K243701 · Nanospectra Biosciences, Inc. · GEX · Jul 2, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243701
Device Name
AuroLase® System
Applicant
Nanospectra Biosciences, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AuroLase System is intended for use in ablation, incision, excision, coagulation and vaporization of soft tissues in open and endoscopic surgical procedures.
Device Story
AuroLase System is a portable, cart-based, liquid-cooled, dual-channel 810nm Class IV GaAs diode laser. It delivers 1-9W of continuous-wave near-infrared optical power via flexible optical fibers (Nanospectra Laser Delivery Device) to soft tissues. System includes an integrating sphere optometer for power calibration, a peristaltic pump for sterile saline coolant, ultrasonic flow sensors, and external thermocouple ports. Operated by healthcare professionals in clinical settings via a footswitch (tap-on/tap-off). Software monitors laser status, coolant flow, and system alerts. Laser energy induces photothermal ablation; liquid cooling moderates fiber surface temperature. Output allows for precise soft tissue management; benefits include controlled thermal effects during open or endoscopic procedures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and verification/validation activities.
Technological Characteristics
Dual-channel 810nm Class IV GaAs diode laser; 1-9W output per channel; 0.25W increments. Compatible with SMA-905 terminated fibers (≥600μm diameter, ≤0.37 NA). Integrated liquid cooling system (saline, 12 mL/min/channel). Includes integrating sphere optometer, ultrasonic flow sensors, and thermocouple ports. Cart-based form factor with medical-grade power supply. Software-controlled monitoring of laser status, coolant flow, and alerts. Sterilization/biocompatibility addressed via cleared LDD (K202953).
Indications for Use
Indicated for ablation, incision, excision, coagulation, and vaporization of soft tissues in open and endoscopic surgical procedures, including general surgery, ophthalmology/oculoplastic, urology, gastroenterology, gynecology, otorhinolaryngology, pulmonary/thoracic, dermatology/plastic surgery, neurosurgery (coagulation only), and orthopedic.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2025
Nanospectra Biosciences, Inc.
% Shepard Bentley
Principal Consultant
Bentley Biomedical Consulting, LLC
29461 Troon Street
Laguna Niguel, California 92677
Re: K243701
Trade/Device Name: AuroLase® System
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: June 2, 2025
Received: June 2, 2025
Dear Shepard Bentley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243701 - Shepard Bentley
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243701 - Shepard Bentley
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE
Digitally signed by
TANISHA L. HITHE
Date: 2025.07.02
19:47:12 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243701 | |
| Device Name AuroLase System | |
| Indications for Use (Describe) The AuroLase System is intended for use in delivering up to 9 Watts of continuous wave radiation to a flexible optical fiber for use in ablation, incision, excision, coagulation and vaporization of soft tissues in open and endoscopic surgical procedures including in general surgery, ophthalmology/oculoplastic, urology, gastroenterology, gynecology, otorhinolaryngology, pulmonary/thoracic, dermatology/plastic surgery, neurosurgery (coagulation only), and orthopedic. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.
The assigned 510(k) Number: K243701
1. Date of Preparation 07/01/2025
2. Applicant
Name: Nanospectra Biosciences, Inc.
Address: 8285 El Rio Street, Suite 150, Houston, Texas 77054
Contact Person: David Jorden, CEO
Telephone: 713.842.2720
FAX: 713.440.9349
Email: david.jorden@nanospectra.com
3. Identification of the Proposed Device
Trade/Device Name: AuroLase System
Common Name: Powered Laser Surgical Instrument
Classification Name: Powered Laser Surgical Instrument
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: GEX
4. Identification of Predicate Device
510(k) Number: K051996
Product Name: Diomed Delta 15
Manufacturer: Diomed, Ltd.
5. Device Description
The AuroLase® System (AS) is a portable, cart-based integrated liquid cooled laser system for use in professional healthcare facilities. The AuroLase System is used to generate photothermal ablation of selected soft tissues using 1-9W of near infrared optical power via 1 or 2 optical fibers either individually or simultaneously while providing liquid coolant to moderate the temperature at the fiber surface. The laser is delivered to the tissue of interest via an optical fiber system, the Nanospectra Laser Delivery Device (LDD), cleared via K202953. It is for prescription use only.
Central to the AuroLase System is the AuroLase Laser Device (ALD), a dual channel, continuous-wave, class IV GaAs diode laser (Figure 1 - red section) that provides the optical energy used to excite to soft tissues to ablative temperatures. The ALD incorporates an integrating sphere power meter (optometer) which is used to adjust the laser output that is delivered to the patient to the power shown on the display. Enabling the ALD's functionality is the AuroLase Control Unit (Figure 1 - green section) which incorporates: a) a peristaltic pump for providing sterile coolant to the Laser Delivery Devices, b) dual ultrasonic sensors to assure that coolant flows through the LDDs, c) a footswitch to control laser activation, d) temperature monitors for up to two optional external thermocouples, and e) a software application that controls, integrates, and monitors these functions. A medical-grade power supply is near the base of the unit to provide electrical service to the rest of the system as well as mechanical stability.
The AS is designed and tested for operation in a professional healthcare facility.
# AuroLase System Incorporates:
{5}
510(k) Summary
## AuroLase Laser Device (ALD)
- The ALD is a dual-channel 810nm laser system capable of 1-9W of output power per channel, adjustable in 0.25W increments.
- The optical outputs of the ALD are configured to insert their optical energy into mated SMA-905 terminated optical fibers with diameters of ≥600μm and numerical apertures of ≤ 0.37.
- An integrating sphere optometer head is used to calibrate the laser power extant at the termination of the diffuser-tipped optical fibers.
- Separate laser channel display information, e.g., channel standby/ready status, power output, and system alerts and alarms.
- A laser ON tone which is either continuous or periodic to distinguish between the two laser channels.

Figure 1. The AuroLase System rendering (left) shows the appearance and layout of the system components. The schematic (above) shows how the constituent component are related functionally.

- A laser “watchdog” timer in order to allow for re-positioning of the laser catheters without the laser returning to standby mode for up to 10min.
## AuroLase Control Unit (ACU)
- Conveys essential information to the OR staff, including:
- separate laser ready status
- coolant flow status
- system warnings and alerts
## Fluid Handling Unit
- houses the coolant pump and electronics.
- houses the two coolant flow sensors and electronics.
- houses two thermocouple ports and electronics.
## Storage
- The ACU has covered drawers for storing:
{6}
510(k) Summary
- safety eyewear, 8-10 pair
## External Mounting Hardware
- A hook for mounting 0.5-3.0L saline or water coolant supply bags.
- A hook for mounting two 1.5L Coolant Recovery Bags.
## Laser Actuation Footswitch
- Footswitch actuation is a momentary switch to enable “tap-on/tap-off” operation rather than continuous depression.
- The footswitch simultaneously actuates the laser, coolant pump, and countdown timer.
## Medical Grade Power Supply
- Located in base to provide mechanical stability.
## 6. Indications for Use
**Intended Use**
The AuroLase System is intended for use in ablation, incision, excision, coagulation and vaporization of soft tissues in open and endoscopic surgical procedures.
**Indications for Use**
The AuroLase System is intended for use in delivering up to 9 Watts of continuous wave radiation to a flexible optical fiber for use in ablation, incision, excision, coagulation and vaporization of soft tissues in open and endoscopic surgical procedures including in general surgery, ophthalmology/oculoplastic, urology, gastroenterology, gynecology, otorhinolaryngology, pulmonary/thoracic, dermatology/plastic surgery, neurosurgery (coagulation only), and orthopedic.
## 7. Substantially Equivalent (SE) Comparison
Comparison of the Indications for Use: Indications for use statement of the subject and predicate devices are comparable.
Comparison of Technology:
| Device & Predicate Device(s): | K243701 | K051996 |
| --- | --- | --- |
| General Device Characteristics | | |
| Laser Source | Diode | Diode |
| Wavelength (nm) | 810±10 | 810±20 |
| Power (W) | 1.0-9.0 W/fiber
Up to 18 W total | ≤ 15 |
| Pulse width (s) | NA (CW) | 0.1-CW |
| Repetition Rate (s) | NA (CW) | 0.1-CW |
| Length of the laser irradiating part of the applicator (mm) | 10/20/30
(isotropic) | 10 mm or end firing |
| Fiber Cooled | Yes | No |
| Coolant | Saline | NA |
| Flow Rate | 0 or 12 mL/min/channel
0 or 24 mL/min for two channels | NA |
| No. of deployable fibers | 1 or 2
(see below) | 1 |
| Aiming Beam | 650 nm | 650 nm |
The technological features of the subject device are comparable to the corresponding technological
{7}
510(k) Summary
features of the predicate device. Any difference in the technological features do not raise different questions of safety and effectiveness.
8. Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The following tests were conducted:
- IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance;
- IEC 60601-1-2:2014 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility;
- Software Validation & Verification Test;
- Bench Testing to verify the performance.
9. Clinical Testing
No clinical study is included in this submission.
10. Conclusion
Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.