K243692 · Advanced Technology Research (A.T.R.) S.R.L. · EKX · Aug 22, 2025 · Dental
Device Facts
Record ID
K243692
Device Name
EnDrive (EnDriveUS)
Applicant
Advanced Technology Research (A.T.R.) S.R.L.
Product Code
EKX · Dental
Decision Date
Aug 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
EnDrive is a battery-operated electronic medical device used for the treatment of root pulp infection, which allows root canal preparation by means of continuously or alternately rotating endodontic instruments with torque control, or retreatment of an already treated tooth, with integrated apex locator.
Device Story
EnDrive is a cordless, battery-operated dental handpiece with an integrated apex locator for endodontic procedures. It accepts NiTi endodontic files to shape root canals via rotary or reciprocating motion. The device uses a brushless DC motor with 3 phases and 3 Hall sensors to control speed and torque. The apex locator function uses a lip clip and file clamp to measure the distance to the root apex via electrical impedance. The system includes a control unit with a touchscreen interface and a handpiece that connects via Bluetooth. Dentists use the device to remove infected tissue and debris while monitoring file position in real-time. Output parameters (speed, torque, apex position) are displayed on the control unit. The device aids in determining working length and performing canal preparation, potentially improving treatment efficiency and accuracy. It is intended for use by qualified dental professionals in clinical environments.
Clinical Evidence
Bench testing only. Performance evaluated for motor speed (250-1000 rpm), torque (0.4-5.0 Ncm), and apex locator accuracy (±0.5mm). Biocompatibility testing performed per ISO 10993-1, -5, -18, and -23. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Reprocessing validated per ISO 17665-1 and ISO 17664-1.
Technological Characteristics
Portable, battery-powered (NiMH control unit, Li-ion handpiece) dental handpiece. Brushless DC motor with 3 Hall sensors. Touchscreen interface. Bluetooth connectivity between handpiece and control unit. Wi-Fi for firmware updates. Patient-contacting materials (stainless steel lip clip) tested for biocompatibility. Sterilizable accessories (contra-angle, file clamp, lip clip). Complies with IEC 60601-1, IEC 60601-1-2, IEC 62133-2, ISO 14457, ISO 17665-1, ISO 17664-1.
Indications for Use
Indicated for patients requiring root canal preparation or retreatment of infected root pulp. Used by qualified dental personnel in hospital, clinic, or dental office settings.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 22, 2025
Advanced Technology Research (A.T.R.) s.r.l.
Lorenzo Baroncelli
Quality and Regulatory Manager
Via San Donato 345
Pistoia, PT 51100
ITALY
Re: K243692
Trade/Device Name: EnDrive (EnDriveUS)
Regulation Number: 21 CFR 872.4200
Regulation Name: Dental Handpiece And Accessories
Regulatory Class: Class I, reserved
Product Code: EKX, LQY
Dated: November 29, 2024
Received: November 29, 2024
Dear Lorenzo Baroncelli:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243692 - Lorenzo Baroncelli
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243692 - Lorenzo Baroncelli
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243692
Device Name
EnDrive (EnDriveUS)
Indications for Use (Describe)
EnDrive is a battery-operated electronic medical device used for the treatment of root pulp infection, which allows root canal preparation by means of continuously or alternately rotating endodontic instruments with torque control, or retreatment of an already treated tooth, with integrated apex locator.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) #: K243692
510(k) Summary
Prepared on: 2025-08-20
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | Advanced Technology Research (A.T.R.) s.r.l. |
| Applicant Address | Via San Donato 345 Pistoia PT 51100 Italy |
| Applicant Contact Telephone | +390573535861 |
| Applicant Contact | Mr. Lorenzo Baroncelli |
| Applicant Contact Email | lorenzo.baroncelli@atrdental.com |
| Device Name | 21 CFR 807.92(a)(2) |
| Device Trade Name | EnDrive (EnDriveUS) |
| Common Name | Dental handpiece and accessories |
| Classification Name | Handpiece, Direct Drive, Ac-Powered |
| Regulation Number | 872.4200 |
| Product Code(s) | EKX, LQY |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) |
| K220829 | Endo Motor |
| Product Code | EKX |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
| EnDrive is a battery-operated electronic medical device used for the treatment of root pulp infection, which allows root canal preparation by means of continuously or alternately rotating endodontic instruments with torque control, or retreatment of an already treated tooth, with integrated apex locator. We can identify two different functional principles: one is the motor function which serves to shape the root canal the other is the apex locator function to determine the apex approach when the file is in the root canal. The motor function, when the file has a movement transmitted from the motor through the contra-angle, allows the proper shaping of a root canal with infected pulp through rotary or reciprocating movement, helping to remove debris and infected tissue from the canal. The motor, controlled in speed and torque, uses brushless DC technology with 3 phases and 3 Hall sensors, driven by an embedded software. The apex locator function allows to determine when, during an endodontic treatment, the tip of an endodontic instrument, in steel or nickel-titanium, inserted by the dentist inside the root canal, is at the apex of the canal. Two electrodes are applied to the patient, a lip clip on the lip and a file clamp to the endodontic instrument positioned inside the canal. When the device signals the apex has been reached, the dentist positions the instrument stopper on the dental crown and indirectly goes to measure the working length of the canal using a ruler for endodontic instruments, so the length from which the dentist must remove the infected pulp. | |
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| EnDrive is a battery-operated electronic medical device used for the treatment of root pulp infection, which allows root canal preparation by means of continuously or alternately rotating endodontic instruments with torque control, or retreatment of an already treated tooth, with integrated apex locator. | |
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| The intended use of the two devices are equivalent, infact: | |
{5}
EnDrive intended use: EnDrive is a battery-operated electronic medical device used for the treatment of root pulp infection, which allows root canal preparation by means of continuously or alternately rotating endodontic instruments with torque control, or retreatment of an already treated tooth, with integrated apex locator.
Endo Motor intended use: The Endo Motor device is an endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length.
# Technological Comparison
21 CFR 807.92(a)(6)
The EnDrive motor is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. The product is a portable device powered by built-in lithium battery used for the handpiece and NiMH battery for the control unit; the control unit is charged by a medical adapter. LCD displays parameters such as selected NiTi file, speed, torque, apex position, etc. Users can also set and modify the settings using the touch screen.
The handpiece connects automatically to the control unit via Bluetooth.
The EnDrive is an endodontic treatment motorized equipment with root canal measurement capability. It can be used to enlarge the canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel. The accessories of EnDrive (contra-angle, file clamp and lip clip) are intended to be sterilized prior to use.
The material composition of root canal files that can be used with the EnDrive is Nickel titanium (NiTi). These files are not included in the submission.
There are three working modes as:
- Endo motor mode: root canal preparation.
- Apex locator mode: root canal length measurement.
- Multi-function mode: undertake root canal measurement and preparation.
The subject device, and the predicate device, C-SMART-I Pilot (K220829), share the same intended use as motorized endodontic handpieces with integrated apex locator capability. Both are intended for root canal preparation and canal length measurement in dental clinical settings, operated by qualified dental professionals. The anatomical site of use (root canal and softened dentin) and the clinical environment (dental clinics, hospitals, university settings) are identical. The two devices are based on the same fundamental technology, consisting of a cordless motorized handpiece powered by an internal rechargeable battery, coupled with a control unit that provides real-time display of operating parameters such as speed, torque, and apex position. Both devices support continuous rotation and reciprocating movements, as well as multiple operating modes: endo motor, apex locator, and combined multifunction mode. Safety mechanisms are equivalent, including automatic reverse rotation when preset torque thresholds are reached. The subject device and predicate device are highly similar in their technical performance characteristics. Both achieve root canal length measurement accuracy within $\pm 0.5\mathrm{mm}$ and use nickel-titanium (NiTi) files. Accessories such as contra-angle, file clamp, and lip clip are designed for sterilization prior to use, in line with standard infection control requirements. Both devices require the usage of a barrier sleeve on the handpiece body to prevent cross-contamination.
Some minor differences exist between the devices:
- Power supply: the EnDrive control unit uses a NiMH battery at 12V, compared to the 7.4V Li-ion battery of the predicate.
- User interface: EnDrive incorporates a touch screen interface with a single power button, while the predicate combines a touchscreen with additional physical buttons.
- Physical dimensions differ slightly for both control unit and handpiece, but these variations do not alter function.
- Performance ranges: EnDrive operates at 250–1000 rpm (vs. 150–1000 rpm) and 0.4–5.0 N·cm torque (vs. 0.6–3.9 N·cm). These ranges overlap and ensure equivalent clinical performance.
- Charging method: EnDrive charges via cable connection between the control unit and handpiece, while the predicate supports wireless charging. Both methods have been demonstrated to be safe and effective.
- Spray nozzle: EnDrive includes a nozzle for lubrication of the contra-angle, while the predicate also incorporates a spray nozzle; these differences are accessory-related and have no impact on performance.
- Patient-contacting materials: EnDrive maintains direct patient contact only through the stainless-steel lip clip, while indirect patient contact may occur via the contra-angle and file clamp. Biocompatibility testing has been completed for the direct and indirect patient-contacting components in accordance with ISO 10993-1, -5, and -18.
- Applied standards: minor variations exist in referenced biocompatibility and performance standards, but both devices comply with the relevant ISO and IEC standards for electrical, mechanical, and thermal safety, as well as EMC requirements.
- Firmware updates: EnDrive supports software updates via Wi-Fi, while the predicate does not specify such a feature.
These differences are minor in nature, have been verified through bench testing and design validation, and do not raise new safety or effectiveness concerns.
{6}
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Based on the bench test reports assessing the apex locator, motor speed and torque accuracies, the subject device is subsanically equivalent based on the following:
- The motor speed range for the subject device is 250-1000 rpm, which is within the range of the predicate device (150-1000 rpm)
- The accuracy of the root apex locator function for the subject and predicate devices is $\pm 0.5\mathrm{mm}$
- The torque is between $0.4 - 5.0\mathrm{Ncm}$ for the subject device is slightly different from that of the predicate $(0.6 - 3.9\mathrm{Ncm})$, but this difference is minor and does not impact substantial equivalence between the two devices.
In addition to these key features that have been tested, biocompatibility, EMC, reprocessing, and device specific performance tests were conducted for the subject in accordance with the following standards:
- ISO 10993-1:20128
- ISO 10993-5:2009
- ISO 10993-18:2020
- ISO 10993-23:2021
- IEC 60601-1:2020
- IEC 60601-1-2:2020
- IEC 62133-2:2017
- ISO 14457:2017
- ISO 17665-1:2006
- ISO 17664-1:2021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.