K243712 · Abanza Tecnomed S.L · MBI · Jan 30, 2025 · Orthopedic
Device Facts
Record ID
K243712
Device Name
WasherCap™ Mini Fixation System (Model 45)
Applicant
Abanza Tecnomed S.L
Product Code
MBI · Orthopedic
Decision Date
Jan 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The WasherCap™ Mini Fixation System is intended for fixation of suture/tape (soft tissue) to bone.
Device Story
WasherCap™ Mini Fixation System is an implantable orthopedic device for soft tissue-to-bone fixation. It comprises a cap, threaded body, handled inserter, screwdriver, drill guide, drill bit, and suture threader. The surgeon uses the drill guide and bit to create a bone socket; the suture/tape is passed through the cap using the threader; the screw is then tightened with the screwdriver to set tension. Used in shoulder, knee, and hand/wrist procedures. The device is provided sterile for single-use. It benefits patients by providing secure fixation of soft tissue to bone during reconstructive surgeries.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including cyclic loading, elongation, insertion, interconnection, and bone anchor pullout force, compared against the predicate device. Bacterial endotoxin testing per EP 2.6.14/USP <85> was performed.
Technological Characteristics
Materials: PEEK and titanium alloy. Components: cap and threaded body. Preloaded on insertion tool. Sterilization: Ethylene Oxide. Single-use. Mechanical fixation principle via threaded component.
Indications for Use
Indicated for fixation of suture/tape (soft tissue) to bone in the shoulder, knee, and hand/wrist in skeletally mature pediatric and adult patients for: Knee (meniscal root repair, meniscal transplant, ACL/PCL repair, ACL reconstruction); Shoulder (rotator cuff repair); Hand/Wrist (triangular fibrocartilage complex).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
January 30, 2025
Abanza Tecnomed S.l. % Jessica Czamanski Regulatory Consultant Presicion Life Science Partners 300 Creek View Road Newark, Delaware 19711
Re: K243712
Trade/Device Name: WasherCapTM Mini Fixation System (Model 45) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: December 2, 2024 Received: December 2, 2024
Dear Jessica Czamanski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# CHRISTOPHER FERREIRA -S
Christopher Ferriera, MS, RAC Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243712
Device Name
WasherCap™ Mini Fixation System (Model 45)
Indications for Use (Describe)
WasherCap™ Mini Fixation System is intended for fixation of suture/tape (soft tissue) to bone in the shoulder, knee and hand/wrist, in skeletally mature pediatric and adult patients for the following procedures:
• Knee:
- Meniscal root repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- Meniscal transplant: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- ACL/PCL Repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- ACL Reconstruction: only with WasherCap™ Mini Fixation System 4.5mm size and using 2 nonabsorbable UHMWPE 1.4mm tapes or 1 non-absorbable UHMWPE 2.2mm tape.
### · Shoulder:
- Rotator Cuff Repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
• Hand/Wrist:
- Triangular fibrocartilage complex (TFCC): only with WasherCap™ Mini Fixation System 3.5mm size and using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
Type of Use (Select one or both, as applicable)
cription Use (Part 21 CFR 801 Subpart D)
Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Avanza. The logo consists of a stylized blue graphic on the left, resembling a curved shape with intersecting lines. To the right of the graphic is the word "ABANZA" in blue, with each letter clearly visible. The logo has a clean and modern design.
## K243712 – 510(k) Summary
This 510(k) Summary is provided per the requirements of section 21 CFR 807.92.
#### l. Submitter
| Submitter Name: | Abanza Tecnomed, S.L. |
|----------------------|-----------------------------------------------|
| Contact Person: | Ander Menaut<br>RD Manager |
| Address: | Calle Nueva 29<br>Multiva, ES, 31192<br>Spain |
| Telephone: | (+34) 948-044-643 |
| Email: | regulatory@abanzamed.com |
| Date of Preparation: | January 30, 2025 |
#### II. Application Correspondent
| Contact's Name: | Precision Life Sciences Partners LLC |
|-----------------|-------------------------------------------------|
| Contact Person: | Jessica Czamanski |
| Address: | 300 Creek View Road<br>Newark, DE, 19711<br>USA |
| Telephone: | (754) 422-9101 |
| Email: | jczamanski@plsp.pro |
#### III. Device
| Trade Name: | WasherCap™ Mini Fixation System (Model 45) |
|-------------------------|------------------------------------------------|
| Common Name: | Non-Degradable Soft Tissue Fixation Device |
| Classification Name: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
#### IV. Predicate Devices
| Primary Predicate | |
|-------------------|---------------------------------|
| Manufacturer: | Arthrex |
| Device Name: | Arthrex SwiveLock Suture Anchor |
| 510(k) Number: | K203495 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Avanza. The logo consists of a blue abstract shape on the left and the word "ABANZA" in blue on the right. The abstract shape is made up of curved lines that intersect each other. The word "ABANZA" is written in a sans-serif font.
| Product Classification: | Class II |
|-------------------------|-----------------|
| Regulation Number: | 21 CFR 888.3030 |
| Product Code: | MAI, MBI |
### Additional Predicates:
| Manufacturer: | Parcus Medical, LLC |
|-------------------------|----------------------------|
| Device Name: | X-Twist PEEK Suture Anchor |
| 510(k) Number: | K221135 |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
| Manufacturer: | Abanza Tecnomed, S.L. |
| Device Name: | WasherCap™ Fixation System |
| 510(k) Number: | K212197 |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR 888.3040 |
| Product Code: | MBI |
#### V. Device Description
The WasherCap™ Mini Fixation System consists of an implantable device, a handled inserter, a handled screwdriver, a drill guide, a drill bit and a suture threader. The WasherCap™ Mini implantable device is pre-loaded on the handled inserter with the suture threader.
The drill guide and the drill bit should be used to perform a drilling guided specifically to the angulation necessary to create the bone socket where the screw of the implant should be located. The suture/tape is passed through the cap with the suture threader and the screw is threaded with the handled screwdriver, setting the tension of the suture/tape.
The WasherCap™ Mini Fixation System is provided sterile and individually packaged. All components are for single use only.
The WasherCap™ Mini Fixation System is available in different sizes and should be implanted following the corresponding surgical procedure.
#### VI. Intended Use
The WasherCap™ Mini Fixation System is intended for fixation of suture/tape (soft tissue) to bone.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Avanza. The logo consists of a stylized blue graphic on the left, resembling a wing or a stylized checkmark with intersecting lines. To the right of the graphic is the word "ABANZA" in blue, with each letter in a sans-serif font. The overall design is clean and modern.
#### VII. Indications for Use
WasherCap™ Mini Fixation System is intended for fixation of suture/tape (soft tissue) to bone in the shoulder, knee and hand/wrist, in skeletally mature pediatric and adult patients for the following procedures:
- Knee:
- -Meniscal root repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- -Meniscal transplant: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- -ACL/PCL Repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
- -ACL Reconstruction: only with WasherCap™ Mini Fixation System 4.5mm size and using 2 nonabsorbable UHMWPE 1.4mm tapes or 1 non-absorbable UHMWPE 2.2mm tape.
- Shoulder:
- Rotator Cuff Repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable -UHMWPE 1.4-2.2mm tapes.
- Hand/Wrist:
- ၊ Triangular Fibrocartilage complex (TFCC): only with WasherCap™ Mini Fixation System 3.5mm size and using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes.
#### VIII. Comparison of Technological Characteristics
The proposed WasherCap™ Mini Fixation System is similar to its primary predicate (K203495) in that both fixate sutures/tapes to bone via a threaded component which holds the suture/tape in place. The main difference in technological characteristics is that the proposed WasherCap™ Mini Fixation System is composed of PEEK and titanium alloy, while the primary predicate (K203495) is only composed of PEEK, which leads to a different type of surface interfacing with the bone. Additionally, the threaded titanium component of the subject device is mainly screwed onto the PEEK component which is implanted into the bone and partially screwed into the bone, while the SwiveLock predicate device's threaded PEEK œmponent is entirely screwed into the bone. While both subject and predicate devices are preloaded onto their respective insertion tools, the predicate device is also pre-loaded with sutures that may or may not be used for fixation, while the subject device is not preloaded with sutures.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for ABANZA. The logo consists of a stylized blue icon on the left and the word "ABANZA" in blue text on the right. The icon appears to be an abstract representation of interconnected shapes.
The following table (Table 1) provides an overview of the comparison of the indications and technological characteristics of the subject and primary predicate device:
| Product<br>Features | Proposed<br>Abanza Tecnomed's<br>WasherCap Mini Fixation System<br>(K243712) | Primary Predicate<br>Arthrex's<br>Arthrex SwiveLok Anchor<br>(K203495) | Substantial Equivalence<br>Determination |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design/<br>Technological<br>Characteristics | - Components: cap and threaded<br>body<br>- Preloaded onto insertion tool<br>- Not preloaded with sutures | - Components: tip and<br>threaded body<br>- Preloaded onto insertion tool<br>- Preloaded with sutures that<br>may be implanted<br>Image: Arthrex SwiveLok Anchor | The subject device is composed of a<br>cap and screw that act together to<br>fixate the suture/tape to the bone<br>with the threaded metallic<br>component mainly screwed onto the<br>peek component and partially<br>screwed into the bone. The<br>threaded component of the<br>predicate device is entirely screwed<br>into the bone.<br>The differences in<br>design/technological characteristics<br>did not lead to different questions<br>for safety and effectiveness.<br>Therefore, the devices are<br>substantially equivalent. |
| Classification | Class II | Same | Substantially equivalent |
| Product Code | MBI | MAI, MBI | Substantially equivalent |
| Indications for<br>Use | WasherCap™ Mini Fixation System is intended for fixation of<br>suture/tape (soft tissue) to bone in<br>the shoulder, knee and hand/wrist,<br>in skeletally mature pediatric and<br>adult patients for the following<br>procedures:<br>- Knee:<br>Meniscal root repair: using<br>non-absorbable UHMWPE USP<br>2-0 sutures or non-absorbable<br>UHMWPE 1.4-2.2mm<br>tapes.<br>Meniscal transplant: using<br>non-absorbable UHMWPE USP<br>2-0 sutures or non-absorbable<br>UHMWPE 1.4-2.2mm<br>tapes.<br>- ACL/PCL Repair: using non-<br>absorbable UHMWPE USP 2-0<br>sutures or non-absorbable<br>UHMWPE 1.4-2.2mm tapes.<br>- ACL Reconstruction: only with<br>WasherCap™ Mini Fixation<br>System 4.5mm size and using 2<br>non-absorbable UHMWPE<br>1.4mm tapes or 1 non-<br>absorbable UHMWPE 2.2mm<br>tape. | The Arthrex SwiveLock Anchor is intended for fixation of suture<br>(soft tissue) to bone in the<br>shoulder, foot/ankle, knee,<br>hand/wrist, elbow, and hip in<br>skeletally mature pediatric and<br>adult patients for the following<br>procedures:<br>- Shoulder: Rotator Cuff<br>Repair, Bankart Repair, SLAP<br>Lesion Repair, Biceps<br>Tenodesis, Capsulolabral<br>Reconstruction, Acromio-<br>Clavicular Separation Repair,<br>Deltoid Repair, Capsular Shift<br>or Capsulo labral<br>Reconstruction.<br>- Knee: Anterior Cruciate<br>Ligament Repair (4.75- 5.5<br>SwiveLock Only), Medial<br>Collateral Ligament Repair,<br>Lateral Collateral Ligament<br>Repair, Patellar Tendon<br>Repair, Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis, Quadriceps<br>Tendon Repair (4.75<br>SwiveLock C Only), Meniscal<br>Root Repair (4.75 SwiveLock<br>C Only), Secondary or adjunct<br>fixation for ACL/PCL<br>reconstruction or repair<br>(4.75-5.5 SwiveLock only),<br>MPFL Repair/Reconstruction | The indications for use of the<br>primary predicate device include<br>additional anatomical sites which<br>have not been validated for the<br>subject device, such as hip, foot and<br>ankle, and elbow applications. Both<br>subject and predicate device<br>indications include applications for<br>the knee, shoulder, and hand/wrist,<br>where the predicate exceeds the<br>indications of the subject device for<br>these anatomical sites. Therefore,<br>the indications for the subject device<br>do not raise new questions for<br>safety or effectiveness and the<br>devices are substantially equivalent. |
| Product<br>Features | Proposed<br>Abanza Tecnomed's<br>WasherCap Mini Fixation System<br>(K243712) | Primary Predicate<br>Arthrex's<br>Arthrex SwiveLok Anchor<br>(K203495) | Substantial Equivalence<br>Determination |
| | - Rotator Cuff Repair: using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes. | Hand/Wrist: Scapholunate Ligament Reconstruction and Ulnar/Radial Collateral Ligament Reconstruction | |
| | • Hand/Wrist:<br>- TFCC: only with WasherCap™ Mini Fixation System 3.5mm size and using non-absorbable UHMWPE USP 2-0 sutures or non-absorbable UHMWPE 1.4-2.2mm tapes. | - Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction, and Lateral Epicondylitis repair<br>- Hip: Capsular Repair, Acetabular labral repair, Gluteus Medius Repair (4.75 - 5.5 mm PEEK SwiveLock suture anchors only), and Proximal Hamstring Repair (4.75 - 5.5 mm PEEK SwiveLock suture anchors only).<br>- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament/Tendon Repair, and Bunionectomy | |
### Table 1 Comparison of Subject and Primary Predicate Device
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Avanza. The logo consists of a stylized blue graphic on the left, resembling a curved shape with intersecting lines. To the right of the graphic is the word "ABANZA" in a bold, sans-serif font, also in blue. The overall design is clean and modern.
#### IX. Sterilization
There are no differences in the sterilization process between the subject and primary predicate device as both are sterilized via Ethylene Oxide.
#### X. Performance Data
Bench testing was performed on the subject device to evaluate cyclic loading and elongation, insertion, interconnection, and bone anchor pullout force. Bench testing was performed on the subject device and compared to the predicate device: cyclic loading and elongation, and bone anchor pullout force. The results of the testing met the defined acceptance criteria and were assessed against the results obtained with the predicate devices to determine substantial equivalence for design and performance.
Final devices were subjected to Bacterial endotoxin testing per EP 2.6.14/USP <85> to demonstrate that the device meets pyrogen limit specifications.
### Performance Testing - Animal
This submission does not include any animal performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ABANZA. The logo consists of a stylized, abstract shape on the left, resembling a curved or folded form with intersecting lines, rendered in a gradient of blue shades. To the right of the shape is the word "ABANZA" in capital letters, also in a matching blue color. The font appears to be sans-serif and modern.
### Performance Testing – Clinical
This submission does not include any clinical performance testing. It was determined that no such testing is required to demonstrate substantial equivalence.
#### XI. Conclusion
The proposed WasherCap™ Mini Fixation System has the same intended use and operating principle, the primary predicate device. Although the indications for use are different, the indications for the primary predicate exceed those for the proposed WasherCap™ Mini Fixation System. Performance testing was completed on both the subject and primary predicate device and results demonstrated equivalence. The differences in technological characteristics and materials did not raise new questions of safety or effectiveness. Therefore, the information provided in this submission demonstrated that the subject device, WasherCap™ Mini Fixation System, is substantially equivalent to its predicate, Arthrex SwiveLock Anchor K203495.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.