | --- | --- | | Fixture Length (mm) | ∅4.2: 7.0, 8.5,10,11.5,13 ∅4.5: 7.0, 8.5,10,11.5,13 ∅5.0: 7.0, 8.5,10,11.5,13 ∅5.5: 7.0, 8.5,10,11.5,13 ∅6.0: 7.0, 8.5,10,11.5,13 | 8, 10, 11.5, 13, 16 | 7.5, 8.5, 10, 11.5, 13, 15 | Different: The subject device and predicate and reference device have similar lengths. While the shortest length of the subject device is outside the length of the predicate and reference | 8 {12} | | | | | device (7.0mm compared to 8mm, 7.5mm), the shorter length does not affect the safety or efficacy of the device. | | --- | --- | --- | --- | --- | | Presence of Bottom Cavity | The HEXIM implant has a cavity which extends through the bottom of the implant and connects to 5 channels running perpendicular to the implant body long axis to aid in intra bone fixation. The bottom cavity is not directly interconnected to the internal hex connection nor to the internal threading feature used to fix abutments to the implant. | The Tapered Screw-Vent® M Implants have a cylindrical cavity that extends to the bottom of the implant. | Absence of Bottom Cavity | Similar: The subject device and predicate device both have cavities that extend through the bottom of the implant. The reference device does not have a bottom cavity. | | Similarities | • The HEXIM implant system matches Tapered Screw-Vent® M Implants in base material and connection type. • The HEXIM implant system matches the AR_N SLA Type Implant System in material, surface finishing, sterilization method, and connection type. | | | | | Differences | • The HEXIM implant system’s diameters do not exactly match the diameters of the predicate or reference device, but its size ranges do fall within the range of those predicate and reference device and do not raise any issues regarding the safety and efficacy of the device. • The lengths of the HEXIM implant system do not exactly match the lengths of the predicate and reference device. HEXIM offers one length shorter than the predicate and reference at 7.0mm and does not offer a 15mm or 16mm length. The shorter length does not affect the safety or efficacy of the device. | | | | 9 {13} Table 2: Comparison of the Subject and Primary Predicate for Samwon General Abutment Cover Screw | | Subject Device: Samwon General abutment for use with the HEXIM Implant | Primary Predicate S-Plant Dental Implant System (K221866) | Comparison | | --- | --- | --- | --- | | Device Name | Samwon General Abutment Abutment Cover Screw | Cover Screw for S-Plant Dental Implant System | N/A | | Manufacturer | Samwon | IDIS Co, Ltd. | N/A | | Classification Name | Abutment, Implant, Dental, Endosseous | Abutment, Implant, Dental, Endosseous | Same | | Product Code | NHA | NHA | Same | | Device Class | II | II | Same | | OTC or Rx | Rx | Rx | Same | | Indications for use | Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used | Dual abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used | Same | {14} | | to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | | | --- | --- | --- | --- | | Principle of Operation | After establishing the fixation device in the alveolar bone, the screw is a standard M2.0 screw. | Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | Similar - The wording of the principle of operation achieves the same end use, but are worded differently. | | Material | Ti-6Al-4V ELI of ASTM F136 | Ti-6Al-4V ELI of ASTM F136 | Same | | Surface Treatment | None | None | Same | | Sterilization | End User (Steam) | End User (Steam) | Same | | Connection Type | Threaded (M2) | Threaded | Similar – All devices are threaded and are screwed into the Endosseous implant. | | Fixture Body Diameter (mm) | Ø 3.35 | Ø 3.6 | Different - While the fixture diameters are not the same, the smaller diameter of the | 11 {15} Table 3: Comparison of the Subject and Primary Predicate for Samwon General Abutment Healing Abutment | | Subject Device: Samwon General abutment for use with the HEXIM Implant | Predicate Device S-Plant Dental Implant System (K221866) | Reference Device SNUCONE Tissue Level Implant System (K222792) | Comparison | | --- | --- | --- | --- | --- | | Device Name | Samwon General Abutment Healing Abutment | Healing Abutment for the S-Plant Dental Implant System | InOcta Healing Abutment for SNUCONE Tissue Level Implant System | N/A | | Manufacturer | Samwon | IDIS Co, Ltd. | SNUCONE Co., LTD. | N/A | | Classification Name | Abutment, Implant, Dental, Endosseous | Abutment, Implant, Dental, Endosseous | Abutment, Implant, Dental, Endosseous | Same | | Product Code | NHA | NHA | NHA | Same | | Device Class | II | II | II | Same | | OTC or Rx | Rx | Rx | Rx | Same | | Indications for Use | Couple Abutments are intended for use with a dental implant to provide support for | Dual abutments are intended for use with a dental implant to provide support for | SNUCONE Tissue Level Implant System is indicated for use in partially or fully | Subject device is the same as the predicate device. The indications are | | | Dental Implant System | Dental Implant System | Dental Implant System | Dental Implant System | {16} | | prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or intermediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions. | different from the reference device because the reference device indications for use does not specify indications for the abutments. | | --- | --- | --- | --- | --- | | Principle of Operation | Before attaching the upper structure, it serves as a structure to create a more natural and aesthetically pleasing formation of the gingival tissue. | Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. | This product is healing Abutment to formation appropriate gingival shape during the soft tissue healing period combined with implant. This product should be removed when the superstructure is set up. | Same | | Material | Ti-6Al-4V ELI of ASTM F136 | Titanium Alloy | Ti-6Al-4V ELI of ASTM F136 | Same | 13 {17} | | | (Ti-6Al-4V, ASTM F136) | | | | --- | --- | --- | --- | --- | | Surface Treatment | None | None | Anodized | The subject device’s surface treatment is the same as the predicate device and different from the Reference Device. | | Sterilization | End User Sterilized (Steam) | End User Sterilized (Steam) | End User Sterilized | Same | | Connection Type | Threaded (M2) | Threaded | Threaded | Similar – both devices have male threading and are screwed into existing female threading in the endosseous implant. | | Fixture Body Diameter (mm) | Ø4.5, Ø4.8, Ø5.0, Ø5.5, Ø6.5 mm | Ø4.0, Ø4.5, Ø5.0, Ø6.0, Ø7.0 | Ø 5.5, Ø 6.1, Ø 6.3 | Different – The size range of the abutments. The differences do not impact the safety or efficacy of the device. | | Gingival Height | 3.5 – 7.5mm | None Listed | 4.0~6.5 …… | Different – A reference device was added to give an example of gingival heights from cleared devices. The subject devices additional gingival heights do not impact the safety or efficacy of the device; these additional heights | 14 {18} Table 4: Comparison of the Subject and Primary Predicate for Samwon General Abutment Screw | | Subject Device: Samwon General abutment for use with the HEXIM Implant System | Predicate Device: S-Plant Dental Implant System (K221866) | Comparison | | --- | --- | --- | --- | | Device Name | Samwon General Abutment Screw | Abutment screw for S-Plant Dental Implant System | N/A | | Manufacturer | Samwon | IDIS Co, Ltd. | N/A | | Classification Name | Abutment, Implant, Dental, Endosseous | Abutment, Implant, Dental, Endosseous | Same | | Product Code | NHA | NHA | Same | | Device Class | II | II | Same | {19} | OTC or Rx | Rx | Rx | Same | | --- | --- | --- | --- | | Indications for use | Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | Dual abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | Same | | Principle of Operation | As an intermediary connecting the fixation device with Couple Abutment the screw thread is a standard M2.0 screw | Abutment Screw is used to connect an abutment to the fixture | Same | | Material | Ti-6Al-4V ELI of ASTM F136 | Ti-6Al-4V ELI of ASTM F136 | Same | | Surface Treatment | None | None | Same | | Sterilization | End User (Steam) | Not Specified | Same | | Connection Type | Threaded (M2xP0.4) | Threaded | Same | 16 {20} Table 5: Comparison of the Subject and Primary Predicate for Samwon General Abutment Hex-type Couple Abutment | | Subject Device: Samwon General abutment for use with the HEXIM Implant | Predicate Device S-Plant Dental Implant System (K221866) | Comparison | | --- | --- | --- | --- | | Device Name | Hex-type Couple Abutment for the Samwon General abutment for use with the HEXIM Implant System | Dual Abutment for the S-Plant Dental Implant System | N/A | | Manufacturer | Samwon | IDIS Co, Ltd | N/A | | Product | HEXIM Implant | IDIS Co, Ltd | N/A | | Product Type | HEXIM Implant | IDIS Co, Ltd | N/A | | Product Type | HEXIM Implant | IDIS Co, Ltd | N/A | {21} | Classification Name | Abutment, Implant, Dental, Endosseous | Abutment, Implant, Dental, Endosseous | Same | | --- | --- | --- | --- | | Product Code | NHA | NHA | Same | | Device Class | II | II | Same | | OTC or Rx | Rx | Rx | Same | | Indication for use | Couple Abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | Dual abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete. | Same | | Principle of Operation | It can be used with an internal Hex fixture (Fixture) with a contracting part of 3.33mm and an internal angle of 11 degrees (included angle). Connection with the fixture is | Using making for general cement-type Prosthesis | Similar – These abutments support tooth prosthesis and are affixed to an endosseous implant with a screw. | 18 {22} 19 | | made using an Abutment Screw. | | | | --- | --- | --- | --- | | Material | Ti-6Al-4V ELI of ASTM F136 | Ti-6Al-4V ELI of ASTM F136 | Same | | Surface Treatment | None | None | Same | | Sterilization | End User (Steam) | End User (Steam) | Same | | Connection Type | Internal Hex | Tapered conical Hex | Similar - The subject device and K221866 have the same connection type. | | Body Diameter (mm) | Ø4.5, Ø4.8, Ø5.0, Ø5.5, Ø6.5 | Ø3.4, Ø3.6, Ø3.8, Ø4.2, Ø5.2, Ø7.0 | Different- The subject device’s range of diameters is within the S-Plant predicates range. | | G/H (mm) | 0.9,1.9,2.9,3.9 | 1,2,3,4,5 | Different – The Gingival height of the subject device is within the range of the predicate devices. | | Post Height (mm) | 4.0 | 4,5.5,7 | Similar – Both predicate devices have 4mm post heights. | | Similarities | The subject device has similar material, sterilization method, connection type, indications for use, and principle of operation. Both are supplied non-sterile. | | | | Differences | The Subject and predicate devices do have different diameters, but the subject device’s diameters are within the range of the predicates, so there is no risk to the safety or efficacy of the product. The subject device has different G/H measurements, but these differences do not affect the safety or efficacy of the device. | | | V. STERILIZATION AND SHELF-LIFE TESTING {23} Sterilization Validation testing for the HEXIM implants (fixtures) has been performed in accordance with: - ISO 11137:2006 Sterilization of health care products - ISO 11137-1:2006/Amd.2:2018 Sterilization of health care products - Requirements for validation and routine control - Radiation - Part One: Requirements for development, validation and routine control of a sterilization process for medical devices - ISO 11137-2:2019 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose for gamma sterilization for gamma sterilization. - Shelf life was established via accelerated aging per ASTM F 1980-2021 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices at 5 years. - LAL bacterial endotoxin testing per USP 41 <85>, rev. 05/2018, USP 41 <161>, rev. 05/2018, ANSI/AAMI ST72:2011. This testing recurs with each batch of implants during production. Steam Sterilization validation for non-sterile devices (abutments) has been performed in accordance with: - ANSI/AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilities - ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices - ISO 17665-2:2009 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1 The Samwon general abutments are supplied non-sterile with no stated shelf life. ## VI. BIOCOMPATIBILITY Biocompatibility has been conducted in accordance with the special controls document, Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. Table 6: Materials of the Proposed Device | Device Part | Material | | --- | --- | | HEXIM Implant | CP Ti Gr 4 ASTM F67 | | Samwon General Abutments | Ti 6Al-4V ELI ASTM F136 | ## HEXIM Implant Biocompatibility testing was conducted on representative final manufactured samples in accordance with: - ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Testing was performed according to ISO 10993-18, ISO 10993-5 Tests for in vitro cytotoxicity, - ISO 10993-18:2020 Biological evaluation of medical devices Part 18: Chemical characterization of medical device materials within a risk management process - ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity - ISO 10993-6:2016 Biological evaluation of medical devices Part 6: Tests for local effects after implantation - ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization - ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity - ISO 10993-23:2021 ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation 20 {24} 21 # Samwon General Abutments Biocompatibility testing was conducted on representative final manufactured samples in accordance with: - ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. - ISO 10993-18:2020 Biological evaluation of medical devices Part 18: Chemical characterization of medical device materials within a risk management process - ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity - ISO 10993-6:2016 Biological evaluation of medical devices Part 6: Tests for local effects after implantation - ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization - ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity ## VII. PERFORMANCE TESTING ### Bench Non-clinical validation testing was performed to demonstrate the safety and effectiveness of the device in accordance with the guidelines outlined in the FDA document "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The following tests were performed: - Fatigue Testing was performed according to ISO 14801:2007 using worst case geometries. - Implant to abutment compatibility was determined through mechanical testing. - A chemical and visual analysis of the HEXIM Implant's SLA blasted surface was performed to demonstrate no blasting particles or cleaning chemicals remain on the surface of the implant. - LAL bacterial endotoxin testing per USP 41 <85>, rev. 05/2018, USP 41 <161>, rev. 05/2018, ANSI/AAMI ST72:2011. This testing recurs with each batch of implants during production. The results of these comprehensive tests have met the criteria set by industry standards, establishing substantial equivalence with the predicate device. Non-clinical worst-case MRI review was performed to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental abutments, and fixation screws) and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque. ### Animal No animal or clinical testing was performed in support of this submission. ### Clinical No animal or clinical testing was performed in support of this submission. ## VIII. SUBSTANTIAL EQUIVALENCE CONCLUSION The HEXIM implant with Samwon General Abutments constitutes a substantially equivalent medical device. Both the implant and the abutments are similar to the predicate device in terms of design, dimension, material, surface treatment, intended use and technological characteristics. The differences identified in this analysis and listed below do not constitute substantial equivalence concerns. {25} The subject HEXIM implants have different diameters and lengths than the predicate device (K111889). The subject diameters and lengths fall within the range of the predicate device (K111889), so they do not constitute a substantial equivalence concern. The subject Cover Screw's body diameter is 3.35mm while the predicate's (K221866) is 3.6mm. This smaller diameter does not constitute a substantial equivalence concern. The subject Healing Abutment has different diameters and G/H measurements than the predicate device (K221866) and reference device (K222792). The subject device's diameters are within the range of the predicate's (K221866) and have been mechanically tested according to the special controls document, Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance for Industry and FDA Staff, so there is no risk to the safety or efficacy of the product. The subject device's G/H measurements are within the range of the a legally marketed device, reference device (K222792), so the difference does not constitute a substantial equivalence concern. The subject Abutment Screw and the predicate (K221866) device have different diameters and lengths. These differences do not impact the safety and efficacy of the product because the subject device's dimensions are compatible with the subject device's couple abutments and because the device has been mechanically tested according to special controls document Root-form Endosseous Dental Implants and Endosseous Dental Abutments – Class II Special Controls Guidance for Industry and FDA Staff therefore the difference does not constitute a substantial equivalence concern. The subject Hex Style Couple Abutment does have different diameters and G/H measurements than the predicate (K221866). The subject device's diameters are within the range of the predicate's, so it does not constitute a substantial equivalence concern. The difference between the subject device and predicates G/H have similar ranges with the subject device's length being 0.1mm smaller than the predicate's. This difference in length is insignificant with respect to the function of the device and does not constitute a substantial equivalence concern. 22
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...